180 mm CONTINUOUS-TRACE RECORDER INSTALLATION AND OPERATION MANUAL
HA247652
Issue 8 Jun 98
Section 1
Page 1 - 3
SECTION 1 INSTALLATION
1.1 INSTALLATION PROCEDURE
The installation procedure is as follows:
1. Unpack the recorder (section 1.2)
2. If the recorder is to be panel mounted, fit the recorder into the panel (section 1.3)
3. Connect supply and signal wiring (sections 1.4 and 1.5)
4. Fit chart and printhead (sections 1.7 and 1.8)
5. Switch on. Configure the recorder.
1.2 UNPACKING THE RECORDER
The recorder is shipped in a special pack, designed to ensure adequate protection during transit. Should the outer
box show signs of abnormal wear or damage, it should be opened immediately and the recorder examined. If there is
evidence of damage, the instrument must not be operated and the local representative contacted for instructions. Af-
ter the recorder has been removed from its packing, the packing should be examined before it is discarded, to ensure
that all accessories and documentation hhave been removed.
1.3 MECHANICAL INSTALLATION
As shown in figure 1.3 below, the recorder is intended for installation in a cutout 281 mm square, in a panel which
is either vertical or sloping up to 30 degrees. Particular attention should be paid to the recommended minimum dis-
tances between adjacent recorders. These distances are advised in order to retain ease of access, to allow full door
opening and for panel strength considerations.
The instrument is inserted through the panel cutout, from the front. With the weight of the recorder supported, the
panel clamps are clipped into the holes located, one each side of the recorder. The jacking screws are now tightened
until the springs of the panel jacks are compressed to approximately half of their free length.Once the recorder is se-
curely mounted, the door should be opened (by lifting and then turning the latch clockwise) and the internal packing
removed.
1.4 SUPPLY VOLTAGE CONNECTION
All connections are made at the rear of the recorder. Supply voltage wiring, Supply voltage selection and fuse re-
placement can all be carried out without the removal of the terminal cover.
1.4.1 Supply voltage connection
As shown in figure 1.4.1 below, the supply voltage set-
ting switch is located to the right of the fuse drawer asso-
ciated with the IEC plug.
The switch is operated upwards (230V indicated) to se-
lect 180 to 264 V, or downwards (115V indicated) to
select 90 to 132V.
230
115
OR
Down for 90
to 130 V
Up for 180
to 264V