Record and storage of measured data
01) To save real-time data in the meter, press the “FAST∕SLOW” button for 2 seconds, until the “RECORD”
symbol appears on the display. At the same time, the meter will start to record real-time data until its
memory is full. Then the display will show a “FULL” symbol, indicating that the meter cannot continue
saving the data.
02) To Delete the stored data in the Sound Level Meter’s memory, press the “LEVEL ▲” button for 2
seconds until the display shows the “CEL” symbol. Then press the “LEVEL▼” button for 2 seconds, until
the “CEL” symbol flashes. After a few seconds, the display will return to normal, indicating that the
data has been deleted.
Installing operation software in a computer
01) Insert the accessory CD into the CD-ROM driver of a computer.
02) Click “setup.EXE” to install the software onto your hard disk.
03) When the installation has finished, click “MS6701 DIGITAL SOUND LEVELMETER” in the program menu.
04) An operation window will open on your computer.
Linking and communication between the meter and a computer
01) Connect the infrared output socket of the Sound Level Meter with the RS232 interface of your
computer with the included accessory cable.
02) Run “MS6701 DIGITAL SOUND LEVELMETER”, click “START” to receive and display data.
03) By pressing the “RS232” button on the meter for 2 seconds, the real-time sound level data is
transmitted to your computer. The “SENDING” symbol appears on the display. To stop the data
transmission, press the “RS232” button again for 2 seconds.
04) By pressing the “A / C” button for 2 seconds, the data in the memory of the sound meter is
transmitted to your computer. The “SENDING MEMO” symbol will appear on the display. During
transmission the computer shows “Down loading please wait…”. When the transmission is complete,
a dialog box will open. You need to enter a filename, then you can save the data onto your
computer.
05) To open a saved file on the computer, run the software of the sound level meter, click “FILE”. Then
click “OPEN” on the menu. Select and open the corresponding filename in the dialog box.
Set time and date
01) First press the “MAX” button and then turn on the Sound meter. The display will show time(hour,
minute, second) and the second digitals are flashing. Press “LEVEL▲” or “LEVEL▼” to increase or
decrease the time; Press “MAX” to set minute, hour, mouth, year. After the setup has been
completed, turn off the Sound meter.
Calibrating Sound level Meter
01) When the Sound Level Meter is used for a long time, it’s accuracy may reduce. You need to check
and calibrate the sound meter, once a year. Calibration needs a standard sound source, please
contact us about the proper method.
Caution
01) Do not use the meter in high temperatures or in wet places.
02) When not using the Sound meter for a long time, please take out the batteries to avoid damaging
the meter.
03) Auto-range (30-130dB) is unfit for measuring instantaneous and high-impact noises.
04) To measure a sound level in a windy environment, please put a windscreen on the microphone to
avoid wind noises.
05) If mark “ ”is shown on the display, it means the voltage is low. You must replace battery, we advise
you to use alkalescent batteries.
Replacing batteries
01) Place 6 x 1,5V AAA alkaline batteries in the battery housing on the back.
02) You can monitor the battery state by looking at the battery icon on the display.