High-Pressure Generator Type 6906
Page 8 002-171e-10.01 (B16.6906e)
4 Operation
4.1 Start-up
1. Screw the generator to the workbench with the screw
clamps supplied.
2. Screw on the rotary lever.
3. Remove the stoppers, open the stop valve and move
the pressure piston against the limit stop by rotating the
spindle clockwise.
4. Add Type 1053 oil up to the mark in the sight glass.
Close the stop valve.
5. Screw the sensor to be calibrated into the adapter (e.g.
Type 69... / 79...) with the prescribed torque. Using a
syringe, fill the holes of the adapter with oil. Also fill the
adapter of the reference sensor with oil. Screw in the
two adapters and tighten them lightly by hand.
6. Open the stop valve. Suck in oil by rotating the spindle
back to the limit stop. Close the stop valve.
7. Loosen one adapter slightly and rotate the spindle to
vent the pressure around the hole. Small air bubbles are
expelled at the adapter thread. Continue until only oil
appears. Open the stop valve and tighten the adapter
with a torque of 10 Nm. Then bleed the air from the
second adapter.
8. The pressure generator is now ready to operate.
Attention:
In order to prevent an OVERLOAD on the built-in sensors,
ensure that the charge amplifiers and evaluation units con-
nected are correctly connected, adjusted and switched to
OPERATE.
• When operating with 6907B1:
Increase the pressure only until the maximum value se-
lected in the Type 6907B... calibrator. is reached.
• Pressure sensors with a pressure range <500bar must
not be calibrated with the pressure generator, since the
danger of an OVERLOAD is too great.
4.2 Safety note
The pressure generator can generate extremely high pres-
sures. It must therefore be operated only with the protec-
tive hood closed.