  • Hello! I have reviewed the installation instructions for the Minebea Intec YCC02-RB02 connecting cable and the YRB10Z rechargeable battery pack. This document outlines how to connect the battery pack to various Combics indicators using the provided cable, which includes a 5-pin round connector and cable gland. I understand the wiring and cable specifications and am ready to answer your questions about this device.
  • What is the purpose of this cable?
    What type of connector does this cable have?
    What is the cable length?
Installation Instructions | Installationsanleitung
Minebea Intec YCC02-RB02|Option: L2
Connecting Cable with 5-pin Round Male Connector and Cable Gland for Connecting a
YRB10Z External Rechargeable Battery Pack to a Combics
CIS1..|CIS2..|CIS3..| CW1S..|2S..|3S..|CAIS...|CAW..|CAH Indicator
Anschlusskabel mit 5-pol. Rundstecker und Kabelverschraubung zum Anschluss
des Akkupacks YRB10Z an Combics-Auswertegeräte
CIS1..|CIS2..|CIS3..| CW1S..|2S..|3S..|CAIS...|CAW..|CAH | CAS...
Open cable ends Cable gland 5-contact round
with ID labels male connector
Offene Kabelenden Kabelverschraubung 5-pol. Rundstecker
mit Kennzeichnung
Cable length: 2 m / Kabellänge: 2 m
Stranded wires that are not connected are cut to 20 mm and the end of the cables are iso-
lated with shrink-down tubing.
Nicht angeschlossene Litzen auf 20 mm abgeschnitten und mit Schrumpfschlauch isoliert.
2
Printed in Germany on paper that has
been bleached without any use of chlorine
RDB · KT
Publication No.: WYC6024-a04041
Open cable ends Cable gland 5-pin round male
with ID labels connector
Offene Kabelenden Kabelverschraubung 5-pol. Rundstecker
mit Kennzeichnung
+5 V 2 +5V SWITCH
GND 3 GND_LINE
L-BATT 4 LOW_BATT
ONOFF 1 ON_OFF
LINEB 5 LINE_B
Cable length: 2 m
Kabellänge: 2 m
Stranded wires that are not connected are cut to 20 mm and the ends of the cables are
isolated with shrink-down tubing.
Nicht angeschlossene Litzen auf 20 mm abgeschnitten und mit Schrumpfschlauch isoliert.
Connection|Anschluss
Indicator Terminal strip
Option Auswertegerät Klemmleiste Pin Label
L2 CIS1..|2..|3.. Akku 1 GND
CW1S..|2S..|3S.. LV3 2 LINEB
Connection:|Anschluss: 3 L-BATT
battery|Akku LV3 4 ONOFF
LV2 4 +5V
When used in conjunction with Option M8, two stranded
wires must be connected to LV2 pin 4.
In Verbindung mit Option M8 müssen zwei Litzen auf LV2
Pin 4 angeschlossen werden.
98647-003-23
98647-003-23
Indicator / Auswertegerät Terminal strip
Connection / Anschluss Klemmleiste Pin Label
Option L2 Battery | Akku
CIS2..|CIS3.. Akku 1 GND
CW.. LV3 2 LINEB
3 L-BATT
4 ONOFF
LV2 4 +5V
More information in the relevant operating instructions.
Weitere Informationen in der entsprechenden Betriebsanleitung.
Minebea Intec Bovenden GmbH & Co. KG
Leinetal 2, 37120 Bovenden, Germany
Phone: 49.551.309.83.0, Fax: 49.551.309.83.190
www.minebea-intec.com
Specifications subject to change without notice.
RS · KT
Publication No.: WYC6024-a17024
/