ELMACRON 1000TR Instructions Manual

Category
Measuring & layout tools
Type
Instructions Manual
1000TR
ORP%
Instructions
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 1 -
- Senast utskrivet 2006-10-27
11:41
CONTENTS
1. INTRODUCTION ............................................................................................................................................. 2
1.1 COMMON INTRODUCTION...................................................................................................................... 2
1.2 PARTS & ACCESSORIES ........................................................................................................................... 2
2. INSTALLATION .............................................................................................................................................. 3
2.1 CASING ........................................................................................................................................................ 3
2.2 MOUNTING .................................................................................................................................................3
2.3 ELECTRICAL INSTALLATION................................................................................................................. 3
2.3.1 Connection of the power supply .............................................................................................................4
2.3.2 Connection of the electrode.................................................................................................................... 4
2.3.3 Connection of a recorder........................................................................................................................ 4
2.4 CHECK THE CONNECTIONS.................................................................................................................... 4
3. FUNCTIONS ..................................................................................................................................................... 5
3.1 COMMON FUNCTIONS ............................................................................................................................. 5
3.2........................................................................................................................................................................ 5
3.2.1 MEASURE.............................................................................................................................................. 5
3.2.2 CAL. 1..................................................................................................................................................... 5
3.2.3 OUTPUT ................................................................................................................................................ 5
3.2.4 ALARM ................................................................................................................................................... 5
3.3 KEY FUNCTIONS .......................................................................................................................................6
4. PARAMETER SETTINGS .............................................................................................................................. 7
4.1 DEFAULT SETTINGS ................................................................................................................................. 7
4.2 PARAMETER SETTINGS ........................................................................................................................... 7
4.2.1 Set the output signal ............................................................................................................................... 7
4.3 CALIBRATION............................................................................................................................................ 7
4.4 OPERATION ................................................................................................................................................ 7
4.5 CHANGING THE PARAMETER SETTINGS ............................................................................................ 8
4.5.1 Changing the output signal..................................................................................................................... 8
5. MAINTENANCE .............................................................................................................................................. 9
5.1 HARDWARE CHECK ................................................................................................................................. 9
5.2 PRE-CALIBRATION ................................................................................................................................... 9
5.2.1 Pre-calibration .......................................................................................................................................9
5.3 RESET OF THE DEFAULT SETTINGS ................................................................................................... 10
6. TROUBLE SHOOTING................................................................................................................................. 11
7. SPECIFICATIONS ......................................................................................................................................... 11
APPENDIX
Calibration instruction
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 2 -
- Senast utskrivet 2006-10-27
11:41
1. INTRODUCTION
1.1 COMMON INTRODUCTION
1000TR is a new CE-approved series of transmitters from Elmacron AB. In the series, there are
instruments for pH, Redox%, Redox mV and temperature measurement.
1000TR is easy to program, calibrate and use.
The layout on the instrument is clear; a large display with four LED-segments, diode indication for the
menu and four large function keys. All settings are made through the function keys on the front panel.
When connecting an electrode to 1000TR-ORP% the display shows the actual value in the solution
measured.
1000TR-ORP% is provided with an isolated output 0/4 - 20 mA, proportional to the measured value,
for connection to, for example, a computer, printer or other recording equipment. When measuring is
interrupted, the output stays frozen at the last measured value.
1.2 PARTS & ACCESSORIES
1000TR is delivered without connection cables and electrodes.
Article Function Code number
Power supply cord, standard Power supply 20-R037-000
Low noise coax cable with BNC-connection
( xx* = 01, 02, 03, 05, 07 or 10 meter )
Connection of electrode/probe 60-T810-0xx*
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 3 -
- Senast utskrivet 2006-10-27
11:41
2. INSTALLATION
2.1 CASING
The casing is made of die-cast polystyrol with a transparent front cover that is closed with a snaplock.
The connection terminals are located in a separate partition at the lower part of the transmitter.
At the bottom of the instrument there are four cable glands ( two 15,2 and two 18,6 ) for the electrical
connections.
The protection class is IP65.
2.2 MOUNTING
The instrument is designed for wall mounting. Be sure that the instrument is mounted at a non-
vibrating place.
1000TR is mounted vertically with two fastening screws ( 7) through the mounting brackets on the
instrument.
2.3 ELECTRICAL INSTALLATION
It is recommended that each instrument is provided with a separate power switch.
The electrode coax cable must be protected by a screen and may not be installed near power cables.
Avoid extension of cables.
The connection terminals are located in a separate space at the lower part of the transmitter.
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 4 -
- Senast utskrivet 2006-10-27
11:41
2.3.1 Connection of the power supply
Connect the power supply to terminal 1 (earth), terminal 2 (zero) and terminal 3 (live).
2.3.2 Connection of the electrode
Electrode cabling is a critical part of the whole system.
Use a low noise coax cable between the electrode and the input terminal on the controller.
Low noise cables has, in general, a black semi-conductive sheath between the centrewire
insulation and the shield. Remove this sheath.
Keep the cable away from power wires.
Max recommended length of cable is 10 meters.
Connect the shield ( reference ) of the coax cable to terminal 4 ( EL - ).
Connect the centre ( metall ) of the coax cable to terminal 5 ( EL + ).
NOTE! When using electrode PtPK/G:
connect the shield ( reference ) of the coax cable to terminal 5 ( ELECTR+ )
connect the centre ( metal pin ) of the coax cable to terminal 4 ( ELECTR- )
2.3.3 Connection of a recorder
Connect the recorder to terminal 10 ( REC + ) and terminal 11 ( REC - ).
2.4 CHECK THE CONNECTIONS
Before the power supply is turned on; check that the connections are mechanical and electrical
correct.
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 5 -
- Senast utskrivet 2006-10-27
11:41
3. FUNCTIONS
3.1 COMMON FUNCTIONS
All settings are made through the function keys on the front panel. The keys + and are accelerating
in three steps when they are kept pressed.
Chosen function is indicated by a green diode in the menu, at alarm a red diode is activated at the
same time as the diode at the actual function activates. When pressing M you step down in the menu.
When calibration or setting mode is entered the output signal stays "frozen" at the latest value, until =
is pressed.
3.2
MEASURE
CAL. 1
OUTPUT
ALARM
Offset
0/4 - 20 mA
3.2.1 MEASURE
When power supply is connected the measuring starts..
3.2.2 CAL. 1
Offset calibration. Pre-set value: 5%.
3.2.3 OUTPUT
Can be set to 0 - 20 mA or 4 - 20 mA.
3.2.4 ALARM
When an error occurs in the measuring mode, the alarm diode activates at the same time as the error
code is shown on the display. At serious errors, the error code is flashing on the display.
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 6 -
- Senast utskrivet 2006-10-27
11:41
3.3 KEY FUNCTIONS
KEY FUNCTION 1 ( at parameter setting mode ) FUNCTION 2 ( in measuring mode )
Steps down the menu
Accelerating key.
Increase the value in setting mode.
Used when making a choice according to the
display.
Interrupt measuring together with
Accelerating key.
Decrease the value in setting mode.
Interrupt measuring together with
Confirm performed settings and choices.
Starts measuring.
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 7 -
- Senast utskrivet 2006-10-27
11:41
4. PARAMETER SETTINGS
4.1 DEFAULT SETTINGS
When delivered, 1000TR have the following default settings.
Output 0 - 20 mA
Offset 5%
Slope 100% ( 20 °)
4.2 PARAMETER SETTINGS
All settings has to be confirmed with = to be saved. To leave the setting mode without saving the
settings: press M.
Before starting up the transmitter the following parameters has to be set: Temperature compensation
and output range.
Turn the power on.
4.2.1 Set the output signal
1. Interrupt the measuring by pressing + and at the same time until the measuring is
interrupted and the diode at CAL.1 is activated.
2. Press M until the diode at OUTPUT. is activated. At the display: 0 - 20 .
3. Choose between 0 - 20 and 4 - 20 mA with +.
4. Press = to confirm. The diode at MEASURE is activated and the display shows Hold.
5. Press = to start the measurement.
4.3 CALIBRATION
It is recommended that the instrument is calibrated regularly, to get the highest possible accuracy in
the measuring. The instrument is preset for calibration at 5%.
During the calibration, the latest output signal stays "frozen".
See attachment "Calibration instructions" for calibration.
4.4 OPERATION
Measuring is started when the power supply is turned on. After an interruption off the measuring it can
be restarted by pressing = when the MEASURE diode is activated.
The display shows the actual value in the solution measured at the same time as the corresponding
mA-signal is sent to the output.
During the operation, the diode at MEASURE is activated.
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 8 -
- Senast utskrivet 2006-10-27
11:41
4.5 CHANGING THE PARAMETER SETTINGS
All settings has to be confirmed with = to be saved. To leave the setting mode without saving the
settings: press M.
4.5.1 Changing the output signal
1. Interrupt the measuring by pressing + and at the same time until the measuring is
interrupted and the diode at CAL.1 is activated.
2. Press M until the diode at OUTPUT. is activated.
3. Choose between 0 - 20 and 4 - 20 mA with +.
4. Press = to confirm. The diode at MEASURE is activated and the display shows Hold.
5. Press = to start the measurement.
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 9 -
- Senast utskrivet 2006-10-27
11:41
5. MAINTENANCE
5.1 HARDWARE CHECK
A check of the hardware is necessary only if you suspect that there is something wrong with your
controller.
1. Disconnect the power supply.
2. Short-circuit the electrode input.
3. Connect a mA-meter to the REC-output.
4. Press M and connect the power supply, keep M pressed during 5 seconds until the four
upper diodes are activated.
5. The transmitter performs a self-test according to table 5.1.1
5.1.1 Hardware check
CH01 Test of LED on display All 4 LED-segments are activated
CH02 Test of function keys Press the keys one by one and the corresponding
sign is shown on the display
CH03 Test of output signal Press =. 20.00 is shown on the display at the
same time as the corresponding signal is sent to
the output. Press =.04.00 is shown on the
display at the same time as the corresponding
signal is sent to the output. Press =.
CH04 test of diodes All the menu diodes are activated
The version number is shown, then the instrument starts to measure.
5.2 PRE-CALIBRATION
The transmitter is always pre-calibrated when delivered. Pre-calibration is only necessary when the
instrument has been shut of for a longer period.
It is recommended that the pre-calibration is carried out by means of the calibrator 1000CAL.
Connect the calibrator to the electrode input.
5.2.1 Pre-calibration
1. Turn the power off.
2. Connect the calibrator to the electrode input.
2. Press and turn on the power supply. Keep the key pressed until the display shows
CAL1.
CAL1.CAL1.
CAL1. (10 seconds). Every second diode is activated.
3. Set the calibrator at 5%. Press = to continue.
4. On the display is E05.0
E05.0E05.0
E05.0 shown during 5 s.
5. On the display CAL2.
CAL2.CAL2.
CAL2.
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 10 -
- Senast utskrivet 2006-10-27
11:41
6. Set the calibrator at 85%. Press = to continue.
7. On the display is 085.0
085.0085.0
085.0 shown during 5 s.
8. The instrument returns to measuring mode, the measuring starts automatically.
5.3 RESET OF THE DEFAULT SETTINGS
The settings are reset according to chapter 4.1.
1. Turn off the power supply.
2. Press + and turn on the power supply. Keep the key pressed until the display shows C
CC
C
l
lll
r
rr
r
O
OO
O(10 seconds). The four lower diodes are activated.
3. Press = to continue.
4. The display shows the version number and then the measuring starts up.
G:\Instruktioner\Instruktioner i W ord\1000TR\Engelska\1000TR % Eng.doc
- 11 -
- Senast utskrivet 2006-10-27
11:41
6. TROUBLE SHOOTING
CODE INDICATES PROBABLE CAUSE CORRECTION
E-01 The deviation is > +30% from
the preset value. *
Faulty buffer solution
Old/bad electrode
Check the buffer
solution
Replace the electrode
E-02 The deviation is < -30% from
the preset value*
Faulty buffer solution
Old/bad electrode
Check the buffer
solution
Replace the electrode
* The deviation is calculated in relation to the offsetvalue ( the zero point ). The accepted range for
calibration at 5% is -25% to 35%.
7. SPECIFICATIONS
Instrument
Version
Dimensions 186x131x103
Weight ca 1,0 kg
Max length, electrode cable 10 meter
Display Four 7-segment LED
Keyboard 4 push buttons
Connection Terminals
Power supply 230 VAC, 50 Hz
Backup > 10 years
Range, % 0 - 100%
Accuracy, % ± 1 %
Resolution, % ± 1 %
Offset Range ± 30%
Calibration Offset, 5%
Input ORP-inp z = 1012
Outputs
Output 0 - 20 mA / 4 - 20 mA
Max load, output 270
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12

ELMACRON 1000TR Instructions Manual

Category
Measuring & layout tools
Type
Instructions Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI