MAINTENANCE
Replacing the battery
When replacing battery, turn the power switch to the OFF position. Remove the battery before long-
term storage. See Specifications for replacement type. If unit fails to operate after replacing the
battery, turn power switch to OFF position, remove battery, press and release the MEASURE
/HOLD button, then reconnect the battery.
Cleaning
To obtain accurate, drift-free readings, the sensor plate and especially the area around the aperture
must be kept absolutely clean at all times. Be sure to always keep the instrument in its protective case
when not actually taking measurements. Never touch the aperture with anything – not even cotton
swabs. To remove dust or other particulate matter use low-pressure instrument-grade air. To remove
severe contamination spray or flush with the smallest practical amounts of clean technical-grade
isopropyl alcohol. Then stand the instrument on end and allow it to dry for several hours. DO NOT
clean with any solvent on a regular basis.
CALIBRATION
The ACL 450B Ionized Environment Electrostatic Locator is factory-calibrated prior to shipment. Recalibration
should be performed annually, or more frequently if specified by contract or company policy. Your instrument
should also be recalibrated any time it has been repaired or tampered with. ACL will recalibrate your instrument
at a reasonable cost or provide information and procedures on calibration upon request. We also offer calibration
services with traceability to NIST (level 1) and with before and after data (level 2).
WARRANTY
ACL, Inc. warrants to the owners, each instrument and sub-assembly manufactured by them to be free from
defects in material and workmanship for a period of one year after shipment from factory. This warranty is
applicable to the original purchaser only. Liability under this warranty is limited to service, adjustment or
replacement of defective parts (other than fuses or batteries) on any instrument or sub-assembly returned to the
factory for this purpose, transportation charges prepaid. This warranty does not apply to instruments or sub-
assemblies subjected to abuse, abnormal operating conditions, or unauthorized repair or modification. Since ACL,
Inc. has no control over conditions of use; no warranty is made, or implied as to the suitability of our product for
the customer's intended use.
RETURN POLICY TO FACTORY
Materials returned to ACL must have a Return Material Authorization number. To obtain a RMA
number, contact our meter department at Customers have 30 days to determine that the product ordered fills
their need and performs as described in ACL literature. Units returned for approved repair or credit must be in the
original packaging including all parts and paperwork plus be in very good physical condition. If not, the customer is
billed the cost to refurbish the unit and for missing accessories and merchandise. No products may be returned for
exchange or credit after 12 months of the shipment date. ACL, Inc. reserves the right to repair
or replace units under warranty.