7
GG-SO2
Operation
Start-Up
Allow 60 seconds for power up time delay of sensor
to end (green power LED will ash during power up).
Sensor can be response tested after power-up, but
best to wait 1 hour to zero or span sensor after power
up.
Start-Up Test:
One person exposes each sensor to calibration
gas. The second person stays at the control unit to
determine that each sensor, when exposed to the
target gas, is connected to the proper input and
responds, causing appropriate alarm functions.
LED functions (see Figure 3 on page 8)
Calibration
The GG-SO2 sensor comes factory calibrated and
should require only minimal adjustments after
installation. There are two pots on the preamp that
are used for calibration (see Figure 3).
Note: Never measure sensor output in mA. Always
use mVDC or VDC voltmeter settings.
Zero Calibration: After the sensor is installed and
has been powered up for 1 hour, the unit can be zero
calibrated by the following:
• Be sure the unit is in clean air. If unsure, apply Zero
Air gas at 0.5 to 0.8 L/min.
• Adjust the Zero pot until the sensor outputs 40 mV
from Test [-] to Test [+] (see Figure 3).
Span Calibration: If span adjustment is required, use
the following procedure:
• To enter calibration mode (disables all ltering and
averaging), turn the Zero pot clockwise 1/4 turn and
then back again. Successful entry into calibration
mode will be indicated by the Power (green) LED
blinking twice per second. Calibration mode will
time out automatically after 4 minutes.
• Apply span gas at 0.5 to 0.8 L/min (span gas must
be in air, not nitrogen or other carrier).
• Sensor should react to gas within 15 seconds.
• Once the output signal has peaked (or 3 minutes
maximum) adjust the Span pot until the correct
output is achieved (see Figure 3). With full-scale
span gas, the calculated span value is 200 mV.
((span gas / sensor range * 16 + 4) (mA output))
Note: Below are a few response characteristics which
may be an indication that the gas sensor is at or near
the end of its useful life. If any of these are observed,
the cell should be replaced:
• Slow response to / recovery from calibration gas.
• Failure of the output to reach 50% of the calibration
gas value prior to span adjustment if last calibration
was 6 months ago or less.
• Unable to achieve correct output during span.