Getting Started
Specifications
For more specifications, refer to the PWS2185, PWS2323, PWS2326, and PWS2721 Technical Reference.
Table 2: Electrical ratings for the power connection
Model
Line Selector
Switch Setting Frequency Fuse Rating Max Power
PWS2185 110 V / 220 V 50 / 60 Hz
For 110 V: 5 A TH, 250 V
For 220 V: 2.5 A TH, 250 V
350 VA
PWS2323 110 V / 220 V 50 / 60 Hz
For 110 V: 5 A TH, 250 V
For 220 V: 2.5 A TH, 250 V
350 VA
PWS2326 110 V / 220 V 50 / 60 Hz
For 110 V: 6.3 A TH, 250 V
For 220 V: 3.15 A TH, 250 V
675 VA
PWS2721 110 V / 220 V 50 / 60 Hz
For 110 V: 5 A TH, 250 V
For 220 V: 2.5 A TH, 250 V
350 VA
Table 3: Environmental performance
Parameter PWS2185 PWS2323 PWS2326 PWS2721
Temperature
Operating: +0 °C to +40 °C
Humidity
Operating:
5% to 95% relative humidity (% RH) at up to +40 °C, non-condensing
Altitude
Operating: 100% capability up to 2,000 meters
Pollution Degree 2, For indoor use only
Nominal Vo ltage Ratings
There are two ranges for all models, selectab le by the line-voltage selector switch. Check the voltage select switch on the
bottom before turning on: 110 V setting, 110 / 115 / 120 V
AC; 220 V setting, 220 / 230 / 240 VAC. Fluctuations must
not exceed 10% of the nominal rated voltage.
Float Voltage Rating
Float voltage rating: up to 100 V (DC + peak AC) between earth ground and any output terminal.
Operating Requirements
1. Place the instrument on a bench or similar surface.
2. Before operating, ensure that the ambient temperature is between +0 °C and +40 °C (+32 °F to +104 °F).
WARNING. To ensure proper cooling, keep the fron
t, sides, and rear of the instrument clear of obstructions. Do not
operate the instrument on a table without the handle in the down and locked position. This allows air flow to the bottom of
the instrument.
2 PWS2185, PWS2323, PW S2326, and PWS2721 User Manual