iGear iG-PO1 User manual

Type
User manual

iGear iG-PO1 is a compact and lightweight fingertip pulse oximeter designed to measure blood oxygen saturation (SpO2) and pulse rate accurately. Its advanced DSP algorithm minimizes the effects of motion artifacts, ensuring precise readings even under low perfusion conditions. With a colorful LED display and four display directions, it allows for easy viewing from different angles. The device operates on two AAA batteries, providing up to 16 seconds of continuous use before automatically turning off to conserve power.

iGear iG-PO1 is a compact and lightweight fingertip pulse oximeter designed to measure blood oxygen saturation (SpO2) and pulse rate accurately. Its advanced DSP algorithm minimizes the effects of motion artifacts, ensuring precise readings even under low perfusion conditions. With a colorful LED display and four display directions, it allows for easy viewing from different angles. The device operates on two AAA batteries, providing up to 16 seconds of continuous use before automatically turning off to conserve power.

Model No.: iG-PO1
General instructions
The Fingertip Pulse Oximeter is used to detect the
blood oxygen saturation and pulse rate of the
human body. This product is based on digital blood
oxygen technology and uses an advanced DSP
algorithm to reduce the effects of motion artifacts
and improve the measurement accuracy under
weak perfusion.
Technical Parameters
Display mode: Colourful LED display, 4 display
directions.
Power supply: 1.5 V (AAA size) alkaline batteries x2
Working current: 30mA
Measurement performance under weak perfusion
Conditions: 0.3%
Measuring Specifications:
Oxygen Saturation (SpO2)
Measuring range 0-100%
Resolution 1%
Accuracy 70~100%, 2%, 0-69%
Pulse Rate
Measuring range 25-250 bpm
Resolution 1 bpm
Accuracy 2 bpm
Features
The product is simple and convenient to operate.
Small in size and light in weigh (including the total
weight of the battery is about 50 grams), easy to carry.
The power consumption is low. Can use Two AAA
batteries.
• Show current battery usage. When the voltage is
very low, the normal operation of the oximeter may
be affected.
When the finger is not in the measurement
position for more than 16 seconds, the product will
automatically turn off the power.
Product use range
The Fingertip Pulse Oximeter is used to measure the
blood oxygen saturation and pulse rate of the human
body through fingers. The product is suitable for
families, hospitals (including medical and surgical
departments, anaesthesiology, pediatrics, intensive
care, etc.), oxygen bars, community medical, sports and
health care.
This device is not intended for continuous monitoring.
Operation guide
Press the power switch on the front panel to turn on
the power pf Oximeter.
Pinch the clip of the oximeter, and then put one of
the fingers completely into cavity of the oximeter.
Press the power switch again to change the display
direction.
Read the data on the display.
When the finger is placed in the oximeter, the
nail surface must be upward.
If your fingers are not fully inserted, it may cause
measurement errors.
Disclaimer: Before and after each test, please use
medical alcohol to clean the oximeter rubber parts
that touch your fingers (The rubber in the oximeter
is medical rubbber, which is non-toxic and
harmless to human skin).
Control panel instructions
The height of the bar graph can indicate the
strength of the pulse.
Product accessories
Lanyard x 1
User manual x1
Battery installation
Put the AAA batteries x2 into the battery box
with correct polarity.
Close the battery cover.
Note: The battery polarity must be installed
correctly. Otherwise, the machine may be damaged.
If the oximeter is not used for a long time, the
battery must be removed.
Lanyard installation
The thin thread at the end of the lanyard should
pass through the perforation used to assemble
the lanyard.
Pass the thick thread at the head of the lanyard
through the thin thread at the tail and tighten it.
Maintenance and storage
When the battery is low, please replace the
battery in time.
Please clean the oximeter before diagnosing the
patient.
It is best to place the oximeter in an
environment with a temperature of-10 to 400C
and a humidity of 10%-95%.
Avoid direct sunlight and exposure to strong
radioactive infrared or ultraviolet rays.
It is recommended to put the product in a dry
environment. A humid environment may affect
its life and even damage the product.
Please dispose of used batteries in accordance
with local government laws.
Dimensions
Length: 58mm
Width: 31mm
Height: 29mm
Weight: 50grams (including AAA batteries x2)
Environmental requirements
Working temperature: 5-400C
Storage temperature: 10-500C
Ambient humidity: 15%-80% work
10%-95% storage
Faults and solutions
There are no user-serviceable parts in the
oximeter, and only qualified personnel can
perform repairs.
If you are not sure whether the measurement is
accurate, you can first check the vital signs of
the patient through other methods, and then
ensure that the patient through other methods,
and then ensure that the oximeter is in a normal
working state.
DO not pour, or spill any liquid into the accessories,
connectors, switches, or openings in the oximeter
because it will damage the oximeter.
Questions
Blood oxygen
saturation and
pulse rate
cannot be
displayed
normally
The blood
oxygen
saturation and
pulse rate
display are
unstable
Cannot turn
on the power
The screen
suddenly
turns off the
display
Possible reason
1. Incorrect finger
insertion.
2. The patient’s
perfusion is too low.
1. The finger may not
be fully inserted
2. Fingers are shaking
or the patient’s body
is moving
1. The battery is low or
the battery is not
installed
2. The battery is not
installed correctly.
3. The oximeter is
damaged
1. Shut down without
detecting a finger for
a long time
2. The battery is
exhausted
Solution
1. Retry to insert your
finger
2. Try several more
measurements, if you
are sure that the
product does not have
any problems, please
go to the hospital for
an accurate diagnosis
in time
1. Try to insert fingers
again
2. Don’t move
1. Please replace the
battery
2. Reinstall the battery
with the correct
polarity
3. Please contact your
local customer service
center
1. Normal
2. Replace the batteries
Sign
%SpO2
/Min
Definition
Class BF equipment
(refer to IEC 60601-1: 1995)
Attention! Please refer to this manual
Arterial oxygen saturation
Pulse rate
1. Symbol for electrical and electronic
equipment according to 200296/EC
2. The waste after use of the
equipment, accessories and
packaging should be properly
handled, please follow the local
disposal regulations or rules
Note: The oximeter applies to this rule
Sign definition
Manufactured and Marketed By:
iGear Tech Pvt. Ltd.
204, Sai Corporate Park, Laxmi Nagar, Off Link
Road, Goregaon West, Mumbai - 400104.
In case of consumer complaints:
Toll free number - 1800 22 8495
(Between 10am to 5pm, Mon- Fri)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8

iGear iG-PO1 User manual

Type
User manual

iGear iG-PO1 is a compact and lightweight fingertip pulse oximeter designed to measure blood oxygen saturation (SpO2) and pulse rate accurately. Its advanced DSP algorithm minimizes the effects of motion artifacts, ensuring precise readings even under low perfusion conditions. With a colorful LED display and four display directions, it allows for easy viewing from different angles. The device operates on two AAA batteries, providing up to 16 seconds of continuous use before automatically turning off to conserve power.