Fluke 1555 FC 10 kV Insulation Tester User manual

Category
Multimeters
Type
User manual
September 2010 Rev. 2, 6/20
© 2010-2020 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
1550C/1555
Insulation Tester
Calibration Manual
Limited Warranty and Limitation of Liability
Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The
warranty period is three years and begins on the date of shipment. Parts, product repairs, and services are warranted for
90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does
not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered,
neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that
software will operate substantially in accordance with its functional specifications for 90 days and that it has been
properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without
interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have
no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is
purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves
the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is
submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or
replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization
information, then send the product to that service center, with a description of the difficulty, postage and insurance
prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will be
returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by neglect,
misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures
caused by use outside the product’s specified rating, or normal wear and tear of mechanical components, Fluke will
provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product
will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation
charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY
CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of
incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any
provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction,
such holding will not affect the validity or enforceability of any other provision.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
11/99
i
Table of Contents
Title Page
Introduction ........................................................................................................ 1
Contact Fluke ..................................................................................................... 2
Safety Information .............................................................................................. 2
Required Equipment .......................................................................................... 2
Performance Test Procedures ........................................................................... 4
IR Port Verification Test ................................................................................. 4
Button Test .................................................................................................... 5
Display Test ................................................................................................... 6
Charging Test ................................................................................................ 7
Insulation Accuracy Test ................................................................................ 7
Output Voltage Test ....................................................................................... 9
Short Circuit Current Test .............................................................................. 9
Voltage Measurement Accuracy .................................................................... 10
Adjustment Procedure ....................................................................................... 10
Interface Connection ...................................................................................... 10
Instrument Setup ........................................................................................... 11
Normalizing the HV Probe and Digital Multimeter ......................................... 11
HV Adjustment ............................................................................................... 11
Current Adjustment ........................................................................................ 12
Charge Adjustment ........................................................................................ 12
Additional Procedures ........................................................................................ 12
Identification (Id) ............................................................................................ 12
Clear Non-volatile Memory ............................................................................ 12
Restart DUT ................................................................................................... 12
Shutdown DUT .............................................................................................. 13
Get Diagnostics ............................................................................................. 13
Query Constants ............................................................................................ 13
Save / Print html Page ................................................................................... 13
Battery Replacement Procedure ........................................................................ 14
Disassembly .................................................................................................. 15
Re-assembly .................................................................................................. 15
Cleaning ............................................................................................................. 16
Replacement Parts/Accessories ........................................................................ 16
Specifications ..................................................................................................... 17
General Specifications ................................................................................... 17
Electrical Specifications ................................................................................. 18
Principle of Measurement and Resistance ........................................................ 19
1550C/1555
Calibration Manual
ii
1
Introduction
XW Warning
To prevent possible electrical shock, fire, or personal injury:
Do not perform the verification tests or calibration
procedures described in this manual unless qualified to do
so.
Read all safety information before you use or service the
product.
The Calibration Manual for the 1550C/1555 Insulation Tester (the Product)
provides the following information.
Fluke Contact Information
Precautions and Safety Information
Performance Test Procedures
Adjustment Procedure
Battery Replacement Procedure
Replaceable Parts/Accessories
Specifications
For complete operating instructions and additional safety information, see the
1550C/1555 Users Manual located at www.fluke.com.
1550C/1555
Calibration Manual
2
Contact Fluke
Fluke Corporation operates worldwide. For local contact information, go to our
website: www.fluke.com.
To register your product, view, print, or download the latest manual or manual
supplement, go to www.fluke.com.
Fluke Corporation
6920 Seaway Blvd
Everett, WA 98203
+1-425-446-5500
Safety Information
General Safety Information is in the printed Safety Information document that
ships with the Product and at www.fluke.com. More specific safety information is
listed where applicable.
Required Equipment
Equipment required to perform the procedures in this manual is listed in
Table 1. If the recommended models are not available, equipment with
equivalent specifications may be substituted. Go to www.fluke.com to download
the Users Manual for specific operating instructions.
XW Warning
For safe operation and maintenance of the product, have an
approved technician repair the product.
W Caution
Do not attempt to use the 5500A, 5520A, or other standard
calibrator for insulation and continuity resistance tests.
Calibrator damage will result.
Insulation Tester
Required Equipment
3
Table 1. Required Equipment
Equipment Minimum Required Characteristics
Recommended
Model
HV Probe
6 kV, ±1 % (1000:1 Divider)
11 kV, ±2 % for the 1555
Fluke 80K-6
80K-15
Digital Multimete
r
500 mVdc to 1 V: ±0.02 % Fluke 8508
Load with Guard Terminal
[1]
Resistances
200 kΩ, ±1.25 %, 500 V
500 kΩ, ±1.25 %, 500 V
1 MΩ, ±1.25 %, 1 kV
2.5 MΩ, ±1.25 %, 2.5 kV
5 MΩ, ±1.25 %, 5 kV
10 MΩ, ±1.25 %, 10 kV
1 GΩ, ±1.25 %, 10 kV
100 GΩ, ±1.25 %, 10 kV
200 GΩ, ±1.25 %, 10 kV
500 GΩ, ±5 %, 10 kV
1 TΩ, ±5 %, 10 kV
2 TΩ, ±5 %, 10 kV
Combinations of:
Welwyn F Series,
Welwyn MFP2 Series
And Vishay HTS-523
Capacitors with Bleeder
Resistors
[2]
0.1
μ
F, ±5 %, 500 V, Polypropylene
1 μF, ±5 %, 2.5 kV, Polypropylene
Calibrator
DC current: 2 mA
Accuracy: ±1.25 %
DC Voltage: 0 - 550 V
Accuracy: ±0.005 %
AC Voltage: 0 - 240 V, 60 Hz
Accuracy: ±1.25 %
Fluke 5080,
Fluke 5520A
IR Cable Assembly Fluke P/N 2166275
Calibration Software
[3]
Snorre
Ammete
r
Fluke 8508
Personal computer
IBM compatible, with Microsoft Windows XP
SP2 or later + .NetFramework 2.0 or later
[1] Resistors must have a voltage coefficient consistent with the test voltage used.
[2] Can use (3) each, 0.033 μF, 2 kV capacitors in series and (8) each, 8 μF, 450 V capacitors in series to obtain
required values. The 0.033 μF capacitors should have a 33 MΩ bleeder resistor across each capacitor. The 8 μF
capacitors should have a 15 MΩ bleeder resistor across each capacitor.
[3] Available from www.fluke.com under 1550C/1555 Product Information.
1550C/1555
Calibration Manual
4
Performance Test Procedures
XW Warning
To prevent possible electrical shock, fire, or personal injury, do
not contact the output terminals while performing the following
procedures. There are potentially dangerous voltages at the
output terminals when the product is in the MΩ TEST function.
The following performance tests should be completed yearly to ensure that the
product, referred to as “the DUT” (Device Under Test) in this section of the
manual, is in proper operating condition and meets the published accuracy
specifications. If the DUT fails any of the performance test steps, repair or
adjustment is needed. Refer to How to Contact Fluke for service information.
IR Port Verification Test
To verify operation of the IR Communications Port:
1. Using a Windows PC, connect the IR adapter cable from the product IR port
to the computer COM port.
2. Activate the Snorre program from the Windows Start menu.
3. Select Diagnostic.
4. Select Identification (ID). The PC sends out an ID command and the product
responds to it.
Insulation Tester
Performance Test Procedures
5
Button Test
Use the pushbuttons to control the product, view test results, and scroll through
chosen test results. Table 2 is a list of the pushbuttons and their functionality.
Table 2. Pushbuttons
ENTER
TEST
1 2 3 4 5 6 7
ghh02.emf
Item Description
Turn on and turn off the Product.
Push to go to the Function menu. Push again to exit the Function menu. To scroll within the
Function menu, use the arrow pushbuttons.
Scrolls through test voltages, stored test results, timer duration, and changes test tag ID
characters. Also used to answer “yes” to yes/no prompts.
After a memory location is set, displays the test parameters, test results stored in memory.
These include voltage, capacitance, polarization index, dielectric absorption ratio, and current.
Use to scroll through test voltages, stored test results, timer duration, and memory locations. Also
used to answer “no” to yes/no prompts.
Use for Test Voltage mode to start incrementally setting the test voltage between 250 V and
10 000 V.
Starts and stops a test. Push and hold for 1 second to start a test. Push again to stop a test.
1550C/1555
Calibration Manual
6
Display Test
Turn on the DUT several times while observing the display during power up.
Compare the display with the example in Figure 1. Check all segments for clarity
and contrast.
1
2
3
4
10
1
2
1
1
5
6
7
9
8
gHH03f.EMF
Interference present. Readings could be out of specified accuracy range.
Polarization Index
Dielectric Absorption Ratio
Electrical breakdown in Ramp mode
Ramp mode indicator
Possible hazardous voltage is at the test terminals.
XW Warning
To prevent personal injury, before and after testing, confirm that
the product does not indicate the presence of hazardous voltage.
If the product beeps continuously and there is hazardous
voltage, disconnect test leads and remove power from the circuit
under test.
Voltage sourced by the product or from the circuit under test at terminals of the product.
Test voltage selection (250 V, 500 V, 1000 V, 2500 V, 5000 V, or 10,000 V)
Battery charge status
Bar graph display of insulation resistance
Digital display of insulation resistance
Text display. Shows voltage, test current, capacitance, programmable test voltages, and menu
options.
Figure 1. Display Features
Insulation Tester
Performance Test Procedures
7
Charging Test
1. With the product switched off, connect a mains supply to the ac supply
receptacle and check that the DUT display shows Charging.
2. Disconnect the mains supply and check that the DUT turns off.
3. Turn on the product and see that all the battery symbol segments display as
shown in Figure 1.
Note
A fully-charged battery is indicated when the battery symbol shows
all segments. Recharge the battery as necessary to obtain all
segments. A full charge may require 12 hours.
Insulation Accuracy Test
Using the various resistances shown in Table 3, perform the DUT insulation
accuracy test. Push
for 2 seconds to start or discontinue a test.
Notes
For best results, allow for settling of up to 60 seconds when
measuring high-value resistances (100 GΩ and above) and take
care to avoid stray currents. Perform the test on a conductive
work surface that is connected to the DUT’s GUARD terminal
and the load GUARD terminal.
Motion/body capacitance can affect the stability of the reading at
higher resistances. When taking the measurements above 1 GΩ,
remain as motionless as possible.
1550C/1555
Calibration Manual
8
The capacitance reading is obtained by pressing after a test has started.
Table 3. Insulation Accuracy Test
Step Voltage Range Resistance
DUT Display Limits
Minimum Maximum
1 250 V 0.1 μF 0.055 0.145
2 500 V 250 kΩ 237 kΩ 263 kΩ
3 500 V 1 GΩ 0.95 GΩ 1.05 GΩ
4 500 V 100 GΩ 80 GΩ 120 GΩ
5 1 kV 1 GΩ 0.95 GΩ 1.05 GΩ
6
[1]
2.1 kV 1 μF 0.82 1.18
7 2.5 kV 1 GΩ 0.95 GΩ 1.05 GΩ
8 5 kV 1 GΩ 0.95 GΩ 1.05 GΩ
9 5 kV 100 GΩ 95 GΩ 105 GΩ
10 5 kV 1 TΩ 0.80 TΩ 1.20 TΩ
11 5 kV 5 MΩ 4.75 MΩ 5.25 MΩ
12
[2]
10 kV 1 GΩ 0.95 GΩ 1.05 GΩ
13
[2]
10 KV 200 GΩ 190 GΩ 210 GΩ
14
[2]
10 KV 2 TΩ 1.6 TΩ 2.4 TΩ
15
[2]
10 KV 10 MΩ 9.5 MΩ 10.5 MΩ
[1] Use “Programmable Test Voltage” mode by pushing .
[2] 1555 only
Insulation Tester
Performance Test Procedures
9
Output Voltage Test
In Table 4, the DUT output voltage is checked with various loads applied. In this
test a voltmeter with a high-voltage probe must be connected to the load resistor
to measure the DUT output voltage. Use 15 80K-6 for voltages below 6 kV.
Table 4. Output Voltage Test
Step Voltage Range Load Resistor
Reading Limits
Minimum Maximum
1 250 V 250 kΩ 250 V 275 V
2 250 V No Load 250 V 275 V
3 500 V 500 kΩ 500 V 550 V
4 500 V No Load 500 V 550 V
5 1 kV 1 MΩ 1000 V 1100 V
6 1 kV No Load 1000 V 1100 V
7 2.5 kV 2.5 MΩ 2500 V 2750 V
8 2.5 kV No Load 2500 V 2750 V
9 5 kV 5 MΩ 5000 V 5500 V
10 5 kV No Load 5000 V 5500 V
11
[1]
10 kV No Load 10000 V 11000 V
12
[1]
10 kV 10 MΩ 10000 V 11000 V
[1] 1555 only. Requires 80K-15 probe.
Short Circuit Current Test
To verify the DUT short circuit current, use the following procedure:
1. Connect an ammeter between the DUT + and - terminals.
2. Turn on the DUT and allow to startup.
3. Wait for Test Voltage to appear on the display and set the test voltage to
5000 V by pushing
.
4. Push
and note that the ammeter reading is within the reading limits
referred to in Table 5.
5. Push
to discontinue the test.
Table 5. Short Circuit Current Test
DUT Voltage Range
Reading Limits
Minimum Maximum
1550C 5000 V 1.20 mA 1.80 mA
1555 10000 V 1.20 mA 1.80 mA
1550C/1555
Calibration Manual
10
Voltage Measurement Accuracy
To verify voltage measurement accuracy of the Live Circuit Warning function,
apply the voltages listed in Table 6 to the + and - terminals of the DUT.
Verify:
DUT reading is within the display limits of Table 6.
DUT is beeping at a 1-second interval.
is flashing on the display.
Table 6. Voltage Measurement Test
Step
Voltage Source
Output
DUT Display DUT Tone
DUT Display Limits
Minimum Maximum
1 -38 V dc Flashing Hazard Beeps 30 V 46 V
2 240 V ac, 60 Hz Flashing Hazard Beeps 202 V 278 V
Adjustment Procedure
The product should be performance tested yearly to ensure compliance with its
specifications. When required, use the following adjustment procedure to bring
the DUT within its nominal accuracy specifications.
Interface Connection
Perform adjustment with software using a computer and IR (infrared) adapter.
Connect the Infrared Cable Assembly to the DUT IR Port and COM port of the
computer. Refer to Figure 2.
1
2
ghh04.emf
IR Port
IR Device
Figure 2. IR Port
Insulation Tester
Adjustment Procedure
11
Instrument Setup
Turn on the product and wait for Test Voltage to appear on the display. From the
computer terminal, activate the Snorre program from the Windows Start menu.
On the Setup tab, confirm the selected COM port settings.
XW Warning
To prevent possible electrical shock, fire, or personal injury, do
not contact the output terminals or test equipment terminals
while performing the following procedures. Potentially
dangerous voltages can occur when the DUT is in the "Calibrate
HV Output and Measurement" mode.
Normalizing the HV Probe and Digital Multimeter
1. Connect the HV probe and digital multimeter to the 5520A NORMAL output
terminals, observing polarity. Manually set the multimeter for a range that has
a 10 MΩ input impedance, for example, 100 V, and provides a maximum
resolution for a 500 mV and 5000 mV input.
2. Set the 5520A output to 506 V and note the digital multimeter reading.
Record this value.
3. Set the 5520A output to 1000 V and note the voltmeter reading. If the error is
>0.025 % from the nominal value, convert the error from nominal to
percentage. Multiply 5005 V by this percentage and algebraically add to
5005 V. Record this value.
4. Set the 5520A to standby and disconnect the HV probe and digital
multimeter.
HV Adjustment
1. Select the CAL HV tab.
2. Connect the HV probe and digital multimeter to the output terminals of the
DUT, as shown in the connection diagram.
3. Press the START button to begin adjustment. The DUT briefly displays HV
OFFSET then flashes with PWM 600, while emitting a beep at 1-second
intervals.
4. Use and on the terminal to modify the DUT output value to as close
as possible to the value recorded in step 2 of Normalizing the HV Probe and
Digital Multimeter. The nominal value for this adjustment is between 502 and
510 V.
5. Press the Cal 500 button. The DUT now increases its output of the 1550C to
nominally 5000 V, the output of the 1555 goes to a nominal 10,000 V.
6. Use and on the terminal to modify the DUT output value to as close
as possible to the value obtained in step 3 of Normalizing the HV Probe and
Digital Multimeter. The nominal value for the 1550C adjustment is between
5000 V and 5010 V, the target range for the 1555 adjustment is 10 000 V and
10 020 V.
7. Press the Cal 500 button. The HV generation and measurement functions are
now calibrated.
8. Disconnect the HV probe and digital multimeter from the DUT.
1550C/1555
Calibration Manual
12
Current Adjustment
1. Select the Cal Current tab.
2. Attach a 2 mA current source to the LO and GUARD terminals of the DUT,
connecting the current source LO to DUT GUARD terminal, as shown in the
connection diagram.
3. Apply 2 mAdc to the DUT.
4. Press the START button and wait until the adjustment is complete.
5. The current measurement is now adjusted. Disconnect the current source.
Charge Adjustment
1. Select the Cal Charge tab.
2. Attach a 2 mA current source to the LO and GUARD terminals of the DUT,
connecting the current source LO to DUT Guard.
3. Apply 2 mAdc to the DUT.
4. Press the Start button and wait until the adjustment is complete; progress is
displayed.
5. The charge measurement is now adjusted. Set the current source to
STANDBY and disconnect it from the DUT.
This completes the Adjustment Procedure.
Additional Procedures
Note
The following additional procedures are used during factory
calibration and repair but should not be performed in the field. They
are included for information only.
Various diagnostics are available from the Diagnostic tab as follows:
Identification (Id)
This button returns the model number and firmware version of the unit.
Clear Non-volatile Memory
To delete all saved test results:
1. Push
to call the Function menu.
2. Push or to select the menu item DELETE RESULT.
3. Push
to call the menu item.
4. Push . REALLY DEL? shows on the display.
5. Push to confirm the deletion.
Note
The Delete function deletes all stored test results. Individual test
locations cannot be deleted but are overwritten.
Restart DUT
This button first sets the DUT to CAL_DIAGS mode the sends out the restart
hardware command. The DUT is then restarted.
Insulation Tester
Additional Procedures
13
Shutdown DUT
This button puts the DUT into CAL_DIAGS mode and sends out the power down
command.
Get Diagnostics
Pressing this button continually gets Raw ADC values from the DUT and updates
the Raw ADC Counts boxes (v_counts, i_counts, q_counts), pressing the button
again turns off this feature.
Query Constants
Pressing this button provides an html screen dump of the present opvars
obtained from the DUT. You will need to check that numerical values are
reported for all nine variables. See Figure 3.
snorre 3.bmp
Figure 3. Query Constant Result
Save / Print html Page
The Save button brings up a Windows Save Dialog Box so that the html
document being displayed can be saved to file.
The Print button brings up a Windows Print Dialog Box so that the rendered html
document may be printed.
The window at the bottom of the page is a scrollable log of the Snorre methods
issuing commands to the DUT and the corresponding responses received back
from the DUT.
1550C/1555
Calibration Manual
14
Battery Replacement Procedure
XW Warning
For safe operation and maintenance of the product:
Batteries contain hazardous chemicals that can cause burns
or explode. If exposure to chemicals occurs, clean with
water and get medical aid.
Remove all probes, test leads, and accessories before the
battery door is opened.
Remove all probes, test leads, and accessories before the
case is opened.
The battery door must be closed and locked before you
operate the product.
Use only specified replacement fuses and batteries.
Do not disassemble the battery.
W Caution
To prevent possible damage to the product or to equipment
under test:
Do not attempt to repair or service the product unless
qualified to do so and you have the relevant calibration,
performance test, and service information.
Remove batteries to prevent battery leakage and damage to
the product if it is not used for an extended period.
Be sure that the battery polarity is correct to prevent battery
leakage.
Repair the product before use if the battery leaks.
Do not disassemble the battery.
Do not short the battery terminals together.
Keep cells and battery packs clean and dry. Clean dirty
connectors with a dry, clean cloth.
Do not disassemble or crush battery cells and battery
packs.
Do not keep cells or batteries in a container where the
terminals can be shorted.
Do not put battery cells and battery packs near heat or fire.
Do not put in sunlight.
This Product uses a rechargeable 12 V lead-acid battery for power. Storing
rechargeable lead-acid batteries in a low-charged state could decrease their life
and cause damage. Fully charge the battery before storing it for extended
periods and examine the charge at regular intervals.
Charge the 12 V lead-acid battery with the ac power cord. Expect up to 12 hours
to fully charge the battery. Do not charge in very high or low temperatures.
Insulation Tester
Battery Replacement Procedure
15
Charge the battery if the Tester is not used for extended periods.
If the battery needs to be replaced, use the following procedure to replace the
battery. Spent batteries should be disposed of by a qualified recycler or
hazardous materials handler. Contact your authorized Fluke Service Center for
disposal and recycling information.
Disassembly
W Caution
To prevent possible damage to the product or to equipment
under test, disassembly must be performed using proper ESD
handling techniques. Place the product on an anti-static mat
and use a grounded wrist strap during the following procedure.
1. Disconnect the test leads from any live source and power off the product.
2. Remove the mains supply cable leads from the instrument.
3. Turn over the product and place it on a level surface with feet up.
4. Remove the 4 screws from the case. This frees the top assembly from the
base. The battery is attached to the base.
5. Lift the base from the top assembly and set it on its side next to the top
assembly.
6. Disconnect the red and black leads for the battery.
7. Set the base on its feet and remove the 4 screws from the battery bracket.
8. Remove the bracket.
9. The battery (PN 2803592) can now be removed.
Re-assembly
1. Place the new battery assembly in position and then reinstall the battery
bracket.
2. Reverse steps 2 through 9 of the disassembly procedure to re-assemble.
When reconnecting the red and black battery leads, the red wire must be
connected to the + terminal of the battery. Connect the black wire to the –
terminal.
1550C/1555
Calibration Manual
16
Cleaning
XW Warning
For safe operation and maintenance of the product, remove
excess water from the cloth before cleaning the product to
ensure that water does not enter any terminal.
Periodically wipe the case with a damp cloth and mild detergent. Do not use
abrasives or solvents to clean the product.
Replacement Parts/Accessories
Table 7 is a list of replacement parts.
Table 7. Replacement Parts
Parts Part Number
Test Lead Set Fluke 1550 3477137
10 kV Clip 3611951
1550C 1555 Safety Input Cover 3529198
1550C Top Case 3622602
1555 Top Case 3624655
Input Jacks Decal 3624643
Case Screws
3552926
IR Cable Assembly
2166275
Battery Hold Bracket 3540654
Case Bottom 3524293
Battery 2803592
Rubber Foot 3777953
AC Power Cord (S. Africa) 1552363
AC Power Cord (Australia) 658641
AC Power Cord (UK) 769455
AC Power Cord (Continental Europe) 789422
AC Power Cord (North America) 284174
Soft Carrying Case 3592805
Extended Lead Set (5 kV rating) 2032761
1550C/1555 Quick Reference Card 3592822
ir3000 FC 1550 BLE-IR Adapter (FC kits only)
4460451
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24

Fluke 1555 FC 10 kV Insulation Tester User manual

Category
Multimeters
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI