  • Hello! I am an AI chatbot trained to assist you with the Taylor Technologies K-9036 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
InstructionNumber:5147
Component Description
> COMBINATIONMIDGET
> CHLORINE/pHTEST
COMPONENTS:
Chlorine
1xR‐0001DPDReagent#1,DB
1xR‐0002DPDReagent#2,DB
1xR‐0003DPDReagent#3,DB
pH
1x4028 Pipet,Calibrated0.5mL,plasticw/cap
1xR‐1003JpHIndicatorSolution(PhenolRed)
APPARATUS:
1x3243 Cap,TestCell,11.5mL,rectangular,plastic
1x4024 TestCell,Calibrated11.5mL,plastic
1x5147 Instruction
1x6002 Brush,CellCleaner
1x9044 MidgetCombo.,ChlorineDPD,0.2‐2.0ppm/pH6.8‐8.0
TOORDERREPLACEMENTPARTSANDREAGENTSCALLTOLL‐FREE
800‐TESTKIT(800‐837‐8548).
PROCEDURE:
CAREFULLYREADANDFOLLOWPRECAUTIONSONREAGENTLABELS.
KEEPREAGENTSAWAYFROMCHILDREN.
FreeChlorine(FC)Test
1. Rinseandfill11.5mLtestcell(#4024)tomarkwithwater
tobetested.
2. Add5dropsR‐0001DPDReagent#1and5dropsR‐0002DPD
Reagent#2.Capandmix.
3. WipedryandplaceincomparatorWITHFROSTEDSIDEFACING
OPERATOR.
4. Matchcolorintestcellwithacolorstandard.Recordas
partspermillion(ppm)freechlorine(FC).
5. Savesamplefortotalchlorine(TC)test.
TotalChlorine(TC)Test
1. UsetreatedsamplefromFCtest.
2. Add5dropsR‐0003DPDReagent#3.Capandmix.
3. WipedryandplaceincomparatorWITHFROSTEDSIDEFACING
OPERATOR.
4. Matchcolor.Recordasppmtotalchlorine(TC).
CombinedChlorine(CC)Test
1. SubtractFCfromTC.Recordasppmcombinedchlorine(CC).
Formula:TC‐FC=CC.
pH
1. Rinseandfill11.5mLtestcell(#4024)to11.5mLmarkwith
watertobetested.
2. Usinga0.5mLpipet(#4028),add0.5mLpHIndicator.Cap
andmix.
3. WipedryandplaceincomparatorWITHFROSTEDSIDEFACING
OPERATOR.
4. Matchcolorintestcellwithacolorstandard.Recordas
pHunits.
NOTE:Ifsamplecolorisbetweentwovalues,pHisaverageof
thetwo.
NOTE:IfdeterminingAcidorBaseDemand,savesampleandproceed
asdirectedonseparateinstruction.
6/00