RKI GX-2012 User guide

  • Hello! I am an AI chatbot trained to assist you with the RKI GX-2012 User guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
are turned on: LNCH BRK, ID DISP, CL RMNDR, and BP RMNDR Set to
CONFIRM
a. AttachtherubbernozzleorsamplehosetotheGX-2012’s quickconnectinletfitting.
b. Ifahoseisused,attachprobetohose.
c. ToturnontheGX-2012,pressandbrieflyholdthePOWERENTERbuttonuntilyouheara
beep.TheGX-2012willbegin itswarmupsequence.Thefollowingscreensmayappear
duringthewarmupsequence:
CALTIMEREMAINING(factorysettoappear)
Iftheinstrumentifoverdueforcalibration,theGX-2012 displaysCAL.PresstheRESET
SILENCEbuttontocontinue.
Calibrationshouldbeperformedassoonaspossible.
BUMPTIMEREMAINING(doesnotappearwithfactorysetting)
Iftheinstrumentisoverdueforbumptesting,the
GX-2012displays“tEStB--Limit”.PresstheRESETSILENCEbuttontocontinue.
Abumptestshouldbeperformedassoonaspossible.
SENSORFAILURE
IftheGX-2012experiencesasensorfailureduringstartup,theLCDwilldisplayFAILand
indicatewhichsensor(s) failed.TocontinueoperationpressandreleaseRESETSILENCEto
acknowledgethefailure.Gasreadingsforthatsensorwillbereplacedby“---”.
Itisnecessarytoreplacethesensorandrecalibrate beforetheGX-2012isused.
2. MeasuringModeScreen
a. AfterwarmuptheGX-2012willdisplaythefollowing:CH40%LEL,OXY20.9%,CO0ppm,
H2S0.0ppm(somereadings maynotappearifsensorsarenotinstalled).
b. Ifthereadingsarenotdisplayingfreshairvaluesasabove, youmustperformaFreshAir
Adjustment.
QuickReferenceGuideForModel
GX-2012
Note: Turn on and adjust the GX-2012 gas monitor in a known
fresh air area.
1.TurningtheGX-2012ON
Note: This start up sequence assumes the following maintenance mode items
www.calcert.com sales@calcert.com1.888.610.7664 0
5
10
15
20
25
30
3. PerformingaFreshAirAdjustment
a. PerformingaFreshAirAdjustmentwillsettheCH4,CO,and H2Schannelsto0andthe
OXYchannelto20.9%.
b. Taketheinstrumenttoafreshairareawheretheairisfreeoftoxicorcombustiblegasand
hasnormaloxygen(20.9%).
c. PressandholdtheAIRbutton.Thedisplaywillpromptyou tocontinueholdingtheAIR
button.
d. ThedisplaywillpromptyoutoreleasetheAIRbutton.TheGX-2012willautomaticallyset
thefreshairreadingsforall channels.
e. TheGX-2012isnowreadyforuse.
Note: If fresh air adjustment fails, refer to user’s manual.
4. FlowIntegrityTest
a. Placefingeroverendofprobeorrubbertip.TheGX-2012 shouldgointoflowfailinafew
seconds.
b. PresstheRESETSILENCEbuttontorestartthepump.
c. Iftheinstrumentfailstoindicateflowfailure,checkprobe andhoseconnectionsforleaks.
5. BreathTest(forinstrumentsequippedwith oxygensensor)
a. Ifconnected,removehoseandprobe.
b. Exhaleneartheinletfittingoftheunit.Observethattheoxygenreadingdropsbelowthe
alarmsettingof19.5%and theaudibleandvisualalarmsareactivated.
c. PresstheRESETSILENCEbuttontoresetthealarmsthe reconnecthoseandprobe.
6. TurningtheGX-2012OFF
PressandholdthePOWERENTERbuttonuntilTURNOFF
has disappeared from the bottom of the screen. The GX-2012 is now off.
PN71-0249RKRev0
120110/1000