SKC 225-9007 User manual

Type
User manual

SKC 225-9007 preloaded coated filters simplify air sampling for mercaptans and phenyl-mercaptans. These filters come ready to use, eliminating the need for assembly. They feature a glass fiber filter coated with mercuric acetate and are designed for use with a two-piece 37-mm cassette in a closed-face sampling configuration. With a storage temperature of ≤ 39.2°F (4°C) for up to 8 months and sample stability of up to 15 days when stored at the same temperature, these filters offer reliable and convenient air sampling solutions.

SKC 225-9007 preloaded coated filters simplify air sampling for mercaptans and phenyl-mercaptans. These filters come ready to use, eliminating the need for assembly. They feature a glass fiber filter coated with mercuric acetate and are designed for use with a two-piece 37-mm cassette in a closed-face sampling configuration. With a storage temperature of ≤ 39.2°F (4°C) for up to 8 months and sample stability of up to 15 days when stored at the same temperature, these filters offer reliable and convenient air sampling solutions.

Form 40036 Rev 211118
Preloaded Coated Filters
Cat. No. 225-9007
Operating Instructions
SKC coated filters are shipped preloaded in the cassettes with
end plugs and shrink bands in place. No assembly is required;
they are ready to use.
Method & Chemical: OSHA 26 for methyl-mercaptans; NIOSH
2542 for mercaptans (methyl-, ethyl-, n-butyl-); OSHA PV2075 for
phenyl-mercaptans
Filter & Coating: Glass fiber filter coated with mercuric acetate.
Two-piece 37-mm cassette. Closed-face sampling configuration
Prior to Sampling: Store at ≤ 39.2 F (4 C) for up to 8 months.
Sample Stability: Store samples at 39.2 F (4 C) for up to
15 days.
For information on other available coated filters, go to
www.skcinc.com.
Form 40036 Rev 211118
How to Use SKC Preloaded Coated Filters
Ensure that filter cassettes are at room temperature
before sampling.
1. Select one coated filter cassette for calibrating the flow rate. A red plug
secures the inlet, which is clearly marked “inlet,” and a blue plug
secures the outlet. Set up the sampling train for calibration with the
representative filter cassette in line. For “closed-face” sampling, remove
the plugs and connect the cassette to the sampling train. For “open-
face” sampling, remove the outlet plug and the cassette inlet section and
then connect the cassette to the sampling train. For details on setting up
a sampling train, refer to the SKC Sample Setup Guide “Sampling Train
Filters” at www.skcinc.com/knowledgecenter.
2. Calibrate the pump to the desired flow rate using a calibrator.
3. Replace the cassette used for calibration with a fresh coated filter
cassette for sample collection. Attach the cassette to a worker’s collar,
as close to the breathing zone as possible, by using a filter cassette
holder SKC Cat. No. 225-1. Sample for the specified time interval and
record the time.
4. Remove the filter cassette at the end of the specified sampling period
and replace both end plugs (and the inlet if necessary). Recheck the
flow rate using the same cassette and calibrator used for calibration in
Step 2 to ensure that the flow rate has not changed by more than 5%.
5. Store samples at 39.2 F (4 C) for up to 15 days. Appropriately
package samples, blanks, and all pertinent data to send to a laboratory
for analysis. Expedite shipping of samples to laboratory. Friday
shipments are not recommended.
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy,
which provides SKC’s sole liability and the buyer’s exclusive remedy. To
view the complete SKC Limited Warranty and Return Policy, go to
skcinc.com/warranty.
  • Page 1 1
  • Page 2 2

SKC 225-9007 User manual

Type
User manual

SKC 225-9007 preloaded coated filters simplify air sampling for mercaptans and phenyl-mercaptans. These filters come ready to use, eliminating the need for assembly. They feature a glass fiber filter coated with mercuric acetate and are designed for use with a two-piece 37-mm cassette in a closed-face sampling configuration. With a storage temperature of ≤ 39.2°F (4°C) for up to 8 months and sample stability of up to 15 days when stored at the same temperature, these filters offer reliable and convenient air sampling solutions.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI