Taylor Technologies K-1443 Owner's manual

Type
Owner's manual
InstructionNumber:5312
Component Description
> DROPTEST
> HYDROGENPEROXIDE(1drop=0.5or1%)
COMPONENTS:
1x4026 Dipper,2g,plastic
1x4029 Pipet,Calibrated0.5&1.0mL,plastic
1x4078 Pipet,Graduated,3mL(0.5mLdiv.),
plastic
1x5312 Instruction
2x9198 SampleTubes,Graduated,25mL,plastic
w/cap
1xR‐0601MolybdateReagent,DB
1xR‐0729SulfuricAcid50%*,DB
1xR‐0765PotassiumIodideCrystals
1xR‐0885ThiosulfateReagent,DB
TOORDERREPLACEMENTPARTSANDREAGENTSCALLTOLL‐FREE
800‐TESTKIT(800‐837‐8548).
PROCEDURE:
CAREFULLYREADANDFOLLOWPRECAUTIONSONREAGENT
LABELS.KEEPREAGENTSAWAYFROMCHILDREN.
1drop=0.5%HydrogenPeroxide
1. Usinga3mLpipet(#4078),add2mLwatertobe
testedto25mLsampletube(#9198).Diluteto25
mLmarkwithdistilled,deionized,orhydrogen
peroxide‐freewater.Swirltomix.
2. Usinga2gdipper(#4026),add1dipperR‐0765
PotassiumIodideCrystalstosecond25mLsample
tube.Diluteto10mLmarkwithdistilled,
deionized,orhydrogenperoxide‐freewater.Swirl
untildissolved.
3. Add10dropsR‐0729SulfuricAcid50%*tosecond25
mLsampletube.Swirltomix.
CAUTION:R‐0729SulfuricAcid50%isastrongacid.
Handlecarefully.
4. Usinga1.0mLpipet(#4029),transfer1.0mLfrom
first25mLsampletubetosecond25mLsample
tube.
5. Add10dropsR‐0601MolybdateReagent.Swirlto
mix. Samplewillturnyellow(yellow‐orange)if
hydrogenperoxideispresent.
6. AddR‐0885ThiosulfateReagentdropwise,swirling
andcountingaftereachdrop,untilcolorchanges
fromyellow(yellow‐orange)tocolorless.Always
holdbottleinverticalposition.
7. MultiplydropsofR‐0885ThiosulfateReagentby
0.5.Recordaspercent(%)hydrogenperoxide.
1drop=1%HydrogenPeroxide
1. Usinga3mLpipet(#4078),add1mLwatertobe
testedto25mLsampletube(#9198).Diluteto25
Instructions
mLmarkwithdistilled,deionized,orhydrogen
peroxide‐freewater.Swirltomix.
2. Usinga2gdipper(#4026),add1dipperR‐0765
PotassiumIodideCrystalstosecond25mLsample
tube.Diluteto10mLmarkwithdistilled,
deionized,orhydrogenperoxide‐freewater.Swirl
untildissolved.
3. Add10dropsR‐0729SulfuricAcid50%*tosecond25
mLsampletube.Swirltomix.
CAUTION:R‐0729SulfuricAcid50%isastrongacid.
Handlecarefully.
4. Usinga1.0mLpipet(#4029),transfer1.0mLfrom
first25mLsampletubetosecond25mLsample
tube.
5. Add10dropsR‐0601MolybdateReagent.Swirlto
mix. Samplewillturnyellow(yellow‐orange)if
hydrogenperoxideispresent.
6. AddR‐0885ThiosulfateReagentdropwise,swirling
andcountingaftereachdrop,untilcolorchanges
fromyellow(yellow‐orange)tocolorless.Always
holdbottleinverticalposition.
7. RecorddropsofR‐0885ThiosulfateReagentas
percent(%)hydrogenperoxide.
*WARNING:SulfuricAcid50%(R‐0729)contains62%
sulfuricacid(w/w),acorrosiveacid.
©2016TaylorTechnologies,Inc.•AllRightsReserved•ViewPrivacyPolicy