Taylor Technologies K-1680 Owner's manual

Type
Owner's manual

Taylor Technologies K-1680 is an essential tool for water testing and analysis. It comes with five different easy-to-use drop tests to determine the levels of five key water quality parameters: chlorine, pH, alkalinity, hardness, and stabilizer.

With the Taylor Technologies K-1680, you can quickly and easily:

  • Measure the chlorine concentration in your pool or spa to ensure proper sanitation.
  • Determine the pH level of your water to ensure it's within the ideal range for swimming and equipment protection.
  • Test the alkalinity of your water to maintain the pH balance and prevent corrosion.

Taylor Technologies K-1680 is an essential tool for water testing and analysis. It comes with five different easy-to-use drop tests to determine the levels of five key water quality parameters: chlorine, pH, alkalinity, hardness, and stabilizer.

With the Taylor Technologies K-1680, you can quickly and easily:

  • Measure the chlorine concentration in your pool or spa to ensure proper sanitation.
  • Determine the pH level of your water to ensure it's within the ideal range for swimming and equipment protection.
  • Test the alkalinity of your water to maintain the pH balance and prevent corrosion.
COMPONENTS:
1 x 5011 Instruction
1 x 9198R Sample Tube, Graduated (25 mL) w/ cap & red dot, plastic
1 x R-0819-C Ferroin Indicator, 2 oz, DB
2 x R-0820-C CAN Solution, 2 oz, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
Sodium Nitrite Test
1.Rinse and ll 25 mLsample tube (#9198R) to 5 mL mark with cooled (room
temperature)watertobetested(Fig.1).
2.Add4dropsR-0819FerroinIndicator.Swirltomix.Sampleshouldturnreddish
orange(Fig.2).
3.AddR-0820CANSolutiondropwise,swirlingandcountingaftereachdrop,until
colorchangesfromreddishorangetoblue(Fig.3).
4.MultiplydropsofR-0820CANSolutionby40.Recordaspartspermillion(ppm)
sodiumnitrite.
NOTE: Forresultsasnitrite,multiplysodiumnitriteconcentrationby0.67.
DROP TEST
SODIUM NITRITE (1 drop = 40 ppm)
Fig. 1
Fig. 2
Fig. 3
Instr. #5011
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 8/16
COMPONENTS:
1 x 5229C Instruction
1 x 9198G Sample Tube, Graduated (25 mL) w/ cap and green dot, plastic
1 x R-0638G-C Phenolphthalein Indicator, 2 oz w/ green cap, DB
1 x R-0645-C Total Alkalinity Indicator, 2 oz, DB
1 x R-0724-C Hydrochloric Acid .12N, 2 oz, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
P/T Alkalinity (HCI) Test
1.Rinseandll25mLsampletube(#9198G)to25mLmarkwithwatertobetested.
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
2.Add3dropsR-0638GPhenolphthaleinIndicator.Swirltomix.Samplewillturn
pink(Fig.1)ifPalkalinityispresent—proceedtoStep3.Ifsampleiscolorless,
proceedtoStep4.
3.AddR-0724HydrochloricAcid.12Ndropwise,swirlingandcountingaftereach
drop,untilcolorjustchangesfrompinktocolorless.RecorddropsasPreading.
DROP TEST
P/T ALKALINITY (HCl) (1 drop = 10 ppm)
4.Add5dropsR-0645TotalAlkalinityIndicator.Swirltomix.Samplewillturngreen
(Fig.2).
5.AddR-0724HydrochloricAcid.12Ndropwise,swirlingandcountingaftereach
drop,untilcolorchangesfromgreentored(Fig.3).Recordtotaldrops(Steps3
&5)asTreading.
6.Multiply P reading by 10. Record as parts per million (ppm) P alkalinity as
calcium carbonate. Multiply T reading by 10. Record as ppm T alkalinity as
calciumcarbonate.
NOTE: For 14.6 mL sample, record P reading as grains per gallon (gpg) P
alkalinityascalciumcarbonate.RecordTreadingasgpgTalkalinityas
calciumcarbonate.
Fig. 1
Fig. 2
Fig. 3
Instr. #5229C
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 8/16
COMPONENTS:
1 x 5212C Instruction
1 x 9198O Sample Tube, Graduated (25 mL) w/ cap & orange dot, plastic
1 x R-0630-C Chromate Indicator, 2 oz, DB
1 x R-0638O-A Phenolphthalein Indicator, .75 oz w/ orange cap, DB
1 x R-0706-C Silver Nitrate Reagent, 2 oz, DB
1 x R-0736O-C Sulfuric Acid .6N, 2 oz w/ orange cap, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
Chloride Test
NOTE: Whensultecontentofsamplewatertobetestedexceeds10ppm,the
sulteshouldbeoxidizedtopreventinterferenceintest.A25mLsample
isrstadjustedtotheappropriatepH,then1mL(or25drops)ofR-0649
Hydrogen Peroxide Solution (sold separately) is added and thoroughly
mixed.Continuewiththerestoftheprocedure.
1.Rinseandll25mLsampletube(#9198O)to25mLmarkwithwatertobetested.
2.Add2dropsR-0638OPhenolphthaleinIndicator.Ifsampleturnspink(Fig.1),add
R-0736OSulfuricAcid.6Ndropwise,swirlingaftereachdrop,untilcolorchanges
frompinktocolorless.Ifsampleiscolorless,proceedtoStep3.
DROP TEST
CHLORIDE (1 drop = 10 ppm)
3.Add5dropsR-0630ChromateIndicator.Swirltomix.Samplewillturnyellow
(Fig.2).
4.AddR-0706SilverNitrateReagentdropwise,swirlingandcountingaftereach
drop,untilcolorchangesfromyellowtoamilkysalmon(brick)red(Fig.3).
NOTE: DonotaddenoughR-0706SilverNitrateReagenttogiveabrowncolor.
Firstchangefromyellowtoamilkysalmon(brick)redistheendpoint.
5.MultiplydropsofR-0706SilverNitrateReagentby10.Recordaspartsper
million(ppm)chloride.
Fig. 1
Fig. 2
Fig. 3
Instr. #5212C
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 9/16
COMPONENTS:
1 x 5203 Instruction
1 x 9198W Sample Tube, Graduated (25 mL) w/ cap & white dot, plastic
1 x R-0638W-C Phenolphthalein Indicator, 2 oz w/ white cap, DB
1 x R-0699-C Iodide Iodate Reagent, 2 oz, DB
1 x R-0725-I Acid Starch Indicator Powder, 10 g
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
Sodium Sulfite Test
NOTE: Sample must be cooled to less than 100ºF (38ºC) to prevent high test
results. Sample must be protected from air contact while cooling to prevent
low test results.
1.Collectwatertobetestedinaclean,preferablylarge-mouthed,bottletooverowing.
Immediately cap and cool to room temperature.
2.Rinseandll25mLsampletube(#9198W)to25mLmarkwithcooled(room
temperature) water to be tested (Fig. 1).
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
DROP TEST
SODIUM SULFITE (1 drop = 10 ppm)
3.Add1dropR-0638WPhenolphthaleinIndicator.Swirltomix.Samplewillturn
pink(Fig.2).
4.AddR-0725AcidStarchIndicatorPowderadipperatatime,swirlingaftereach
dipper,until color changes frompink to colorless. Add 2 moredippers. Swirl
until dissolved.
5.AddR-0699IodideIodateReagentdropwise,swirlingandcountingaftereach
drop, until sample changes from colorless to a faint but permanent blue (Fig. 3).
6.Multiply drops of R-0699 Iodide Iodate Reagent by 10. Record as parts per
million(ppm)sodiumsulte.
NOTE: For14.6mLsample,recorddropsasgrainspergallon(gpg)sodiumsulte.
NOTE:Forresultsassulte,multiplysodiumsulteconcentrationby0.64.
Fig. 1
Fig. 2
Fig. 3
Instr. #5203
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 9/16
COMPONENTS:
1 x 5202 Instruction
1 x 9198B Sample Tube, Graduated (25 mL) w/ cap & blue dot, plastic
1 x R-0619B-C Hardness Buffer, 2 oz w/ blue cap, DB
1 x R-0620B-I Hardness Indicator Powder, 10 g w/ blue dot
1 x R-0683-C Hardness Reagent, 2 oz, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
Total Hardness Test
1.Rinseandll25mLsampletube(#9198B)to25mLmarkwithwatertobetested
(Fig.1).
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
2.Add5dropsR-0619BHardnessBuer.Swirltomix.
3.Add1dipperR-0620BHardnessIndicatorPowder.Swirluntildissolved.Sample
willturnred(Fig.2)ifhardnessispresent.
4.AddR-0683HardnessReagentdropwise,swirlingandcountingaftereachdrop,
untilcolorchangesfromredtoblue(Fig.3).
DROP TEST
TOTAL HARDNESS (1 drop = 10 ppm)
5.MultiplydropsofR-0683HardnessReagentby10.Recordaspartspermillion
(ppm)totalhardnessascalciumcarbonate.
NOTE: For14.6mLsample,recorddropsasgrainspergallon(gpg)totalhardness
ascalciumcarbonate.
Fig. 1
Fig. 2
Fig. 3
Instr. #5202
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 7/16
COMPONENTS:
1 x 5051 Instruction
1 x 9198P Sample Tube, Graduated (25 mL) w/ cap & purple dot, plastic
1 x 9315 Test Paper, pH, 1.8-3.8, 200 strips
1 x R-0686P-C Sulfuric Acid N, 2 oz w/ purple cap, DB
1 x R-0697-C Thiosulfate N/10, 2 oz, DB
1 x R-0802P-I Xylenol Orange Indicator Powder, 10 g
1 x R-0803-C Phosphonate Titrating Solution, 2 oz, DB
1 x R-0805-C Fluoride Masking Agent, 2 oz, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
NOTE: Iron can cause negative interference at a level greater than 5 ppm.
Orthophosphate and polyphosphate can cause positive interference at all
levels.
NOTE: Run a blank using tap water. Normal blank requires about 2 drops of R-0803
Phosphonate Titrating Solution to reach endpoint.
1.Rinseandll25mLsampletube(#9198P)to25mLmarkwithwatertobetested
(Fig.1).
2. Add:
1dropR-0697ThiosulfateN/10
10dropsR-0805FluorideMaskingAgent
1 level dipper R-0802P Xylenol Orange Indicator Powder
Swirl to mix.
DROP TEST
PHOSPHONATE EQUIVALENCE (PPM) ATMP (1.0); Na
5
ATMP (1.3); HEDP (0.9); K
6
HDTMP (1.2); DTPMP (1.45), Na
5
DTPMP (1.7)
3. Adjust pH between 2.6 and 3.0:
Add1dropR-0686PSulfuricAcidN.Swirltomix.Diptestpaper(#9315)into
sample,indirectionofarrow,for3seconds,withallcolorzonesimmersed.Match
indicatorzone(unnumberedsquarebetween2.7and3.0colorstandards)with
color scale. Read printed pH value. If necessary, continue adding R-0686P Sulfuric
Acid N dropwise, swirling and checking pH with a new test paper after each drop,
untilapHbetween2.6and3.0isobtained.Sampleshouldbeyellow(Fig.2).
4. Add R-0803 Phosphonate Titrating Solution dropwise, swirling and counting after
eachdrop,untilcolorchangesfromyellowtopurplishpink(Fig.3).
5. Subtract drops of R-0803 Phosphonate Titrating Solution in blank from drops in
sample(Step4).Multiplybyappropriateconversionfactor(seeCONVERSION
FACTORS).Recordaspartspermillion(ppm)phosphonate.
CONVERSION FACTORS:
To express phosphonate as: Multiply drops by:
Aminotri(methylenephosphonicacid)(ATMP) ........................... 1.0
Aminotri(methylenephosphonicacid),
pentasodiumsalt(Na
5
ATMP) ................................................. 1.3
1-Hydroxyethylidene-1,1-diphosphonicacid(HEDP) ............... 0.9
Hexamethylenediaminetetra(methylenephosphonicacid),
hexapotassiumsalt(K
6
HDTMP) ............................................ 1.2
Diethylenetriaminepenta(methylenephosphonicacid)
(DTPMP) ................................................................................. 1.45
Diethylenetriaminepenta(methylenephosphonicacid),
pentasodiumsalt(Na
5
DTPMP) .............................................. 1.7
Fig. 1
Fig. 2
Fig. 3
Instr. #5051
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 7/16
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6

Taylor Technologies K-1680 Owner's manual

Type
Owner's manual

Taylor Technologies K-1680 is an essential tool for water testing and analysis. It comes with five different easy-to-use drop tests to determine the levels of five key water quality parameters: chlorine, pH, alkalinity, hardness, and stabilizer.

With the Taylor Technologies K-1680, you can quickly and easily:

  • Measure the chlorine concentration in your pool or spa to ensure proper sanitation.
  • Determine the pH level of your water to ensure it's within the ideal range for swimming and equipment protection.
  • Test the alkalinity of your water to maintain the pH balance and prevent corrosion.