Keysight Technologies 16065C Operation And Service Manual

Type
Operation And Service Manual

Keysight Technologies 16065C External Bias Adapter is an accessory designed to operate specifically with the E4980AL LCR Meter and E4981A Capacitance Meter. When used in conjunction with an external dc voltage source, the 16065C can supply a bias voltage across components of up to ±40 V dc through a four-terminal pair type test fixture. This allows users to apply a bias voltage to components under test, enabling them to characterize the component's behavior under different bias conditions.

Keysight Technologies 16065C External Bias Adapter is an accessory designed to operate specifically with the E4980AL LCR Meter and E4981A Capacitance Meter. When used in conjunction with an external dc voltage source, the 16065C can supply a bias voltage across components of up to ±40 V dc through a four-terminal pair type test fixture. This allows users to apply a bias voltage to components under test, enabling them to characterize the component's behavior under different bias conditions.

Operation and
Service Manual
Keysight 16065C External Bias Adapter
-
Notices
© Keysight Technologies
1991 - 2019
No part of this manual may be repro-
duced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement
and written consent from Keysight
Technologies, Inc. as governed by
United States and international
copyright laws.
Manual Part Number
16065-90020
Edition
Edition 3, August 23, 2019
Printed in Malaysia
Published by:
Keysight Technologies International
Japan G.K,
1-3-3 Higashikawasaki-cho
Chuo-ku
Kobe-shi, Hyogo, Japan
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT LIM-
ITED TO THE IMPLIED WARRANTIES
OF MERCHANTABILITY AND FIT-
NESS FOR A PARTICULAR PUR-
POSE. KEYSIGHT SHALL NOT BE
LIABLE FOR ERRORS OR FOR INCI-
DENTAL OR CONSEQUENTIAL DAM-
AGES IN CONNECTION WITH THE
FURNISHING, USE, OR PERFOR-
MANCE OF THIS DOCUMENT OR
ANY INFORMATION CONTAINED
HEREIN. SHOULD KEYSIGHT AND
THE USER HAVE A SEPARATE WRIT-
TEN AGREEMENT WITH WARRANTY
TERMS COVERING THE MATERIAL
IN THIS DOCUMENT THAT CON-
FLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPA-
RATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are fur-
nished under a license and may be
used or copied only in accordance
with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight prod-
ucts may be downloaded from the
Web. Go to http://www.key-
sight.com/go/conformity. You can
then search by product number to
find the latest Declaration of Confor-
mity.
U.S. Government Rights
The Software is “commercial com-
puter software,” as defined by Fed-
eral Acquisition Regulation (“FAR”)
2.101. Pursuant to FAR 12.212 and
27.405-3 and Department of Defense
FAR Supplement (“DFARS”)
227.7202, the U.S. government
acquires commercial computer soft-
ware under the same terms by which
the software is customarily provided
to the public. Accordingly, Keysight
provides the Software to U.S. gov-
ernment customers under its stan-
dard commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of which
can be found at http://www.key-
sight.com/find/sweula
. The license
set forth in the EULA represents the
exclusive authority by which the U.S.
government may use, modify, dis-
tribute, or disclose the Software. The
EULA and the license set forth
therein, does not require or permit,
among other things, that Keysight:
(1) Furnish technical information
related to commercial computer soft-
ware or commercial computer soft-
ware documentation that is not
customarily provided to the public; or
(2) Relinquish to, or otherwise pro-
vide, the government rights in
excess of these rights customarily
provided to the public to use, modify,
reproduce, release, perform, display,
or disclose commercial computer
software or commercial computer
software documentation. No addi-
tional government requirements
beyond those set forth in the EULA
shall apply, except to the extent that
those terms, rights, or licenses are
explicitly required from all providers
of commercial computer software
pursuant to the FAR and the DFARS
and are set forth specifically in writ-
ing elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise modify
the Software. With respect to any
technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the
U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data.
Safety Notices
A CAUTION notice denotes a haz-
ard. It calls attention to an oper-
ating procedure, practice, or the
like that, if not correctly per-
formed or adhered to, could
result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated condi-
tions are fully understood and
met.
A WARNING notice denotes a haz-
ard. It calls attention to an oper-
ating procedure, practice, or the
like that, if not correctly per-
formed or adhered to, could
result in personal injury or death.
Do not proceed beyond a WARN-
ING notice until the indicated
conditions are fully understood
and met.
Contents
Keysight 16065C External Bias Adapter 3
1. General Information
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Using the 16065C. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Accessories Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operating and Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2. Operating and Safety Precautions
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Operating and Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Connecting the Adapter for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Repackaging the Adapter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3. Operation
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
OPEN/SHORT Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
DC Bias . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
You Can NOT Use the Internal DC BIAS Source of Your Instrument. . . . . . . . . . . . . . . . . 15
Requirement for External Voltage Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Current Limiting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Capacitive Component Requires Time to Charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Additional Error for D Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4. Service
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Replaceable Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4 Keysight 16065C External Bias Adapter
Contents
Keysight 16065C External Bias Adapter
Operation and Service Manual
1 General Information
Introduction
The purpose of this manual is to enable you to use your 16065C External Bias
Adapter efficiently and confidently. This manual contains both general and
specific information. To use the 16065C to perform a specific function (without
having to read the entire manual), follow the directions in "Using the 16065C".
Using the 16065C
To install the 16065C, turn to Chapter 2.
To operate the 16065C, turn to Chapter 3.
To order replaceable parts for the 16065C, turn to "Replaceable Parts" in
Chapter 4.
Product Description
The 16065C has been designed to operate specifically with the E4980AL LCR
Meter and E4981A Capacitance Meter. By connecting an external dc voltage
source to the 16065C, the 16065C can supply a bias voltage across
components of up to ±40 V dc through a four-terminal pair type test fixture.
Accessories Supplied
The following accessories are supplied with the 16065C:
Table 1-1 Furnished Accessories
Description Part Number Quantity
Operation and Service Manual (this manual) 16065-90020 1
6 Keysight 16065C External Bias Adapter
General Information
Operating and Safety Precautions
1-
Operating and Safety Precautions
Operating
You need observe only normal precautions when handling and operating the
16065C. Do not exceed the operating input power, voltage, and current level
and signal type appropriate for the measurement instrument being used, refer
to your instrument's operation manual.
Service
The voltage levels in this adapter do not warrant more than normal caution for
operator safety. Nevertheless, service should be performed only by qualified
personnel.
Electrostatic discharge (esd) can damage the highly sensitive microcircuits
in your instrument. ESD damage is most likely to occur as the 16065C is
being connected or disconnected. Protect the 16065C from ESD damage
by wearing a grounding strap that provides a high resistance path to
ground. Alternatively, ground yourself to discharge any static charge
build-up by touching the outer shell of any grounded instrument chassis
before touching the test port connectors.
Never touch a test port connector's center contacts.
Use a work station equipped with an anti-static work surface.
Keysight 16065C External Bias Adapter 7
General Information
Specifications
Specifications
This section lists the complete 16065C specifications. These specifications are
the performance standards and limits against which the 16065C is tested.
When shipped from the factory, the 16065C meets the following specifications:
External DC Bias
Up to ±40 V can be applied to the DC BIAS INPUT BNC
connector.
Frequency Range
100 Hz to 1 MHz
Applicable Instrument
E4981A, E4980AL
Applicable Fixture
16034E/G/H, 16047A/E, 16048A/D, 16089A/B/C
Input Resistance
2 k±10%
Weight
450 g
Operating Temperature
0 to 55 °C
Operating Humidity
~95% RH (@40 °C)
Non-operating Temperature
–40 to 70 °C
Non-operating Humidity
~95% RH (@65 °C)
Dimensions
160 mm (w) × 50 mm (h) × 150 mm (d)
Cable Length
210 mm
8 Keysight 16065C External Bias Adapter
General Information
Specifications
1-
Keysight 16065C External Bias Adapter
Operation and Service Manual
2 Operating and Safety Precautions
Introduction
This chapter explains how to install the 16065C External Bias Adapter. The
topics covered include initial inspection, ambient environmental
considerations, connecting the adapter for use, and repackaging the adapter.
Initial Inspection
The adapter has been carefully inspected electrically and mechanically before
being shipped from the factory. It should be in perfect physical condition, no
scratches, dents or the like, and it should be in perfect electrical condition.
Verify this by carefully performing an incoming inspection to check the adapter
for signs of physical damage and missing contents. If any discrepancy is found,
notify the carrier and Keysight Technologies. Your Keysight Technologies sales
office will arrange for repair and replacement without waiting for the claim to
be settled.
Step 1. Inspect the shipping container for damage, and keep the shipping materials
until the inspection is completed.
Step 2. Verify that the shipping container contains everything shown in Figure 2-1
and listed in Table 2-1.
Step 3. Inspect the exterior of the 16065C for any signs of damage.
10 Keysight 16065C External Bias Adapter
Operating and Safety Precautions
Environmental Considerations
2-
Figure 2-1 Product Overview
Environmental Considerations
Operating and Storage
The 16065C should be operated within an ambient temperature range of 0 °C
to 55 °C and relative humidity up to 95% at 40 °C (non-condensing).
The 16065C may be stored within a temperature range of –40 °C to +70 °C,
and at a relative humidity of up to 95% at +65 °C (non-condensing).
Table 2-1 Contents
Description Keysight Part Number Quantity
External BIAS Adapter
16065-60370
1
(16065C)
1. Keysight internal-only part number. Purchase the whole unit of the 16065C External
Bias Adapter for replacement.
1
Operation and Service Manual
2
2. Operation and Service Manual is not shown in Figure 2-1.
16065-90020 1
Keysight 16065C External Bias Adapter 11
Operating and Safety Precautions
Connecting the Adapter for Use
Connecting the Adapter for Use
Figure 2-2 Connecting the Adapter
Repackaging the Adapter
If shipping to a Keysight Technologies service center is required, each adapter
should be repackaged using the original factory packaging materials.
Alteratively, comparable packaging materials may be used. Wrap the adapter
in heavy paper and pack in anti-static plastic packing material. Use sucient
shock absorbing material on all sides of the 16065C to provide a thick, firm
cushion and to prevent movement. Seal the shipping container securely and
mark it FRAGILE.
Make sure that the voltage source is turned off before you start making
connections.
12 Keysight 16065C External Bias Adapter
Operating and Safety Precautions
Repackaging the Adapter
2-
Keysight 16065C External Bias Adapter
Operation and Service Manual
3 Operation
Introduction
This chapter describes the features of the 16065C (Figure 3-1), the connection
to an external voltage source and test fixture, and compensation techniques.
Figure 3-1 Adapter Features
1. Four-terminal-pair test leads. These are connected to the UNKNOWN
terminals of an LCR meter or impedance analyzer.
2. Four-terminal-pair BNC terminals. These terminals are connected to a
four-terminal-pair type fixture.
3. DC Bias Input BNC. This terminal is connected to an external voltage
source in order to apply dc bias voltage to the DUT.
14 Keysight 16065C External Bias Adapter
Operation
OPEN/SHORT Correction
3-
OPEN/SHORT Correction
The 16065C has inherent stray capacitance, residual inductance, and residual
resistance which, if not properly compensated for before a measurement is
made, will degrade measurement accuracy. To cancel these residuals and thus
minimize measurement error, the measuring instrument's OPEN/SORT
correction capability must be used. The procedure is described in the
instrument's operation manual.
Operation
The step-by-step instructions on how to make a measurement with the
16065C are:
Step 1. Connect the test leads of 16065C to the measuring Instrument's UNKNOWN
terminals.
Step 2. Connect the test fixture directly to the 16065C.
Step 3. Connect the dc voltage source to the 16065C's DC BIAS INPUT BNC.
Step 4. Perform OPEN and SHORT correction as described in the measuring
instrument manual.
Step 5. Insert the component to be tested into the test fixture.
Do NOT turn on the voltage source while correction is in progress.
Do NOT short the high and low terminals when the voltage source is turn
on.
When a positive bias vol tage is used, the positive terminal of an electrolytic
capacitor must be connected to the instruments high terminal. When using
a negative bias voltage, connect an electrolytic capacitor's negative
terminal to the instrument's high terminal.
When measuring high value capacitors, allow suffcient time for the
capacitor to charge to the applied voltage.
Keysight 16065C External Bias Adapter 15
Operation
DC Bias
DC Bias
You Can NOT Use the Internal DC BIAS Source of Your Instrument
The 16065C contains a 100 µF capacitor, series connected between the H
cur
terminal and the component under test. Its function is to block the applied dc
voltage from flowing back into the measuring instrument. Also, because of its
location, this capacitor makes it impossible to bias samples from the
measuring instrument's internal bias source. Thus, the 16065C can not be used
for applications in which the instrument's internal bias source is used.
Requirement for External Voltage Source
The external dc voltage source used for biasing samples connected to the
16065C must be
capable of delivering 20 mA at 40 V.
Current Limiting
The 16065C's DC Bias Input has a 2 k current limiting resistor in series with
the component under test.
Capacitive Component Requires Time to Charge
The 16065C DC BIAS INPUT has a 2 k current limiting resistor in series with
the component under test. The time required for a capacitive component to
charge through this resistance is calculated as:
T(s) = 1.6 + (0.01 × C) [sec]
Where C is the capacitance of the sample, in µF.
16 Keysight 16065C External Bias Adapter
Operation
DC Bias
3-
Additional Error for D Measurement
Because of the series-connected capacitor between the H
pot
terminal and the
component under test, there is a test frequency dependent phase shift
between the H
pot
terminal and the L
cur
terminal. The input impedance of the
instrument causes additional error in some measurement modes. The
additional error for phase measurements (Ep) is derived by the following
equation:
E
p
_ 70 / f (%)
(Typical)
where f is measurement frequency and unit is Hz.
The additional error for D measurements (E
D
) is derived by the following
equation:
E
D
_ 0.7 / f
(Typical)
The compensation function of the E4981A and E4980AL can correct the
additional error caused by the 16065C. For more information on
compensation and correction, refer your instrument's manual.
Keysight 16065C External Bias Adapter
Operation and Service Manual
4 Service
Introduction
This chapter gives service information for the 16065C. The Replaceable Parts
List and the Schematic Diagram are included.
Serial numbers for Non-RoHS 16065C: “MY431xxxxx and below” /
“SG431xxxxx and below
Serial numbers for RoHS 16065C: “MY432xxxxx and above” / “SG432xxxxx and
above”
Replaceable Parts
Figure 4-1 and Table 4-1 identify the replaceable mechanical parts and their
respective RoHS compliant replacement support part. RoHS conversion
involves design and dimension change, which results in the RoHS support part
backward incompatible with the non-RoHS 16065C. Special handling is
needed while using the RoHS replacement part on the non-RoHS 16065C. The
original support part number is replaced by the respective “RoHS Compliant
Keysight Part Number”. Once the original support part is depleted, please
proceed to obtain the RoHS compliant support part. The parts listed can be
ordered from your nearest Keysight Technologies office. Ordering information
should include the Keysight part number and the quantity required.
18 Keysight 16065C External Bias Adapter
Service
Replaceable Parts
4-
Figure 4-1 Replaceable Parts Identiers
Table 4-1 Replaceable Mechanical Parts
Reference
Designator
Keysight
Part Number
RoHS
Compliant
Keysight
Part Number
Qty Descrption
1 1250-0083 1250-0083 1 Connector BNC
(for "Ext DC Bias")
2190-0016 2190-0016 1
Washer
1
2950-0043 2950-0043 1
Nut
1
2 16065-04002 16065-04602 1 Chassis
3 16065-04001 16065-04601 1 Chassis Top Cover
Keysight 16065C External Bias Adapter 19
Service
Replaceable Parts
4 16065-61602 16065-61672 1 Cable Assembly "LC"
5 16065-61603 16065-61673 1 Cable Assembly "LP"
6 16065-61605 16065-61675 1 Cable Assembly "HP"
7 16065-61604 16065-61674 1 Cable Assembly "HC"
8 0515-0914 0515-1946 4 Screw Flat Head M3 L6
9 94801-40042 0400-0010 4 Grommet
10 1400-0249 1400-0249 4 Cable Tie
11 0515-1550 0515-0372 4 Screw Pan Head M3×0.5 L8 w/
Washer
12 16065-66503 16065C 1 PC Board Assembly
13 0515-1550 0515-0372 3 Screw Pan Head M3×0.5 L8 w/
Washer
14 16065-01201 16065-01271 1 Angle
15 1250-0118 1250-0118 4 Connector BNC
16047-40002 16047-40002 8
Insulator
2
2190-0016 2190-0016 4
Washer
2
0360-1190 0360-1190 4
Terminal Solder Lug
2
2950-0043 2950-0043 4
Nut
2
0403-0427 0403-0427 6
Bumper Foot
2
1. These parts are for 1.
2. These parts are for 15.
Table 4-1 Replaceable Mechanical Parts (Continued)
Reference
Designator
Keysight
Part Number
RoHS
Compliant
Keysight
Part Number
Qty Descrption
20 Keysight 16065C External Bias Adapter
Service
Replaceable Parts
4-
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21

Keysight Technologies 16065C Operation And Service Manual

Type
Operation And Service Manual

Keysight Technologies 16065C External Bias Adapter is an accessory designed to operate specifically with the E4980AL LCR Meter and E4981A Capacitance Meter. When used in conjunction with an external dc voltage source, the 16065C can supply a bias voltage across components of up to ±40 V dc through a four-terminal pair type test fixture. This allows users to apply a bias voltage to components under test, enabling them to characterize the component's behavior under different bias conditions.