2 Installation Instructions and Safety Information Installation Instructions and Safety Information 3
Contents Warnings and Safety
Intended Use
Signum Models with Option Y2
Signum scales meet the highest require-
ments placed on the accuracy and
reliability of weighing results. Equipped
with Option Y2, Signum models are
suitable for use as category II 3 GD
electrical apparatus in hazardous areas
designated zone 2 (gases) or 22 (dusts).
Note:
Read these installation instructions
and safety information carefully before
connecting the Signum equipment to
power and putting it into operation.
Technical Advice on Applications/
Hotline
Phone: +49.40.67960444
Fax: +49.40.67960474
E-mail: technical.support.hh
@minebea-intec.com
Contents
General View of the Equipment ........... 2
Intended Use ......................................... 3
Warnings and Safety Information ....... 3
Installation Instructions ....................... 4
Getting Started ...................................... 5
Ground (Earth) ...................................... 6
Connecting the Scale to AC Power ...... 6
Care and Maintenance .......................... 7
Recycling ............................................... 8
Declaration of Conformity ................... 9
Safety instructions ............................... 12
Symbols
The following symbols are used in these
instructions:
§ indicates required steps
$ indicates steps required only under
certain conditions
> describes what happens after you have
performed a particular step
indicates an item in a list
! indicates a hazard
English
In cases involving questions of
interpretation, the German-language
version shall prevail.
Installation Instructions and Safety Information 3
Contents Warnings and Safety
Warnings and Safety
Precautions
Note:
Improper use or handling can result in
damage and/or injury. The scale may be
installed and operated by qualified per-
sonnel only. Make sure you observe the
warnings and safety precautions in their
entirety during installation and operation,
as well as while performing maintenance
and repair work on the equipment. All
laws, regulations, standards, accident
prevention rules and environmental pro-
tection requirements applicable in your
country must be followed. It is important
that all personnel using the equipment
understand these warnings and safety pre-
cautions, and have access to the relevant
documents at all times. Furthermore, the
warnings and safety precautions sup-
plied with any electrical equipment con-
nected, such as peripheral devices, must
be observed as well. These warnings and
safety precautions may have to be sup-
plemented by the equipment operator. All
operating personnel must be informed of
any additions to these instructions. Make
sure the equipment is accessible at all
times.
Make absolutely sure to disconnect the
scale from power (unplug the power
cord) before you connect or disconnect a
peripheral device (printer or PC) to or from
the interface port).
If you use cables purchased from another
manufacturer, check the pin assignments
in the cable against those specified by
Minebea Intec before connecting the
cable to Minebea Intec equipment, and
disconnect any wires that are assigned
differently. The operator shall be solely
responsible for any damage or injuries that
occur when using cables not supplied by
Minebea Intec.
General Provisions for Installing Signum
Models Equipped with Option Y2
Signum models equipped with Option Y2
meet the requirements of EC Directive
2014/34/EU for group II devices and are
suitable for use in zone 2 and 22 hazard-
ous areas in accordance with the declara-
tion of conformity. Furthermore, Signum
models meet the EC Directives for electro-
magnetic compatibility (see “Documents”).
The area of use for which Signum models
are intended is defined in the documents
supplied. Make sure the restrictions defined
in those documents are observed.
Operation outside of these restrictions is
not permitted, and is considered use of
the equipment for other than its intended
purpose.
Note:
Make sure the safety instructions described
in the drawing (for details, see “Docu-
ments:” Drawing no. 36287-740-16) are
followed.
The device is protected against penetra-
tion by solid objects (IP65 protection on
all models that have Option Y2).
The protection rating for the display and
control unit is ensured only if the rubber
gasket is installed and all connections are
fastened securely (including the caps on
unused sockets).
Weighing platforms must be installed, and
the weighing system inspected, by
a certified electrical technician.
If you install an interface port or battery
connector after setting up your Signum,
keep the protective cap(s) in a safe place
to be used for protecting the interface
port when not in use, or prior to shipment.
This will protect the data interface or from
vapors, moisture and dust or dirt. Do not
leave interface ports uncovered.
4 Installation Instructions and Safety Information Installation Instructions and Safety Information 5
Installation Instructions Getting Started
Installation Instructions
Setting up the Scale
Choose a suitable location where your scale will
not be exposed to drafts, heat radiation, moisture
or vibrations. Use a spirit level when adjusting the
weighing platform. Make sure to read these operat-
ing instructions carefully before connecting the
scale to power.
! Observe all warnings and safety precautions.
Using the Equipment in Legal Metrology:
If the scale is to be verified, make sure to observe
the applicable regulations regarding verification.
If any of the verification seals are damaged, make
sure to observe the national regulations and stand-
ards applicable in your country in such cases.
In some countries, the equipment must be
re-verified.
§ The air bubble must be centered within the circle on
the level indicator.
°C
Installation Instructions and Safety Information 5
Installation Instructions Getting Started
Getting Started
§ Use the connecting cable from the weighing
platform to connect the display and control unit.
Note:
The cable gland is installed at the factory.
Please use extreme caution when performing any
work on the equipment that affects this cable gland.
Use a torque wrench
and tighten the cable gland to 5 Nm.
§ Prepare and install the cable as follows:
Guide the cable through the cable gland.
Close and tighten the cable gland in accordance
with the applicable standards.
Remove the casing from a section of the cable end
(see illustration). The shield (1) must have contact
with the clamps (2).
Expose approximately 15 cm (4 inches) of the wires
(3) for connection to the terminals.
Guide the cable through the cable gland.
It is important to make sure that the shield is in con-
tact with the clamps, because the cable is grounded
by the shield.
6 Installation Instructions and Safety Information Installation Instructions and Safety Information 7
Getting Started Care and Maintenance
Grounding
For models with analog platform (option A15) and
internal battery pack (option L9) only
§ Ground the weighing platform.
Securely connect the cable to the grounding terminal.
Assembly:
Toothed lock washer, cable, toothed lock washer
Screw, or nut with locknut
§ Additionally, secure the cable with a clip
(strain relief).
For SIW..DCP models:
§ Ground the weighing platform with the column.
Securely connect the cable to the grounding terminal.
Assembly:
Toothed lock washer, cable, toothed lock washer
Screw, or nut with locknut
§ Additionally, secure the cable with a clip
(strain relief).
For SIWS..BBP models:
§ Ground the indicator.
Connect the cable to the grounding terminal.
Assembly:
Toothed lock washer, cable, toothed lock washer
– Nut
– Locknut
Power Connection
! Make sure that the voltage rating printed on this
unit matches the voltage at the place of installation.
If the voltage specified on the label does not match
the rating or standard you use, please contact your
local Minebea Intec office or dealer.
! Make sure to observe the safety instructions.
§ Connect the equipment to AC power.
Installation Instructions and Safety Information 7
Getting Started Care and Maintenance
Care and Maintenance
Cleaning
!Never use concentrated acids, alkali
solutions, solvents or pure alcohol to
clean the equipment.
$ Do not allow liquids to penetrate the
scale housing.
$ Use a brush or a soft, dry, lint-free cloth
to clean the scale.
Storage and Shipping Conditions
$ The packaging used for shipping
your Minebea Intec equipment is opti-
mally designed to prevent damage dur-
ing transport. It is a good idea to save
the box and all parts of the packaging
for future storage or shipment of the
equipment. Only the original packaging
provides the best protection for
shipment.
$ Allowable storage temperature:
–20°C to +75°C (–4°F to +167°F)
$ Allowable humidity during storage:
max. 90%
$ Please see also the information under
“Safety Inspection”, below.
Safety Inspection
If there is any indication that safe
operation is no longer warranted;
for example:
If there is visible damage to the
connecting cable,
If the device no longer functions
properly,
If the equipment has been stored for a
relatively long period under unfavorable
conditions,
If the equipment has been subjected to
rough handling during shipment,
§ make sure all warnings and safety pre-
cautions are observed, and notify your
nearest Minebea Intec Service Center
or the International Technical Support
Unit based in Goettingen, Germany.
Maintenance and repair work may be
performed only by authorized
Minebea Intec service technicians who
have access to the required maintenance
manuals and have received the neces-
sary training.
!The seals affixed to this equipment
indicate that only authorized service
technicians are allowed to open the
equipment and perform maintenance
work so that safe and trouble-free oper-
ation of the equipment is ensured and
the warranty remains in effect.
8 Installation Instructions and Safety Information Installation Instructions and Safety Information 9
Recycling
Recycling
In Germany and many other countries
(see www.minebea-intec.com > Ser-
vices > Downloads for details), Minebea
Intec or the organization contracted by
us takes care of the return and legally
compliant disposal of its electrical and
electronic equipment on its own.
In countries that are not members
of the European Economic Area (EEA)
or where no Minebea Intec subsidiaries
or dealerships are located, please con-
tact your local authorities or a commer-
cial disposal operator.
These products may not be placed
with the household waste or brought
to collection centers run by local public
disposal operations – not even by small
commercial operators.
For disposal in Germany and in the
other member nations of the European
Economic Area (EEA), please contact
our Service technicians on location
or our Service Center in Bovenden,
Germany:
Minebea Intec Bovenden GmbH & Co. KG
Leinetal 2
D-37120 Bovenden, Germany
WEEE-Reg.-Nr. DE 58091735
Prior to disposal and/or scrapping
of the equipment, any batteries should
be removed and disposed of in local
collection boxes.
Minebea Intec will not take back equip-
ment contaminated with hazardous
materials (ABC contamination) – either
for repair or disposal.
Please refer to the accompanying leaf-
let/manual or visit our Internet
website (www.minebea-intec.com)
for comprehensive information that
includes our service addresses to con-
tact if you plan to send your equipment
in for repairs.
If you no longer need the packag-
ing after successful installation of the
equipment, you should return it for
recycling. The packaging is made from
environmentally friendly materials and
is a valuable source of secondary raw
material.
The equipment, including accessories
and batteries, does not belong in your
regular household waste.
Installation Instructions and Safety Information 9
Recycling
10 Installation Instructions and Safety Information Installation Instructions and Safety Information 11
Certificate Certificate
Installation Instructions and Safety Information 11
Certificate Certificate
12 Installation Instructions and Safety Information Installation Instructions and Safety Information 13
Safety information Safety information
Sheet 3
of
1
36287-740-16
Safety information
25.07.2012
Klausgrete
Drawing No.
Title
Revision 01
PC, printer,
PLC, etc.
For data transfer only and, if required, supply
for the connected device (e.g., printer), there
is no supply voltage going from the
connected device to the complete scale
These devices may also be installed in
Zone 2 or Zone 22 if they are suitable
for Category 3 as per the ATEX
Directive.
Gas: Group IIC, temperature class T4
Dust: Maximum device surface temperature 80°C
Ambient temperature: -10°C to +40°C
Protect connector from
equipment supplied from
being disconnected. Make
sure the emergency stop
works.
Connector from
equipment supplied
Connection
alternatives
Use an explosion-
protected connector
Equipotential bonding
conductor
Equipotential bonding
conductor
6)
6)
(load cells) of
series
or other weighing platforms suitable for Zone
2 or 22 (category 3G or 3D) that can be
supplied via a of at
least .
with rated bridge
supply voltage
Equipotential
bonding conductor
6)
6)
Installation Instructions and Safety Information 13
Safety information Safety information
Sheet 3
of
2
36287-740-16
Safety information
25.07.2012
Klausgrete
Drawing No.
Title
Revision 01
1. This device may be used in potentially explosive atmospheres of Zones 2 and 22. The permitted
ambient temperature range is -10°C to +40°C. The device may only be used indoors.
2. Do not use as a portable instrument.
3. Installation, operation, maintenance, and repairs should only be performed by an authorized
specialist, in accordance with applicable laws, rules and regulations, ordinances, and standards. In
particular, the EN 60079-14 standard must be taken into account. Installation, maintenance,
cleaning, and repair work may only take place with all power disconnected from the device.
4. The information on installation, operation, maintenance, and repair given in the operating
instructions supplied with the equipment must be observed.
5. Tighten the cable entry glands using a torque of 5 Nm. Tighten the cable entry gland for the mains
cable with a torque of 3 Nm only. Install the external connecting cables firmly to avoid damage
and strain. The cable connections inside the explosion-risk area must be secured against loosening.
6. All metal parts (housing, support arm, etc.) must be electrically connected to the terminal for the
equipotential bonding conductor so that any electrostatic charges can be conducted away from
the equipment. For this purpose, the equipment operator is obligated to connect a conductor with
a gauge of at least 4 mm² (cross section) to the equipotential bonding terminal (indicated by the
ground symbol) located on the back of the housing. During on-site installation, check if this
connection to the equipotential bonding conductor is of low resistance. Secure the connection
against being loosened by pulling the grounding cable.
if no metal connection
(e.g., support arm) is used. The standard indicator (which does not have a ground connection) is
wired to the equipotential bonding conductor internally.
7. Only use a damp cloth to clean the equipment. The equipment
operator assumes responsibility for preventing any risks caused by electrostatic charging.
8. When using a dust cover, avoid generating static electricity. Clean the device using a damp cloth
only.
9.
10. The data cables connected to the appliance are classed as flammable (Ex nA) electrical circuits. The
connections are protected against involuntary disconnection and may only be connected and
disconnected when the equipment is switched off. Block unused outlets to guarantee the IP65
level of protection. Keep any transitory voltage phenomenon away from the device.
11. Use the instrument only within the specified temperature limits. Avoid exposure to inadmissible
sources of heat or cold. Heat build-up should be avoided. Ensure that the device is well ventilated.
12. During installation, take suitable steps to prevent stray electrical interference (e.g., due to
magnetic fields).
13. Ensure the device is not exposed to strong jolts or vibrations.
14. The indicator (indicator of the complete scale) should be installed so that there is only a low risk of
mechanical danger to the IP protection.
14 Installation Instructions and Safety Information Installation Instructions and Safety Information 15
Safety information
Sheet 3
of
3
36287-740-16
Safety information
25.07.2012
Klausgrete
Drawing No.
Title
Revision 01
15. Accordingly, the instrument must be installed and operated in a manner that avoids exposure to
ultraviolet light (no direct sunlight).
16. The power connection must meet the regulations applicable in the country of operation. A correct
power connection must be ensured.
17. Ensure that the area is not potentially explosive when commissioning the device for the first time.
18. If cables are connected subsequently, make sure that the connections are not corroded.
19. All external cables are designed for fixed installation only and must be permanently laid.
Otherwise, use screwed connections designed in accordance with EN 60079-0 and rounded at an
angle of at least 75º and a radius equal to at least one-quarter of the diameter of the cables
without exceeding 3 mm.
20. Cables from third-party manufacturers (subject to the user’s responsibility) must be tested for
suitability: A traction strain of 10 times the diameter of the cable (in mm) in N, and at least 100 N
over 6 hours, should not stretch the cable by more than 6 mm. Follow the correct pin assignment.
Observe the wiring diagram. Remove unnecessary connections.
21. Unused openings must be sealed using suitable cover caps (dummy plugs). Do not remove while it
is carrying current.
22. When using external devices in Zone 2 explosion-risk areas, observe the gas group and
temperature class. The outputs must include the Ex nA electrical circuits.
23. Chemicals that can attack housing gaskets and cable sheathings must be kept away from the
instrument. These include oil, grease, benzine, acetone, and ozone. If you are uncertain, contact
the manufacturer.
24. The installation must be inspected for correct function and safety by a trained and qualified
person at appropriate intervals.
25. In the event of repair, use only original spare parts supplied by the manufacturer.
26. If the installation does not operate properly, disconnect it from the supply voltage immediately
and secure it against further use.
27. Any modifications to the instrument (except by persons authorized by Sartorius) will mean that it
no longer conforms to the requirements for use in Zone 2 and Zone 22 explosion-risk areas and
that all guarantee claims are null and void. Similarly, the device may only be opened by qualified
and authorized personnel.
28. Modifications (also those by Sartorius personnel) are subject to written approval.
29. The seals to the opened housing parts specified for IP65 protection must be replaced when the
instrument is repaired or serviced.
Installation Instructions and Safety Information 15
Safety information
Minebea Intec GmbH & Co. KG
Leinetal 2
37120 Bovenden, Germany
Phone +49.551.309.83.0
Fax +49.551.309.83.190
www.minebea-intec.com
Copyright by Minebea Intec,
Bovenden, Germany.
No part of this publication may be
reprinted or translat Swithout the prior
written permission of Minebea Intec. All
rights reserved. The status of the informa-
tion, specifications and illustrations in
this manual is indicated by the date given
below. Minebea Intec reserves the right to
make changes to the technology, features,
specifications and design of the equipment
without notice.
Status: September 2019
Printed in Germany on paper that has been
bleached without any use of chlorine.
MS · KT
Publication No.: WSI6009-e19096
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16

Minebea Intec Signum: Option Y2 for Use in Zone 2 and 22 Hazardous Areas 98648-016- Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI