Signia MOTION SP 3PX User guide

Type
User guide
1 / 2
We declare the following hearing instruments:
Pure 312 1X, Pure 312 2X,
Pure 312 3X, Pure 312 5X, Pure 312 7X,
Pure 312 sDemo DX, CROS Pure 312 X
(DTI 20ATEX0139X)
Pure 312 1Nx, Pure 312 2Nx,
Pure 312 3Nx, Pure 312 5Nx, Pure 312 7Nx,
Pure 312 sDemo DNx, CROS Pure 312 Nx,
Pure 10 1Nx, Pure 10 2Nx
Pure 10 3Nx, Pure 10 5Nx, Pure 10 7Nx,
Pure 10 sDemo DNx
(DTI 17ATEX0085X)
Motion 13 BT 7PX,
Motion 13 1Nx, Motion 13 2Nx,
Motion 13 3Nx, Motion 13 5Nx, Motion 13 7Nx,
Motion 13 sDemo DNx,
Motion 13P 1Nx, Motion 13P 2Nx,
Motion 13P 3Nx, Motion 13P 5Nx, Motion 13P 7Nx,
Motion 13P sDemo DNx
(DTI 18ATEX0086X)
Pure 13 3Nx, Pure 13 5Nx, Pure 13 7Nx,
Pure 1px, Pure 2px, Motion SP 1px,
Motion SP 2px, Motion SP 3px, Motion SP 5px,
Motion SP 7px, Motion P 1px, Motion P 2 px,
Motion SA 1 px, Motion SA 2px, Contrast HP+,
Contrast R+, Contrast S+, Contrast SP+
(BVS 16 ATEX E 108)
are designed and manufactured in conformity with the
European Council Directive 2014/34/EU and has been
found to comply with the Essential Health and Safety
Requirements relating to the design and construction
of products intended for use in potentially explosive
atmospheres given in Annex II to the Directive.
Group / Class
II 2G Ex ia IIC T6 Gb
I M2 Ex ia I Mb
Applied Standards
EN 60079-0:2012 +
A11:2013
General Requirements 0123
0158
EN 60079-11: 2012
Intrinsic Safety “i”
Marking:
Hearing instruments conforming to Directive 2014/34/EU will bear only the
mark CE0158 referring to the registration number of DEKRA EXAM GmbH
which is the notified body for the QUALITY ASSURANCE OF THE
PRODUCTION PROCESS according to ANNEX IV of the Directive. ATEX
group / class is printed on the ATEX card, which has to be available close
at hand when the hearing instruments are used in explosive atmospheres.
EU-Type Examination Certificate:
- BVS 16 ATEX E 108 by DEKRA EXAM GmbH
- DTI 17ATEX0085X by Danish Technological Institute
- DTI 18ATEX0086X by Danish Technological Institute
- DTI 20ATEX0139X by Danish Technological Institute
User instructions:
When used in potentially explosive atmospheres, the following
additional instructions to the standard user manual apply:
For standard batteries: only use CE-marked zinc-air batteries,
IEC type: PR 41 (size: 312), IEC type: PR 44 (size: 675) or
IEC type: PR 48 (size: 13)
Do not use rechargeable batteries
Do not connect external audio-signal sources (via Audio-input)
Do not connect programming units
Do not use remote control (miniTek, easyPocket, easyTek,
miniPocket, Streamline Mic, VoiceLink, Smartphone)
Do not use the charger
Always carry the ATEX card with you and show it upon request.
Electrical Key Parameters
1. For all the above mentioned hearing instruments:
Power supply
Voltage UiDC 1.6 V
Current In3 mA
Data communication
Transmitter-/ receiver frequency 3.2768 MHz
Radiated power Po65 µW
Bluetooth communication
Transmitter- / receiver frequency range 2.402 – 2.480 GHz
Radiated power Po1 mW
2. Supply battery size: 312 (IEC type: PR 41)
Voltage UoDC 1.45 V
Current In138 mAs
Rated capacity 180 mAh
3. Supply battery size: 675 (IEC type: PR 44)
Voltage UoDC 1.45 V
Current In413 mAs
Rated capacity 650 mAh
4. Supply battery size: 13 (IEC type: PR 48)
Voltage UoDC 1.45 V
Current In145 mAs
Rated capacity 310 mAh
User Guide Supplement for ATEX
Hearing inruments in potentially explosive atmospheres
2 / 2
Document No. 03557-99T5-7600
Order/Item No. 109 591 70
© 09.2020, Signia GmbH
All rights reserved
Legal Manufacturer
Signia GmbH
Henri-Dunant-Strasse 100
91058 Erlangen, Germany
Phone +49 9131 308 0
www.signia-hearing.com
  • Page 1 1
  • Page 2 2

Signia MOTION SP 3PX User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI