1: Introduction __________________________________________________________________1
1.1 Intended Use............................................................................................................................................................ 1
1.2 Intended Use for the User Guide .......................................................................................................................... 1
1.3 Product Complaints ................................................................................................................................................ 1
1.4 Technical Support ................................................................................................................................................... 1
1.5 Quality Control ....................................................................................................................................................... 1
1.6 Installation Instructions ......................................................................................................................................... 1
1.7 User Profiles............................................................................................................................................................. 2
1.7.1 Mammography Technologist ..................................................................................................................... 2
1.7.2 Radiologists, Surgeons ................................................................................................................................ 2
1.7.3 Medical Physicist ......................................................................................................................................... 2
1.8 Training Requirements........................................................................................................................................... 3
1.9 Terms and Definitions ............................................................................................................................................ 3
1.10 International Symbols ............................................................................................................................................ 4
1.11 Warnings, Cautions, and Notes ............................................................................................................................ 5
2: General Information ___________________________________________________________7
2.1 System Description ................................................................................................................................................. 7
2.2 How to Handle the Biopsy Guidance Module.................................................................................................... 8
2.3 Biopsy Guidance Module Components ............................................................................................................... 9
2.4 Biopsy Control Module Components ................................................................................................................ 10
2.5 Safety ...................................................................................................................................................................... 11
2.6 Warnings and Precautions ................................................................................................................................... 11
2.7 Compliance Information ...................................................................................................................................... 13
2.7.1 Requirements .............................................................................................................................................. 13
2.7.2 Label Locations........................................................................................................................................... 13
3: How to Install or Remove the System ___________________________________________15
3.1 Installation of Components ................................................................................................................................. 15
3.1.1 How to Attach the Biopsy Guidance Module ........................................................................................ 15
3.1.2 How to Attach the Biopsy Control Module ........................................................................................... 16
3.2 Installation and Removal of Accessories ........................................................................................................... 18
3.2.1 Biopsy Compression Paddles ................................................................................................................... 18
3.2.2 Biopsy Device Holder ................................................................................................................................ 18
3.2.3 Needle Guides ............................................................................................................................................ 19
3.2.4 Affirm Table-top Stand ............................................................................................................................. 20
3.3 Removal of Main Components ........................................................................................................................... 21
3.3.1 Biopsy Control Module ............................................................................................................................. 21
3.3.2 Biopsy Guidance Module ......................................................................................................................... 21
Affirm User Guide
Table of Contents
vi MAN-03209 Revision 005
4: How to Use the System _______________________________________________________23
4.1 System Verifications ............................................................................................................................................. 23
4.1.1 Confirm the Host Connection .................................................................................................................. 23
4.1.2 The QAS Test .............................................................................................................................................. 23
4.2 Biopsy Control Module Screens .......................................................................................................................... 26
4.2.1 Home Screen ............................................................................................................................................... 26
4.2.2 Target Guidance Screen ............................................................................................................................ 26
4.2.3 Jog Mode Screen ......................................................................................................................................... 30
4.2.4 Select Target Screen ................................................................................................................................... 32
4.3 Selenia Dimensions ............................................................................................................................................... 33
4.3.1 Biopsy Views .............................................................................................................................................. 33
4.3.2 C-Arm Stereo Modes ................................................................................................................................. 35
4.3.3 Biopsy Tab ................................................................................................................................................... 38
4.3.4 Stereotactic Lesion Targeting ................................................................................................................... 41
4.3.5 Lesion Targeting with Tomosynthesis .................................................................................................... 42
4.3.6 Post Biopsy .................................................................................................................................................. 43
4.3.7 Printing Stereo Images .............................................................................................................................. 43
5: Care and Cleaning____________________________________________________________45
5.1 General Information ............................................................................................................................................. 45
5.2 Preventive Maintenance Schedule ...................................................................................................................... 45
5.3 For General Cleaning ............................................................................................................................................ 46
5.3.1 How to Clean the Biopsy Control Module Screen ................................................................................. 46
5.3.2 To Prevent Possible Injury or Equipment Damage ............................................................................... 47
5.4 Geometry Calibration ........................................................................................................................................... 47
5.4.1 Geometry Calibration Procedure ............................................................................................................. 47
6: Troubleshooting _____________________________________________________________49
6.1 Audible Alerts ....................................................................................................................................................... 49
6.2 Error Messages ...................................................................................................................................................... 50
Appendix A: System Specifications ______________________________________________51
A.1 Affirm Measurements .......................................................................................................................................... 51
A.2 Biopsy Guidance Module .................................................................................................................................... 51
A.3 Biopsy Control Module ........................................................................................................................................ 51
Appendix B: Forms _____________________________________________________________53
B.1 QAS Test Checklist ............................................................................................................................................... 53
Appendix C: CNR Correction for Biopsy __________________________________________55
C.1 CNR Correction for Stereotactic Biopsy ............................................................................................................. 55
C.1.1 AEC Table 0 (Standard Stereotactic Biopsy Dose) ................................................................................. 55
C.2 CNR Correction for Biopsy Under Tomosynthesis Option............................................................................. 55
C.2.1 AEC Table 0 (Tomosynthesis Option: Standard Tomo Dose) .............................................................. 55
Index _________________________________________________________________________57
Affirm User Guide
Table of Contents
MAN-03209 Revision 005 vii
List of Figures
Figure 1: Affirm on the Selenia Dimensions .................................................................................................................. 7
Figure 2: How to Lift the Biopsy Guidance Module ..................................................................................................... 8
Figure 3: How to Store the Biopsy Guidance Module .................................................................................................. 8
Figure 4: Biopsy Guidance Module ................................................................................................................................. 9
Figure 5: Biopsy Control Module .................................................................................................................................. 10
Figure 6: Label Location .................................................................................................................................................. 13
Figure 7: Installation of the Biopsy Guidance Module ............................................................................................... 15
Figure 8: Attachment of the Biopsy Control Module .................................................................................................. 16
Figure 9: How to Install the Needle Guides ................................................................................................................. 19
Figure 10: Affirm Table-top Stand ................................................................................................................................. 20
Figure 11: Home Screen on the Biopsy Control Module ............................................................................................ 23
Figure 12: Admin Screen ................................................................................................................................................. 24
Figure 13: Device Field in the Biopsy Tab .................................................................................................................... 24
Figure 14: The Home Screen........................................................................................................................................... 26
Figure 15: Target Guidance Screen ................................................................................................................................ 27
Figure 16: Green Differential Cells ................................................................................................................................ 28
Figure 17: Yellow and Red Cells .................................................................................................................................... 28
Figure 18: Alert Sounds are Audible ............................................................................................................................. 29
Figure 19: Alert Sounds are Muted ............................................................................................................................... 29
Figure 20: Jog Mode Screen ............................................................................................................................................ 31
Figure 21: Select Target Screen ....................................................................................................................................... 32
Figure 22: Add a Biopsy View Screen ........................................................................................................................... 33
Figure 23: Edit Biopsy View Screen............................................................................................................................... 34
Figure 24: Stereo Mode ................................................................................................................................................... 36
Figure 25: C-Arm Mode .................................................................................................................................................. 36
Figure 26: System Status Icon ......................................................................................................................................... 37
Figure 27: System Status Menu ...................................................................................................................................... 37
Figure 28: System Defaults Screen ................................................................................................................................. 37
Figure 29: The Biopsy Tab .............................................................................................................................................. 38
Figure 30: Slice Indicator................................................................................................................................................. 38
Figure 31: Function Buttons and Data on the Biopsy Tab .......................................................................................... 39
Figure 32: Stereo Pair Print Screen ................................................................................................................................ 43
Affirm User Guide
Table of Contents
MAN-03209 Revision 005 ix
List of Tables
Table 1: Components of the Biopsy Guidance Module ................................................................................................ 9
Table 2: Components of the Biopsy Control Module.................................................................................................. 10
Table 3: How to Use the Sound Button ........................................................................................................................ 29
Table 4: The C-Arm Stereo Mode Button ..................................................................................................................... 36
Table 5: How to Select the C-Arm Stereo Mode .......................................................................................................... 37
Table 6: Radiologic Technologist Preventive Maintenance Schedule....................................................................... 45
Table 7: Affirm Audible Alerts ...................................................................................................................................... 49
Table 8: Affirm Error Messages ..................................................................................................................................... 50
Affirm User Guide
Chapter 1: Introduction
MAN-03209 Revision 005 Page 1
1: Introduction
1.1 Intended Use
United States federal law restricts this device to use by, or on the order of, a physician.
The Affirm™ breast biopsy guidance system is an optional accessory for the Selenia®
Dimensions® digital mammography system. Its function is to localize lesions accurately
in the breast in three dimensions. It is intended to provide targeting guidance for
interventional procedures such as biopsy, presurgical localization, or treatment devices.
1.2 Intended Use for the User Guide
Always refer to the User Guide for instructions on using the system.
1.3 Product Complaints
Report any complaints or problem in the quality, reliability, safety, or performance of this
product to Hologic. If the device has caused or added to patient injury, immediately
report the incident to Hologic. (See the title page for contact information.)
1.4 Technical Support
Refer to the title page of this manual for contact information for product support.
1.5 Quality Control
Facilities that are ACR accredited must follow the 1999 Stereotactic Breast Biopsy Quality
Control Manual. Facilities that are not ACR accredited can follow the above manual or
perform the QAS Test described in this manual at the required interval.
1.6 Installation Instructions
Installation instructions are available in the Service Manual.
Chapter 1:
Affirm User Guide
Chapter 1: Introduction
Page 2 MAN-03209 Revision 005
1.7 User Profiles
1.7.1 Mammography Technologist
• Meets all requirements that apply to the location in which the Mammography
Technologist operates.
• Completed training on the mammography system.
• Has training in mammography positions.
• Knows about Stereotactic breast biopsy procedures.
• Knows how to operate a computer and its peripherals.
• Can lift 20 pounds to shoulder height with two hands (necessary for upright
stereotactic systems).
• Understands sterile procedures.
1.7.2 Radiologists, Surgeons
• Meets all requirements that apply to the location in which the Physician operates.
• Knows about Stereotactic breast biopsy procedures.
• Knows how to operate a computer and its peripherals.
• Understands sterile procedures.
• Gives local anesthesia.
• Knows about basic surgical procedures for core biopsy.
1.7.3 Medical Physicist
• Meets all requirements that apply to the location in which the Medical Physicist
operates.
• Knows about mammography.
• Has experience with digital imaging.
• Knows how to operate a computer and its peripherals.
Affirm User Guide
Chapter 1: Introduction
MAN-03209 Revision 005 Page 3
1.8 Training Requirements
In the United States, users must be Registered Radiologic Technologists meeting criteria
to perform mammography. The mammography users must meet all applicable MQSA
personnel requirements under FDA guidelines for conventional and digital
mammography.
The user has options available for training, which include but are not limited to:
• Onsite applications training by a Hologic Clinical Services Specialist
• Onsite on the job training also known as peer training
Additionally, the user manual is a guide for directions on how to use the system.
All users must make sure that they receive training on correct operation of the system
before use on patients.
Hologic does not accept the responsibility for injury or damage from wrong system
operation.
1.9 Terms and Definitions
Affirm
The breast biopsy guidance system for the Selenia Dimensions
Biopsy Control
Module
The user control device for the Affirm breast biopsy guidance
system
Biopsy Guidance
Module
Holds and operates the biopsy device. Responds to commands
from the Biopsy Control Module to move the device into position
and do the biopsy.
Comm
Communication
Diff
Differential
C-Arm Mode
Lets the C-arm and Tube Arm to move together to the Needle
Approach angle for the Localization procedure.
Exposure
Technique
Combination of x ray parameters (kVp, mAs, filter) for an acquired
image
Needle
Approach Angle
The angle of incidence of the needle to the breast
QAS
Quality Assurance Standard
Safety Margins The minimum space allowed between the biopsy device needle
that is installed and components of the Selenia Dimensions system.
Stereo Mode
Lets the Tube Arm rotate for acquisition of stereotactic images
while the C-arm stays in position.
Stereotactic
Procedure
A type of examination that allows stereotactic views at the
Acquisition Workstation.
Stereotactic View
A specialized image view that causes the application to capture a
stereotactic images.
Stroke Margin
The safety margin (in mm) that remains between the fired needle
position and the breast platform.
Affirm User Guide
Chapter 1: Introduction
Page 4 MAN-03209 Revision 005
Tomosynthesis
An imaging procedure which combines a number of projections
taken at different angles. The tomosynthesis images can be
reconstructed to show focal planes (slices) within the object
(Tomosynthesis option).
View
The combination of one x-ray image and a specified set of
conditions for image acquisition.
X-axis • The (lateral) axis from left to right across the biopsy
window
Y-axis • The (longitudinal) axis from front to back above the biopsy
window
Z-axis • The (vertical) axis through the biopsy window
1.10 International Symbols
This section describes the International Symbols on this system.
Potential Equalization terminal
Protective Earth terminal
"On" and "Off" (power) for the computer and display.
Discard electrical and electronic equipment separately from standard waste.
Send decommissioned material to Hologic or contact your service
representative.
Manufacturer
Date of Manufacture
Affirm User Guide
Chapter 1: Introduction
MAN-03209 Revision 005 Page 5
1.11 Warnings, Cautions, and Notes
Descriptions of Warnings, Cautions, and Notes used in this manual:
WARNING! The procedures that you must follow accurately
to prevent possible dangerous or fatal injury.
Warning: The procedures that you must follow accurately to
prevent injury.
Caution: The procedures that you must follow accurately to prevent
the damage to equipment, loss of data, or damage to files
in software applications.
Note Notes show additional information.
Affirm User Guide
Chapter 2: General Information
MAN-03209 Revision 005 Page 7
2: General Information
2.1 System Description
The Affirm attaches to the Selenia Dimensions. A biopsy device attaches to the Affirm. X-
and Y-axes motors in the Affirm move the biopsy device left or right and forward or back.
Z-axis movement is manual. The Affirm Biopsy system has two main components:
• Biopsy Guidance Module
• Biopsy Control Module
The Tube Arm on the Selenia Dimensions moves separately from the Compression Arm
to allow the acquisition of stereotactic images for the procedure. Refer to the Selenia
Dimensions User Guide for complete information about that system.
Affirm licensing displays on the Acquisition Workstation screen as "Stereo Licensed".
Refer to Licensing Setup in System Tools of the Operating System.
Figure 1: Affirm on the Selenia Dimensions
Figure Legend
1. Biopsy Guidance Module
2. Biopsy Control Module
Chapter 2:
Affirm User Guide
Chapter 2: General Information
Page 8 MAN-03209 Revision 005
2.2 How to Handle the Biopsy Guidance Module
Caution: To prevent damage or alignment problems with the
Needle Guidance Stage, be careful when you move the
Biopsy Guidance Module.
Caution: The Affirm Biopsy Guidance Module weighs 15 pounds.
When you move it, be sure to have a secure grip on the
handles.
• Only lift the Biopsy Guidance Module with the handles.
Figure 2: How to Lift the Biopsy Guidance Module
• When the Biopsy Guidance Module is not in use, put the device on its back.
Figure 3: How to Store the Biopsy Guidance Module
Affirm User Guide
Chapter 2: General Information
MAN-03209 Revision 005 Page 9
2.3 Biopsy Guidance Module Components
The Biopsy Guidance Module installs on the front of the Selenia Dimensions. A lock lever
(item 8) secures this module in position. A cable (item 7) connects to the Selenia
Dimensions for operation of the biopsy guidance system.
Figure 4: Biopsy Guidance Module
Table 1: Components of the Biopsy Guidance Module
# Name Description
1
Two on each side hold the Biopsy Guidance Module on the Selenia
Dimensions gantry.
2
One on each side. Hold both handles to lift the Biopsy Guidance
Module.
3
Rotate either knob to move the biopsy device along the Z-axis.
4
Holds the biopsy device holder and provides the track for Z-axis
movements.
5
Attaches to the Needle Guide Mount on the Biopsy Device Holder.
6
Holds the biopsy device. Moves along the Z-axis Slide Rail when a
Z-axis Control Knob is rotated.
7
Connects to Selenia Dimension to bring power to the Affirm.
8
One on each side. Engage both levers to lock the Biopsy Guidance
Module in position and on the Selenia Dimensions gantry.
9 Receptacle Accepts the cable from the Biopsy Control Module.
Affirm User Guide
Chapter 2: General Information
Page 10 MAN-03209 Revision 005
2.4 Biopsy Control Module Components
The Biopsy Control Module attaches to either the left or right handle on the Biopsy
Guidance Module with a bracket (item 5). The display screen (item 2) is a touch screen for
the user to perform the desired tasks. Motor Enable buttons (item 3) on either side of this
module (and at the rear) activate motorized movement of the biopsy device.
Figure 5: Biopsy Control Module
Table 2: Components of the Biopsy Control Module
# Name Description
1
Cable
Connects to Biopsy Guidance Module.
2
Display Screen
Shows targets, system status, name of the biopsy device, and safety
margins. Touch buttons allow option selection.
3
Motor Enable Buttons
Front and back button pairs on either side of the display. Press the
front and back buttons of either side at the same time to activate a
motor movement.
4
Articulating Arm Lock
and Release
Rotate to release the lock and adjust the module. Rotate in the
opposite direction to lock the arm and hold the module in the new
position.
5
Attachment Bracket
Attaches to either handle of the Biopsy Guidance Module.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70

Hologic Affirm breast biopsy guidance system 1.7 User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI