1. Connect the power adaptor to a power outlet.
2. Press the power button to turn on the appliance.
3. Immerse the electrodes into the water and stir them gently. The
reading on the display will show up.
4. Press “Mode” Button consecutively to switch the between EC,
CF, PPM display.
5. After use, turn off the monitor and rinse the electrodes with
alcohol.
PH Calibration
1. Prepare 3 small beakers of ph 6.86 , 4.00 and 9.18 solution.
2. Turn on the monitor and immerse the ph electrode into the 6.86
solution. Stir it gently and wait until the reading on display
stabilizes.
3. If the reading is not 6.86 now, use the screwdriver to adjust the PH
calibration button until the reading becomes 6.86.
4. Take out the ph electrode from the solution and rinse it with
alcohol.
5. Put the ph electrode into the 4.00 or 9.18 solution to see if the
reading is correct. If not, readjust the reading to the correct value.
EC/CF/PPM Calibration
1. Prepare 2 small beakers of solution of 12.88 EC and 14.1EC.
2. Turn on the monitor and press the “MODE” button to choose the
“EC” mode.
3. Immerse the EC/CF electrode into the 12.88 EC solution. Stir it
gently and wait until the reading on display stabilizes.
4. If the reading is not 12.88 now, use the screwdriver to adjust the
EC/CF/PPM calibration button until the reading becomes 12.88.
5. Take out the EC/CF/PPM electrode from the solution and rinse it
with alcohol.
6. Put the EC/CF/PPM electrode into the 14.1EC solution to see if the
reading is correct. If not, readjust the reading to the correct value.