NOTE:
● Use only the parts listed above that are distributed by
ForaCare; use of other parts may constitute a hazard.
● The item to , are regarded as consumable accesso-
ries. Replace them with new ones if they are still dirty or
obstructed after cleaning.
● The nebulizer kit is for a course of treatment only.
● Expected service life of the main unit is 1 year.
Adults / Children Masks (optional)
Cap
Spray Nozzle
Medicine Cup
Mouthpiece
Nebulizer Kit (assembly of items
to )
Main Unit
Air Hose
Cleaning Steps
1. Pull out the mouthpiece or
mask ( ), and detach the
nebulizer kit from the main
unit ( ).
4. Attach the mouthpiece (or mask) to
the cap ( ). You can directly
connect the nebulizer kit to the
main unit ( ), or connect the air
hose to the nebulizer kit ( ) and
the main unit ( ) if needed.
5. Close your lips around the
mouthpiece and press the power
button to activate the nebulizer.
Hold the main unit vertically
during your treatment. Never lie
down while using the nebulizer.
6. Inhale slowly and deeply, and breathe out naturally. Inhalation
should be stopped when the medicine has been fully used.
7. After treatment, press the power button to turn off the device
and clean as directed in “CLEANING AND MAINTENANCE”.
311-7015700-003
Version 2.0
2022/12
NBL300 Series
Compressor Nebulizer
Operation Instructions
Read instruction before use.
Distributed by ForaCare, Inc.
893 Patriot Drive Suite D
Moorpark, CA 93021 USA
Products made in Taiwan
Toll Free: 1-888-307-8188
(8:30 am-5:00 pm PST, Mon.-Fri.)
For assistance outside of these hours,
please contact your healthcare
professional.
www.foracare.com
DEAR SootheNeb NBL300g OWNER:
Thank you for purchasing the SootheNeb NBL300g Compressor
Nebulizer. This instruction provides important information to help
you use the system properly. Before using this product, please
read the following contents thoroughly and carefully.
The nebulizer accessories and power plug (adapter) can be
purchased separately. If you have other questions regarding this
product, please contact the place of purchase or call the
Customer Service Line at 1-888-307-8188 (8:30 am-5:00 pm
PST, Mon.-Fri.). For assistance outside of these hours, please
contact your healthcare professional.
INTENDED USE
SootheNeb NBL300g Compressor Nebulizer is designed to
provide a compressed air source to aerosolize physician-pre-
scribed liquid medication when used in combination with the
packaged nebulizer accessories, except for Pentamidine. For
reasons of hygiene, the packaged nebulizer accessories are
intended for single use by single patient. SootheNeb NBL300g
Compressor Nebulizer is intended for use with children and adult
patients in the homecare settings.
Caution: Federal law restricts this device to sale by or on the
order of a physician.
IMPORTANT INFORMATION ABOUT INHALATION THERAPY
The device is ideally suitable for inhalation at home or when
travelling. It guarantees highly effective, fast-acting inhalation
treatment for children and adult patients.
Carry out inhalation in a quiet and relaxed state, and inhale
slowly and deeply, so that the medication can penetrate into the
fine, deep bronchial tubes. Exhale normally.
ITEMS SUPPLIED AND DESCRIPTIONS OF PARTS
Before you start to use the SootheNeb NBL300g Compressor
Nebulizer, please check if the following items are included in the
package.
SPECIFICATIONS
Model No.: SootheNeb NBL300g
Dimension & Weight: 99.1 mm x 81.4 mm x 246.4 mm, 305g
Charger output power: 5V (volts) / 2A (amps), USB Type C
Battery: Li-Polymer, non-removable
IP classification: IP22
Mode of operation: 30 mins on/ 30 mins off
Nebulization Rate: ≥ 0.25 ml/min
Medication Capacity: 10ml
MMAD: 3 microns
Power consumption: 7W
Noise level: 60 dBA
Compressor Pressure Range: 13 to 20 Psi
Operating Pressure Range: 5 to 10 Psi
Maximum Flow Rate: ≥3 LPM (I/min)
Aerosol Output: 1.17 ± 0.13 (ml)*
Aerosol Output Rate: 0.35 ± 0.04 (ml/min)**
Operating conditions: 41°F to 104°F (5°C to 40°C), 15% to 93%
relative humidity, 700 hPa to1060 hPa
Storage conditions: -13°F to 158°F (-25°C to 70°C), 10% to 95%
relative humidity
*:Continue the treatment until the medicine cup is empty or the
mist stops.
**:The treatment time for aerosol output rate is 1 min.
1
2
1
2
3
4
1
1
2
2
4
3
4
5
5
6
7
8
8
9
10
11
12
13
14
15
16
USB Type C Cable
Air Opening
Air Inlet
Power Button
4G Connection Indicator
Power Button Indicator
Battery Compartment
USB Type C Port (for recharge)
NOTE:
Please contact your dealer or Customer Service Line for
assistance with any other difficulties.
Notice what can be immersed (left) and cannot be
immersed (right) in the water.
IMPORTANT SAFETY PRECAUTIONS
● Use the equipment only for its intended use as described in
this instruction. Do not use attachments not recommended by
the manufacturer.
● Use the equipment with medications only under the instruction
of your physician.
● Do not use the equipment if it has any damaged parts, or it
has fallen into water.
● To use the equipment for the first time, or after storing it for an
extended period, be sure to clean all necessary parts as
described in the cleaning instructions.
● Do not use while bathing.
● The unit should not be left unattended while plugged in or
turned on.
● Keep the unit out of reach of small unsupervised children. The
small parts detached from the device may result children
choking from inhaling or swallowing.
● The accessible materials used in the device will not cause the
potential allergic reactions to skin.
● Do not try to modify the device to prevent any danger.
● Do not place the device in liquid, nor put it where it could fall
into liquid. If the device becomes wet, unplug the device
before touching it.
● Do not use the device if it is not working properly, or if it has
suffered any damage.
● Do not subject the nebulizer to any impact and do not drop it.
● After a period of no use, and after every use, please clean the
main unit.
● Do not expose the nebulizer to direct sunlight, high tempera-
tures or humidity.
START YOUR TREATMENT
Before Use
● Ensure that all parts are clean and dry.
● Before starting treatment, talk to your physician about the
duration, dosage and frequency of use of the nebulizer.
● The device is designed for an intermittent use: 30 minutes ON
/ 30 minutes OFF.
● Do not remove the nebulizer kit from the main unit during
nebulization.
● Pour the distilled water (2 - 5 ml) into the medicine cup. Press
the power button to turn on the device, check if the nebuliza-
tion is taking place properly. Then, press the power button to
turn off and discard any remaining water.
● The prescribed volume of the drug should be adhere to.
● The device does not heat up the medicine cup during opera-
tion, so there is no loss of solution due to evaporation.
How to Use
● Before first use, pull out the insulation
mylar embedded in the battery
compartment and recharge the main
unit for 5 hours. Suggestion for
using 5V / 2A power plug for
recharge.
● Once the nebulizer is plugged into the
power plug at any time, it will go into
recharge mode and stop nebulization.
1. Align the tab of the spray
nozzle with the notch of the
cap and assemble.
2. Add the appropriate amount
of medication to the medicine
cup, as directed by your
physician. Do not exceed
the line marked “MAX”.
3. Assemble the nebulizer kit.
● Turn clockwise, and the
junction of the nebulizer kit
will become one full circle.
● Align and insert the
assembled cap on the
medicine cup.
CLEANING AND MAINTENANCE
Please clean your device after each use according to the instruc-
tions below.
Before Cleaning
Make sure that the nebulizer is turned off before disassembly.
TROUBLESHOOTING
Maintenance for Storage
1. Always handle your nebulizer with care.
2. Keep your nebulizer out of children’s reach. The device may
contain small parts that can be swallowed.
3. If you store your nebulizer, do not bend the air hose.
4. If you store your nebulizer, keep it in the following environmen-
tal range:
a) Temperature: -13°F to 158°F (-25°C to 70°C),
b) Relative humidity: 10% to 95%
5. If possible, store your nebulizer in a well-ventilated room.
Disposal
Dispose of the device, components and optional accessories
according to your local regulations.
2. Turn couterclockwise, and
depart the cap with spray
nozzle from the medicine
cup.
4. Rinse the mouthpiece, medicine cup,
spray nozzle and cap thoroughly with
running water.
*Use clean cloth only to wipe the
masks.
*Do not use brushes and thin, sharp
objects (i.e., pins, needles, etc.) to
clean out the clog in the gap.
*Once the mouthpiece and nebulizer
kit are dirty or clogged after
rinsing, immerse them in water.
5. Shake off excess water,
and left them air dry.
6. Use clean cloth to wipe the
outer casing of the main unit
and air hose, then let them air
dry.
3. Take the spray nozzle and
cap apart.
1
4
3
Press and hold the power button
for 5 seconds in standby mode.
Press the power button during
nebulization.
Flashing yellow and turning off
automatically in 10 seconds.
Initiate the 4G
connection
Data upload failed
The indicator change from flashing
blue to solid blue.
Press the power button during 4G
connection. Flashing blue turns
quicker and fades away in 8
seconds, then the nebulizer turns off.
4G connection
succeeded
Forced shutdown 4G
connection
Solid blue and turning off automatically.
Data upload
completed.
Message Probable Cause Solution
LED INDICATORS
LED Indicator
4G Power Nebulizer Status
Ready for use /
Standby mode
Nebulization mode.
Low battery
Recharge the battery
First time activate the
nebulizer.
The nebulizer is on standby,
ready to start at any time.
Press the power button to activate
the nebulizer.
Battery recharge is in progress.
The battery is fully recharged.
The nebulizer is functional, but it is
time to recharge.
The battery is too low. Plug in the
power plug to recharge.
Pull out the mylar. Both 4G and
Power indicators will light up
simultaneously and change
color in sequence as shown.
Description
The device does not
work or the device is
on but nebulizes
weakly.
The device is on but
does not nebulize.
The nebulizer kit does not
assemble well.
Re-assemble the nebulizer
kit.
Plug in the power plug to
recharge the main unit. The
nebulizer will not activate
while charging.
Remove any bend or kink in
the air hose.
Place the spray nozzle in its
correct position.
The battery is getting too
low.
The spray nozzle does
not place in its position.
Add the appropriate amount
of medicine prescribed by
your physician to the
medicine cup.
No medicine is left.
Compressor switch is off.
Air hose is bent.
Turn the switch on.
Plot of cumulative size distribution of results for particle size
Cumulative Fraction (%)
Mean Mass Fraction (%)
AeroDynamic Diameter (µm)
0
0.1 110 100
20
40
60
80
100
0
10
30
20
40
50
individual tests
mean
The device has been certified to meet the electrical and safety
requirements of: IEC 60601-1, IEC 60601-1-2.
>10s
1
2
1
2
3
4
5
6
7
8
10
11
12
14
13
15
16
9
2
1