Milwaukee MW803 User manual

  • Hello! I am an AI chatbot trained to assist you with the Milwaukee MW803 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
USER MANUAL
MW803 • MW804
Pocket-size
pH/EC/TDS/Temperature
Meters
SPECIFICATIONS
Range
0.00 to 14.00 pH
0to3999μS/cm(MW803)
0to2000ppm(MW803)
0.00to20.00mS/cm(MW804)
0.00to10.00ppt(MW804)
0.0to50.0°C/32.0to122.0°F
Resolution
0.01 pH
1μS/cm/1ppm(MW803)
0.01mS/cm/0.01ppt(MW804)
0.1°C/0.1°F
Accuracy (@25°C)
0.05 pH
2%FS(EC/TDS)
0.5°C/±1°F
TDS Factor
0.45to1.00(CONV)
Temperature Compensation
Automatic,withß=0.0to2.4%/°C
Calibration
Automatic,1pointforECand1or
2pointsforpH
Probe Replaceable MI60P
Environment
0to50°C;100%RHmax.
Battery Type
4x1.5V;IECLR44,A76
Battery Life
Approx.100hoursofuse
Auto-off
After8min.ofnon-use
Dimensions / Weight
200xdia38mm/100g
Makesurethegasketisinplacebeforescre-
wing back the cap. Batteries should only be
replacedinasafeareausingthebatterytype
speciedinthisinstructionmanual.
ACCESSORIES
MI60P Replaceableprobefor
MW803&MW804
M10030B12.88mS/cmcalibration
solution,20mlsachet,25pcs
M10031B1413μS/cmcalibrationsolution,
20mlsachet,25pcs
M10032B1382ppmTDSCalibrationSolution,
20mlsachet,25pcs.
M10038B6.44pptTDSCalibration,
20mlsachet,25pcs.
MA906012.88mS/cmcalibrationsolution,
230mlbottle
MA9061 1413μS/cmcalibrationsolution,
230mlbottle
MA9062 1382ppmTDSCalibration
Solution,230mLbottle
M10016BCleaningsolution,20mlsachet,
25pcs
M10000BRinsesolution,20mlsachet,
25pcs
M10004BpH4.01buffer,20mlsachet,
25pcs.
M10007BpH7.01buffer,20mlsachet,
25pcs.
M10010BpH10.01buffer,20mlsachet,
25pcs.
MA9004pH4.01buffer,230mlbottle
MA9007pH7.01buffer,230mlbottle
MA9010pH10.01buffer,230mlbottle
MA9006 pH6.86buffer,230mlbottle
MA9009 pH9.18buffer,230mlbottle
MA9015Electrodestoragesolution,230ml
ThesecondaryLCDdisplays“ESC”for1se-
condandthenthemeterreturnstothenormal
measurementmode.
To reset to the default values and clear a
previous calibration, press the SET/HOLD
button after entering the calibration mode
andbeforetherstpointisaccepted.
The secondary LCD displays “CLR” for 1
second,themeterresetstothedefaultcalib-
rationandtheCALtagontheLCDturnsoff.
If avalid buffervalueis detected,the REC
messageisdisplayedandthemetercomp-
letesthecalibrationprocedure.
TheLCDshowstheacceptedvaluewiththe
“OK2”messageandtheinstrumentreturns
to the normal measurement mode. Other-
wise,ifnovalidbufferisdetected,themeter
displaystheWRNGmessage.
PROBE REPLACEMENT
Removetheprotectivecapandunscrewthe
plasticringonthetopoftheprobe.
PullouttheMI60Pprobeandreplaceitwith
anewone.
Make sure the gaskets are in place before
screwingbackthering.
BATTERY REPLACEMENT
Whenthebatteriesbecomeweak,thebattery
symbolontheLCDwilllightuptoadvisethat
onlyafewhoursofworkingtimeisremaining.
ThemeterisalsoprovidedwithBEPS(Battery
Error Prevention System), which avoids any
erroneous readings due to low battery level
byautomaticallyswitchingthe meter off.It is
recommendedtoreplacethebatteriesimme-
diately.
To replace the batteries unscrew the battery
compartmentcapandreplaceallfour1.5Vbat-
terieswhilepayingattentiontotheirpolarity.
OPERATIONAL GUIDE
Removetheprobecapandturnthemeteron
bypressingtheON/OFF/CALbutton.Allthe
usedsegmentsontheLCDwillbevisiblefor
1secondoraslongasthebuttonispressed.
Immersetheprobeinthesolutiontobetes-
tedand select either pH, ECorTDS mode
withSET/HOLDbutton.
Stirgentlyandwaitforthereadingtostabilize,
i.e.thestabilityindicatorontheLCDturnsoff.
The pH and EC (or TDS) values are auto
-
maticallycompensatedfortemperatureand
will be displayed on the main LCD, while
thetemperatureisshownonthesecondary
LCD.
To freeze the display,whileinmeasurement
mode, press and hold the SET/HOLD but
-
ton. The “HOLD” message appears on the
secondary display and the reading will be
frozenontheLCD.
Press any button to return to normal
mode.
To turn the meter off,presstheON/OFF/CAL
button. The “OFF” message will appear on
thesecondarydisplay.Releasethebutton.
Notes:
Before taking any measurement, make sure
themeteriscalibrated(theCALtagison).
After use alwaysturn the meter OFF, rinse
theprobewithwaterandstoreitwiththepro-
tectivecap.
Measuring unit
on main LCD
Main LCD
Temperature reading
Secondary LCD
Low battery indicator
Automatic temperature
compensation
Calibration indicator
Stability indicator
SETUP
Setup mode allows the selection of tempera-
ture (°C or °F), calibration buffer set, in pH
modeonly,TDSconversionfactor(CONV)and
temperature coefcient (BETA), in EC mode
only.ToentertheSetup mode,presstheON/
OFF/CALbuttonuntil“CAL”onthesecondary
LCD is replaced by “TEMP” and the current
temperatureunit(e.g.TEMP°C).Then:
In EC and pH mode:
for °C/°F selection:
usetheSET/HOLDbutton,thenpresstheON/
OFF/CALbuttontogointhesettingsmenuand
toreturntothenormalmeasurementmode.
InECmodeonly:
to change the TDS factor value:
aftersettingthetemperatureunit,pressON/
OFF/CALbuttononcetoshowthecurrentval-
ue(e.g.0.50CONV).Selectthedesiredvalue
byusingtheSET/HOLDbutton,thenpressON/
OFF/CALbutton twice toreturnto the normal
measurementmode.
to change the temperature coefcient:
after setting the TDS factor, press ON/OFF/
CAL button to show the current value of the
temperaturecoefcientß(e.g.2.1BETA).Use
theSET/HOLDbuttontosetthedesiredvalue,
thenpressON/OFF/CALbuttontoreturntothe
normalmeasurementmode.
InpHmodeonly:
• to change the calibration buffer set:
after setting the temperature unit, press ON/
OFF/CALbuttononceandselectthebufferset
(“pH7.01BUFF”forpH4.01/7.01/10.01or“pH
6.86 BUFF” for NIST set 4.01/6.86/9.18) by
using the SET/HOLD button. Press ON/OFF/
CALbutton to return to the normal measure-
mentmode.
EC MEASUREMENT AND CALIBRATION
PROCEDURE
Select the EC or TDS mode with the SET/
HOLDbutton.
Submerge the electrode in the solution to be
tested. The measurements should be taken
whenthestabilityindicatoronthetopleftof
theLCDdisappears.
For better accuracy, frequent calibration of
the tester is recommended. Calibration is
alsonecessaryafterprobereplacement,after
testingaggressive chemicalsand where ex-
tremeaccuracyisrequired.FromnormalEC
operationmode,pressandholdtheON/OFF/
CALbuttonuntilthe“OFF”messageonthe
secondaryLCDissubstitutedby“CAL”.Re-
leasethebutton.
Immerse the probe in the proper calibration
solution: M10031 (1413 μS/cm) for MW803
andM10030(12.88mS/cm)forMW804.
Once thecalibrationhas beenautomatically
performed,theLCDwillshow“OK”for1sec-
ondandthemeterwillreturntonormalmea-
surementmode.
Sincethereisaknownrelationshipbetween
ECandTDSreadings,itisnotnecessaryto
calibratethemeterinTDSifitwaspreviously
calibratedinECmode.
pH MEASUREMENT AND CALIBRATION
SelectthepHmodewiththeSET/HOLD
button.Submergetheelectrodeinthesolution
tobetested.Themeasurementsshouldbeta-
kenwhenthestabilityindicatoronthetopleft
oftheLCDdisappears.
Forbetteraccuracy,frequentcalibrationofthe
tester is recomanded. Calibration is also ne-
cessaryafterelectrodereplacement,aftertes-
tingaggressivechemicalsandwhereextreme
accuracyisrequired.
From normal operation mode, press and hold
theON/OFF/CALbuttonuntiltheOFFmessage
onthesecondaryLCDissubstitutedby“CAL“.
Releasethebutton.
The instrument enters the calibration mode
by displaying “pH 7.01 USE“ (or ‘‘pH 6.86
USE“iftheNISTbuffersetwasselected).
For a single-point calibration, immerse the
electrodeinanybuffer,i.e.pH4.01,7.01(or
6.86),10.01(or9.18).
The meter activates the automatic buffer
recognition.Ifnovalidbufferisdetected,the
meterkeepstheUSEindicationactivefor12
seconds,andthenreplacesitwithWRNGin-
dicating that the sample being measured is
notavalidbuffer.Otherwise,ifavalidbuffer
is detected, then its value is shown on the
main display, and REC appears on the se-
condaryLCD.
IfthepH7.01(or6.86)wasused,pressthe
SETbuttontoexittheCalibrationmodeand
the “OK 1” message willappear on thedis-
play. The calibration point is stored and the
meterreturnstonormalmeasurementmode.
Forbetteraccuracy,itisalwaysrecommended
toperforma2-pointcalibration.
For a 2-point calibration, immerse the elec-
trodeinpH7.01(or6.86)buffersolution.
After the rst point has been accepted, the
meterwillthenaskforthesecondbufferand
themessage“pH4.01USE”willappear.
Rinsetheelectrodeandimmerseitinthese-
condsolution(pH4.01,10.01or9.18).
Notes:
When the calibration procedure is completed,
theCALtagisturnedon.
To quit the procedure and return to the last
calibrationdata,afterenteringthecalibration
modepresstheON/OFF/CALbutton.
/