Invacare Platinum IRC9LXO2AWQ User manual

Type
User manual
Invacare® Platinum™ Series
IRC5LXAW, IRC5LXO2AW, IRC5LXO2AWQ, and
IRC9LXO2AWQ
en HomeFill® System Compatible Oxygen
Concentrators with SensO
2
User Manual ................................ 3
da HomeFill® System Kompatibel
iltkoncentratorer med SensO
2
Brugsanvisning .............................. 35
sv HomeFill® systemet kompatibelt
syrgaskoncentrerare med SensO
2
Bruksanvisning .............................. 67
no HomeFill®System Kompatibla
konsentratoren med SensO
2
Bruksanvisning .............................. 99
 HomeFill® System yhteensopiva Happi
keskittimet kanssa SensO
2
Käyttöohje .................................133
This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
© 2017Invacare Corporation
Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout
priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™ and®.Alltrademarksare
ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
Pine-solandLestoilaretrademarksofTheCloroxCompany.
DawnisatrademarkofTheProctorandGambleCompany.
Making Life's Experiences Possible is a registered trademark in the U.S.A.
Contents
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuture
reference.
1General........................................4
1.1Symbols.....................................4
1.2IntendedUse..................................6
1.3Description...................................7
1.4Contraindications..............................7
1.5OptionalAccessories............................7
2Safety..........................................8
2.1LabelLocations................................8
2.2GeneralGuidelines.............................10
2.3RadioFrequencyInterference......................13
3Components....................................14
3.1ComponentIdentification.........................14
4Setup..........................................15
4.1ShippingandHandling............................15
4.2Unpacking....................................15
4.3Inspection....................................15
4.4Storage......................................15
5Usage..........................................16
5.1Introduction..................................16
5.2SelectaLocation...............................16
5.3SettingUptheConcentrator......................18
5.3.1SettingUptheHumidifier......................18
5.3.2TurningtheConcentratorOn...................20
5.3.3Flowrate..................................20
5.3.4SensO
2
OxygenPurityIndicator.................21
5.3.5UnitswithoutSensO
2
........................22
5.4MainPowerLossAlarmSystem....................22
5.5ElapsedTimeMeter(HourMeter)..................22
6Maintenance....................................23
6.1CleaningtheCabinet............................23
6.2CleaningtheCabinetFilter........................24
6.3HumidifierCleaningandThermicDisinfection..........25
6.4CleaningandDisinfectionBetweenPatients............25
6.5PreventiveMaintenanceChecklist...................26
7AfterUse.......................................27
7.1RecyclingInformation............................27
7.2WearandTear................................27
8Troubleshooting.................................28
8.1Troubleshooting...............................28
9Technicaldata...................................30
9.1TypicalProductParameters.......................30
Invacare®Platinum™Series
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious
injury.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious
injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage
orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor
efficient,trouble-freeuse.
SymbolsinDocumentation
Readmanual
NoSmoking
NoOpenFlame
ClassII,DoubleInsulated
AlternatingCurrent
TypeBFequipment
Recycle
DONOTdisposeofinhouseholdwaste
ElectricalHazard
Keepdry
ECRepresentative
Manufacturer
4 1195661-A
General
Callsupplier
**ThisproductcomplieswithDirective
93/42/EECconcerningmedicaldevices.
Thelaunchdateofthisproductisstatedinthe
CEdeclarationofconformity.
FormoredetailsaboutCEmarks,refertoTypicalProduct
Parameters.
SymbolsonProduct
Unitrunning
Unitnotrunning
O
2
Indicators
SymbolO
2
PurityIndicatorLights(LED)
O
2
SYSTEMOKAY
O
2
over85%
GREENIndicatorLight
O
2
between73%to
85%
YELLOWIndicatorLight
A.YELLOWSolid
B.YELLOWFlashingSensor
Failure.Callaqualified
technician
SYSTEMFAILURE
O
2
Below73%
REDIndicatorLight
RefertoTroubleshooting.
LXIndicators
SymbolO
2
PurityIndicatorLights(LED)
I/O
SYSTEMOKAYGREENIndicatorLight
SYSTEMFAILURE
REDIndicatorLight
ContinuousAudibleAlarm
Sieve—GARD™
CompressorShutdown
Callaqualifiedtechnician.
1195661-A 5
Invacare®Platinum™Series
1.2IntendedUse
Youroxygenconcentratorisintendedforindividualusebypatients
withrespiratorydisorderswhorequiresupplementaloxygen.The
deviceisnotintendedtosustainorsupportlife.Theconcentratoris
intendedforusewithinahomeorinstitutionalenvironment.
DANGER!
RiskofInjuryorDeath
Thisproductistobeusedasanoxygensupplementand
isnotintendedtobelife-supportingorlife-sustaining.
ONLYusethisproductifthepatientiscapableof
spontaneousbreath,abletoinhaleandexhalewithout
theuseofamachine.
–DONOTuseinparallelorserieswithotheroxygen
concentratorsoroxygentherapydevices.
DANGER!
RiskofInjuryorDeath
Dependingontheirmedicalcondition,patientsonflow
ratesgreaterthan5l/mmaybeatanincreasedriskfor
seriousinjuryordeathintheeventoffailure.
–ALWAYSdiscussthisincreasedriskwithyourhealth
careproviderBEFOREusingthisproductifyouare
prescribedaflowrategreaterthan5l/m.
DANGER!
RiskofInjuryorDeath
WhileInvacarestrivestoproducethebestoxygen
concentratorinthemarkettoday,thisoxygen
concentratorcanfailtoproduceoxygenduetopower
failureordevicemalfunction.
–ALWAYShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduceoxygen,
theconcentratorwillbrieflyalarmsignalingthepatient
toswitchtotheirbackupsourceofoxygen.Referto
Troubleshootingformoredetail.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduseand
productparametershasnotbeentestedandmayleadto
productdamage,lossofproductfunction,orpersonal
injury.
–DONOTusethisproductinanywayotherthan
describedintheproductparametersandintendeduse
sectionsofthismanual.
6 1195661-A
General
1.3Description
TheInvacarePlatinumconcentratorisusedbypatientswith
respiratorydisorderswhorequiresupplementaloxygen.Thedevice
isnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasrangesfrom87%
to95.6%.Theoxygenisdeliveredtothepatientthroughtheuse
ofanasalcannula.
TheInvacarePlatinumconcentratorusesamolecularsieveand
pressureswingadsorptionmethodologytoproducetheoxygen
gasoutput.Ambientairentersthedevice,isfilteredandthen
compressed.Thiscompressedairisthendirectedtowardoneof
twonitrogenadsorbingsievebeds.
Concentratedoxygenexitstheoppositeendoftheactivesievebed
andisdirectedintoanoxygenreservoirwhereitisdeliveredtothe
patient.
TheInvacarePlatinumconcentratoriscapableofoperationbythe
patientinahomeenvironmentorinaninstitutionalenvironment.
Deviceoperatesatanominal230VAC/50Hertzsupply.
Serviceinformationwillbeavailableuponrequesttoqualified
technicalpersonnelonly.
1.4Contraindications
InvacareisnotawareofanycontraindicationsfortheInvacare
PlatinumConcentratorSeries.
1.5OptionalAccessories
Therearemanydifferenttypesofhumidifiers,oxygentubing,cannulas
andmasksthatcanbeusedwiththisdevice.Youshouldcontactyour
localhomecareproviderforrecommendationsonwhichofthese
deviceswillbebestforyou.Theyshouldalsogiveyouadviceonthe
properusage,maintenance,andcleaning.
Thesupplyaccessories(nasalcannula,mask,tubing,humidifier,etc.)
usedtodeliveroxygentothepatientneedtoincludeameansto
reducethepropagationoffireintheaccessoriesforthesafetyofthe
patientandothers.Ifacommerciallyavailable,fire-activatedflow
stopdeviceisusedintheaccessoriessetup,itshouldbeplacedas
closetothepatientaspossible.
1195661-A 7
Invacare®Platinum™Series
2Safety
2.1LabelLocations
A
SerialNumberLabelislocatedontheresonator
intakeassembly.
B
SpecificationLabelislocatedonthebackofthe
concentratoratthebase.
C
DoubleInsulationLabelislocatedonthebackof
theconcentratoratthebase.
8 1195661-A
Safety
IRC5LXO2AW,IRC5LXO2AWQ,IRC9LXO2AWQIRC5LXAW
D
E
1195661-A 9
Invacare®Platinum™Series
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyandoperationofthe
concentratortheseinstructionsMUSTbefollowed.
DANGER!
RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath,injuryor
damage.Thissectioncontainsimportantinformationfor
thesafeoperationanduseofthisproduct.
–DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand
understandingtheseinstructionsandanyadditional
instructionalmaterialsuchasusermanuals,service
manualsorinstructionsheetssuppliedwiththis
productoroptionalequipment.
–Ifyouareunabletounderstandthewarnings,cautions
orinstructions,contactahealthcareprofessional,
dealerortechnicalpersonnelbeforeattemptingto
usethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproductisnot
workingcorrectly,contactatechnicianorInvacare
forrepair.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,greasy
substancesandothercombustiblesareeasilyignited
andburnwithgreatintensityinoxygenenrichedairand
whenincontactwithoxygenunderpressure.Toavoid
fire,death,injuryordamage:
–DONOTSMOKEwhileusingthisdevice.
–DONOTnearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
–NOSMOKINGsignsshouldbeprominentlydisplayed.
–Avoidcreationofanysparknearoxygenequipment.
Thisincludessparksfromstaticelectricitycreatedby
anytypeoffriction.
–Keepallmatches,lightedcigarettesorothersources
ofignitionoutoftheroominwhichthisconcentrator
islocatedandawayfromwhereoxygenisbeing
delivered.
–Keeptheoxygentubing,cord,andconcentratorout
fromundersuchitemsasblankets,bedcoverings,
chaircushions,clothing,andawayfromheatedorhot
surfacesincludingspaceheaters,stoves,andsimilar
electricalappliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesaletoor
ontheorderofamedicalpractitionerlicensedbya
governmentalagencywherehe/shepractices.
–ONLYalicensedmedicalpractitionermayorderthe
purchaseoruseofthisdevice.
10 1195661-A
Safety
DANGER!
RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,deathor
injurytopersons:
–DONOTdisassemble.Referservicingtoqualified
servicepersonnel.Therearenouserserviceable
parts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the
concentratormustbelocatedinanotherroomat
least7ft(2.5m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereitcan
dropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfalleninto
water.UnplugIMMEDIATELY.
–DONOTusefrayedordamagedACadaptercords.
WARNING!
RiskOfInjuryOrDamage
–Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare
accessories.Accessoriesdesignedbyother
manufacturershavenotbeentestedbyInvacareand
arenotrecommendedforusewithInvacareproducts.
–Therearemanydifferenttypesofhumidifiers,oxygen
tubing,cannulasandmasksthatcanbeusedwiththis
device.Youshouldcontactyourlocalhomecare
providerforrecommendationsonwhichofthese
deviceswillbebestforyou.Theyshouldalsogiveyou
adviceontheproperusage,maintenance,andcleaning.
DANGER!
RiskofInjuryorDeath
Toavoidchokingoringestionofchemicalsfromairway
contamination:
–DONOTusetheconcentratorinthepresenceof
pollutants,smoke,fumes,flammableanesthetics,
cleaningagents,orchemicalvapors.
WARNING!
RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratorisusedby
ornearchildrenorphysically-challengedindividuals.
–Monitorpatientsusingthisdevicewhoareunableto
hearorseealarmsorcommunicatediscomfort.
WARNING!
RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing
entanglement:
–Keepchildrenandpetsawayfromnasalcannulaand
tubing.
–Closesupervisionisnecessarywhenthenasalcannula
isusedbyornearchildrenand/ordisabledpersons.
WARNING!
RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
–Replacethenasalcannulaonaregularbasis.Check
withyourequipmentproviderorphysicianto
determinehowoftenthecannulashouldbereplaced.
–DONOTsharecannulasbetweenpatients.
1195661-A 11
Invacare®Platinum™Series
WARNING!
RiskofInjury
Achangeinaltitudemayaffecttotaloxygenavailableto
you.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingtohigheror
loweraltitudestodetermineifyourflowsettings
shouldbechanged.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorbypulling
onthecord.
–DONOTuseextensioncordswithACpower
adaptersprovided.
–Properlystoreandpositionelectricalcordsand/or
tubingtopreventatrippinghazard.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhenplugged
in.
–Makesureconcentratorisoffwhennotinuse.
WARNING!
RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned
tominimizeroutinepreventivemaintenance.Toprevent
injuryordamage:
–Onlyprofessionalsofthehealthcarefieldor
personsfullyconversantwiththisprocesssuchas
factorytrainedpersonnelshouldperformpreventive
maintenanceorperformanceadjustmentsonthe
oxygenconcentrator,exceptfortasksdescribedin
thismanual.
–UsersshouldcontactyourdealerorInvacarefor
service.
CAUTION!
RiskofDamage
Topreventdamagefromliquidingress:
–Iftheconcentratorisnotworkingproperly,ifit
hasbeendroppedordamaged,ordroppedinto
water,callequipmentprovider/qualifiedtechnicianfor
examinationandrepair.
–NEVERdroporinsertanyobjectorliquidintoany
opening.
–Invacarerecommendstheconcentratornotbeused
intherain.
12 1195661-A
Safety
2.3RadioFrequencyInterference
WARNING!
RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamagefrom
interferencewithwirelessequipment:
–Keepconcentratoratleast9.8ft(3.0m)awayfrom
wirelesscommunicationequipmentsuchaswireless
homenetworkdevices,mobilephones,cordless
phonesandbasestations,walkie-talkies,etc.
ThisequipmenthasbeentestedandfoundtocomplywithEMClimits
specifiedbyIEC/EN60601-1-2.Theselimitsaredesignedtoprovide
areasonableprotectionagainstelectromagneticinterferenceina
typicalmedicalinstallation.
Otherdevicesmayexperienceinterferencefromeventhelowlevels
ofelectromagneticemissionspermittedbytheabovestandards.
Todetermineiftheemissionsfromtheconcentratorarecausing
theinterference,turntheconcentratorOff.Iftheinterference
withtheotherdevice(s)stops,thentheconcentratoriscausingthe
interference.Insuchrarecases,interferencemaybereducedor
correctedbyoneofthefollowingmeasures:
•Reposition,relocate,orincreasetheseparationbetweenthe
equipment.
•Connecttheequipmentintoanoutletonacircuitdifferentfrom
thattowhichtheotherdevice(s)isconnected.
1195661-A 13
Invacare®Platinum™Series
3Components
3.1ComponentIdentification
A
OxygenOutlet
B
Flowmeter
C
OxygenPurityIndicatorLights/FaultandPower
IndicatorLights
D
PowerSwitch
E
CircuitBreaker
F
ElapsedTimeMeter
G
CabinetFilter
RearView
H
PowerCord
I
HomeFillOutletFitting
Accessories(NotShown):HomeFIllhomeoxygen
compressor—IOH200AW
TheHomeFilloutletfittingIistobeusedonlyfor
fillingoxygencylinderswiththeHomeFillhomeoxygen
compressor.Theoutletfittingdoesnotaffectconcentrator
performance.RefertotheHomeFIllusermanualfor
connectionandoperatinginstructions.Whennotinuse,the
plugprovidedwiththeconcentratorshouldbeinsertedinto
theoutletfitting.FormoreinformationabouttheHomeFill,
contactyourInvacaredealer.
14 1195661-A
Setup
4Setup
4.1ShippingandHandling
Beforeyouinstallandoperatetheconcentrator,pleasereferto
PlatinumSeriesUserManual.
ThePlatinumconcentratorshouldALWAYSbekeptintheupright
positiontopreventcabinetdamagewhilebeingtransported.The
shippingcontainerhasbeendesignedtoassuremaximumprotection
oftheconcentrator.
Iftheconcentratoristobereshippedbycommoncarrier,itshould
bepackedinanewcarton.Additionalcartonsareavailablefrom
Invacare.
Theaircompressorsuspensionsystemhasbeenengineeredto
withstandseveremotionandorientation.
4.2Unpacking
1.Checkforanyobviousdamagetothecartonoritscontents.If
damageisevident,notifythecarrier,oryourlocaldealer.
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallthecomponentsfromthecarton.
Unlesstheoxygenconcentratoristobeused
IMMEDIATELY,leaveconcentratorinitspackagingfor
storageuntiluseisrequired.
4.3Inspection
1.Inspect/examineexterioroftheconcentratorfornicks,dents,
cracks,scratchesorotherdamage.
2.Inspectallcomponents.
4.4Storage
1.Storetherepackagedoxygenconcentratorinadryarea.Referto
TypicalProductParametersforstoragetemperatureparameters.
2.DONOTplaceobjectsontopofrepackagedconcentrator.
1195661-A 15
Invacare®Platinum™Series
5Usage
5.1Introduction
Youroxygenconcentratorisintendedforindividualuseindoors.
Itisanelectronicallyoperateddevicethatseparatesoxygenfrom
roomair.Itprovideshighconcentrationofoxygendirectlytoyou
throughanasalcannula.Clinicalstudieshavedocumentedthat
oxygenconcentratorsaretherapeuticallyequivalenttoothertypesof
oxygendeliverysystems.
Yourproviderwillshowyouhowtouseyouroxygenconcentrator.
He/Sheshouldbecontactedwithanyquestionsorproblems
regardingyouroxygenconcentrator.Thisusermanualwilltellyou
aboutyourconcentratorandwillserveasareferenceasyouuse
yourconcentrator.
5.2SelectaLocation
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamagefromairbornepollutants
and/orfumesandforoptimalperformance:
–Locateandpositiontheconcentratorinawell
ventilatedspacesothattheairintakeandtheair
exhaustsarenotobstructed.
–NEVERblocktheairopeningsoftheconcentrator
orplaceitonasoftsurface,suchasabedorcouch,
wheretheairopeningmaybeblocked.
–Keeptheopeningsfreefromlint,hairandsimilar
foreignitems.
–Keepconcentratoratleast12inches(30,5cm)away
fromwalls,draperiesandfurniture.
–Avoiduseinpresenceofpollutants,smokeorfumes,
flammableanesthetics,cleaningagentsorchemical
vapors.
–DONOTuseinacloset.
WARNING!
RiskofInjury
Toavoidinjuryduringtherapy:
–TheoxygenconcentratorMUSTbeplacedonalevel
surfaceforuse.
–DONOTrelocatetheoxygenconcentratorwhilein
use.
Youmayselectaroominyourhousewhereusingyouroxygen
concentratorwouldbemostconvenient.Yourconcentratorcanbe
easilyrolledfromroomtoroomonitscasters.
16 1195661-A
Usage
Youroxygenconcentratorwillperformbestwhenoperatedunder
theconditionsoutlinedinthetablebelow.
Usageinenvironmentsotherthanthosedescribedmayresultinthe
needforincreasedequipmentmaintenance.Theairintakeofthe
unitshouldbelocatedinawellventilatedareatoavoidairborne
pollutantsand/orfumes.
RecommendedGuidelinesforOptimumPerformance
Temperature:
10°C-35°C(50°F-95°F)
Electrical:Noextensioncords.
Placement:
Nocloserthan30,5cm(12
in)fromthewall,furniture,
draperies,orsimilarsurfaces.
TubingandCannula:IRC5LXAW,IRC5LXO2AW,
IRC5LXO2AWQ
2,1m(7ft)cannulawitha
maximum15,2m(50ft)of
Crush-ProofTubing(DONOT
pinch.)
IRC9LXO2AWQ
Recommendeduseupto15.2
m(50ft)highflowtubingwith
highflowcannulaatallflow
rates.
Environment:
Smokeandsoot-free.No
confinedspaces
(Example:Noclosets).
RelativeHumidity:
20to60%
TimeofOperation:Upto24hoursperday.
FlowRate
IRC5LXAW,IRC5LXO2AW,
IRC5LXO2AWQ
0.5L/min.to5L/min.Flow
rateslessthan1L/min.,arenot
recommended.
IRC9LXO2AWQ
1L/min.to9L/min.Flowrates
lessthan1L/min.arenot
recommended.
MinimumOperatingTime:30Minutes
Ensurethatyourconcentratorisatleast30,5cm(12in)
awayfromwalls,draperiesorfurnituretoassuresufficient
airflow.Avoiddeeppilecarpetsandheaters,radiatorsor
hotairregisters.
1195661-A 17
Invacare®Platinum™Series
5.3SettingUptheConcentrator
1.Pluginpowercordtoanelectricaloutlet.
2.ConnectHumidifier(ifsoprescribed).
5.3.1SettingUptheHumidifier
WARNING!
RiskofInjuryorDamage
Toavoidburnsfromsteamorhotwater,inhalationof
waterand/orwaterdamagetotheconcentrator:
–DONOTfillhumidifierbottlewithhotwater.Allow
boiledwatertocooltoroomtemperaturebefore
filling.
–DONOToverfillhumidifier.
–Replacethehumidifiercapandsecurelytighten.
Confirmthatthecapisnotcross-threadedonthe
humidifierbottle.
–DONOTreversetheoxygeninputandoutput
connections.Waterfromthehumidifierbottlewill
travelthroughthecannulabacktothepatientifinput
andoutputconnectionsarereversed.
–Whenusingtubingconnectionslongerthan2.1m
(7feet)inlength,positionthehumidifierasclose
tothepatientaspossibletoallowformaximum
humidificationoutput.
WARNING!
RiskofInjuryorDamage
Failuretoproperlyinstallthehumidifierbottleorother
accessoriestotheconcentratorwillimpacttheflowof
oxygen.
–Tocheckforproperoperationoftheoxygen
concentratorandaccessories,placetheendofthe
nasalcannulaunderthesurfaceofahalf-fullcupof
waterandlookforthebubbles.
–Iftherearenobubbles,checkallconnections
(includinghumidifierbottleandotheraccessories,if
applicable)andrepeat.Contactyourdealerorservice
providerimmediatelyifbubblesstilldonotappear.
FillingtheHumidifierBottleA A
A
1.Removecapfrombottle.
2.FillhumidifierbottleAwithboiledtapwaterorbottledwater
tothelevelindicatedbythemanufacturer.Boiltapwaterfor
approximately10minutesandcooltoroomtemperatureprior
touse.
18 1195661-A
Usage
CAUTION!
RiskofDamage
–Confirmthehumidifiercapisnotcross-threaded
onthehumidifierbottle.
3.InsertaflatheadscrewdriverintheplategrooveBonthetop
edgeofthefilteraccessdoorCandgentlyprythefilteraccess
dooroff.
4.PullupandremovethehumidifierbottleadapterD(nextto
inletfilter).
WARNING!
RiskofInjury
Toavoidimpactingtheflowofoxygen:
–Installthehumidifierbottleand/orother
accessoriesproperly.
5.Attachthehumidifierbottleadaptertothehumidifierbottleby
turningthewingnutonthehumidifierbottlecounterclockwise
untilitissecurelyfastened.
6.WhenplacingthehumidifierbottleinthecompartmentEon
theconcentrator,attachthehumidifierbottle/adaptertubeFto
theoxygenoutletconnectorGontheconcentrator.
Whenusingtubingconnectionslongerthan2.1m(7
feet)inlength,positionthehumidifierasclosetothe
patientaspossibletoallowformaximumhumidification
output.
7.Attachthepatients’nasalcannulasupplytubetothehumidifier
bottleoutletH.
8.Afterassembly,ensurethatoxygenisflowingthroughthenasal
cannula.
9.Replacethefilteraccessdooronthesideofconcentratorby
snappingitbackintotheplategroove.
1195661-A 19
Invacare®Platinum™Series
5.3.2TurningtheConcentratorOn
1.PresspowerswitchAtoOnposition.
Allthepanellightsandtheaudiblealarmwillcomeon
foronesecond,indicatingthattheunitisfunctioning
properly.
5.3.3Flowrate
WARNING!
RiskofInjury
Toavoidinjuryfromexcessoxygenoradeficitof
oxygen:
–DONOTchangetheL/minsettingontheflowmeter
unlessachangehasbeenprescribedbyyourphysician
ortherapist.
–DONOTsettheflowgreaterthan9L/min.
1.TurntheflowrateknobAtothesettingprescribedbyyour
physicianortherapist.
ToproperlyreadtheflowmeterB,locatethe
prescribedflowratelineontheflowmeter.Next,turn
theflowknobuntiltheballCrisestotheline.Now,
centertheballontheL/minlineprescribed.
2.Iftheflowrateontheflowmetereverfallsbelow0.5L/minfor
morethanaboutoneminute,thePotentialObstructionAlert
willbetriggered.Thisisarapidbeepingoftheaudiblealarm.
Checkyourtubingoraccessoriesforblockedorkinkedtubing
oradefectivehumidifierbottle.Afterratedflowisrestoredto
between0.5L/minand0.75L/min.,thePotentialObstruction
Alertwillturnoff.
APotentialObstructionAlertindicatesaconditionthat
maybeassociatedwithapartialorcompleteobstruction
ofoxygenoutput.
TheuseofsomeaccessoriessuchasthePreciseRxâ„¢pediatric
flowmeterandtheHomeFillcompressorwilldeactivatethe
PotentialObstructionAlert.
20 1195661-A
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168

Invacare Platinum IRC9LXO2AWQ User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI