TRANSCUTANEOUS PACER ANALYZER
The unit can test external transcutaneous pacemakers. It has a wide variety of loads and
can measure the Pacer Pulse, Demand Sensitivity and Refractory Periods (Pacing and
Sensing):
LOAD:
RANGE: 50, 100, 150, 200, 300, 400, 500, 600, 700, 800, 900, 1000,
1100, 1200, 1300, 1400, 1500, 1600, 1700, 1800, 1900, 2000, 2100,
2200, 2300 Ω
PULSE:
PULSE CURRENT: 4 TO 300 mA (100 ohm load)
RATE: 30 TO 800 ppm
WIDTH: 0.6 to 80 ms
DEMAND SENSITIVITY:
WAVEFORMS:
SELECTIONS: SQUARE, TRIANGLE, HAVERSINE
WIDTH: 10, 25, 40, 100, 200 ms
ECG:
AMPLITUDE – OUT: 0 to 4 mV
PACER INPUT (50 TO 400 Ω):
AMPLITUDE – OUT: 0 to 10 mV / 50 Ω
RATE – IN: 30 to 100 ppm
PACER INPUT (500 TO 2300 Ω & OPEN):
AMPLITUDE – OUT: 0 to 100 mV
RATE – IN: 30 to 100 ppm
DEFIBRILLATOR PLATES:
AMPLITUDE – OUT: 0 to 10 mV
RATE – IN: 30 to 100 ppm
REFRACTORY PERIOD:
PACING: 20 to 500 ms
SENSING: 20 to 500 ms
50/60 HZ INTERFERENCE TEST SIGNAL:
ECG OUTPUT: 0, 0.4, 0.8, 1.2, 1.6, 2.0, 2.4, 2.8, 3.2, 3.6, 4.0 mV
PACER INPUT 50 OHM: 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 mV
PACER INPUT 100 OHM: 0, 2, 4, 6, 8, 10, 12, 14, 16, 18, 20 mV
PACER INPUT 150 OHM: 0, 3, 6, 9, 12, 15, 18, 21, 24, 27, 30 mV
PACER INPUT 200 OHM: 0, 4, 8, 12, 16, 20, 24, 28, 32, 36, 40 mV
PACER INPUT 300 OHM: 0, 6, 12, 18, 24, 30, 36, 42, 48, 54, 60 mV
PACER INPUT 400 OHM: 0, 8, 16, 24, 32, 40, 48, 56, 64, 72, 80 mV
PACER INPUT ≥ 500 OHM: 0, 10, 20, 30, 40, 50, 60, 70, 80, 90, 100
mV
DEFIBRILLATOR PLATES: 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 mV
INPUT CIRCUITRY PROTECTION
INPUT CIRCUITRY IS PROTECTED AGAINST DAMAGE IN THE
EVENT OF AN ACCIDENTAL DEFIBRILLATOR DISCHARGE INTO
THE PACEMAKER TEST INPUT TERMINALS