QH7 User Manual
XIII
Using the impact device
The procedure explained herein is applicable to all impact device
types except for impact device type DC as it does not use a spring
loading mechanism.
Loading the impact device
Place the impact device over the surface you want to measure and
load by gently pushing the moving body of the impact device in the
same direction as the test piece surface. Keep pushing the moving
body until you reach the bottom limit and then bring the moving
body back to its initial position.
The impact device is now loaded and ready to be used.
Release and measure
Once the impact device has been loaded use one hand to firmly hold
the bottom of the impact device (the part that touches the surface
you want to measure) against the surface, and use your other hand
to press the release button located on top of the impact device.
After pressing the button, the impact tip will hit the test piece surface
and a hardness value will immediately be displayed.
Repeat this same procedure for each measurement.
Important
The Leeb method for hardness testing has certain test
piece requirements that must be met in order to obtain
reliable and accurate measurements.