1–4 Getting Started
12. Turn the
SENSITIVITY knob counterclockwise 2 clicks, to
10 mV. Verify the Y Output meter displays less than ±20%
deviation. Repeat trimming frequency and depth with the
INPUT FILTER knob to minimize the meter deflection. The
meter should be less than ±10% deviation.
13. Turn
SENSITIVITY 3 clicks clockwise, back to 100 mV. Press
[Y mode] to return both outputs to Lock-In mode.
14. Press the [Type] button twice, to select Band Pass. The X Output
meter should be near −100% deviation, and the Y Output meter
near 0%.
15. Perform an auto-phase adjustment by pressing the [Phase] but-
ton within the AUTO block. After a brief delay, the REFER-
ENCE display should show near 180 deg, and the X Output
panel meter should show +100% deviation.
16. Press the [Type] button again to select High Pass filter. The X
Output panel meter should move near 0 deflection, and the Y
Output meter near −100% deviation.
17. Perform another auto-phase adjustment by pressing [Phase] in
the AUTO block. After the pause, the phase should show a
value near 90 deg, the X Output meter should return to near
+100% deflection, and the Y Output meter should be near zero.
18. Press the [Type] button again to select Low Pass filter. The X
Output meter should swing to near −100% deflection, with the
Y Output meter near zero.
19. Perform another auto-phase adjustment by pressing [Phase] in
the AUTO block. After the pause, the phase should show a
value near 270 deg, and the X Output meter should return to
near +100% deflection. The Y Output meter should remain
near zero.
1.5 Phase sensitive detector
This section walks you through a demonstration of how the phase
sensitive detector (square wave mixer) operates. You will need a
2-channel oscilloscope to view the output waveforms.
A BNC cable should connect the Reference Output (Ref. Out) to
the A/I signal input. Connect the X Out BNC to channel 1 of the
oscilloscope, and the Y Out BNC to channel 2.
1. Restore default settings by pressing [Recall]; turn
REFERENCE if necessary to display ”deFLt”, and then
press [Recall] again.
SR2124 Dual-Phase Analog Lock-In Amplifier