SPECIFICATIONS
S
ervice: Compatibleliquids.
W
etted Materials:
B
ody:316SSorbrass;
Spring:302SS;
Rangespring:302SS;
Magnet:PTFE-coated;
Meteringcone:AcetalorPTFE;
Seals:Buna.
Temperature Limits: -40to200°F(-40to93°C).
Pressure Limit: DTFW-3S:1500psig(100bar);Allothermodels:3000psig(200
bar).
Accuracy: ±2%FS.
Repeatability: ±1%FS.
Size: Diameterdialface2.5˝(63.5mm).
Process Connection: Seemodelchart.
Weight:
DTFW-1Band1S:3lb(1.36kg);
DTFW-2Band2S:5lb(2.27kg);
DTFW-3S:10lb(4.54kg).).
T
he Series DTFW Variable-Area Flowmeter for Liquids measures water flow
rates with ±2% of full-scale accuracy at an affordable price. Durable metal
construction ensures great compatibility and the strength to withstand system
pressures of up to 3000 psig (200 bar). With a wide variety of flow ranges
available,theSeriesDTFWcanbeusedwithmanydifferentapplicationsallatan
e
conomicalprice.
C
alibration
AllSeriesDTFWVariable-AreaFlowmetersforLiquidsarecalibratedatthefactory
for mounting in the horizontal position and should be installed that way for
maximumaccuracy.
Theunitwillnormallyretainitsaccuracytolerancefortheusefullifeofthedevice.
Ifyouwanttochecktheaccuracy,dosoonlywithinstrumentsofcertifiedaccuracy.
Do not attempt to check accuracy with other flowmeters piped in series. Even
minorvariationsinpipingandbackpressurecancausesignificantdifferencesin
indication.Ifindoubt,returntheflowmetertothefactoryforacalibrationcheck.
1
-1/2
[
38.10]
7
[177.80]
2.5˝ B-STYLE CASE DTFW OR DTFO 1˝ IN/OUT
3-5/32
[80.26]
Ø
2
[
50.80]
3
[76.20]
1
-9/64
[
28.96]
3
[
76.20]
Ø
2-59/64
[
74.17]
1
˝ NPT
INLET & OUTLET
1-1/2
[38.10]
1
1/16
[17.53]
1
/2˝ NPT
I
NLET & OUTLET
Ø
2-59/64
[
74.17]
5-47/64
[145.54]
1
[25.40]
3/4
[19.05]
3
-19/32
[
91.44]
3
[
76.20]
Ø
2-59/64
[
74.17]
11/16
[17.53]
1
/4˝ NPT
I
NLET & OUTLET
2
[
50.80]
2
.5˝ B-STYLE CASE DTFW OR DTFO 1/4˝ IN/OUT
2.5˝ B-STYLE CASE DTFW OR DTFO 1/2˝ IN/OUT
R
1
[
25.40]
Connection
1/4˝NPT
1/4˝NPT
1/4˝NPT
1/4˝NPT
1/4˝NPT
1/4˝NPT
1/4˝NPT
1/4˝NPT
1/4˝NPT
1/4˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1/2˝NPT
1˝NPT
1˝NPT
1˝NPT
1˝NPT
1˝NPT
Model
DTFW-1B-1W
DTFW-1B-2W
DTFW-1B-3W
DTFW-1B-4W
DTFW-1B-5W
DTFW-1S-1W
DTFW-1S-2W
DTFW-1S-3W
DTFW-1S-4W
DTFW-1S-5W
DTFW-2B-1W
DTFW-2B-2W
DTFW-2B-3W
DTFW-2B-4W
DTFW-2B-5W
DTFW-2B-8W
DTFW-2B-10W
DTFW-2S-1W
DTFW-2S-2W
DTFW-2S-3W
DTFW-2S-4W
DTFW-2S-5W
DTFW-2S-8W
DTFW-2S-10W
DTFW-3S-10W
DTFW-3S-15W
DTFW-3S-20W
DTFW-3S-25W
DTFW-3S-30W
Range, GPM
Water
0to1
0to2
0to3
0to4
0to5
0to1
0to2
0to3
0to4
0to5
0to1
0to2
0to3
0to4
0to5
0to8
0to10
0to1
0to2
0to3
0to4
0to5
0to8
0to10
0to10
0to15
0to20
0to25
0to30
Body
Brass
Brass
Brass
Brass
Brass
SS
SS
SS
SS
SS
Brass
Brass
Brass
Brass
Brass
Brass
Brass
SS
SS
SS
SS
SS
SS
SS
SS
SS
SS
SS
SS
Metering
Cone
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
Acetal
PTFE
PTFE
PTFE
PTFE
PTFE
L
ocation
S
electalocationwheretheflowmetercanbeeasilyreadandwherethetemperature
w
ill not exceed 200°F (93°C). The mounting surface and piping to the flowmeter
shouldbefreefromexcessivevibration.Ifpulsingfloworvibrationcausesexcessive
pointeroscillation,contactfactoryforwaystoprovideadditionaldampening.
B
ecausethedeviceoperatesthroughamagneticcouplingsystem,magneticmaterial
n
eartheflowblockcanaffectaccuracy.Tominimizetheeffectofoutsideinterference,
k
eepmagneticmaterialsatleast3inches(7.6cm)awayfromtheunit.
Inlet Piping
It is a good practice to have the inlet port of the flowmeter be as free of elbows,
restrictions and size changes as possible. Inlet piping should be as close to the
f
lowmeterconnectionsizeaspossibletoavoidpossibleturbulencewhichcanoccur
w
ithdrasticsizechanges.Thelengthoftheinletpipinghasverylittleeffectonnormal
p
ressurefedflowmeters.
Forvacuumservice,theinletpipingshouldbeasshortandopenaspossibletoallow
operation at or near atmospheric pressure, thus maintaining the accuracy of the
device.Notethatforvacuumservice,anyflowcontrolvalvemustbeinstalledonthe
d
ischargesideoftheflowmeter.
O
utlet Piping
Pipingontheoutletportshouldbeatleastaslargeastheflowmeterconnection.
O
peration
I
fthe mediaislikelytobedirty,itisrecommendedthata50–micronfilterisplaced
u
pstreamoftheinletporttoassureacleanfluidmediumflowsthroughtheflowmeter.
After all connections are complete, introduce flow as slowly as possible to avoid
damage.Withliquidflow,it maybenecessaryto purgeairfromthe system before
readingsstabilize.Nolubricationorperiodicservicingisrequired.Keepcaseexterior,
i
ncludingfrontcover,cleanforbestvisibility.
M
AINTENANCE/REPAIR
U
ponfinalinstallationoftheSeriesDTFW,noroutinemaintenanceisrequired.The
SeriesDTFWisnotfieldserviceableandshouldbereturnedifrepairisneeded.Field
repairshouldnotbeattemptedandmayvoidwarranty.
W
ARRANTY/RETURN
R
eferto“TermsandConditionsofSales”inourcatalogandonourwebsite.Contact
c
ustomerservicetoreceiveaReturnGoodsAuthorizationnumberbeforeshippingthe
p
roductbackforrepair.Besuretoincludeabriefdescriptionoftheproblemplusany
additionalapplicationnotes.
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: [email protected]
©Copyright2012 DwyerInstruments,Inc. PrintedinU.S.A.6/12 FR# R4-443965-00Rev.3
  • Page 1 1
  • Page 2 2

Dwyer Series DTFW User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI