Hologic ThinPrep Genesis Processor Owner's manual

Type
Owner's manual
ThinPrep™ Genesis™ Processor
Operator’s Manual
For Use with Version 1.x.y Software MAN-08098-002
Hologic, Inc.
250 Campus Drive
Marlborough, MA
01752 USA
Tel: 1-800-442-9892
1-508-263-2900
Fax: 1-508-229-2795
Web: www.hologic.com
Hologic BV
Da Vincilaan 5
1930 Zaventem
Belgium
Australian Sponsor:
Hologic (Australia and
New Zealand) Pty Ltd
Suite 302, Level 3
2 Lyon Park Road
Macquarie Park
NSW 2113
Australia
Tel: 02 9888 8000
UK Responsible
Person:
Hologic, Ltd.
Oaks Business Park
Crewe Road
Wythenshawe
Manchester
M23 9HZ
United Kingdom
Caution:
Federal law restricts this device to sale by or on the order of a physician, or any other practitioner
licensed by the law of the State in which the practitioner practices to use or order the use of the device and are
trained and experienced in the use of the ThinPrep™ Genesis™ Processor
Preparation of microscope slides using the ThinPrep™ Genesis™ Processor should be performed only by
personnel who have been trained by Hologic or by organizations or individuals designated by Hologic.
Evaluation of microscope slides produced with the ThinPrep™ Genesis™ Processor should be performed only by
cytotechnologists and pathologists who have been trained to evaluate ThinPrep prepared slides by Hologic or by
organizations or individuals designated by Hologic.
© Hologic, Inc., 2023. All rights reserved.
Although this guide has been prepared with every precaution to ensure accuracy, Hologic assumes no liability for
any errors or omissions, nor for any damages resulting from the application or use of this information.
This product may be covered by one or more U.S. patents identified at http://hologic.com/patentinformation
Hologic, Aptima, CytoLyt, Genesis, Genius, PreservCyt and ThinPrep are trademarks and/or registered
trademarks of Hologic, Inc. or its subsidiaries in the United States and other countries. All other trademarks are
the property of their respective owners.
Caution:
Changes or modifications to this unit not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
Document number: AW-23046-002 Rev. 002
1-2023
Revision History
Revision Date Description
AW-23046-002 Rev. 001 1-2022 Clarify instructions. Add instructions regarding reporting
serious incidents. Remove urine collection kit instructions.
Add annual preventive maintenance requirement. Add more
slide printer maintenance.
AW-23046-002 Rev. 002 1-2023 Describe the “Clear Lines” and filter detection features.
Administrative changes.
This page intentionally left blank.
Instructions For Use
Instructions For Use
ThinPrep™ Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 1/36
ThinPrep™ Genesis™ Processor
Instructions for Use
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 2/36
INTENDED USE
The ThinPrep Genesis Processor is part of the ThinPrep System. It is used to prepare
ThinPrep microscope slides from ThinPrep PreservCyt vials for use as a replacement for the
conventional method of Pap smear preparations for screening for the presence of atypical cells,
cervical cancer, or its precursor lesions (Low-grade Squamous Intraepithelial Lesions, High-grade
Squamous Intraepithelial Lesions), as well as all other cytologic categories as defined by The
Bethesda System for Reporting Cervical Cytology1.
Also, for preparation of ThinPrep microscope slides from non-gynecologic (non-gyn) samples,
including urine samples, and can be used to pipette an aliquot from the sample vial to the
specimen transfer tube. For professional use.
SUMMARY AND EXPLANATION OF THE SYSTEM
The ThinPrep process begins with the patient’s gynecologic sample being collected by the
clinician using a cervical sampling device which, rather than being smeared on a microscope
slide, is immersed and rinsed in a vial filled with 20 ml of PreservCyt Solution (PreservCyt). The
ThinPrep sample vial is then capped, labeled, and sent to a laboratory equipped with a ThinPrep
Genesis processor.
At the laboratory, the PreservCyt sample vial is placed into a ThinPrep Genesis processor. A
laboratory can elect to set up the ThinPrep Genesis processor to track the chain of custody for
the sample, and to set up printing IDs on each glass microscope slide. A gentle dispersion step
mixes the cell sample by currents in the fluid that are strong enough to separate debris and
disperse mucus, but gentle enough to have no adverse effect on cell appearance.
The cells are then captured on a gynecological ThinPrep Pap test filter that is specifically
designed to collect cells. The ThinPrep Genesis processor constantly monitors the rate of flow
through the ThinPrep Pap test filter during the collection process in order to prevent the cellular
presentation from being too scant or too dense. A thin layer of cells is then transferred to a glass
slide in a 20 mm-diameter circle, and the slide is automatically deposited into a fixative solution.
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 3/36
The ThinPrep Sample Preparation Process
(1) Dispersion (2) Cell Collection (3) Cell Transfer
The ThinPrep Pap test filter rotates
within the sample vial, creating
currents in the fluid that are strong
enough to separate debris and
disperse mucus, but gentle enough
to have no adverse effect on cell
appearance.
A gentle vacuum is created within
the ThinPrep Pap test filter, which
collects cells on the exterior surface
of the membrane. Cell collection is
controlled by the ThinPrep Genesis
processor’s software that monitors
the rate of flow through the
ThinPrep Pap test filter.
After the cells are collected on the
membrane, the ThinPrep Pap test
filter is inverted and gently pressed
against the ThinPrep microscope
slide. Natural attraction and slight
positive air pressure cause the cells
to adhere to the ThinPrep
microscope slide resulting in an
even distribution of cells in a defined
circular area.
As with conventional Pap smears, slides prepared with the ThinPrep® Genesis processor are
examined in the context of the patient’s clinical history and information provided by other
diagnostic procedures such as colposcopy, biopsy, and human papillomavirus (HPV) testing, to
determine patient management.
The PreservCyt® Solution component of the ThinPrep Genesis processor is an alternative
collection and transport medium for the testing of Human Papilloma Virus (HPV) and sexually
transmitted infections (STIs) in gynecological specimens, including, but not limited to:
Chlamydia trachomatis and Neisseria gonorrhoeae (Aptima Combo 2® assay),
Chlamydia trachomatis (Aptima® CT assay),
Neisseria gonorrhoeae (Aptima® GC assay),
Mycoplasma genitalium (Aptima® Mycoplasma genitalium assay),
Trichomonas vaginalis (Aptima® Trichomonas vaginalis assay),
Human papillomavirus (Aptima® HPV assay), and
Human papillomavirus (Aptima® HPV 16 18/45 genotype assay)
Refer to the respective manufacturer’s package inserts for instructions for using PreservCyt
Solution for collection, transport, storage, and preparation of specimens for use in those systems.
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 4/36
In addition to preparing a slide from a PreservCyt sample vial, the ThinPrep Genesis processor
has the ability to remove a 1-ml aliquot from the sample vial and transfer the aliquot to a
specimen transfer tube.
If any serious incident occurs related to this device, or any components used with this device,
report it to Hologic Technical Support and the competent authority local to the user and/or
patient.
LIMITATIONS
Gynecologic samples collected for preparation using the ThinPrep Genesis processor
should be collected using a broom-type or endocervical brush/plastic spatula combination
collection devices. Refer to the instructions provided with the collection device for
warnings, contraindications, and limitations associated with specimen collection.
Preparation of microscope slides using the ThinPrep Genesis processor should be
performed only by personnel who have been trained by Hologic or by organizations or
individuals designated by Hologic.
Evaluation of microscope slides produced with the ThinPrep Genesis processor should be
performed only by cytotechnologists and pathologists who have been trained to evaluate
ThinPrep prepared slides by Hologic or by organizations or individuals designated by
Hologic.
Supplies used by the ThinPrep Genesis processor are those designed and supplied by
Hologic specifically for the ThinPrep Genesis processor. These include PreservCyt
Solution vials, ThinPrep Pap test filters, ThinPrep microscope slides, and tubes for the
aliquot. Alternative collection media, filters, and slides have not been validated by Hologic
and may lead to erroneous results. Hologic does not provide a warranty for results using
any of these alternatives. Product performance may be compromised if supplies that have
not been validated by Hologic are used. After use, supplies should be disposed of in
accordance with local, state, and federal regulations.
A ThinPrep Pap test filter must be used only once and cannot be reused.
A ThinPrep microscope slide can be used only once. The slide can only have cells
transferred onto it once.
Aliquots taken by the ThinPrep Genesis processor have not been evaluated for specific
assays. Please refer to the instructions provided with a specific assay.
The performance of HPV and STI ancillary testing on sample vials reprocessed using
glacial acetic acid has not been evaluated.
WARNINGS
For In Vitro Diagnostic Use
Danger. PreservCyt Solution contains methanol. Toxic if swallowed. Toxic if inhaled.
Causes damage to organs. Flammable liquid and vapor. Keep away from heat, sparks,
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 5/36
open flames and hot surfaces. PreservCyt Solution should be stored and disposed of in
accordance with all applicable regulations.
Alternative collection media, filters, and slides have not been validated by Hologic and
may lead to erroneous results. Hologic does not provide a warranty for results using any
of these alternatives.
PRECAUTIONS
This equipment generates, uses and can radiate radio frequency energy, and if not
installed and used in accordance with the operator’s manual, may cause interference to
radio communications. Operation of this equipment in a residential area is likely to cause
harmful interference, in which case the user will be required to correct the interference at
his/her own expense.
PreservCyt Solution with cytologic sample intended for ThinPrep Pap testing must be
stored between 15oC (59oF) and 30oC (86oF) and tested within 6 weeks of collection.
Testing for certain sexually transmitted infections (STIs) and for Human Papilloma Virus
(HPV) in conjunction with cytology may be performed. Refer to assay specific guidance for
the collection, transport, and storage conditions of specimens for use in those systems.
PreservCyt Solution was challenged with a variety of microbial and viral organisms. The
following table presents the starting concentrations of viable organisms, and the log
reduction of viable organisms found after 15 minutes in the PreservCyt Solution. As with
all laboratory procedures, universal precautions should be followed.
Organism Initial Concentration Log Reduction After
15 Minutes
Candida albicans 5.5 x 105 CFU/ml ≥4.7
Candida auris 2.6 x 105 CFU/ml ≥5.4
Aspergillus niger 4.8 x 105 CFU/ml 2.7*
Escherichia coli 2.8 x 105 CFU/ml ≥4.4
Staphylococcus aureus 2.3 x 105 CFU/ml ≥4.4
Pseudomonas aeruginosa 2.5 x 105 CFU/ml ≥4.4
Mycobacterium tuberculosis 9.4 x 105 CFU/ml 4.9**
Rabbitpox virus 6.0 x 106 PFU/ml 5.5***
HIV-1 3.2 x 107 TCID50/ml ≥7.0***
Hepatitis B virus 2.2 x 106 TCID50/ml ≥4.25
SARS-CoV-2 virus 1.8 x 106 TCID50/ml ≥3.75
* After 1 hour 4.7 log reduction
** After 1 hour 5.7 log reduction
*** Data is for 5 minutes
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 6/36
Organism Initial Concentration Log Reduction After
15 Minutes
Organisms were tested with similar organisms from the same genus to
assess antimicrobial effectiveness
Note: All log reduction values with a ≥ designation yielded undetectable microbial
presence after exposure to PreservCyt Solution. The listed values represent the
minimum allowable claim given the initial concentration and the detection limit of
the quantitative method.
PERFORMANCE CHARACTERISTICS: REPORT OF CLINICAL STUDIES
The ThinPrep Genesis processor uses similar cell collection and slide preparation technology as
the ThinPrep 2000 system. The performance characteristics of the ThinPrep Genesis processor
are predicated on those of the ThinPrep 2000 system. Both clinical studies for the ThinPrep
2000 system and those comparing the ThinPrep Genesis processor to the ThinPrep 2000
system are described in the following sections.
ThinPrep 2000 System Compared to Conventional Pap Smear
A prospective multi-center clinical study was conducted to evaluate the performance of the
ThinPrep 2000 system in direct comparison to the conventional Pap smear. The objective of the
ThinPrep clinical study was to demonstrate that gynecologic specimens prepared using the
ThinPrep 2000 system were at least as effective as conventional Pap smears for the detection of
atypical cells and cervical cancer or its precursor lesions in a variety of patient populations. In
addition, an assessment of specimen adequacy was performed.
The initial clinical study protocol was a blinded, split sample, matched pair study, for which a
conventional Pap smear was prepared first, and the remainder of the sample (the portion that
normally would have been discarded) was immersed and rinsed into a vial of PreservCyt Solution.
At the laboratory, the PreservCyt sample vial was placed into a ThinPrep 2000 system and a slide
was then prepared from the patient’s sample. ThinPrep and conventional Pap smear slides were
examined and diagnosed independently. Reporting forms containing patient history as well as a
checklist of all possible categories of The Bethesda System were used to record the results of
the screening. A single independent pathologist reviewed all discrepant and positive slides from
all sites in a blinded fashion to provide a further objective review of the results.
Since the time of the ThinPrep 2000 system study, terminology in The Bethesda System
categories was revised. The data below retains the terminology from the original study.
LABORATORY AND PATIENT CHARACTERISTICS
Cytology laboratories at three screening centers (designated as S1, S2, and S3) and three
hospital centers (designated as H1, H2, and H3) participated in the clinical study. The screening
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 7/36
centers in the study serve patient populations (screening populations) with rates of abnormality
(Low-grade Squamous Intraepithelial Lesion [LSIL] and more severe lesions) similar to the United
States average of less than 5%.2 The hospital centers in the study serve a high risk referral
patient population (hospital populations) characterized by high rates (>10%) of cervical
abnormality. Data on race demographics was obtained for 70% of the patients that participated in
the study. The study population consisted of the following race groups: Caucasian (41.2%), Asian
(2.3%), Hispanic (9.7%), African American (15.2%), Native American (1.0%) and other groups (0.6%).
Table 1 describes the laboratories and the patient populations.
Table 1: Site Characteristics (ThinPrep 2000 System Study)
Laboratory Characteristics Clinical Study Demographics
Site Type of
Patient
Population
Laboratory
Volume -
Smears per Year
Cases Patient
Age Range
Post-
Menopausal
Previous
Abnormal Pap
Smear
Convent.
Prevalence
LSIL+
S1 Screening 300,000 1,386 18.0 - 84.0 10.6% 8.8% 2.3%
S2 Screening 100,000 1,668 18.0 - 60.6 0.3% 10.7% 2.9%
S3 Screening 96,000 1,093 18.0 - 48.8 0.0% 7.1% 3.8%
H1 Hospital 35,000 1,046 18.1 - 89.1 8.1% 40.4% 9.9%
H2 Hospital 40,000 1,049 18.1 - 84.4 2.1% 18.8% 12.9%
H3 Hospital 37,000 981 18.2 - 78.8 11.1% 38.2% 24.2%
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 8/36
CLINICAL STUDY RESULTS
The diagnostic categories of The Bethesda System were used as the basis of the comparison
between conventional and ThinPrep® findings from the clinical study. The diagnostic classification
data and statistical analyses for all clinical sites are presented in Tables 2 through 11. Cases with
incorrect paperwork, patient’s age less than 18 years, cytologically unsatisfactory slides, or
patients with a hysterectomy were excluded from this analysis. Few cases of cervical cancer
(0.02%3) were represented in the clinical study, as is typical in the United States patient
population.
Table 2: Diagnostic Classification Table, All Categories (ThinPrep 2000 System Study)
Conventional
NEG ASCUS AGUS LSIL HSIL SQ CA GL CA TOTAL
ThinPrep NEG 5224 295 3 60 11 0 0 5593
ASCUS 318 125 2 45 7 0 0 497
AGUS 13 2 3 0 1 0 1 20
LSIL 114 84 0 227 44 0 0 469
HSIL 11 15 0 35 104 2 0 167
SQ CA 0 0 0 0 0 1 0 1
GL CA 0 0 0 0 0 0 0 0
TOTAL 5680 521 8 367 167 3 1 6747
Abbreviations for Diagnoses: NEG = Normal or negative, ASCUS = Atypical Squamous Cells of
Undetermined Significance, AGUS = Atypical Glandular Cells of Undetermined Significance, LSIL = Low-
grade Squamous Intraepithelial Lesion, HSIL = High-grade Squamous Intraepithelial Lesion, SQ CA =
Squamous Cell Carcinoma, GL CA = Glandular Cell Adenocarcinoma
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 9/36
Table 3: Three Category Diagnostic Classification Table (ThinPrep 2000 System Study)
Conventional
NEG ASCUS/AGUS+ LSIL+ TOTAL
ThinPrep NEG 5224 298 71 5593
ASCUS/AGUS+ 331 132 54 517
LSIL+ 125 99 413 637
TOTAL 5680 529 538 6747
Table 4: Two Category Diagnostic Classification Table,
LSIL and More Severe Diagnoses (ThinPrep 2000 System Study)
Conventional
NEG/ASCUS/
AGUS+
LSIL+ TOTAL
ThinPrep NEG/ASCUS/
AGUS+
5985 125 6110
LSIL+ 224 413 637
TOTAL 6209 538 6747
Table 5: Two Category Diagnostic Classification Table,
ASCUS/AGUS and More Severe Diagnoses (ThinPrep 2000 System Study)
NEG ASCUS/AGUS+ TOTAL
ThinPrep NEG 5224 369 5593
ASCUS/AGUS+ 456 698 1154
TOTAL 5680 1067 6747
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 10/36
The diagnostic data analysis from the sites is summarized in Table 6 and 7. When the p-value is
significant (p < 0.05), the method favored is indicated in the tables.
Table 6: Results by Site, LSIL and More Severe Lesions (ThinPrep 2000 System Study)
Site Cases ThinPrep
LSIL+
Convent.
LSIL+
Increased
Detection*
p-Value Method
Favored
S1 1,336 46 31 48% 0.027 ThinPrep
S2 1,563 78 45 73% <0.001 ThinPrep
S3 1,058 67 40 68% <0.001 ThinPrep
H1 971 125 96 30% <0.001 ThinPrep
H2 1,010 111 130 (15%) 0.135 Neither
H3 809 210 196 7% 0.374 Neither
*Increased detection = ThinPrep® LSIL+ - Conventional LSIL+ x 100%
Conventional LSIL+
For LSIL and more severe lesions, the diagnostic comparison statistically favored the ThinPrep®
method at four sites and was statistically equivalent at two sites.
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 11/36
Table 7: Results by Site, ASCUS/AGUS and More Severe Lesions
(ThinPrep 2000 System Study)
Site Cases ThinPrep
ASCUS+
Convent.
ASCUS+
Increased
Detection*
p-Value Method
Favored
S1 1,336 117 93 26% 0.067 Neither
S2 1,563 124 80 55% <0.001 ThinPrep
S3 1,058 123 81 52% <0.001 ThinPrep
H1 971 204 173 18% 0.007 ThinPrep
H2 1,010 259 282 (8%) 0.360 Neither
H3 809 327 358 (9%) 0.102 Neither
*Increased detection = ThinPrep® ASCUS+ - Conventional ASCUS+ x 100%
Conventional ASCUS+
For ASCUS/AGUS and more severe lesions, the diagnostic comparison statistically favored the
ThinPrep method at three sites and was statistically equivalent at three sites.
One pathologist served as an independent reviewer for the six clinical sites, receiving both slides
from cases where the two methods were either abnormal or discrepant. Since a true reference
cannot be determined in such studies and therefore true sensitivity cannot be calculated, the use
of an expert cytologic review provides an alternative to histologic confirmation by biopsy or
human papillomavirus (HPV) testing as a means for determining the reference diagnosis.
The reference diagnosis was the more severe diagnosis from either of the ThinPrep or
conventional Pap slides as determined by the independent pathologist. The number of slides
diagnosed as abnormal at each site, compared to the reference diagnosis of the independent
pathologist, provides the proportion of LSIL or more severe lesions (Table 8) and the proportion
of ASCUS/AGUS or more severe lesions (Table 9). The statistical analysis allows a comparison of
the two methods and a determination of which method is favored when using the independent
pathologist for expert cytologic review as the adjudicator of the final diagnosis.
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 12/36
Table 8: Independent Pathologist Results by Site, LSIL and More Severe Lesions
(ThinPrep 2000 System Study)
Site Cases
Positive by
Independent
Pathologist
ThinPrep
Positive
Conventiona
l Positive
p-Value Method
Favored
S1 50 33 25 0.0614 Neither
S2 65 48 33 0.0119 ThinPrep
S3 77 54 33 <0.001 ThinPrep
H1 116 102 81 <0.001 ThinPrep
H2 115 86 90 0.607 Neither
H3 126 120 112 0.061 Neither
For LSIL and more severe lesions, the diagnostic comparison statistically favored the ThinPrep method
at three sites and was statistically equivalent at three sites.
Table 9: Independent Pathologist Results by Site, ASCUS/AGUS and More Severe Lesions
(ThinPrep 2000 System Study)
Site Cases
Positive by
Independent
Pathologist
ThinPrep®
Positive
Conventiona
l Positive
p-Value Method
Favored
S1 92 72 68 0.0511 Neither
S2 101 85 59 0.001 ThinPrep
S3 109 95 65 <0.001 ThinPrep
H1 170 155 143 0.090 Neither
H2 171 143 154 0.136 Neither
H3 204 190 191 1.000 Neither
For ASCUS/AGUS and more severe lesions, the diagnostic comparison statistically favored the
ThinPrep method at two sites and was statistically equivalent at four sites.
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 13/36
Table 10 below shows the summary for all sites of the descriptive diagnosis for all Bethesda System
categories.
Table 10: Summary of Descriptive Diagnosis (ThinPrep 2000 System Study)
Descriptive Diagnosis ThinPrep Conventional
Number of Patients: 6747 N % N %
Benign Cellular Changes: 1592 23.6 1591 23.6
Infection:
Trichomonas Vaginalis 136 2.0 185 2.7
Candida spp. 406 6.0 259 3.8
Coccobacilli 690 10.2 608 9.0
Actinomyces spp. 2 0.0 3 0.0
Herpes 3 0.0 8 0.1
Other 155 2.3 285 4.2
Reactive Cellular Changes
Associated with:
Inflammation 353 5.2 385 5.7
Atrophic Vaginitis 32 0.5 48 0.7
Radiation 2 0.0 1 0.0
Other 25 0.4 37 0.5
Epithelial Cell Abnormalities: 1159 17.2 1077 16.0
Squamous Cell:
ASCUS 501 7.4 521 7.7
favor reactive 128 1.9 131 1.9
favor neoplastic 161 2.4 140 2.1
undetermined 213 3.2 250 3.7
LSIL 469 7.0 367 5.4
HSIL 167 2.5 167 2.5
Carcinoma 1 0.0 3 0.0
Glandular Cell:
Benign Endometrial cells
in Postmenopausal
Women
7 0.1 10 0.1
Atypical Glandular Cells
(AGUS)
21 0.3 9 0.1
favor reactive 9 0.1 4 0.1
favor neoplastic 0 0.0 3 0.0
undetermined 12 0.2 2 0.0
Endocervical
Adenocarcinoma
0 0.0 1 0.0
Note: Some patients had more than one diagnostic subcategory.
Table 11 shows the rates of detection for infection, reactive changes, and the total benign cellular
changes for both the ThinPrep® and conventional methods at all sites.
ThinPrep® Genesis™ Processor Instructions for Use English AW-23047-001 Rev. 001 11-2021 14/36
Table 11: Benign Cellular Changes Results (ThinPrep 2000 System Study)
ThinPrep Conventional
N % N %
Benign
Cellular
Changes
Infection 1392 20.6 1348 20.0
Reactive
Changes 412 6.1 471 7.0
Total* 1592 23.6 1591 23.6
* Total includes some patients that may have had both an infection and reactive cellular change.
Tables 12, 13, and 14 show the specimen adequacy results for the ThinPrep method and
conventional smear method for all of the study sites. Of the 7,360 total patients enrolled, 7,223
are included in this analysis. Cases with patient’s age less than 18 years or patients with a
hysterectomy were excluded from this analysis.
Two additional clinical studies were conducted to evaluate specimen adequacy results when
samples were deposited directly into the PreservCyt® vial, without first making a conventional Pap
smear. This specimen collection technique is the intended use for the ThinPrep 2000 system.
Tables 15 and 16 present the split sample and direct to vial results.
Table 12: Summary of Specimen Adequacy Results (ThinPrep 2000 System Study)
Specimen Adequacy ThinPrep Conventional
Number of Patients: 7223 N % N %
Satisfactory 5656 78.3 5101 70.6
Satisfactory for Evaluation
but Limited by: 1431 19.8 2008 27.8
Air-Drying Artifact 1 0.0 136 1.9
Thick Smear 9 0.1 65 0.9
Endocervical Component
Absent 1140 15.8 681 9.4
Scant Squamous
Epithelial Component 150 2.1 47 0.7
Obscuring Blood 55 0.8 339 4.7
Obscuring Inflammation 141 2.0 1008 14.0
No Clinical History 12 0.2 6 0.1
Cytolysis 19 0.3 119 1.6
Other 10 0.1 26 0.4
Unsatisfactory for
Evaluation: 136 1.9 114 1.6
Air-Drying Artifact 0 0.0 13 0.2
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284
  • Page 285 285
  • Page 286 286
  • Page 287 287
  • Page 288 288
  • Page 289 289
  • Page 290 290
  • Page 291 291
  • Page 292 292
  • Page 293 293
  • Page 294 294
  • Page 295 295
  • Page 296 296
  • Page 297 297
  • Page 298 298
  • Page 299 299
  • Page 300 300
  • Page 301 301
  • Page 302 302
  • Page 303 303
  • Page 304 304
  • Page 305 305
  • Page 306 306
  • Page 307 307
  • Page 308 308
  • Page 309 309
  • Page 310 310
  • Page 311 311
  • Page 312 312
  • Page 313 313
  • Page 314 314
  • Page 315 315
  • Page 316 316
  • Page 317 317
  • Page 318 318
  • Page 319 319
  • Page 320 320
  • Page 321 321
  • Page 322 322
  • Page 323 323
  • Page 324 324
  • Page 325 325
  • Page 326 326
  • Page 327 327
  • Page 328 328
  • Page 329 329
  • Page 330 330
  • Page 331 331
  • Page 332 332
  • Page 333 333
  • Page 334 334
  • Page 335 335
  • Page 336 336
  • Page 337 337
  • Page 338 338
  • Page 339 339
  • Page 340 340
  • Page 341 341
  • Page 342 342
  • Page 343 343
  • Page 344 344
  • Page 345 345
  • Page 346 346
  • Page 347 347
  • Page 348 348
  • Page 349 349
  • Page 350 350