Thermo Fisher Scientific SITE User guide

Type
User guide
Site Guide Addendum
Part Number: 1013117 C Release Date: October 2004
Site Guide Addendum
The information in this addendum applies to all instrument line site guides that are
included on the Operators Documentation CD, part number 1008281.
Source Exhaust
The requirements in this section apply to the API 4000™ and 4000 Q TRAP® instruments.
Source Exhaust Requirement
The source exhaust gas requirement is between a minimum of 55 psi and a maximum
60 psi.
Ventilation
The requirements in this section apply to all instruments.
Venting Device Operation
The devices for venting the instrument must operate properly whenever the instrument
power is on, waste is in the waste container, or reagents are being used on the instrument.
WARNING! The lab ventilation system integrity is vital for safe operation of the
instrument.
Ventilation Device Requirement
The laboratory ventilation device should produce an average airflow velocity of
30 linear m/min, or 100 linear ft/min.
Air Exchange
The air exchange requirement in this section applies to all instruments.
Air Exchange Requirement
For safe operation, the instrument must be placed in a properly ventilated laboratory
environment.
Note: As a guideline for a typical laboratory environment, the outdoor air exchange
requirements for ventilation can be referenced from the ANSI/ASHRAE Standard
62-2001, Ventilation for Acceptable Indoor Air Quality.
Based on an estimated maximum occupancy of 30 people per 1000 ft2, or 100 m2, the
outdoor air exchange requirement is 20 cfm/person.
Site Guide Addendum
2
Room
The room requirement in this section applies to the API 4000 and 4000 Q TRAP
instruments.
Room Requirement
The minimum suggested bench size is 75 cm H × 75 cm W × 150 cm L, or
29 × 29 × 59 in., instead of 150 cm H × 75 cm W × 75 cm L as indicated in the site guide.
Computer
The computer requirement in this section applies to all instruments.
Network Connections Requirement
To ensure a speedier installation, the customer or network administrator should provide a
computer name and domain name to the CSR at time of installation.
Electrical
The electrical requirement in this section applies to all the QSTAR® instruments.
Electrical Requirement
A 207 to 242 VAC 50/60 Hz, 30A grounded, single phase circuit with one receptacle
should be used to supply the power requirements for the QSTAR instrument. The required
receptacle is an IEC 309, 32A, 3-pin, 230 VAC receptacle. It must be located within 1.8 m
(6 ft) of the instrument.
IEC 309 plug
Note: If you require the IEC 309, 30A, 3-pin 250 VAC receptacle (compatible with
QSTAR series instruments), it is available, free of charge, by ordering p/n 1007134
through your sales/service representative.
Site Guide Addendum
3
Electrical
The electrical requirement in this section applies to all API instruments.
Electrical Requirement
Within North America, use NEMA 6-15R receptacles and NEMA 6-15P plugs. Outside of
North America, use local standards. Plugs for these receptacles must be provided by the
customer. Each circuit must have two receptacles.
NEMA 6-15P plug
UPS The UPS information in this section applies to all instruments.
General UPS and Power Conditioner Requirements
See the following tables for general UPS and power conditioner requirements for all
instruments.
General UPS and Power Conditioner Requirements
Output voltage 207–242 (AC) True on line (sometimes called double conversion)
Frequency 50/60 Hz
Waveform Pure Sine Wave
Maximum peak current 3 × Nominal Current (MIN)
Output voltage distortion < 3%
Output protection Circuit Breaker
Minimum power requirement See the following table.
Site Guide Addendum
4
Electrical Requirements
Product
Line
Operating
Voltage
(VAC) Phase Current
Max Frequency Number of
Receptacles Type of
Receptacle Operating
Current
Minimum
Power
Requirement
QSTAR 207–242 Single 30 A 50/60 Hz 1 IEC 309,
32 A, 3 pin,
230 V AC
20 A 8000 VA
4000
QTRAP 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
4.5 A
4400 VA
API 4000 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
4.5 A
4400 VA
API 3000 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
4.5 A
4400 VA
Q TRAP 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
6 A
4400 VA
API 2000 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
6 A
4400 VA
API 365 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
4.5 A
4000 VA
API 300 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
4.5 A
4000 VA
API 350 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
4.5 A
4000 VA
API 165 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
4 A
Roughing pump
4.5 A
4000 VA
API
150EX 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
4.5 A
4000 VA
API
150MCA 207-242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
4 A
Roughing pump
4.5 A
4000 VA
API 100 207–242 Single 15 A 50/60 Hz 2 Nema 6–15R Main Console
6 A
Roughing pump
4.5 A
4000 VA
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

Thermo Fisher Scientific SITE User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI