Specification
Nominal speed: 0-11000 RPM
*
Maximum G-force: 8,200 g
Capacity: 4capillary tubes
Max. Power: 12 volt DC
Height: 11.4cm (4.5 inches)
Width: 15cm (6 inches)
Depth: 20cm (8 inches)
Weight: 1.58kg (3.5 lb.)
Boxed dimensions: 21.59 x 26.67 x 15cm
(8.5”) x (10.5”) x (6”) in
Boxed weight: 2.73kg (6 lbs.)
Average Speed:* These are average speeds, which may vary
slightly between units and with load sizes. Please use a
tachometer for actual RPMs on your unit and for calibration
purposes.
Warranty
Your Clinispin Micro centrifuge is covered by a
limited 1-year warranty on parts and labour.If the
centrifuge or any of its parts should fail during this
period, simply contact Woodley Equipment Company
Ltd. Any attempts to repair the centrifuge or replace
any of its parts without authorisation from Woodley
Equipment Company Ltd may void the warranty.
Instructions for use
1Spin only balanced loads. Make sure that capillary tubes are
placed opposite each other.
2Insert capillary tube(s) and close lid.
3 Set timer for 5 minutes - the unit will immediately begin spinning
atfull speed (11,000 rpm).
4 Do not open the lid at any time whilst the centrifuge is still
spinning - wait for the timer to stop and the audible alarm to
sound.
5 After the unit has stopped, carefully remove the capillary tubes.
NB - Wait until the rotor has stopped completely before
opening the lid.
Reading your Haematocrit results:
1 Place the reader card directly behind the tube and align the top
and bottom of the fluid (not the tube or sealant) with the 0 and
100 lines (see fig. 5).
2 The separation line between the red cells and plasma should be
read on the graph at the point of separation.
3 NB - The separation line will be at a slight angle, therefore you
must read the midpoint of the angle for an accurate percentage.
4 0 - 100 on the reader indicates a percentage (%) result. To express
the result in SI units multiply the % value by 0.01, eg 45% = 0.45
5 SAFETY - Used capillary tubes must be disposed of as clinical
waste in a sharps container.
Care & Maintenance
With proper care and maintenance the clinispin HCT centrifuge will
provide years of service.
A) Clean the inside of the centrifuge. Whenever there is a tube
breakage or spillage, it is recommended that the inside of the
centrifuge is cleaned, and if possible disinfected.
(Do not immerse the electrical components in any liquid).
B) The Clinispin HCT utilises a maintenance-free
brushless motor and its bearings are permanently
lubricated; it should not need servicing for the life of
the centrifuge. Likewise, the electrical components
weredesigned for high reliability and should not
need regular service.