Kistler 6021B Owner's manual

Type
Owner's manual
Instruction
manual
High temperature
pressure sensor
Type 6021B...
Standard version
(non-Ex-version)
and explosion-proof
version (Ex-ia, Ex-nA)
6021B_012-002e-06.22
IECEx
Foreword
6021B_012-002e-06.22 Page 1
Foreword
Thank you for choosing a Kistler quality product
characterized by technical innovation, precision and long
life.
Information in this document is subject to change without
notice. Kistler reserves the right to change or improve its
products and make changes in the content without
obligation to notify any person or organization of such
changes or improvements.
© 2022 Kistler Group. Kistler Group products are
protected by various intellectual property rights. For more
details visit www.kistler.com. The Kistler Group includes
Kistler Holding AG and all its subsidiaries in Europe,
Asia, the Americas and Australia.
Kistler Group
Eulachstre 22
8408 Winterthur
Switzerland
Tel. +41 52 224 11 11
info@kistler.com
www.kistler.com
Content
Page 2 6021B_012-002e-06.22
Content
1. Introduction ............................................................................................................................. 3
2. Important information ............................................................................................................. 4
2.1 Protection and standards ................................................................................................. 4
2.2 Precautions ..................................................................................................................... 5
2.3 Disposal of electronic equipment ..................................................................................... 5
2.4 Personal safety warnings................................................................................................. 5
3. Pressure sensor ...................................................................................................................... 7
3.1 Technical data ................................................................................................................. 7
3.2 Ordering key ................................................................................................................... 7
4. Operation ................................................................................................................................. 8
4.1 Operation in non-hazardous areas ................................................................................... 8
4.1.1 General instructions for operation........................................................................ 8
4.1.2 Sensor execution ................................................................................................ 8
4.1.3 Ground isolated measuring chain ........................................................................ 8
4.1.4 Marking............................................................................................................... 9
4.2 Operation in hazardous areas........................................................................................ 10
4.2.1 Instructions for operation in hazardous areas .................................................... 10
4.2.2 Sensor executions ............................................................................................ 11
4.2.3 Marking on sensor ............................................................................................ 11
4.2.4 Labeling and packaging .................................................................................... 11
4.2.5 Non-Sparking Equipment (Ex-nA) for use only in zone 2 ................................... 11
4.2.6 Intrinsically safe equipment (Ex-ia) .................................................................... 12
4.2.6.1 Intrinsically safe measuring chain, ground isolated
(sensor version Type 6021BE…) ........................................................ 12
4.2.6.2 Entity parameter of the sensor ............................................................ 12
5. Installation, service & maintenance ..................................................................................... 13
5.1 Mounting bore ............................................................................................................... 13
5.2 Mounting/removal .......................................................................................................... 14
5.2.1 Direct mounting for sensors with Lemo connector or open leads ....................... 14
5.2.2 Mounting screw with removal tool...................................................................... 16
5.2.3 Indirect mounting of the sensor with Lemo connector or open leads .................. 18
18
5.2.4 Direct mounting of the sensor with a 7/16“ connector ........................................ 19
5.2.5 Indirect mounting of the sensor with a 7/16“ connector ...................................... 20
5.3 Cable handling .............................................................................................................. 21
5.3.1 Mineral insulated hardline cable ........................................................................ 21
5.3.2 Softline cable .................................................................................................... 21
5.3.3 Hardline cable fixation ....................................................................................... 22
5.3.4 Electrical connections ....................................................................................... 22
5.4 Service and maintenance .............................................................................................. 23
5.5 Measuring chain components and accessories .............................................................. 23
6. Declaration of conformity ..................................................................................................... 24
Total Seiten 24
Introduction
6021B_012-002e-06.22 Page 3
1. Introduction
Please read through this manual thoroughly to enable
you to fully exploit the various features of your product.
To the extent permitted by law Kistler accepts no liability
if this instruction manual is not followed or products other
than those listed under Accessories are used.
These instructions describe the piezoelectric high
temperature pressure sensor Type 6021B.
Our thermoacoustics product and application brochures,
in addition to our detailed data sheets, will provide you
with an overview of our product range.
If you require additional assistance beyond what can be
located online or in this manual, please contact Kistler‘s
extensive support organization or visit us online at
www.kistler.com/thermoacoustics
High temperature pressure sensor Type 6021B...
Page 4 6021B_012-002e-06.22
2. Important information
It is essential to review the following information, which is
intended to ensure your personal safety when working
with high temperature pressure sensors and contribute
toward a long-fault-free service life.
To the extent permitted by law Kistler accepts no liability
if this instruction manual is not followed or products other
than those listed under accessories are used.
2.1 Protection and standards
All equipment is tested and has left the factory in a
perfect, safe condition. In order to
maintain this
condition and to ensure safe operation, it is essential
to follow strictly the instructions and warnings
contained in these operating instructions or printed on
the product itself.
Compliance with local safety regulations that apply to the
use of electrical and electronic equipment is to be
followed.
The high temperature pressure sensor 6021B... is in
compliance with CE and complies with the following
provisions of the directives:
ISO 9001
2011/65/EU (ROHS)
2014/30/EU (EMC)
2014/34/EU (ATEX)
Hazardous area:
The following harmonised standards were applied:
EN 60079-0:2012+A11:2013/IEC 60079-0:2011
EN 60079-11:2012/IEC 60079-11:2011
EN 60079-15:2010/IEC 60079-15:2010
Important information
6021B_012-002e-06.22 Page 5
2.2 Precautions
Compliance with the following precautionary mea-
sures is vital to ensure safe and reliable operation:
The equipment should be used only under the speci-
fied operating conditions.
If there is evidence that safe operation is no longer
possible, the instrument must be powered off and
rendered safe against accidental start-up.
Safe operation is no longer possible when the product
shows visible signs of damage
is no longer operating
has been subjected to prolonged storage under
unsuitable conditions
2.3 Disposal of electronic equipment
Do not discard old electronic equipment in
municipal trash. For disposal at end of life, please
return this product to the nearest authorized
electronic waste disposal service or contact your
Kistler sales office for disposal instructions.
2.4 Personal safety warnings
Hazardous environment.
Read and understand manual before operating!
High temperature pressure sensor Type 6021B...
Page 6 6021B_012-002e-06.22
WARNING: HAZARDOUS ENVIRONMENT
Risk of explosive atmosphere.
Adapters, isolation valves, or sealing rings that are
not compatible with process gasses will corrode and
fail. This failure may
result in gas leaks, fire,
explosion, or projectiles.
Prevent corrosion and failure by verifying that all
components
are compatible as described in the
installation procedures.
Do not connect or disconnect while circuit is live
unless area is known to be nonhazardous.
Important:
The sensor must not be modified in any way.
Otherwise the ATEX,
IECEx and all other Ex
approvals will become invalid.
Substitution of components may impair suitability for
hazardous area.
Explosion Hazard. Do not connect or disconnect
when energized unless the area is free of ignitible
concentrations.
Pressure sensor
6021B_012-002e-06.22 Page 7
3. Pressure sensor
Basic execution
Sensor Type 6021B… is a 2-wire, internally ground
isolated differential piezoelectric pressure sensor. The
sensor is equipped with an integral mineral insulated
hardline cable and is available with different cable
terminations.
Executions for operation in hazardous areas
(Ex-versions)
All sensors of the 6021BE... series are certified for use in
potentially explosive atmosphere” acc. to ATEX directive
and IECEx Standards.
The sensors are certified as Ex ia (intrinsic safety) and
Ex nA (non Sparking).
3.1 Technical data
The sensors may only be used acc. to the technical
specifications made in the respective data sheet.
The technical specifications are given in data sheet Doc.
No. 003-590.
3.2 Ordering key
High temperature pressure sensor Type 6021B...
Page 8 6021B_012-002e-06.22
4. Operation
For proper operation it is necessary to first determine
whether the sensor is installed in a potentially explosive
atmosphere.
4.1 Operation in non-hazardous areas
4.1.1 General instructions for operation
The following instructions regarding the installation
and operation must be followed:
The sensor is hermetically sealed. The integral
hardline cable meets ingress protection IP50 and
must be protected against adverse environmental
effects where necessary.
The sensor may only be used for its intended
purpose. Any violation will waive all warranties and
responsibilities of the manufacturer.
IP protection degree has to be maintained.
The EC
Type examination certificate must be
followed as well as any ‘special conditions’ that
may be specified.
4.1.2 Sensor execution
In non-hazardous areas, the sensor without Ex-certifi-
cation Type 6021B-... can be used.
4.1.3 Ground isolated measuring chain
For measurements of very small pressure fluctuations
and in environments with electromagnetic interference, it
is recommended to operate the measurement chain
completely ground-isolated.
This requires a differential charge amplifier. Depending
on the required application, Kistler recommends the
differential charge amplifier Type 5181, 5183 (Ex ia) or
5185 (Ex nA).
Measuring chain examples are shown in Kistler
document No. 960-201.
Operation
6021B_012-002e-06.22 Page 9
Recommended measuring chain
Fig. 1: Recommended measuring chain for operation in
non-hazardous areas
4.1.4 Marking
The standard execution, ‚non Ex-Version’ is marked as
follows:
Kistler Instrumente AG
Winterthur Switzerland
Sensor Type: 6021B…
SN:
Sensor
Type 6021B
High temperature pressure sensor Type 6021B...
Page 10 6021B_012-002e-06.22
4.2 Operation in hazardous areas
4.2.1 Instructions for operation in hazardous areas
In addition to the instructions for proper operation in
Section 4.1.1
the following instructions must be
followed for installation and operation in potentially
explosive atmospheres:
The sensor can be used in intrinsically safe
circuits in zones 0, 1 and 2.
Installation and operation of the equipment in
hazardous areas must be executed in accordance
to the explosion protection standards of the
respective country.
Installation of intrinsically safe circuits must be
conducted by qualified and authorized personnel
only and in accordance with the instructions of the
contractor.
All electrical connections must be properly
connected.
No modifications are permitted on any equipment
that is certified for use in hazardous areas.
Defective sensors must be returned to the
manufacturer.
Temperature classes and corresponding ambient
temperature ranges:
T6: 55°C = Ta = +75°C
T5: 55°C = Ta = +90°C
T4: 55°C = Ta = +125°C
T3: 55°C = Ta = +190°C
T2: 55°C = Ta = +290°C
T1: 55°C = Ta = +440°C
T710°C: 55°C = Ta = +700°C
X: Specific conditions of use valid for Ex-nA and
Ex-ia:
1. For ambient temperature installation restrictions
of the 6021BE... HT p-Sensor see equipment
description on certificates.
2. Where the integral mineral insulated cable of the
602X HT p-Sensor has ying leads, these must
be provided with a degree of protection of at least
IP20 when installed.
3. Where the integral mineral insulated cable of the
602X HT p-Sensor with type of protection “nA”
has ying leads, these must be provided with a
degree of protection of at least lP54 when
installed.
Operation
6021B_012-002e-06.22 Page 11
II 1G
II 3G
Ex ia IIC T6...T710°C Ga
Ex nA IIC T6...T710°C Gc
SEV 18ATEX0208X
IECEx SEV 18.0038X
4.2.2 Sensor executions
For applications in hazardous areas the sensor version
Type 6021BE… (certified ‚Ex-nA‘ and ‚Ex-ia‘) must be
used.
4.2.3 Marking on sensor
Minimal marking for "Ex-nA" and "Ex-ia"
Due to lack of space the minimal Ex marking is engraved
on the sensor.
Kistler
Type …
SN …
4.2.4 Labeling and packaging
The relevant declaration of conformance and explosion
protection symbols are shown on a label that is affixed to
the product package.
4.2.5 Non-Sparking Equipment (Ex-nA) for use only in zone 2
Fig. 2: Recommended measuring chain for Zone 2 (sensor version Type 6021BE…)
Type 6021BE
Hazardous area (Zone 2)
non-hazardous area
High temperature pressure sensor Type 6021B...
Page 12 6021B_012-002e-06.22
4.2.6 Intrinsically safe equipment (Ex-ia)
4.2.6.1 Intrinsically safe measuring chain, ground isolated (sensor version Type 6021BE)
The sensor can be used in intrinsically safe circuits of the
zones 0, 1 and 2.
For operation in hazardous areas, the sensor must be
operated with an Ex-
certified differential charge
amplifier and a corresponding Ex-certified barrier. The
entity parameters between the sensor and differential
charge amplifier respectively differential charge
amplifier and barrier must be appropriately matched
to each other.
.
4.2.6.2 Entity parameter of the sensor
The entity parameters of the sensors are:
Ui V ≤30
Li mA ≤130
Pi W ≤0,8
Ci pF 15 + 170 pF/m
Li μH 0
Fig. 3: Recommended intrinsically safe measuring chain for Zone 0, 1 and 2
Zone 0,1 or 2
Zone 1 or 2
Hazardous area
Zone 0, 1 or 2
Non-hazardous area
Type 6021BE…
Installation, service & maintenance
6021B_012-002e-06.22 Page 13
5. Installation, service & maintenance
The installation bore as well as the proper installation of
the sensor are essential to guarantee the specifications
of the sensor.
5.1 Mounting bore
Fig. 4: Direct mounting
Fig. 5: Mounting configuration with adapter
Type 6419A21C
* at mounting with
Kistler tools
*
* at mounting with
Kistler tools
High temperature pressure sensor Type 6021B...
Page 14 6021B_012-002e-06.22
Fig. 6: Direct mounting for sensor with 7/16" connector
5.2 Mounting/removal
For the mounting and removal of the sensor generally
accepted engineering practice must be followed.
Particularly when work is performed on electrical and
pneumatic equipment any special applicable safety
regulations must be observed. In Germany this applies
among other for ZH 1/94 "Sicherheitslehrbrief für
Handwerker".
5.2.1 Direct mounting for sensors with Lemo connector or open leads
The sensor is installed with the mounting screw (M12x1)
Type 6419A21A, which is included in the delivery content
of the sensor:
To ensure a leak tight installation the sealing ring Type
1147A21A must be used. It is recommended to use a
new sealing ring each time the sensor is installed.
High temperature screw paste Type 1059 must be
applied to the mounting screw.
* at mounting with Kistler
tools
Installation, service & maintenance
6021B_012-002e-06.22 Page 15
Fig. 7: Blow up drawing for direct installation
(Illustration for the Lemo connector version)
B
i
Fig. 8: Application of torque
High temperature pressure sensor Type 6021B...
Page 16 6021B_012-002e-06.22
Fig. 9: Application of torque in recessed installation
situation
5.2.2 Mounting screw with removal tool
The removal tool Type 6419A21B which facilitates
removal of the sensor is available as an option.
The removal tool can only be applied for the Lemo
connector and the open leads sensor version.
Fig. 10: Pre-assembling the removal tool
Mounting tool, slotted
Type 1251A21A
Installation, service & maintenance
6021B_012-002e-06.22 Page 17
Fig. 11: Torquing of the removal tool pre-assembly
High temperature pressure sensor Type 6021B...
Page 18 6021B_012-002e-06.22
5.2.3 Indirect mounting of the sensor with Lemo connector or open leads
Fig. 12: Mounting with adapter M18x1.5
Fig. 13: Apply final torque
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26