Digester Gas Monitor Maintenance Manual Replacing the H
2
S Dilution Air Flowmeter • 11
15. Reinstall the flowmeter/bracket assembly by using the screws, lock washers, and flat washers
removed in step 6 to attach the bracket to the plate.
Figure 15: Attaching the H
2
S Dilution Sample Flowmeter Bracket to Plate
16. Reconnect the tubing to the fittings.
17. Use pliers to reinstall the clamps over the tubing fitting connections.
18. Plug in or turn on all incoming power to the Digester Gas Monitor.
19. Place the Digester Gas Monitor’s power switch in the ON position, then verify that the PILOT
light is on.
20. Turn the flow regulator knob clockwise to set the sensor flow to 3 SCFH once the system has
started up.
21. Close and secure the housing door.
Replacing the H
2
S Dilution Air Flowmeter
Replace the H
2
S Dilution Air Flowmeter if it becomes dirty or wet and cannot be properly flushed out
or cleaned or if the flowmeter ball sticks so that it is difficult to accurately determine the flowrate.
1. Turn off or unplug all incoming power to the Digester Gas Monitor.
2. Open the Digester Gas Monitor housing door, then place the power switch in the OFF position.
3. Turn the flow regulator knob all the way counterclockwise to close the flow regulator’s output.
Flat Washer
Screw
Lock Washer