RKI Instruments 65-2485RK-02 Owner's manual

Type
Owner's manual
RKI Instruments, Inc.
www.rkiinstruments.com
65-2485RK-02
Multi Point Detector
Operator’s Manual
Part Number: 71-0237RK
Revision: B
Released: 6/17/20
2 • 65-2485RK-02 Multi Point Detector
WARNING
Read and understand this instruction manual before operating detec-
tor. Improper use of the detector could result in bodily harm or death.
Periodic calibration and maintenance of the detector is essential for
proper operation and correct readings. Please calibrate and maintain
this detector regularly! Frequency of calibration depends upon the
type of use you have and the sensor types. Typical calibration frequen-
cies for most applications are between 3 and 6 months, but can be
required more often or less often based on your usage.
65-2485RK-02 Multi Point Detector 3
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in
materials, workmanship, and performance for a period of one year from date of shipment from
RKI Instruments, Inc. Any parts found defective within that period will be repaired or
replaced, at our option, free of charge. This warranty does not apply to those items which by
their nature are subject to deterioration or consumption in normal service, and which must be
cleaned, repaired, or replaced on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or
repair procedures not in accordance with the operator’s manual. This warranty indicates the
full extent of our liability, and we are not responsible for removal or replacement costs, local
repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES
AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER
OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC.
INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS,
INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR
DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR
FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, dealers,
and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this
gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
a) Absorbent cartridges d) Batteries
b) Pump diaphragms and valves e) Filter elements
c) Fuses
4 • 65-2485RK-02 Multi Point Detector
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Junction Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Mounting the Multi-Point Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Wiring the Multi-Point Detector to a Controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Introducing Incoming Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting the Zero (Fresh Air) Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Replacing Components of the Multi-Point Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Calibration, H
2
S and CO Detectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Preparing for Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Setting the Zero Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Setting the Response Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Returning to Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Calibration, Oxygen Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Preparing for Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Setting the Fresh Air Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Setting the Zero Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Returning to Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
65-2485RK-02 Multi Point Detector 5
Overview
This manual describes the 65-2485RK-02 multi point direct connect detector. This manual
also describes how to install, start up, maintain, and calibrate the detector when it is used with
a gas monitoring controller. A parts list at the end of this manual lists replacement parts and
accessories for the detector.
Specifications
WARNING: Do not use this product in a manner not specified in this instruction manual.
Table 1 lists specifications for the multi-point detector.
NOTE: The following symbol on the detector label is a caution to the user to refer to this
documentation for installation and operation instructions:
WARNING: When using the 65-2485RK-02, you must follow the instructions and warnings
in this manual to assure proper and safe operation of the 65-2485RK-02 and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the 65-2485RK-02 as described in this manual.
Table 1: Specifications
Target Gas O
2
/H
2
S/CO
*Note: Since the 65-2485RK-02 is a direct connect
detector, calibration is done at the controller.
Area Classification Explosion proof for Class I, Groups B, C, and D
Sampling Method Diffusion
Detection Range O
2
: 0 to 25 %volume
H
2
S: 0 to 100 ppm (parts per million)
CO: 0 to 300 ppm (parts per million)
Response Time 90% in 30 seconds
Accuracy Oxygen
:
± 0.5% O
2
Hydrogen Sulfide:
± 5% of reading or ± 2 ppm H
2
S (whichever is greater)
Carbon Monoxide
:
± 5% of reading or ± 5 ppm CO (whichever is greater)
Operating Temperature 23°F to 104°F (-5°C to 40°C)
!
6 • 65-2485RK-02 Multi Point Detector
Description
This section describes the multi-point detector. It consists of the detectors, the terminal strips,
the conduit seal, and the junction boxes.
Detectors
This section describes the components of the various detectors that are used with the 65-
2485RK-02.
Oxygen Detector
The detector consists of the oxygen sensor, the detector housing body, detector housing cap,
and cap gasket.
Figure 1: Oxygen Detector Component Location
Detector Housing Body
The detector housing body protects the electronic components within the housing. Use the
mounting threads at the top of the housing to screw the oxygen detector into the 3/4 NPT hub
on the bottom of the junction box. Two wires extend from the top of the detector housing
body. Use these wires to connect the oxygen detector to the terminal block. One of the wires is
white and one of the wires is green.
The housing includes two sockets installed on a circuit board. These sockets accept the plug-
in sensor’s two pins to provide electrical connection for the sensor. The circuit board with the
sockets conditions the sensor’s signal before the signal reaches the controller.
Detector HousingBody
Detector Housing Cap
Flame Arrestor
Guard
Plug-in Oxygen Sensor
Cap Gasket
65-2485RK-02 Multi Point Detector 7
Housing Cap & Cap Gasket
The housing cap screws onto the detector housing. It retains the plug-in sensor and protects it
from damage. A foam gasket is installed inside the housing cap that seals against the sensor
face. The housing cap also includes a flame arrestor which contains any sparks that may occur
within the detector and a flame arrestor guard which protects the flame arrestor from damage.
Unscrew the detector cap to access the plug-in sensor for maintenance or replacement. A cap
gasket seals the interface between the housing and cap.
Plug-in Oxygen Sensor
The plug-in sensor is secured in the detector assembly by the housing cap. It has two pins that
mate with the sockets in the detector housing body. Through a series of chemical and
electrical reactions, the sensor produces an electrical output that corresponds to the detection
range of the transmitter.
CO Detector
The detector consists of the CO sensor, charcoal filter with rubber boot, detector housing
body, detector housing cap, and cap gasket.
Figure 2: CO Detector Component Location
Detector Housing, Housing Cap, & Cap Gasket
The detector housing body protects the sensing components within the housing. Use the 3/4
NPT mounting threads at the top of the housing to screw the CO detector into the bottom
conduit hub of the junction box. Use the removable cap near the bottom of the housing to
access the sensor for maintenance or replacement. The cap protects the sensor from damage
and includes a flame arrestor which contains any sparks which may occur within the detector
housing. A cap gasket seals the interface between the housing and cap. A flame arrestor guard
is permanently bonded to the cap.
Two wires extend from the top of the detector housing body. Use these wires to connect the
Plug-In CO Sensor
Rubber Boot With
Charcoal Filter
Detector Housing Cap
Detector Housing Body
Cap Gasket
Flame Arrestor
Guard
8 • 65-2485RK-02 Multi Point Detector
CO detector to the multi-point detector. The housing includes a four-socket pattern. This
socket pattern accepts the sensor’s four pins to secure the sensor within the detector housing.
A pre-amplifier, located between the sockets and two interconnect wires, conditions the
sensor’s signal before the signal reaches the controller.
Plug-In Sensor
The sensor is secured within the sensor housing by the housing cap. It has four pins that mate
with the sockets in the detector housing body. Through a series of chemical and electrical
reactions, the sensor produces an electrical output that corresponds to the detection range of
the detector.
Charcoal Filter
The disc-shaped charcoal filter is secured to the face of the CO sensor with a rubber boot. The
charcoal filter prevents interference gases (hydrogen sulfide [H
2
S] and certain hydrocarbons)
from producing false CO readings.
H
2
S Detector
The detector consists of the detector housing body, detector housing cap, cap gasket, rubber
boot, spacer, and H
2
S sensor.
Figure 3: H
2
S Detector Component Location
Detector Housing, Housing Cap, & Cap Gasket
The detector housing protects the sensing components within the housing. Use the mounting
threads at the top of the housing to screw the H
2
S detector into a 3/4 NPT hub. Use the
removable cap near the bottom of the housing to access the sensor for maintenance or
replacement. The cap protects the sensor from damage and includes a flame arrestor which
Cap Gasket
Detector Housing Body
Detector
Housing Cap
Flame
Arrestor
Guard
H2S Plug-inSensor
Rubber Boot
Spacer
65-2485RK-02 Multi Point Detector 9
contains any sparks which may occur within the detector housing. A cap gasket seals the
interface between the housing and cap. A flame arrestor guard is permanently bonded to the
cap.
Two wires extend from the top of the detector housing. Use these wires to connect the H
2
S
detector to the multi-point detector. The housing includes a four-socket pattern. This socket
pattern accepts the sensor’s four pins to secure the sensor within the detector housing. A pre-
amplifier, located between the sockets and two interconnect wires, conditions the sensor’s
signal before the signal reaches the controller.
Rubber Boot and Spacer
A rubber boot and spacer are installed between the detector housing cap and the sensor. They
help ensure that the detector remains plugged into the detector housing body.
Sensor
The sensor is secured within the sensor housing by the detector housing cap. It has four pins
that mate with the sockets in the detector housing body. Through a series of chemical and
electrical reactions, the sensor produces an electrical output that is proportional to the
detection range of the detector.
Junction Box
Use the junction box to install the multi-point detector at a mounting site that is remote from
the controller. The junction box protects the terminal strip and wiring connections made to the
terminal strip. Use the top 3/4 NPT conduit hub to connect wiring from the terminal strip to
the controller. Use the cover on the front of the junction box to access the interior of the
junction box. The detectors and terminal strip are factory installed in the junction box.
Installation
This section describes procedures to mount the multi-point detector in the monitoring
environment and wire the detector to a controller.
Mounting the Multi-Point Detector
1. Select a mounting site that is representative of the monitoring environment. Consider the
following when you select the mounting site.
Select a site where the detector is not likely to be bumped or disturbed. Make sure
there is sufficient room to perform start-up, maintenance, and calibration procedures.
Select a site where the target gases are likely to be found first.
10 • 65-2485RK-02 Multi Point Detector
Figure 4: Outline & Mounting Dimensions 65-2485RK-02
2. At the mounting site you select, use #10 screws to mount the detector to a vertical surface.
CAUTION: Mount the multi-point detector with the detectors facing down (see Figure 4).
NOTE: If your application does not require a specific mounting site, mount the detector at
approximately breathing level.
3/4 NPT Hubs (4X)
Oxygen
Detector
3.11
.25
.20 DIA X .45 Slot,(2X) ,mounting
1.10
CO
Detector
2.40
H2S
Detector
3.00
3.15
3.00
1.25
7.75
8.44 max.
65-2485RK-02 Multi Point Detector 11
Wiring the Multi-Point Detector to a Controller
WARNING: Always verify that the power source is OFF before you make wiring
connections.
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Remove the junction box cover from the junction box.
WARNING: To maintain the explosion proof classification of the multi-point detector, a
conduit seal must be used within 2 inches of the junction box conduit hub used
for wiring to the controller.
4. Guide a six-conductor, shielded cable, or six wires in conduit through the top conduit hub
of the junction box. Use appropriate conduit fittings and construction technique for the
environmental rating and hazardous location classification of the junction box. The
junction box is rated NEMA 4X and classified explosion proof for Class I, Groups B, C,
and D.
5. Connect the wires to the multi-point detector using the terminal block in the junction box.
CAUTION: If using shielded cable, leave the drain wire insulated and disconnected at the
multi-point detector. You will connect the opposite end of the cable’s drain wire
at the controller.
6. Secure the junction box cover to the junction box.
7. Route the cable or wires leading from the multi-point detector through one of the conduit
hubs at the controller housing. Use appropriate conduit fittings and construction technique
for the environmental rating of the controller. RKI controllers are typically rated NEMA
4X.
CAUTION: Do not route power and multi-point detector wiring through the same conduit
hub. The power cable may disrupt the transmission of the detector signal to the
controller.
8. Connect the wires to the applicable controller terminal strips. See the controller operator’s
manual and the controller’s detector head specification sheet. Also see Figure 5 for a
wiring diagram.
12 • 65-2485RK-02 Multi Point Detector
Figure 5: Wiring the 65-2485RK-02 to a Controller
9. Connect the cable’s drain wire to an available chassis ground at the controller. RKI
controllers typically have a ground stud that can be used to ground the cable’s drain wire.
Green
White
Black
Red
Terminal Strip
ControllerH2S
Terminals
Black
Red
ControllerOxygen
Terminals
ControllerCO
Terminals
OxygenDetector
Wires
White
Green
OxygenDetector
CO Detector
Wires
Black
Red
Black
H2S Detector
Red
H2S Detector
Wires
CO Detector
65-2485RK-02 Multi Point Detector 13
Start Up
This section describes procedures to start up the multi-point detector and place the detector
into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power wiring to the controller is correct and secure. Refer to the controller
operator’s manual.
3. Turn on power to the controller.
4. Turn on the controller.
5. Verify that the controller is on and operating properly. Refer to the controller operator’s
manual.
CAUTION: Allow the multi-point detector to warm up for 5 minutes before you continue
with the next section, “Setting the Zero (Fresh Air) Reading”.
Setting the Zero (Fresh Air) Reading
WARNING: If the detector installation is explosion proof, do not remove the detector
housing cap or junction box cover while the circuits are energized unless the
area is determined to be non-hazardous. Keep the detector housing cap and
junction box cover tightly closed during operation.
CAUTION: If you suspect the presence of a target gas in the monitoring environment, use
the calibration kit and a zero air calibration cylinder to introduce “fresh air” to
the desired detector and verify an accurate zero setting. See “Calibration, H
2
S
and CO Detectors” on page 21 or “Calibration, O
2
Detector” on page 23 for
instructions to introduce zero air to the desired detector.
1. Verify that the detector is in a fresh air environment (environment known to be free of
toxic gases, and of normal oxygen content, 20.9%).
2. At the controller, verify a reading of:
20.9% for oxygen
0 ppm for CO or H
2
S
If the display reading is the appropriate fresh air reading, start up is complete. The detector
is in normal operation. If the display reading is not the appropriate fresh air reading,
continue with step 3.
3. Perform a zeroing operation at the controller. See the controller operator’s manual for
directions.
14 • 65-2485RK-02 Multi Point Detector
Maintenance
This section describes maintenance procedures. It includes preventive maintenance,
troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance
of the multi-point detector. It includes daily and quarterly procedures.
Daily
At the controller, verify a display reading of:
20.9% for oxygen
0 ppm for CO or H
2
S
Investigate significant changes in the display reading.
Quarterly
Calibrate the multi-point detector as described in the Calibration section of this manual.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for
problems you may encounter with the multi-point detector.
NOTE: This troubleshooting guide describes multi-point detector problems only. See the
controller operator’s manual for problems you may encounter with the controller.
Table 2: Troubleshooting the Multi-Point Detector
Condition Symptom(s) Probable Causes Recommended Action
Fail
Condition
Controller indicates a
fail condition.
The detector wiring is
disconnected or
misconnected.
The CO, O
2
, or H
2
S
plug-in sensor is not
properly plugged into
the sockets in the
detector housing body.
The detector’s zero
reading is low enough
to cause a fail
condition.
The detector is
malfunctioning.
1. Verify that the detector wiring is
correct and secure.
2. Confirm that the CO, O
2
, or H
2
S
plug-in sensor is installed properly.
3. Perform a zero adjustment. A full
calibration is recommended.
4. If the fail condition continues,
replace the CO, O
2
, or H
2
S plug-in
sensor as described later in this
section.
5. If the fail condition continues,
contact RKI for further instruction.
65-2485RK-02 Multi Point Detector 15
Replacing Components of the Multi-Point Detector
This section includes maintenance procedures for the oxygen detector, the CO detector, and
the H
2
S detector.
Oxygen Detector
This section contains instructions to replace the plug-in oxygen sensor and to replace the
entire detector assembly. In most cases it in not necessary to replace the entire detector
assembly.
Replacing the Plug-In Oxygen Sensor
CAUTION: The plug-in sensor contains electrolyte that is an irritant. Do not disassemble
the sensor when replacing it with a new one. If electrolyte comes in contact with
your skin, wash affected area thoroughly with soap and water.
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Unscrew the detector housing cap from the detector housing body. Make sure not to lose
the cap gasket.
4. Unplug and remove the oxygen sensor.
5. Carefully plug the replacement sensor into the socket pattern that is located in the detector
housing.
Slow or No
Response/
Difficult or
Unable to
Calibrate
Unable to accurately
set the zero or
response reading
during calibration.
Detector requires
frequent calibration.
Note: Under
“normal”
circumstances, the
transmitter requires
calibration once every
3 months.
Some applications
may require a more
frequent calibration
schedule.
The calibration
cylinder is low, out-
dated, or defective.
The flame arrestor in
the CO, O
2
, or H
2
S
detector housing cap is
wet or clogged with
dirt or other
particulates.
The calibration gas is
not an appropriate
concentration.
The detector is
malfunctioning.
1. Verify that the calibration cylinder
contains an adequate supply of a
fresh test sample.
2. Check the detector housing cap to
determine if the flame arrestor is wet
or dirty. Clean if necessary.
3. Verify that the calibration gas
concentration is appropriate for each
detector.
4. If the calibration/response difficulties
continue, replace the CO, O
2
, or H
2
S
plug-in sensor as described later in
this section.
5. If the calibration/response difficulties
continue, contact RKI for further
instruction.
Unexplained
Upscale
Readings or
Alarms
Controller indicates
a reading that cannot
be verified.
Alarms occur at the
controller that cannot
be explained.
Charcoal filter in CO
detector is saturated
and no longer
scrubbing out
interfering gases.
1. Replace charcoal filter in the CO
detector.
2. If difficulties continue, contact RKI
for further instructions.
Condition Symptom(s) Probable Causes Recommended Action
16 • 65-2485RK-02 Multi Point Detector
6. Make sure the cap gasket is in place and screw the detector housing cap back onto the
detector housing body.
7. Turn on or plug in power to the controller.
8. Turn on the controller and place into normal operation.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with
the next step.
9. Calibrate the detector as described in “Calibration, O
2
Detector” on page 23.
Replacing the Oxygen Detector
NOTE: In most cases, it is only necessary to replace the oxygen sensor.
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. Remove the junction box cover from the junction box.
4. Disconnect the detector leads from the terminal block in the junction box. Note the
position of the color-coded leads as you remove them.
5. Unscrew the detector from the junction box hub.
6. Guide the detector leads of the replacement detector through the junction box hub then
screw the mounting threads of the detector into the hub. If necessary for environmental
conditions, apply thread sealant or Teflon tape to the hub and/or detector threads to seal
them.
7. Connect the detector leads to the terminal block the same way the old detector was wired.
See the controller operator’s manual and the controller’s detector head specification sheet
for the oxygen detector to verify the connections are correct. Also see Figure 5 for a
wiring diagram.
8. Reinstall the junction box cover.
9. Turn on or plug in power to the controller.
10. Turn on the controller.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue
with the next step.
11. Calibrate the replacement detector as described in “Calibration, O
2
Detector” on page 23.
65-2485RK-02 Multi Point Detector 17
CO Detector
This section includes a procedure to replace the plug-in sensor, one to replace the charcoal
filter, and one to replace the entire detector assembly. In most cases it is not necessary to
replace the entire detector assembly.
Replacing the Plug-In CO Sensor
CAUTION: The sensor contains electrolyte which is a dilute acid. Do not disassemble the
sensor when replacing it with a new one. If sensor electrolyte comes in contact
with your skin, wash affected area thoroughly with soap and water.
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Unscrew the detector housing cap from the detector housing body. Make sure not to lose
the cap gasket.
4. Unplug and remove the CO sensor with the boot and charcoal filter attached.
5. Remove the rubber boot and charcoal filter from the old sensor.
6. Remove the replacement sensor from its packaging and remove the wire jumper. This wire
jumper is installed on the sensor pins for shipment or storage but must be removed for the
sensor to operate properly when installed in a detector.
WARNING: The CO sensor will not operate properly if the wire jumper is not removed.
7. Install the rubber boot with charcoal filter onto the replacement sensor’s face.
8. Carefully plug the replacement sensor into the four-socket pattern that is located in the
detector housing.
9. Make sure the cap gasket is in place and screw the detector housing cap back onto the
detector housing body.
10. Turn on or plug in power to the controller.
11. Turn on the controller and place into normal operation.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with
the next step.
Figure 6: Plug-In Sensor Jumper Removal
Remove jumper
18 • 65-2485RK-02 Multi Point Detector
12. Calibrate the replacement sensor as described in “Calibration, H
2
S and CO Detectors” on
page 21.
Replacing the Charcoal Filter
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Unscrew the detector housing cap from the detector housing body.
4. Unplug and remove the CO sensor with the boot and charcoal filter attached.
5. Remove the rubber boot that secures the charcoal filter to the CO sensor.
6. Remove the charcoal filter from the rubber boot.
7. Place the replacement filter in the rubber boot in the same position as the filter you
removed in the previous step.
8. Reinstall the rubber boot with charcoal filter to the CO sensor.
9. Carefully plug the replacement sensor into the four-socket pattern that is located in the
detector housing.
10. Make sure the cap gasket is in place and screw the detector housing cap back onto the
detector housing body.
11. Turn on or plug in power to the controller.
12. Turn on the controller and place into normal operation.
Replacing the CO Detector
NOTE: In most cases, it is only necessary to replace the CO sensor.
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Remove the junction box cover from the junction box.
4. Disconnect the detector leads from the terminal block in the junction box. Note the
position of the color-coded leads as you remove them.
5. Unscrew the detector from the junction box hub.
6. Guide the detector leads of the replacement detector through the junction box hub then
screw the mounting threads of the detector into the hub. If necessary for environmental
conditions, apply thread sealant or Teflon tape to the hub and/or detector threads to seal
them.
7. Connect the detector leads to the terminal block the same way the old detector was wired.
See the controller operator’s manual and the controller’s detector head specification sheet
to verify the connections are correct. Also see Figure 5 for a wiring diagram.
8. Reinstall the junction box cover.
9. Turn on or plug in power to the controller.
65-2485RK-02 Multi Point Detector 19
10. Turn on the controller and place into normal operation.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue
with the next step.
11. Calibrate the replacement detector as described in the “Calibration, H
2
S and CO
Detectors” on page 21.
H
2
S Detector
This section includes a procedure to replace the H
2
S sensor and one to replace the entire
detector assembly. In most cases it is not necessary to replace the entire detector assembly.
Replacing the Plug-In H
2
S Sensor
CAUTION: The sensor contains electrolyte which is a dilute acid. Do not disassemble the
sensor when replacing it with a new one. If sensor electrolyte comes in contact
with your skin, wash affected area thoroughly with soap and water.
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Unscrew the detector housing cap from the detector housing body. Make sure not to lose
the cap gasket.
4. Unplug and remove the H
2
S sensor with the rubber boot and spacer attached.
5. Remove the rubber boot and spacer from the old sensor.
6. Remove the replacement sensor from its packaging and remove the wire jumper. This wire
jumper is installed on the sensor pins for shipment or storage but must be removed for the
sensor to operate properly when installed in a detector.
WARNING: The H
2
S sensor will not operate properly if the wire jumper is not removed.
7. Install the spacer and rubber boot onto the replacement sensor’s face.
8. Carefully plug the replacement sensor into the socket pattern that is located in the detector
housing.
Figure 7: Plug-In Sensor Jumper Removal
Remove jumper
20 • 65-2485RK-02 Multi Point Detector
9. Make sure the cap gasket is in place and screw the detector housing cap onto the detector
housing body.
10. Turn on or plug in power to the controller.
11. Turn on the controller.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with
the next step.
12. Calibrate the detector as described in “Calibration, H
2
S and CO Detectors” on page 21.
Replacing the H
2
S Detector
NOTE: In most cases, it is only necessary to replace the H
2
S sensor.
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Remove the junction box cover from the junction box.
4. Disconnect the detector leads from the terminal block in the junction box. Note the
position of the color-coded leads as you remove them.
5. Unscrew the detector from the junction box hub.
6. Guide the detector leads of the replacement detector through the junction box hub then
screw the mounting threads of the detector into the hub. If necessary for environmental
conditions, apply thread sealant or Teflon tape to the hub and/or detector threads to seal
them.
7. Connect the detector leads to the terminal block the same way the old detector was wired.
See the controller operator’s manual and the controller’s detector head specification sheet
for the H
2
S detector to verify the connections are correct. Also see Figure 5 for a wiring
diagram.
8. Reinstall the junction box cover.
9. Turn on or plug in power to the controller.
10. Turn on the controller.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue
with the next step.
11. Calibrate the replacement detector as described in “Calibration, H
2
S and CO Detectors”
on page 21.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26

RKI Instruments 65-2485RK-02 Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI