Taylor Technologies K-1530 Owner's manual

Type
Owner's manual
InstructionNumber:5207
Component Description
> DROPTEST
> TOTALALKALINITY(1drop=10ppm)
COMPONENTS:
1x5207 Instruction
1x9198G SampleTube,Graduated,25mL,plastic
w/capandgreendot
1xR‐0645‐CTotalAlkalinityIndicator,2oz,DB
1xR‐0687G‐CSulfuricAcid.12N(greencap),2oz,DB
TOORDERREPLACEMENTPARTSANDREAGENTSCALLTOLL‐FREE
800‐TESTKIT(800‐837‐8548).
PROCEDURE:
CAREFULLYREADANDFOLLOWPRECAUTIONSONREAGENTLABELS.
KEEPREAGENTSAWAYFROMCHILDREN.
1. Rinseandfill25mLsampletube(#9198G)to25mL
markwithwatertobetested.
2. Add5dropsR‐0645TotalAlkalinityIndicator.
Swirltomix.Sampleshouldturngreen.
3. AddR‐0687GSulfuricAcid.12Ndropwise,swirling
andcountingaftereachdrop,untilcolorchanges
fromgreentored.Alwaysholdbottleinvertical
position.
4. MultiplydropsofR‐0687GSulfuricAcid.12Nby10.
Recordaspartspermillion(ppm)totalalkalinity
ascalciumcarbonate.
03/09
©2016TaylorTechnologies,Inc.•AllRightsReserved•ViewPrivacyPolicy