B. Equilibrium Dialysis
The following example protocol is for monitoring the partitioning of a drug between plasma and tissue samples. Optimize or
modify this procedure as required for specific applications and analysis methods.
1. Determine the number of binding targets (such as tissue extracts) to be examined in a competition dialysis experiment
and the corresponding number of inserts and well size required (see Figure 2 in the Important Product Information
section for an example experimental setup).
2. Prepare 10-fold (w/v) tissue homogenates diluted in isotonic dilution buffer. Each dialysis chamber requires 200 μl of
extract. Duplicates can be performed in the same well or in a duplicate experiment in another well.
Note: If performing precipitation to prepare sample for analysis (see Procedure for Sample Analysis – For Unlabeled
Compounds, Step 5) prepare an additional 100 μl of extract.
3. For each test compound, prepare the required volume and concentration by mixing the test compound with plasma
diluted 10-fold with isotonic dilution buffer. The volume required is based on the base plate well size (Table 1). For
tissue binding studies, a drug concentration of 1 μM is common.
Note: If performing precipitation to prepare sample for analysis (see Procedure for Sample Analysis – For Unlabeled
Compounds, Step 4), 50 μl of diluted plasma free of test compound is required for each tissue and buffer sample.
4. Transfer the test compound/plasma mixture prepared in Step 3 to an appropriately sized well. Place lid onto the base plate.
5. Suspend the dialysis chambers (open end up) into the well of the base plate containing the test compound by placing
inserts into the appropriate slots of the base plate lid. To avoid damage, do not touch the dialysis membrane.
6. Place 200 μl of tissue extract, buffer or alternative binding target into a dialysis chamber, which is indicated by a red
ring. Record the appropriate location of each binding target.
7. Cover the unit with sealing tape and incubate at 37°C on an orbital shaker at 100-500 rpm.
Note: Incubating for 4 hours is generally sufficient to achieve equilibrium; however, actual time required may differ
depending on the test compounds and shaker used. For best results, perform a pilot experiment to empirically determine
the time required to reach equilibrium before processing samples. Time periods greater or less than 4 hours can be used;
however, excessively long incubations (i.e., ≥ 6 hours) may promote instability of extracts or test compounds, or result in
a volume change from hydrostatic pressure.
8. Remove equal volumes from each of the dialysis chambers and from the bottom well and place them in separate
microcentrifuge tubes or into a deep-well plate for analysis. Use an insert with a open side (lacking a dialysis membrane)
to directly access the bottom well.
9. Remove and discard used inserts and wash the base plate for reuse.
Note: The inserts can be easily removed with forceps or the Thermo Scientific RED Device Insert Removal Tool
(Product No. 89812), which enables quick removal of eight inserts at once. Alternatively the base plate lid can be
removed and inserts pushed up from the bottom.
Procedure for Sample Analysis
Determine the test compound concentration in the plasma, buffer and tissue samples to determine the in vitro partitioning
ratios (Kp). Some common analysis methods include LC/MS/MS, radioactivity and UV/visible/fluorescent spectrometry.
Modify the example protocols below as required.
• Procedure for radio-labeled compounds
1. Remove 50 μl of each sample and place into a scintillation vial appropriate for available instrumentation. Follow
manufacture’s recommendations for determining total radioactivity counts per sample.
2. Use the ratio of radio counts to determine the in vitro Kp value between the tissue and the plasma. The free drug fraction
can also be obtained from a ratio of the buffer and plasma.
sample plasma of counts radio sample tissue of counts radio
Kp =
100% x
sample plasma of counts radio sample buffer of counts radio
drug Free =
Pierce Biotechnology PO Box 117 (815) 968-0747 www.thermo.com/pierce
3747 N. Meridian Road Rockford, lL 61105 USA (815) 968-7316 fax
3