  • Hello! I am an AI chatbot trained to assist you with the Taylor Technologies K-1645 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
COMPONENTS:
1 x 5067B Instruction
1 x 9198B Sample Tube, Graduated (25 mL) w/ cap & blue dot, plastic
1 x R-0619B-C Hardness Buffer, 2 oz w/ blue cap, DB
1 x R-0620B-I Hardness Indicator Powder, 10 g w/ blue dot
1 x R-0683-C Hardness Reagent, 2 oz, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
Total Hardness Test
1.Rinseandll25mLsampletube(#9198B)to25mLmarkwithwatertobetested
(Fig.1).
2.Add5dropsR-0619BHardnessBuer.Swirltomix.
3.Add1dipperR-0620BHardnessIndicatorPowder.Swirluntildissolved.Sample
willturnred(Fig.2)ifhardnessispresent.
4.AddR-0683HardnessReagentdropwise,swirlingandcountingaftereachdrop,
untilcolorchangesfromredtoblue(Fig.3).
DROP TEST
TOTAL HARDNESS (1 drop = 10 ppm)
5.MultiplydropsofR-0683HardnessReagentby10.Recordaspartspermillion
(ppm)totalhardnessascalciumcarbonate.
Fig. 1
Fig. 2
Fig. 3
Instr. #5067B
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 7/16
COMPONENTS:
1 x 5067W Instruction
1 x 9198W Sample Tube, Graduated (25 mL) w/ cap & white dot, plastic
1 x R-0638W-C Phenolphthalein Indicator, 2 oz w/ white cap, DB
1 x R-0699-C Iodide Iodate Reagent, 2 oz, DB
1 x R-0725-I Acid Starch Indicator Powder, 10 g
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
Sodium Sulfite Test
NOTE: Sample must be cooled to less than 100ºF (38ºC) to prevent high test
results. Sample must be protected from air contact while cooling to prevent
low test results.
1.
Collect water to be tested in a clean, preferably large-mouthed, bottle to overowing.
Immediately cap and cool to room temperature.
2. Rinse and ll 25 mL sample tube (#9198W) to 25 mL mark with cooled (room
temperature) water to be tested (Fig. 1).
NOTE: For results in grains per gallon (gpg), ll to 14.6 mL mark.
DROP TEST
SODIUM SULFITE (1 drop = 10 ppm)
3. Add 1 drop R-0638W Phenolphthalein Indicator. Swirl to mix. Sample will turn
pink (Fig. 2).
4. Add R-0725 Acid Starch Indicator Powder a dipper at a time, swirling after each
dipper, until color changes from pink to colorless. Add 2 more dippers. Swirl until
dissolved.
5. Add R-0699 Iodide Iodate Reagent dropwise, swirling and counting after each
drop, until sample changes from colorless to a faint but permanent blue (Fig. 3).
6. Multiply drops of R-0699 Iodide Iodate Reagent by 10. Record as parts per million
(ppm) sodium sulte.
NOTE: For 14.6 mL sample, record drops as grains per gallon (gpg) sodium sulte.
NOTE: For results as sulte, multiply sulte concentration by 0.64.
Fig. 1
Fig. 2
Fig. 3
Instr. #5067W
31 Loveton Circle, Sparks, MD 21152 U.S.A.
800-TEST KIT (837-8548) • 410-472-4340 8/16
COMPONENTS:
1 x 5067pH Instruction
1 x 5425 Color Card Comparator/Instruction
1 x 9017 Test Cell, Calibrated (5 mL) w/ cap, plastic
1 x R-1003U-C-DB Long Range Indicator, 2 oz, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a vertical position.
pH (long range) Test
1.Rinseandll5mLtestcell(#9017)to5mLmarkwithwatertobetested.
2.Add5dropsR-1003ULongRangeIndicator.Capandmix.
3.PlacetestcellonwhiteareaofColorCardComparator(#5425).Matchcolorintestcellwitha
color standard. Record as pH units.
COLOR COMPARISON TEST
pH (LONG RANGE) (3.0-11.0)
Instr. #5067pH
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 8/16
COMPONENTS:
2 x 2269 Cap, Dispenser, 18 mm, white
1 x 4027 Funnel, 58 mm, plastic
1 x 5067OP Instruction
1 x 6003 Brush, Test Tube
1 x 6009 Filter Paper, 9.0 cm, #610, 100/box
1 x 9021 Test Tube, Mixing, Calibrated (5, 10, 14, 15, 17.5 mL), glass w/ stopper
1 x 9025 2-Standard Comparator, Orthophosphate, Stannous Chloride,
30 & 60 ppm
2 x R-0601-C Molybdate Reagent*, 2 oz
1 x R-0602P-I Stannous Chloride Powder, 10 g
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
Orthophosphate Test
1. Filter water to be tested to clarify.
2.Rinseandlltesttube(#9021)to5mLmarkwithlteredsample.
3.CapR-0601MolybdateReagentwithdispensercap(#2269).AddR-0601MolybdateReagent
to15mLmark.Stopperandmix.Removestopper.
4.Add2leveldippersR-0602PStannousChloridePowder.Stopperandmix.
5.Wipedryandplaceincomparator.WAIT1MINUTE.Readbefore5minutes.
COLOR COMPARISON TEST
ORTHOPHOSPHATE (30 & 60 ppm)
6.Matchcolorintesttubewithacolorstandard.Recordaspartspermillion(ppm)orthophosphate.
*WARNING: MolybdateReagent(R-0601)contains5-10% (w/w)sulfuricacid,acorrosiveacid.
Instr. #5067OP
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 9/16
COMPONENTS:
1 x 5067O Instruction
1 x 9198O Sample Tube, Graduated (25 mL) w/ cap & orange dot, plastic
1 x R-0630-C Chromate Indicator, 2 oz, DB
1 x R-0638O-A Phenolphthalein Indicator, .75 oz w/ orange cap, DB
1 x R-0687O-C Sulfuric Acid .12N, 2 oz w/ orange cap, DB
1 x R-0706-C Silver Nitrate Reagent, 2 oz, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
Chloride Test
NOTE: Whensultecontentofsamplewatertobetestedexceeds10ppm,the
sulteshouldbeoxidizedtopreventinterferenceintest.A25mLwater
sampleis rstadjusted tothe appropriatepH,then 1 mL(or 25 drops)
of R-0649 Hydrogen Peroxide Solution (sold separately) is added and
thoroughlymixed.Continuewiththerestoftheprocedure.
1.Rinseandll25mLsampletube(#9198O)to25mLmarkwithwatertobetested.
2.Add2dropsR-0638OPhenolphthaleinIndicator.Swirltomix.Ifsampleturns
pink(Fig.1),addR-0687OSulfuricAcid.12Ndropwise,swirlingaftereachdrop,
untilcolorchangesfrompinktocolorless.
DROP TEST
CHLORIDE (1 drop = 10 ppm)
3.Add5dropsR-0630ChromateIndicator.Swirltomix.Samplewillturnyellow
(Fig.2).
4.AddR-0706SilverNitrateReagentdropwise,swirlingandcountingaftereach
drop,untilcolorchangesfromyellowtomilkysalmon(brick)red(Fig.3).
NOTE: DonotaddenoughR-0706SilverNitrateReagenttogiveabrowncolor.
Firstchangefromyellowtoamilkysalmon(brick)redistheendpoint.
5.MultiplydropsofR-0706SilverNitrateReagentby10.Recordaspartspermillion
(ppm)chloride.
Fig. 1
Fig. 2
Fig. 3
Instr. #5067O
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 8/16
COMPONENTS:
1 x 5067G Instruction
1 x 9198G Sample Tube, Graduated (25 mL) w/ cap & green dot, plastic
1 x R-0637-C Methyl Orange Indicator, 2 oz, DB
1 x R-0638G-A Phenolphthalein Indicator, .75 oz w/ green cap, DB
1 x R-0645-C Total Alkalinity Indicator, 2 oz, DB
1 x R-0687G-C Sulfuric Acid .12N, 2 oz w/ green cap, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
P/M Alkalinity Test
1.Rinseandll25mLsampletube(#9198G)to25mLmarkwithwatertobetested.
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
2.Add3dropsR-0638GPhenolphthaleinIndicator.Swirltomix.Samplewillturn
pink(Fig.1)ifPalkalinityispresent—proceedtoStep3.Ifsampleiscolorless,
proceedtoStep4.
3.AddR-0687GSulfuricAcid.12Ndropwise,swirlingandcountingaftereachdrop,
untilcolorjustchangesfrompinktocolorless.RecorddropsasPreading.
DROP TEST
P/M & P/T ALKALINITY (1 drop = 10 ppm)
Fig. 1
Fig. 2
Fig. 3
Instr. #5067G
(OVER)
4.Add5dropsR-0637MethylOrangeIndicator.Swirltomix.Samplewillturnyellow
(Fig.2).
5.Add R-0687G Sulfuric Acid .12N dropwise, swirling and counting after
each drop, until color changes from yellow to orange (salmon pink)
(Fig.3).Recordtotaldrops(Steps3&5)asMreading.
6.MultiplyPreadingby10.Recordaspartspermillion(ppm)Palkalinityascalcium
carbonate.MultiplyMreading by10.Recordas ppm Malkalinityascalcium
carbonate.
NOTE: For 14.6 mL sample, record P reading as grains per gallon (gpg) P
alkalinityascalciumcarbonate.RecordTreadingasgpgTalkalinityas
calciumcarbonate.
P/T Alkalinity Test
1.Rinseandll25mLsampletube(#9198G)to25mLmarkwithwatertobetested.
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
2.Add3dropsR-0638GPhenolphthaleinIndicator.Swirltomix.Samplewillturn
pink(Fig.4)ifPalkalinityispresent—proceedtoStep3.Ifsampleiscolorless,
proceedtoStep4.
3.AddR-0687GSulfuricAcid.12Ndropwise,swirlingandcountingaftereachdrop,
untilcolorjustchangesfrompinktocolorless.RecorddropsasPreading.
4.Add5dropsR-0645TotalAlkalinityIndicator.Swirltomix.Samplewillturngreen
(Fig.5).
5.ContinueaddingR-0687GSulfuricAcid.12Ndropwise,swirlingandcounting
aftereachdrop,untilcolorchangesfromgreentored(Fig.6).Recordtotaldrops
(Steps3&5)asTreading.
6.Multiply P reading by 10. Record as parts per million (ppm) P alkalinity as
calcium carbonate. Multiply T reading by 10. Record as ppm T alkalinity as
calciumcarbonate.
NOTE: For 14.6 mL sample, record P reading as grains per gallon (gpg) P
alkalinityascalciumcarbonate.RecordTreadingasgpgTalkalinityas
calciumcarbonate.
DROP TEST
P/M & P/T ALKALINITY (1 drop = 10 ppm)
Fig. 4
Fig. 5
Fig. 6
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 8/16
Instr. #5067G
InstructionNumber:5425
Component Description
> COLORCARDTEST
> LONGRANGEpH(3‐11)
1. Rinseandfill5mLtestcell(#9017)to5mLmarkwith
watertobetested.
2. Add5dropsR‐1003ULongRangeIndicator.Capandmix.
3. LaytestcellflatalongwhiteareaofColorCard(#5425).
Matchcolorintestcellwithacolorstandard.Record
aspHunits.