Page 7
Operating Instructions
Usage Guidance and restrictions
• Indoor use only
• Altitude up to 2000m
• Temperature 5°C to 40°C
• Maximum relative humidity 80% for temperature up to 30°C decreasing
linearly to 50% relative humidity at 40°C
This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.
POLLUTION DEGREE 2, states that: “Normally only non-conductive pollution
occurs.
Occasionally, however, a temporary conductivity caused by condensation
must be expected”.
Adjusting Speed and Time
a. Connect the power supply and switch on the unit. The last run parameters
will be displayed.
b. To adjust the speed, press ( ) button once, the display for speed will start
flashing. Press ( ) or ( ) button,to reach the desired speed (2000-
12000rpm). Please note that the displayed value is in “x1000rpm” (e.g. 10
means 10,000rpm). Press ( ) button again to adjust time. The time display
will start flashing, press( )or( )button to reach the required run time
(1-99min).
Starting and Stopping a Run
a. Make sure the cover is closed fully. Press ( ) button to start the run. The unit
will stop automatically when the set run time is reached.
b. During a run, you can change the speed any time. To interrupt and stop the
run at any time, press ( ) button.
Exchanging Rotors
The centrifuge is supplied with two interchangeable quick release rotors. No
tools are required to exchange the rotors or attach them to the motor shaft.
The rotors are easily exchanged and can be snapped into position by hand.
Ensure the rotor is locked into position by lifting the rotor to check before use.