SOMNOmedics SOMNO HD Eco User manual

Category
Blood pressure units
Type
User manual
SOMNO HD Eco
I
nstruction Manual
SOMNOmedics GmbH Am Sonnenstuhl 63 D-97236 Randersacker
Tel. (+49) 931 / 35 90 94-0 Fax (+49) 931 / 35 90 94-49
ha
E-Mail info@somnomedics.de Internet www.somnomedics.eu
2
SOMNOmedics GmbH Am Sonnenstuhl 63 D-97236 Randersacker
Tel.: (+49) 931 / 35 90 94 - 0 Fax: (+49) 931 / 35 90 94 - 49
E-Mail: info@somnomedics.de - Internet: www.somnomedics.eu
Rev. 1
28.10.2019
SOMNO HD eco is a trade mark of the
SOMNOmedics company.
3
Inhalt
Getting started Quick Guide ......................................................................................... 5
Introduction ..................................................................................................................... 7
2.1 Intended use ............................................................................................................ 7
2.2 Patients ................................................................................................................... 7
2.3 Essential performance features of the device .......................................................... 7
2.4 Safety instructions. .................................................................................................. 8
2.5 System components & accessories ....................................................................... 10
2.6 Labels/Symbols ..................................................................................................... 11
3 Preparing the patient: applying main device, sensors and electrodes ........................... 12
3.1 Connecting the sensor plug to the main device ...................................................... 13
3.2 Sensors for Polysomnography ............................................................................... 14
3.2.1 Attaching the effort belts ................................................................................. 14
3.2.2 Nasal Cannula ................................................................................................ 15
3.2.3 SpO2 sensor................................................................................................... 16
3.2.4 Leg EMG sensor ............................................................................................. 16
3.2.5 Mikrophone ..................................................................................................... 17
3.2.6 Flow sensor (Thermistor) ................................................................................ 18
3.2.7 ECG ............................................................................................................... 18
3.3 EEG/ EOG ............................................................................................................. 19
3.3.1 Attaching the PSG headbox according to AASM Guidelines ........................... 19
3.3.2 Application for 32-channel headbox according to 10-20 System ..................... 20
3.4 CPAP sensor and CPAP device ............................................................................ 20
4 Operating the main device ............................................................................................ 21
4.1 Control Elements ................................................................................................... 21
4.1.1 Switch-on /-off of the device ............................................................................ 22
4.1.2 Patient markers/event markers ....................................................................... 23
4.2 Initialisation of a recording ..................................................................................... 23
4.2.1 Initialisation via DOMINO Software ................................................................. 24
4.2.2 Direct initialisation of the device ...................................................................... 26
4.3 Functionality of the LEDs/buttons during the recording .......................................... 27
4.4 Signal view via tablet ............................................................................................. 27
4.5 Impedance check and biocalibration ...................................................................... 28
4.6 Blood pressure calibration ..................................................................................... 28
4.7 The docking station ................................................................................................ 28
4.7.1 Data transfer via docking station ..................................................................... 29
4.7.2 Recharging of the battery ............................................................................... 29
4
4.7.3 Firmware Update ............................................................................................ 29
5 Troubleshooting ............................................................................................................ 29
6 Cleaning and maintenance ........................................................................................... 31
6.1 Cleaning and disinfection ....................................................................................... 31
6.2 Maintenance interval .............................................................................................. 33
6.3 Usage and maintenance of the rechargeable battery ............................................. 33
7 Technical Data .............................................................................................................. 34
8 Service .......................................................................................................................... 41
8.1 Interference ........................................................................................................... 41
8.2 Product life cycle .................................................................................................... 41
8.3 Storage and transport ............................................................................................ 41
8.4 Storage and usage conditions ................................................................................ 41
8.5 Warrenty ................................................................................................................ 41
8.6 Accessories and replacement parts ....................................................................... 42
8.7 Disposal of application parts and batteries and/or the device ................................. 42
8.8 EMV declaration .................................................................................................... 42
8.9 Service hotline and contact .................................................................................... 46
5
Getting started Quick Guide
Control and display elements of the main device
Control elements which are active and selectable are backlit.
1 Power button: with this button the device can be turned on and off. The button is
illuminated continuously red if the device is on. The button is flashing red if the device
is recording and green if in wait mode.
2 Button for starting a recording. Until serial number 76 the button is not illuminated. From
serial number 77 onwards the button is backlight with a green and a red LED. The
button is illuminated in green if a recording with manual recording start is initialised or
the device is ready to start the recording. If the button is illuminated in red this is a
warning for the user that there is an untransferred recording on the device. The
recording can be started anyway.
3 Marker button for indicating events. Backlit with a red LED. If connection between
device and PC is active the LED is flashing red quickly.
4 Light sensor: detects ambient light
5 Battery status:
yellow battery is being charged
green battery is fully charged, docking station is connected to the power supply
switched off docking station is not connected to the power supply
6 Bluetooth: blue light indicates connection with the paired receiver
7 LEDs for sensor ports: either green or red. LEDs are indicating missing/connected
sensor during start of the recording and during the recording. When switching on the
device all LEDs light up shortly and indicate the operational state and battery capacity
of the device. Afterwards, the LEDs turn red and show which sensors are needed for
the next recording. All LEDs turn off after 5 minutes after the recording has started. If a
sensor is accidentally removed during the recording the LED next to the port turns red
and stays red until the sensor is re-connected again. LEDs for flow (9) and abdomen
(10) are green all the time (as soon as they are part of the montage).
8 LED for headbox connector. Functionality identical to 7.
7
10
7
1
2
3
8
9
6
5
4
6
7
Introduction
Thank you for purchasing the SOMNO HDTM eco from SOMNOmedics. Please read the
following instruction manual carefully before installing and using the device. We gratefully
accept any feedback and helpful suggestions to improve our product or instruction manual.
Please find the safety instructions and important information in the chapter “Safety
instructions”, as well as in the individual chapters where they are labelled as follows:
Note: This is an example.
2.1 Intended use
The SOMNO HDTM eco is a mobile device for recording, displaying and saving biophysical
parameters. The device is used as polysomnography system (PSG) for assisting in the
diagnosis of sleep related disorders. The device is used in hospitals, medical offices, sleep
centres, clinics or other research facilities where adults and/or infants are observed. The
system may also be used in ambulatory settings.
The software DOMINO is part of the system and offers the opportunity to record video signals
or signals of external devices. Those signals are displayed synchronously with the SOMNO
HDTM eco signals.
DOMINO is not providing full, automatic diagnosis. The software is supposed to be used by
trained medical experts. DOMINO generates reports of the recorded and manually scored data
(number, index, average values, maximum and minimum, ranges). All values displayed within
the software and on the device itself must be checked and evaluated by trained medical stuff.
Diagnosis must be given after visual evaluation of all recorded data.
This device is not intended to be used as a life support system or together with other
systems in life support applications. The device does not provide an alarm. The device
is not designed to be used as an automatic apnoea monitor. The device is not designed
for use with HF surgical equipment.
2.2 Patients
The SOMNO HDTM eco and its components may only be used as follows:
- At patients as of the first month of life, no premature infants
2.3 Essential performance features of the device
According to DIN EN 60601-1:2013-12 (Edition 3.1) SOMNO HDTM does not have any
essential performance features if used as polysomnography device. The basic safety of the
device must be ensured without the presence of essential features.
For extension of the intended use of the device to neurological applications then the DIN EN
60601-2-26 is applicable. The following essential performance features are defined:
- Accuracy of signal reproduction 201.12.1.101.1
- Input dynamic range and differential offset voltage 201.12.1.101.2
- Input fault 201.12.1.101.3
- Frequency response 201.12.1.101.4
- common mode rejection 201.12.1.101.5
8
Electromagnetic influences, caused by other devices or device with radio equipment, may
affect the essential performance feature or could lead to a loss of the essential performance
features. The waveforms of EEG signal can be affected by an increased noise level, peaks,
offset or other noise.
2.4 Safety instructions.
Evaluation and functional diagnosis of the recorded data is intended to be done by trained
medical experts. For evaluation of the data it is necessary to consider all data together. Signal
quality, interfering signals, dropouts and impedance must be checked previously to the final
diagnosis by the user.
Applications with children must be performed under continuous control.
The SOMNO HDTM eco system may not be used in the course of life support monitoring.
In order to guarantee the correct interpretation of the recorded data the operator should hold
suitable qualifications recognised in their local country.
We recommend that patients with a cardiac pacemaker should wear the main device on the
abdomen rather than on the thorax. We recommend placing the main device at least 16 cm
away from the pacemaker. When following these instructions, no signs of interference of the
cardiac pacemaker with the main device are expected. Additionally, a safe distance needs to
be maintained between the Bluetooth data receiver and patients with cardiac pacemakers or
other implanted stimulation devices.
It is important to deactivate Continuous Impedance Measurement for all EXG channels in
patients that have a cardiac pacemaker or other implanted stimulation devices. Please find a
detailed description regarding this topic in the instruction manual of the DOMINO software, in
the section called Continuous impedance measurement”.
If a sensor cable or the device housing itself is damaged do not use it any longer. A risk of
injury due to electrical voltage or thermal energy may occur. The device housing, all wires,
sensor cables and plug connections must be checked for damage before every use. Damaged
parts should not be used and must be repaired or replaced by trained experts or by persons
authorized by SOMNOmedics.
Warning: Avoid using the device in close proximity to other equipment or in stacked
combination with other equipment. This could lead to improper operation. However, if
necessary, in the manner described above, this device and the other devices should be
observed to verify that they are working properly.
When using the 4-channel analogue optocoupler to record signals from several external
devices, medical devices and non-medical devices should not be connected at the same time
as there is no galvanic separation between the 4 signal inputs.
The main device is not designed to be used in a potentially explosive or combustible
environment.
The main device is not designed for outdoor use and thus must be protected against
temperatures below 0°C and above 50°C. Ingression of dust, dirt and water may damage the
device.
9
Regarding ingression of humidity and water, the main device complies with protection class IP
22. Consequently, the device must be protected against humidity and water.
Should any liquids enter the device, the device must be cleaned immediately by
SOMNOmedics customer service personnel. Following this, the device will undergo a safety-
related examination. Do not switch on the device if any liquids have entered.
Cleaning should only be conducted with a lint-free and damp cloth. Dont use steam autoclave.
Please refer to the information about disinfection (see chapter 6.1 Cleaning and disinfection).
Adhere to the indicated dosages as well as to the application time. Use disposable protective
gloves when using aggressive disinfectants.
Chemicals required for maintenance have to be stored, prepared and kept ready in appropriate
containers to avoid confusion.
When stored for a longer period, the main device should be kept in a closed room to avoid
water condensation caused by temperature fluctuations at high humidity.
Do not use Radio Transmitters and Receivers, High Frequency Devices, CB-Radio Systems,
Cellular Phones, Microwave Ovens, etc. within a distance of 30 cm (in accordance with IEC
60601-1-2) to SOMNO HD™.
If a battery change is required, only batteries specified by SOMNOmedics may be used.
SOMNOmedics cannot guarantee error-free operation if different batteries are used.
Refrain from removing the battery while the device is running as this can result in data loss
and damage of the storage media.
Insert the battery always with the imprinted arrow pointing up and to the back.
Do not put any labels or stickers onto the battery. This could lead to a malfunction of the battery.
Do not put any kind of lable on the battery. Sticking a lable on the battery could lead to a
malefunction of the battery.
Check on the battery regularly. Do NOT use batteries that appear severely swollen.
Please note that the internal time and date will not be saved when storing the device without
a battery. If the device is stored without a battery, it must be initialized before use via the
DOMINO software in order to synchronize time and date.
Repairs, modifications and opening of the main device must be carried out in the
factory. Otherwise, this will lead to immediate loss of warranty.
Any damage of the warranty seal Warranty void, if seal is broken”, located on the back of the
device case, will lead to immediate loss of warranty.
The main device should only be used with micro SD cards specified by SOMNOmedics.
SOMNOmedics cannot guarantee error-free operation if third-party memory cards are used.
The memory card must not be removed during recording as this may lead to data loss.
Since 1st October 1998, batteries may not be disposed of in household waste. The consumer
is obliged to hand used batteries to the manufacturer, authorized dealers or to local collection
10
points. The foundation Das Gemeinsame Rücknahmesystem (GRS) which was founded in
Hamburg organises safe disposal.
When recharging a partly-charged Li-Ion battery, this counts as one whole recharging cycle
and shortens the durability of the battery.
Always recharge the battery with the SOMNOmedics docking station NGA102 as the battery
can otherwise be damaged.
SOMNO HDTM eco should only ever be used with sensors and electrodes specified and
approved for usage with this device by SOMNOmedics.
The docking station of the main device should only be used with the supplied power supply,
XP-Power VEP15US12.
Connect the docking station to a freely accessible socket. In case of a fault it should be
possible to unplug it easily.
When using SOMNOmedics equipment with an external plug-in power supply unit, the
power supply unit should not be placed in direct proximity to the patient.
2.5 System components & accessories
Main Device
Internal channels: body position, movement, flow/snore, CPAP-pressure, RIP effort
abdomen, ambient light, patient marker, battery status
Additional Sensors
PSG Headbox
EEG32 Headbox
RIP effort sensor thorax
SpO2 sensor (SpO2, pulse, pleth)
Limb EMG sensors left and/or right (for paediatric applications and adults)
Microphone
Thermistor
Pneumotachograph
Dual pressure sensor
Accessories
11
Nasal cannula
RIP effort belt abdomen
RIP effort belt thorax
--- Check the current price list for all available options ---
Cable length
Item
Article number
Cable length
USB cable for docking
station (1-port)
NGA102
180cm
USB cable for docking
station (3-port)
NGA103
180cm
Network cable for Bluetooth
receiver
Komp90
200cm
Network cable for Camera
Komp90
200cm
Electrode cable Gold cup
SEN710
75 cm
Electrode cable snap
fastener
SEN018
80 cm
2.6 Labels/Symbols
Information, Symbols, Icons and Classifications on the Type Label of SOMNO HDTM eco
and its accessories:
HF-Transmitter with integrated Bluetooth Protocol.
Read the instruction manual very carefully before you start working with the
SOMNO HD™ eco.
Protection Class: BF
The CE icon shows that the SOMNO HD™ eco complies with the
applicable regulations of the EU and that the conformity was declared by
the manufacturer.
IP22
This device complies with the IP-Protection-Class 22.
Regulatory Compliance Mark (Australia)
Do not dispose electronic devices in the domestic waste.
The power supply meets the requirements of protection class II and does
not need any additional protective conductor connection.
12
The power supply is intended for indoor use only.
3 Preparing the patient: applying
main device, sensors and
electrodes
Note: The main device should be placed on top of
light clothes (nightshirt, pyjamas) and should not
be in direct contact with the skin.
Note: If applying the main device and the sensor in a
different way then shown in the figure this could lead
to wrong results.
Note: Attach the belts and sensor/electrode wires
securely to the patient to avoid strangulation. Fix
the sensor cable with adhesive tape if needed.
Recordings with children may only be performed
while under continuous observation.
We recommend placing the main device in the
abdominal region of the patient. Alternatively, you
can place the device in the thoracic region.
Please take note of the following images in order
to place the main device and the sensors correctly
on the patient’s body. If the main device and the
sensors are placed differently to that shown in
the images, invalid data may be generated.
13
Main device and sensors on the patient’s body
3.1 Connecting the sensor plug to the main device
Please insert the sensor plugs into the main device as shown in the following image:
Inserting a sensor plug into the main device
The device automatically recognises the connected sensors.
The Luer-lock plug of the nasal cannula is screwed directly into the sensor port:
Connecting the luer-lock plug to the main device
Insert the headbox plug into the designated port and position. Turn the safety screw as
indicated by the arrows in the image and indicated on the plug/device.
14
Connecting the headbox plug to the main device
Insert the SpO2 plug into the bottom as shown in the image.
Connecting the SpO2 plug to the main device
Note: In order to prevent damage of the sensor, patient should take off SpO2 sensor while
washing hands or splashing water.
3.2 Sensors for Polysomnography
3.2.1 Attaching the effort belts
Insert the thin end of the effort belt into the slot on the left side of the main device (seen from
the front). Follow the steps shown in the following image:
15
Connecting the effort belt to the main device using the thin end
Wrap the belt around the patient. Insert the wider side of the effort belt to the right side following
the guiding rail.
Connecting the effort belt with the main device using the wider end
Thread the belt into the sensor.
Attaching the effort belts to the patient
3.2.2 Nasal Cannula
Attach the nasal cannula to the patient’s nose. Make sure that the
holes of the sensor elements (highlighted red in the following image) are placed directly below
the nostrils.
16
Note: Nasal cannulas are single-use items.
Attaching the nasal cannula
3.2.3 SpO2 sensor
The finger sensor should be attached to the most suitable finger. The upper side of the sensor
is marked by the embossing of a finger as shown in the following image. You can use adhesive
tape to secure the cable.
Attaching the finger sensor
3.2.4 Leg EMG sensor
The active electrodes (EMG) should be attached to the skin in the area of the anterior tibialis
muscle of each leg between the knee and ankle (see the following image). Use two adhesive
disposable snap electrodes to place it in the correct position. Connect the electrodes with the
sensor before attaching it to the patient. Use adhesive tape to secure the cable to ensure an
accurate measurement and for strain-relief.
The sensors are labelled L and R and should only be applied to the corresponding leg.
Adjust fixing sleeve
in order to secure the
sensor’s position.
17
Attaching the PLM electrodes to the legs
Note: Do not use dried-out adhesive snap electrodes. Only use electrodes provided by
SOMNOmedics to ensure the best signal quality.
Note: When using the leg EMG sensors, always use a ground lead electrode to ensure optimal
signal quality. You can use the following sensors: PSG-headbox, ECG sensor, combi electrode
EEG/EOG.
3.2.5 Mikrophone
Attach the microphone between the larynx and carotid artery to record the snoring sounds.
(DO NOT place the sensor directly on the larynx and not too close to the carotid artery). Ensure
that the cable does not restrict the patient’s breathing or endanger the patient in any way.
Attach the sensor with the flat side of the microphone towards the skin. The patient’s head
should be as straight as possible (looking straight ahead) when attaching the microphone to
prevent discomfort. Check to make sure the patient can turn their head to the left and right
before applying the securing tape.
Attaching the microphone Fixation of the microphone
Strain relief
18
3.2.6 Flow sensor (Thermistor)
The thermistor measures the air flow via temperature changes and can be worn like a nasal
cannula. See the following image to fit the sensor to the patient’s face. Make sure that both
upper sensor elements (highlighted red in the following image) are located directly below the
nostrils and that the lower element is located directly in front of the mouth. Adjust fixing sleeve
(small rubber sleeve) along the tube in order to ensure the sensor’s position during the whole
measurement. The sensor can be adjusted to the contour of the patient’s face by bending it
gently. When using the nasal cannula and the thermistor simultaneously, they can be
connected by using the nasal cannula adapter. Use adhesive tape to attach the cable to the
cheeks for an optimal fit.
Attaching of the thermistors Adapter for thermistor and nasal cannula
3.2.7 ECG
Note: Use disposable snap electrodes only. Clean the skin before attaching the ECG
electrodes.
Note: The main device must not be placed directly on the skin. The cables of the ECG
electrodes should run under the pyjamas/nightshirt. The connecting cable of the corresponding
ECG sensor should run on top of the pyjamas/nightshirt to the main device.
The ECG is applied using the PSG or EEG32 channel headbox.
Please use two snap electrode cables similar to the following figure. Insert the two ECG cables
into the ports labelled ECG on the digital headbox.
Attaching the ECG cables to the headbox (similar for EEG32 channel headbox)
Adjust fixing jacket in
order to secure the
sensor’s position.
19
3.3 EEG/ EOG
Previous to the application clean all skin areas with Nuprep abrasive gel.
Connect the EEG electrodes in accordance to the labelling and colour-coding to the headbox.
Make sure you are using the same type of electrodes for all derivations (e.g. only Ag/AgCl or
only gold cup).
3.3.1 Attaching the PSG headbox according to AASM Guidelines
Position der Elektroden (PSG nach AASM)
GND: Centre of forehead
REF: Applied to Cz, see 10/20 system
EOG l: 1cm above outer canthus of left eye
EOG r: 1cm below outer canthus of right eye
F3/4: See 10/20-System
C3/4: See 10/20-System
O1/2: See 10/20-System
M1/2: Behind left/right ear (mastoid; select boniest area)
EMGref: centred 1 cm above the inferior edge of the mandible
EMG1/2: 2 cm below the inferior edge of the mandible and 2 cm to the left/right of the center.
20
3.3.2 Application for 32-channel headbox according to 10-20 System
Electrode positions (10-20)
Use the cable link between the reference electrode and the second free EMG port for
application of EMG electrodes according to AASM rules
3.4 CPAP sensor and CPAP device
The PAP device is connected via the dual pressure sensor. Therefore, you need the following
items: dual pressure sensor, pneumotachograph (blue) and connecting tube. The
pneumotachograph is connected directly to the dual pressure sensor. Make sure that the white
Luer-Lock of the pneumotachograph is connected in the direction of the end of the CPAP tube
and the also connected with the positive pole of the dual pressure sensor. Use the small
connecting tube to connect the pneumotachograph to the dual pressure to the PAP device. To
the other end of the pneumotachograph connect the end of the masks’ tube. Afterward,
connect the cable of the dual pressure sensor with the main device.
Cable link
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46

SOMNOmedics SOMNO HD Eco User manual

Category
Blood pressure units
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI