Taylor Technologies K-1443 Owner's manual

  • Hello! I am an AI chatbot trained to assist you with the Taylor Technologies K-1443 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
InstructionNumber:5312
Component Description
> DROPTEST
> HYDROGENPEROXIDE(1drop=0.5or1%)
COMPONENTS:
1x4026 Dipper,2g,plastic
1x4029 Pipet,Calibrated0.5&1.0mL,plastic
1x4078 Pipet,Graduated,3mL(0.5mLdiv.),
plastic
1x5312 Instruction
2x9198 SampleTubes,Graduated,25mL,plastic
w/cap
1xR‐0601MolybdateReagent,DB
1xR‐0729SulfuricAcid50%*,DB
1xR‐0765PotassiumIodideCrystals
1xR‐0885ThiosulfateReagent,DB
TOORDERREPLACEMENTPARTSANDREAGENTSCALLTOLL‐FREE
800‐TESTKIT(800‐837‐8548).
PROCEDURE:
CAREFULLYREADANDFOLLOWPRECAUTIONSONREAGENT
LABELS.KEEPREAGENTSAWAYFROMCHILDREN.
1drop=0.5%HydrogenPeroxide
1. Usinga3mLpipet(#4078),add2mLwatertobe
testedto25mLsampletube(#9198).Diluteto25
mLmarkwithdistilled,deionized,orhydrogen
peroxide‐freewater.Swirltomix.
2. Usinga2gdipper(#4026),add1dipperR‐0765
PotassiumIodideCrystalstosecond25mLsample
tube.Diluteto10mLmarkwithdistilled,
deionized,orhydrogenperoxide‐freewater.Swirl
untildissolved.
3. Add10dropsR‐0729SulfuricAcid50%*tosecond25
mLsampletube.Swirltomix.
CAUTION:R‐0729SulfuricAcid50%isastrongacid.
Handlecarefully.
4. Usinga1.0mLpipet(#4029),transfer1.0mLfrom
first25mLsampletubetosecond25mLsample
tube.
5. Add10dropsR‐0601MolybdateReagent.Swirlto
mix. Samplewillturnyellow(yellow‐orange)if
hydrogenperoxideispresent.
6. AddR‐0885ThiosulfateReagentdropwise,swirling
andcountingaftereachdrop,untilcolorchanges
fromyellow(yellow‐orange)tocolorless.Always
holdbottleinverticalposition.
7. MultiplydropsofR‐0885ThiosulfateReagentby
0.5.Recordaspercent(%)hydrogenperoxide.
1drop=1%HydrogenPeroxide
1. Usinga3mLpipet(#4078),add1mLwatertobe
testedto25mLsampletube(#9198).Diluteto25
Instructions
mLmarkwithdistilled,deionized,orhydrogen
peroxide‐freewater.Swirltomix.
2. Usinga2gdipper(#4026),add1dipperR‐0765
PotassiumIodideCrystalstosecond25mLsample
tube.Diluteto10mLmarkwithdistilled,
deionized,orhydrogenperoxide‐freewater.Swirl
untildissolved.
3. Add10dropsR‐0729SulfuricAcid50%*tosecond25
mLsampletube.Swirltomix.
CAUTION:R‐0729SulfuricAcid50%isastrongacid.
Handlecarefully.
4. Usinga1.0mLpipet(#4029),transfer1.0mLfrom
first25mLsampletubetosecond25mLsample
tube.
5. Add10dropsR‐0601MolybdateReagent.Swirlto
mix. Samplewillturnyellow(yellow‐orange)if
hydrogenperoxideispresent.
6. AddR‐0885ThiosulfateReagentdropwise,swirling
andcountingaftereachdrop,untilcolorchanges
fromyellow(yellow‐orange)tocolorless.Always
holdbottleinverticalposition.
7. RecorddropsofR‐0885ThiosulfateReagentas
percent(%)hydrogenperoxide.
*WARNING:SulfuricAcid50%(R‐0729)contains62%
sulfuricacid(w/w),acorrosiveacid.