AC Input voltage range:
200 ~ 240 Vac
Frequency: 47 ~ 63 Hz
To avoid electrical shock connect the protective
grounding conductor of the AC power cord to
an earth ground.
The power switch that is included in the
instrument is not considered a disconnecting
device.
The permanently connected power input is used
as the disconnecting device and shall remain
readily operable.
a. A switch or circuit-breaker must be included
in the installation
b. It must be suitably located and easily reached
c. It must be marked as the disconnecting
device for the equipment.
d. It shall be located near the equipment
Do not position the equipment so that it is
difficult to operate the disconnecting device.
Ask for professional technician for installation.
It requires 200Vac input condition and the
maximum input current [15A (ASR-3200), 22.5A
(ASR-3300), 30A (ASR-3400/ASR-3400HF)],
which conforms to cord diameter by local
regulations.
Breaker, of which the specification is required to
larger than 20A (ASR-3200), 30A (ASR-3300),
40A (ASR-3400/ASR-3400HF) individually,
should be in the near proximity of unit.