Introduction
1
*
During normal testing, any blood glucose meter or lancing device may
come in contact with blood. All parts of the kit are considered biohazardous
and can potentially transmit infectious diseases from blood-borne
pathogens, even after you have performed cleaning and disinfection.
The meter and lancing device should never be used by more than one
person. Do not share the meter and lancing device with anyone, including
family members, due to the risk of infection from blood-borne pathogens.
Cleaning and disinfecting the meter and lancing device destroys most, but
not necessarily all, blood-borne pathogens.
If the meter is being operated by a second person who is providing testing
assistance to the user, the meter and lancing device should be cleaned
and disinfected prior to use by the second person.
Disinfect the meter and lancing device before allowing anyone else to
to handles them. Do not allow anyone else to test with the meter or lancing
device.
It is important to keep the meter and lancing device clean and disinfected.
Clean and disinfect the lancing device once per week to remove visible dirt or
other material for safe handling and/or prior to disinfecting.
For instructions on how to clean and disinfect the meter and lancing device,
see Chapter Maintenance.
Wash hands thoroughly before and after handling the meter, lancing device,
or test strips.
Choking hazard. Small parts included. Keep away from children.
Strong electromagnetic fields may interfere with the proper operation of the
meter. Do not use this meter close to sources of strong electromagnetic
radiation.
To avoid electrostatic discharge, do not use the meter in a very dry
environment, especially one in which synthetic materials are present.
Important Safety information
Warning
Peel off the insulation film from battery compartment before first use.
Venous whole blood and neonatal blood sample collection and preparation
should be obtained by healthcare professionals.
Thank you for purchasing the Kinetik Wellbeing BG-710 Blood Glucose Monitoring System.
Before you start testing, carefully read this Instruction Manual.
Intended Use
The Kinetik Wellbeing BS-602 Test Strips are used with the Kinetik
Wellbeing BG-710 Blood Glucose Meter to measure blood glucos
with fresh capillary whole blood for self-testing. The blood glucose
concentration of venous blood or neonatal blood is to be measured
by healthcare professionals. The fresh capillary whole blood should be
collected from your fingertip and neonatal blood collected from the
heel. The system is intended for in vitro diagnostic home-use or by
healthcare professionals in a clinical setting as an aid to moni or the
effectiveness of diabetes control. This system is not for use in the
diagnosis of diabetes mellitus.
Test Principle
The strip measures glucose by using amperometric technology
employing a glucose dehydrogenase reaction. When whole blood
or control solution is drawn into the tip of est strip, glucose in
the sample reacts with chemicals and produces an electrical
current.
The meter measures electrical current and calculates amount of
glucose. The glucose result is displayed as a calculated plasma
value.
INCLUDED IN YOUR KIT
c
d
e
Manufacturer of Lancing Device and Sterile Lancet
EC REP
Manufacturer of Lancing Device and Sterile LancetShandong Lianfa
Medical Plastic Products Co.. Ltd.No.1
Shuangshan Sanjian Road, 250200, Zhangqiu City, Jinan, Shandong,
PEOPLE'SREPUBLIC OF CHINA
ShanghaiInternational Holding Corp. GmbH (Europe)EC| REPEiffestrasse 80,
20537 Hamburg Germany
Beijing Ruicheng Medical Suplies Co., Ltd.
No. 558 Zhangzikou, Yangsong Town, Huairou District, 10400 Bejing, China
Lotus NL B.V.
EC REPKoningin Julinaplein 10, le Verd,2595AA, The Hague, Netherlands.
for sterile lancet
CE:for lancing device;
for sterile lancet
CE:for lancing device;
EC REP
WARNING: Keep the meter and testing supplies away from children.
Small items such as the battery cover, batteries, test strips,
lancets, protective covers of the lancets, and control solution vial
cap are choking hazards.
In-door glucose measurement is recommended.
f
Meter (2 x AAA Batteries included)
Test Strips
Lancing Device
Sterile Lancet
Control Solution
USB data cable
b
0123
0197
a
Getting to know your system
7
USB Port
Setting up your meter
OR
Set low blood glucose alarm threshold
Press and release the side UP button or side DOWN
button to select one value.Press until the correct
value appears.Press "M" button to enter high blood
glucose alarm function setting.
OR
OR
OR
Set highbloodglucose alarm function
Press andrelease the
button to select thehighbloodglucosealarmfunction,
if select "On",press "M" buttontoenterhighblood
glucose alarm concentration value setting.If select
"OFF",press "M"buttontogoonto"Alarm clock"
function setting.
side UP button or side DOWN
Set low blood glucose alarm function
Press and release the side UP button or side DOWN
button to select the low blood glucose alarm function.
If select "On",press "M" button to enter low blood
glucose value setting. If select ”OFF”,press "M"
button to go on to high blood glucose alarm function
setting.
Set highbloodglucose alarm threshold
Press andrelease the
button to select onevalue. Press untilthecorrect
value appears. Press "M" buttonto enter"Alarmclock"
function setting.
side UP button or side DOWN
Computer connection
13
The meter can set 6 alarms. These alarms can be used to
remind you to take a glucose measurement.
If not, the device will automatically shut down after
3 minutes without use.
Install GlucoJoy software. You can find this software on
kinetikwellbeing.com/blood-glucose-monitors/blood-glucose-monitoring-system
Open GlucoJoy and follow GlucoJoy User Guide to import data from
glucose meter.
Download the GlucoJoy User Guide
15
26
If your blood glucose is below 5.55mmol/L (100mg/dL), your results generally
should fall within 0.83mmol/L ( 15mg/dL) of the laboratory result.± ±
If your blood glucose is equal to or over 100mg/dL), your results5.55mmol/L (
generally should fall within ±15% of the laboratory result.
To maximize your chances of an accurate comparison between meter
and laboratory results, follow a few basic guidelines: please
Before going to the lab
Perform a control solution test to make sure the meter is working properly.
Do Not eat for at least eight hours before you test your blood glucose.
While at the lab
Perform your meter test within 15 minutes from the lab test.
Use fresh capillary blood obtained from the fingertip, venous blood,
neonatal blood.
glucose test by your meter.
Comparing your meter's results with laboratory results periodically.
The results of the Kinetik Wellbeing BG-710 meter are plasma equivalent.
This method
will help you and your healthcare professional compare your
meter results
with laboratory test results. The
Kinetik Wellbeing
BG-710 meter's
test result and
laboratory
test results both are expressed in plasma-equivalent
units. However, your
blood glucose meter result may differ from your laboratory
result due to normal
variation. Your blood glucose meter results can be affected
by factors and
conditions that do not affect laboratory results in the same way
Follow all instructions in this Instruction Manual to perform a blood
Comparing Your Meter result to a Laboratory result
When to perform a control test
Control Solution contains a specific amount of glucose and is used to
check that the meter and the test strips are working properly.
Do a control solution test:
Whenever you open a new vial of test strips.
Whenever you want to check if you are testing correctly.
If you suspect the meter and test strips are not working properly.
If you have had repeated unexpected blood glucose results.
If you drop or damage the meter.
If you left the test strip container open or you think the test strips
have been damaged.
If the test strips were stored in extreme temperatures and/or humidity
Your test result does not match how you feel.
NOTE:
Only use the
Kinetik Wellbeing
control solution.
Close the control solution bottle tightly after use.
Write the date you open the control solution bottle on the bottle label.
The control solution must be discarded 3 months after the
first open date or the expire date on the vial label, whichever comes
first.
Do not use control solution that exceeds the expired date or discard date.
Refer to the control solution label for control solution storage conditions.
The control solution can stain fabric.
Remove stains by washing with soap and water.
CAUTION: Do Not swallow control solution. It is not for human
consumption. Do Not apply control solution to the skin or eyes
as it may cause irritation.
Control solution testing
The meter will average your results every 7 days. You
will be shown your latest 7 day, 14 day and 28 day
result average.
n = indicates if you are viewing your 7, 14 or 28 day
average.
Press and hold
29
Control solution testing
Understanding out-of-range control test results
Compare the result displayed on the meter
to the control solution range printed on the
test strip vial. Each control solution have a
different control solution range.
Out-of-range results may be due to:
not following the instructions detailed in steps 1–6
expired or contaminated control solution
expired or damaged test strip
a problem with the meter.
CAUTION: The control solution range printed on the test strip
vial is for
Kinetik Wellbeing
Control Solution only. It is not a
recommended
range for your blood glucose level.
CAUTION: If you continue to get control solution test results
that fall outside the range printed on the test strip vial, Do Not use
the meter, the test strips, or the control solution. Contact the
vendor.
use of a test strip or control solution that passed its discard date
CTRL 1
mmol/L
mg/dL
CTRL 2
mmol/L
mg/dL
CTRL 3
mmol/L
mg/dL
2.0-6.0
36-108
8.0-12.0
144-216
14.0-22.0
252-396
32
Remove the old batteries and insert new,
non-rechargeable batteries.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26

Kinetik user manual User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI