Fluke 719 Pressure Calibrator, 714 Thermocouple Calibrator, 715 Volt/mA Loop Calibrator, 717 Pressure Calibrator, 718 Pressure Calibrator User manual

  • Hello! I am an AI chatbot trained to assist you with the Fluke 719 Pressure Calibrator User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Manual Supplement
© 2012-2020 Fluke Corporation. All rights reserved. x
Manual Title: 71X Series Calibration Supplement Issue: 1
Part Number: 686540 Issue Date: 2/20
Print Date: January 1998 Page Count: 1
Revision/Date: 12, 9/19
This supplement contains information necessary to ensure the
accuracy of the above manual.
71X Series Calibration Manual Supplement
2/20 1
Change #1, 34
On page 18, in Table 19, under 718 Version 4.0 or Higher, 718 Ex V 2.0 and Higher:
Change the 4
th
column heading
From: Accuracy
[1]
Pressure accuracy is ± % of positive range
To: Accuracy
[1]
Pressure Accuracy is ± % of Range
On page 20, in Table 27, under the Accuracy
[1]
column:
Change
From: ±0.025 % of positive Range (6 months calibration)
±0.035 % of positive Range (1-year calibration)
To: ±0.025 % of Range (6 months calibration)
±0.035 % of Range (1-year calibration)
/