Thermo Fisher Scientific 8 x 50mL Individually Sealed Rotor User manual

Category
Fridge-freezers
Type
User manual
Thermo Scientific
8 x 50 sealed
Instruction Manual
50121058-d • 07 / 2020
Visit us online to register your warranty:
thermofisher.com/labwarranty
WEEE Conformity
This product is subject to the regulations of the EU Waste Electrical & Electronic
Equipment (WEEE) Directive 2012/19/EU. It is marked by the following symbol:
Thermo Scientific 75003694 8 x 50 sealed i
T
Preface .......................................................................................................................................iii
Scope of Supply ........................................................................................................................iii
Precautions ...............................................................................................................................iii
Chapter 1 Rotor Specifications ..............................................................................................................1-1
Technical Data ........................................................................................................1-2
Chapter 2 Accessories .............................................................................................................................2-1
Chapter 3 AutoLock .............................................................................................................................3-1
Rotor installation ......................................................................................................3-2
Removing the rotor ..................................................................................................3-3
Chapter 4 Rotor loading ...........................................................................................................................4-1
Before a Run ............................................................................................................4-2
Proper Loading .........................................................................................................4-2
Improper Loading ....................................................................................................4-2
Maximum Loading ...................................................................................................4-3
Cycle Counter ..........................................................................................................4-3
Chapter 5 Aerosol-tight Applications ...................................................................................................5-1
Basic Principles .........................................................................................................5-2
Placing O-ring ..........................................................................................................5-2
Fill Level ...................................................................................................................5-2
Checking the Aerosol-Tightness ...............................................................................5-2
Table of Contents
Chapter 6 Maintenance and Care ......................................................................................................... 6-1
Cleaning intervals .................................................................................................... 6-2
Cleaning .................................................................................................................. 6-2
Disinfection ............................................................................................................. 6-3
Decontamination ..................................................................................................... 6-4
Autoclaving .............................................................................................................. 6-5
Service of Thermo Fisher Scientific .......................................................................... 6-5
RCF-Values ...............................................................................................................A-1
Chemical Compatibility Chart ..............................................................................B-1
Thermo Scientific 75003694 8 x 50 sealed iii
P
Preface
Before starting to use the rotor, read through this instruction manual carefully and follow the
instructions.
The information contained in this instruction manual is the property of Thermo Fisher Scientific; it is
forbidden to copy or pass on this information without explicit approval.
Failure to follow the instructions and safety information in this instruction manual will result in the
expiration of the sellers warranty.
Scope of Supply
If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.
Precautions
In order to ensure safe operation of the 8 x 50 sealed, the following general safety regulations must
be followed:
Do not remove the magnet at the rotor bottom;
Do not use rotors which show any signs of corrosion and/or cracks;
Use only with rotors that have been loaded properly;
Never run the rotor with missing stainless-steel tube holders;
Never overload the rotor;
Use only accessories which have been approved by Thermo Fisher Scientific. Exceptions to this
rule are commercially available glass or plastic centrifuge tubes, provided they have been approved
for the speed or the RCF value of the rotor.
Article Number Quantity Check
75003694 8 x 50 sealed 1
50121058 Instruction Manual 1
70009824 Anti-corrosion oil 1
76003500 Rubber seal grease 1
Preface
iv 75003694 8 x 50 sealed Thermo Scientific
Please observe the safety instructions.
Please pay particular attention to the following aspects:
Rotor installation: Check that the rotor is locked properly into place before operating the
centrifuge.
Always balance the samples.
Maximum sample density at maximum speed: 1.2g
ml
------
This symbol refers to general hazards.
ATTENTION means that material damage could occur.
WARNING means that injuries or material damage or contamination could occur.
This symbol refers to biological hazards.
Observe the information contained in the instruction manual to keep yourself and your
environment safe.
Thermo Scientific 75003694 8 x 50 sealed 1-1
1
Rotor Specifications
Contents
Technical Dataon page 1-2
1-2 75003694 8 x 50 sealed Thermo Scientific
1 Rotor Specifications
Technical Data
Technical Data
Table 1-1. 230 V, 50 / 60 Hz ventilated
Centrifuge Multifuge X1 Megafuge 16
Catalog # 75004210 75004230
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6700 6700
maximum RCF value at nmax 7177 7177
max. / min. radius [cm] 14.3 / 6.9 14.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 30 / 35 30 / 35
sample heating at nmax [°C] referred to
ambient temperature of 23 °C, running
time 60 minutes
12 12
Aerosol-tight *
* Tested by HPA, Porton-down, UK
yes yes
permissible temperature range
autoclavable °C
121 121
Table 1-2. 230 V, 50 / 60 Hz refrigerated
Centrifuge Multifuge X1 R Megafuge 16R
Catalog # 75004250 75004270
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6700 6700
maximum RCF value at nmax 7177 7177
max. / min. radius [cm] 14.3 / 6.914.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 30 / 35 30 / 35
Aerosol-tight *yes yes
permissible temperature
range autoclavable °C
121 121
*Tested by HPA, Porton-down, UK
Megafuge 8
75007211
4.7
50,000
30 x 30
4400
3030
14.0 / 8.5
37
16 / 27
14
no
121
Megafuge 8R
75007214
3.3
50,000
18 x 9
5600
5014
14.3 / 6.3
45
35 / 40
yes
121
1 Rotor Specifications
Technical Data
Thermo Scientific 75003694 8 x 50 sealed 1-3
Table 1-3. 120 V, 60 Hz ventilated
Centrifuge Multifuge X1 Megafuge 16
Catalog # 75004211 75004231
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6500 6500
maximum RCF value at nmax 6755 6755
max. / min. radius [cm] 14.3 / 6.9 14.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 40 / 40 40 / 40
sample heating at nmax [°C] referred to
ambient temperature of 23 °C, running
time 60 minutes
11 11
Aerosol-tight *
* Tested by HPA, Porton-down, UK
yes yes
permissible temperature range
autoclavable °C
121 121
Table 1-4. 120 V, 60 Hz refrigerated
Centrifuge Multifuge X1 R Megafuge 16R
Catalog # 75004251 75004271
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6700 6700
maximum RCF value at nmax 7177 7177
max. / min. radius [cm] 14.3 / 6.9 14.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 40 / 40 40 / 40
Aerosol-tight *yes yes
permissible temperature range
autoclavable °C
121 121
* Tested by HPA, Porton-down, UK
Megafuge 8
75007210
4.7
50,000
30 x 30
4400
3030
14.0 / 8.5
37
18 / 26
14
no
121
Megafuge 8R
75007213
3.3
50,000
18 x 9
5600
5014
14.3 / 6.3
45
30 / 40
yes
121
1-4 75003694 8 x 50 sealed Thermo Scientific
1 Rotor Specifications
Technical Data
Table 1-5. 230 V, 50 / 60 Hz ventilated
Centrifuge Sorvall Legend X1 Sorvall ST 16
Catalog # 75004220 75004240
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6700 6700
maximum RCF value at nmax 7177 7177
max. / min. radius [cm] 14.3 / 6.914.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 30 / 35 30 / 35
sample heating at nmax [°C] referred to
ambient temperature of 23 °C, running
time 60 minutes
12 12
Aerosol-tight *
* Tested by HPA, Porton-down, UK
yes yes
permissible temperature
range autoclavable °C
121 121
Table 1-6. 230 V, 50 / 60 Hz refrigerated
Centrifuge Sorvall Legend X1R Sorvall ST 16R
Catalog # 75004260 75004380
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6700 6700
maximum RCF value at nmax 7177 7177
max. / min. radius [cm] 14.3 / 6.914.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 30 / 35 30 / 35
Aerosol-tight *yes yes
permissible temperature
range autoclavable °C
121 121
*Tested by HPA, Porton-down, UK
Sorvall ST 8
75007201
4.7
50,000
30 x 30
4,400
14.0 / 8.5
37
16 / 27
no
121
Sorvall ST 8R
75007204
3030
14
3.3
50,000
18 x 9
5600
5014
14.3 / 6.3
45
35 / 40
yes
121
1 Rotor Specifications
Technical Data
Thermo Scientific 75003694 8 x 50 sealed 1-5
Table 1-7. 120 V, 60 Hz ventilated
Centrifuge Sorvall Legend X1 Sorvall ST 16
Catalog # 75004221 75004241
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6500 6500
maximum RCF value at nmax 6755 6755
max. / min. radius [cm] 14.3 / 6.9 14.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 40 / 40 40 / 40
sample heating at nmax [°C] referred to
ambient temperature of 23 °C, running
time 60 minutes
11 11
Aerosol-tight *
* Tested by HPA, Porton-down, UK
yes yes
permissible temperature range
autoclavable °C
121 121
Table 1-8. 120 V, 60 Hz refrigerated
Centrifuge Sorvall Legend X1R Sorvall ST 16R
Catalog # 75004261 75004381
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6700 6700
maximum RCF value at nmax 7177 7177
max. / min. radius [cm] 14.3 / 6.9 14.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 40 / 40 40 / 40
Aerosol-tight *yes yes
permissible temperature
range autoclavable °C
121 121
* Tested by HPA, Porton-down, UK
Sorvall ST 8
75007200
4.7
50,000
30 x 30
4,400
14.0 / 8.5
37
18 / 26
no
121
3030
14
Sorvall ST 8R
75007203
3.3
50,000
18 x 9
5600
5014
14.3 / 6.3
45
30 / 40
yes
121
1 Rotor Specifications
Technical Data
1-6 75003694 8 x 50 sealed Thermo Scientific
Table 1-9. 100 V, 50 / 60 Hz ventilated
Centrifuge Sorvall Legend 1X Sorvall ST 16
Catalog # 75004223 75004243
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6000 6000
maximum RCF value at nmax 5755 5755
max. / min. radius [cm] 14.3 / 6.914.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 35 / 35 35 / 35
sample heating at nmax [°C] referred to
ambient temperature of 23 °C, running
time 60 minutes
10 10
Aerosol-tight *
* Tested by HPA, Porton-down, UK
yes yes
permissible temperature
range autoclavable °C
121 121
Table 1-10. 100 V, 50 / 60 Hz refrigerated
Centrifuge Sorvall Legend X1R Sorvall ST 16R
Catalog # 75004263 75004383
Weight empty [kg] of rotor 3.3 3.3
max. cycle number 50,000 50,000
Maximum permissible load [g] 8 x 189 8 x 189
maximum speed nmax [rpm] 6500 6500
maximum RCF value at nmax 4997 4997
max. / min. radius [cm] 14.3 / 6.914.3 / 6.9
Pitch angle [°] 45 45
Accel. / braking time [s] 40 / 35 40 / 35
Aerosol-tight *yes yes
permissible temperature range
autoclavable °C
121 121
* Tested by HPA, Porton-down, UK
Sorvall ST 8
75007202
4.7
50,000
30 x 30
4,400
14.0 / 8.5
37
18 / 26
no
121
3030
14
Sorvall ST 8R
75007205
3.3
50,000
18 x 9
5600
5014
14.3 / 6.3
45
35 / 40
yes
121
1 Rotor Specifications
Technical Data
Thermo Scientific 75003694 8 x 50 sealed 1-7
Table 1-11. 230 V, 50 / 60 Hz ventilated
Centrifuge Thermo Scientific SL 16
Catalog # 75004000
Weight empty [kg] of rotor 3.3
max. cycle number 50,000
Maximum permissible load [g] 8 x 189
maximum speed nmax [rpm] 6700
maximum RCF value at nmax 7177
max. / min. radius [cm] 14.3 / 6.9
Pitch angle [°] 45
Accel. / braking time [s] 30 / 35
sample heating at nmax [°C] referred to
ambient temperature of 23 °C, running
time 60 minutes
12
Aerosol-tight *
* Tested by HPA, Porton-down, UK
yes
permissible temperature range
autoclavable °C
121
Table 1-12. 230 V, 50 / 60 Hz refrigerated
Centrifuge Thermo Scientific SL 16R
Catalog # 75004030
Weight empty [kg] of rotor 3.3
max. cycle number 50,000
Maximum permissible load [g] 8 x 189
maximum speed nmax [rpm] 6700
maximum RCF value at nmax 7177
max. / min. radius [cm] 14.3 / 6.9
Pitch angle [°] 45
Accel. / braking time [s] 30 / 35
Aerosol-tight *yes
permissible temperature range
autoclavable °C
121
*Tested by HPA, Porton-down, UK
Thermo Scientific SL 8
75007221
4.7
50,000
30 x 30
4,400
14.0 / 8.5
37
16 / 27
14
no
121
3030
Thermo Scientific SL 8R
75007224
3.3
50,000
18 x 9
5600
5014
14.3 / 6.3
45
35 / 40
yes
121
1 Rotor Specifications
Technical Data
1-8 75003694 8 x 50 sealed Thermo Scientific
Table 1-13. 120 V, 60 Hz ventilated
Centrifuge Thermo Scientific SL 16
Catalog # 75004001
Weight empty [kg] of rotor 3.3
max. cycle number 50,000
Maximum permissible load [g] 8 x 189
maximum speed nmax [rpm] 6500
maximum RCF value at nmax 6755
max. / min. radius [cm] 14.3 / 6.9
Pitch angle [°] 45
Accel. / braking time [s] 40 / 40
sample heating at nmax [°C] referred to
ambient temperature of 23 °C, running
time 60 minutes
11
Aerosol-tight *
* Tested by HPA, Porton-down, UK
yes
permissible temperature range
autoclavable °C
121
Table 1-14. 120 V, 60 Hz refrigerated
Centrifuge Thermo Scientific SL 16R
Catalog # 75004031
Weight empty [kg] of rotor 3.3
max. cycle number 50,000
Maximum permissible load [g] 8 x 189
maximum speed nmax [rpm] 6700
maximum RCF value at nmax 7177
max. / min. radius [cm] 14.3 / 6.9
Pitch angle [°] 45
Accel. / braking time [s] 40 / 40
Aerosol-tight *yes
permissible temperature range
autoclavable °C
121
* Tested by HPA, Porton-down, UK
Thermo Scientific SL 8
75007220
4.7
50,000
30 x 30
4,400
14.0 / 8.5
37
18 / 26
14
no
121
3030
Thermo Scientific SL 8R
75007223
3.3
50,000
18 x 9
5600
5014
14.3 / 6.3
45
30 / 40
yes
121
Thermo Scientific 75003694 8 x 50 sealed 2-1
2
Accessories
Contents
Rotor data on page 2-2
Accessories on page 2-2
2 Accessories
2-2 75003694 8 x 50 sealed Thermo Scientific
Rotor Specifications
* Max Speed may vary depending on centrifuge unit.
RCF/Radius
Maximum
Minimum
* Maximum Speed may vary by centrifuge model.
8 x 50ml Individually Sealed Fixed Angle Rotor
Cat. No. Description
75003694 8 x 50 ml Individually Sealed Fixed Angle Rotor
Accessories
Cat. No. Description
75003789 O-ring Refresher Kit
Tube Ordering Information
Cat. No. Qty
/Pkg Description Cat. No. Qty
/Pkg
No. of
Places
- 50 50 Disposable Conical - - 17 x 121 -- - - -- - -
Accessories Required:
Adapters Required:Sealing Assembly Required:
Radius (cm)
14.3
6.9
Max.
Speed
(rpm)
4107
8 x 50
17 x 121
Capacity (ml)
Tube Dimensions (mm)
Thermo Scientific General Purpose Rotors 8 x 50ml Individually Sealed Rotor
K Factor
Net Weight (kg)
8 x 50 ml Individually Sealed Fixed Angle Rotor
45
6700*
Angle
o
Maximum Speed (rpm)
3.3
Dims Ø x L
(mm)
Cat. No. Tube
Vol. (ml)
Vol. Fill
(ml) Description Qty.
RCF (x g)
7177
3459
Thermo Scientific 75003694 8 x 50 sealed 3-1
3
AutoLock
Contents
“Rotor installationon page 3-2
“Removing the rotor” on page 3-3
3 AutoLock
Rotor installation
3-2 75003694 8 x 50 sealed Thermo Scientific
Rotor installation
This rotor is equipped with an AutoLock-system.
This system is used to automatically lock the rotor to the centrifuge spindle. The rotor does not have
to be bolted onto the centrifuge spindle.
Proceed as follows:
1. Open the door of the centrifuge and if necessary remove any dust, foreign objects or residue
from the chamber.
AutoLock and O-ring must be clean and
undamaged.
Figure 3-1. AutoLock
2. Hold the rotor over the centrifuge spindle and let it slide slowly down the centrifuge spindle.
The rotor clicks automatically into place.
3. Check if the rotor is properly installed by lifting it slightly on the handle. If the rotor can be
pulled up, then it must be reclamped to the centrifuge spindle.
CAUTION
Unapproved or incorrectly combined accessories can cause serious damage to the
centrifuge.
CAUTION
Do not force the rotor onto the centrifuge spindle.
If the rotor is very light, then it may be necessary to press it onto the centrifuge spindle
with a small amount of pressure.
WARNING
If the rotor cannot be properly locked in place after several attempts, then the
AutoLock is defective and you are not permitted to operate the rotor.
Check for any damage to the rotor: Damaged rotors must not be used.
Keep the hub area clear of objects.
AutoLock
O-ring
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47

Thermo Fisher Scientific 8 x 50mL Individually Sealed Rotor User manual

Category
Fridge-freezers
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI