2
Fragment Library Preparation Using the AB Library Builder™ System: SOLiD™4 System Quick Reference
2Shear the DNA with the
Covaris® S2 System
(continued)
d. Shear the DNA using the appropriate program on the Covaris® S2 System:
IMPORTANT! Ensure that the bath temperature limit is set at 15°C, and keep the bath
temperature to ≤ 10°C.
e. Place the Covaris® microTUBE into the loading station, then transfer the 110 µL of sheared DNA
to a new 1.5-mL sample tube provided in the Library Builder™ Fragment Core Kit for SOLiD™ 4.0.
3Set up the AB Library
Builder™ System
Instrument for standard
or express fragment
library preparation
IMPORTANT! If you are preparing barcoded libraries for multiplexed sequencing, for each
sequencing run, use at least one of the following full sets of four barcodes: Barcodes 1–4, 5–8, 9–12,
13–16, 17–20, 21–24, 25–28, 29–32, 33–36, 37–40, 41–44, 45–48, 49–52, 53–56, 57–60, 61–64, 65–68,
69–72, 73–76, 77–80, 81–84, 85–88, 89–92, or 93–96. Use only one of the barcoded-T-0XX adaptors
for each ligation reaction, unless < 4 libraries are being barcoded.
Use the barcodes according to these conditions:
• If <4 libraries are prepared for sequencing, then use multiple barcodes per library in equal ratios.
For example, for 2 libraries, use 2 barcodes for each library. For 3 libraries, use 4 barcode adaptors
for each library for a total of 12 barcodes.
•If ≥4 libraries are prepared for sequencing and libraries are split into sets of 4 to use full sets of
barcodes, then use one set of barcodes for the remaining libraries (1,2,or 3 libraries). There is no
need to use multiple barcodes per library in equal ratios.
a. Thaw the AB Library Builder™ Cartridges at room temperature or on ice for ≤ 2 hours or until
completely thawed.
IMPORTANT! Avoid leaving the cartridges at room temperature for longer than necessary to
completely thaw them. Avoid repeated freeze-thawings of unused cartridges.
IMPORTANT! Before loading the cartridges into the cartridge rack, ensure that the cartridges
are completely thawed, particularly reagents in cartridge wells 2 and 3.
Condition Setting
Number of cycles 6
Bath temperature 5°C
Bath temperature limit 15°C
Mode Frequency
sweeping
Water quality testing function Off
Duty cycle 10%
Intensity 5
Cycles/burst 100
Time 60 seconds