Medisana MTX connect Owner's manual

Category
Blood pressure units
Type
Owner's manual
DE

Oberarmmanschette mit
Luftschlauch

Display

START-Taste

MEMO-Taste

Blutdruck-Indikator

USB-Schnittstelle

BLUETOOTH-/ON/OFF-
Schalter

TIMER-Taste

Anschluss fÃŒr Luftschlauch

Anschluss fÃŒr Netzteil

Benutzerspeicher 1 / 2

Bluetooth-Symbol

Uhrzeit / Datum

Anzeige des systolischen
Drucks

Anzeige des diastolischer
Drucks

Anzeige der Pulsfrequenz

DatenÃŒbertragungs-Symbol

Speicher-Symbol

Batterie schwach
Bitte neue Batterien
einlegen

Alarm-Symbol

Symbol fÃŒr 3MAM-Modus

Count down-Symbol

GerÀt hat Arrhythmie fest-
gestellt

Blutdruck-Balkenanzeige

Puls-Symbol
GB

Upper arm cuff with hose

Display

START button

MEMO button

Blood pressure indicator

USB Interface

BLUETOOTH/ON/OFF
switch

TIMER button

Connector for the hose

Socket for mains adaptor

User memory 1 / 2

Bluetooth symbol

Time / Date

Display of systolic pressure

Display of diastolic pressure

Display of pulse rate

Data transfer symbol

Memory symbol

Batteries low
Please replace batteries.

Alarm symbol

Symbol for 3MAM mode

Countdown symbol

Unit has detected an
arrhythmia

Blood pressure bar display

Pulse symbol
NL

Bovenarm-manchet met
luchtslang

Display

START-toets

MEMO-toets

Bloeddrukindicator

USB Interface

BLUETOOTH-/ON/OFF
schakelaar

TIMER-toets

Aansluiting voor luchtslang

Aansluiting voor voedings
apparaat

Gebruikersgeheugen 1 / 2

Bluetooth symbol

Tijd / datum

Aanduiding van de
systolische druk

Aanduiding van de
diastolische druk

Aanduiding van de polsslag

Gegevensoverdracht

Geheugensymbool

Batterij zwak
Nieuwe batterijen erin
leggen s.v.p.

Alarmsymbool

Symbool voor 3MAM-modus

Count down-symbool

Apparaat heeft aritmie
vastgesteld.

Bloeddrukbalkindicatie

Polsslag-symbool
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:16 U
18
1 Safety Information
GB
IMPORTANT INFORMATION!
RETAIN FOR FUTURE USE!
Explanation of symbols
This instruction manual belongs to this device.
It contains important information about starting up
and operation. Read the instruction manual thoroughly.
Non-observance of these instructions can result in
serious injury or damage to the device.
WARNING
These warning notes must be observed to prevent
any injury to the user.
CAUTION
These notes must be observed to prevent any
damage to the device.
NOTE
These notes give you useful additional information
on the installation or operation.
Classification: type BF applied part
Lot number
Manufacturer
Date of manufacture
Read the instruction manual carefully before using this device, especially the
safety instructions, and keep the instruction manual for future use.
Should you give this device to another person, it is vital that you also pass on
these instructions for use.
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
19
GB
1 Safety Information
WARNING
Do not take any therapeutic measures based on your own
measurements! Never change the quantity of medication
prescribed by your doctor!
•
Irregularity of pulse or arrhythmia can lead to difficulties in recording a
correct reading when measurements are taken using oscillometric blood
pressure devices. This device is elec tronically equipped to detect over 20 of
the most commonly occurring types of arrhythmia and movement artifacts
and indicates this with a symbol in the display.
•
If you should suffer from conditions such as arterial occlusive disease, please
consult your doctor before using the unit.
•
The unit may not be used to check the heart rate of patients with a
pacemaker.
•
Pregnant patients should take their own personal state of health into
consideration and take all necessary precautions before using the unit. If in
doubt, consult your doctor.
•
Should you experience discomfort or other complaints during measurement,
for example pain in the upper arm, please take the following action: Press the
START button

to deflate the cuff immediately. Loosen the cuff and
remove it from your arm. Please contact either your dealer or MEDISANA
directly.
• Patients with a weak pulse wave should raise their arm and open and close
their hand around 10 times before taking a reading. This exercise optimizes
the pulse wave and the measuring process.
• The unit is only to be used for the specific purpose described in this
instruction manual.
• Any misuse will void the warranty.
• On the rare occasion of a fault causing the cuff to remain fully inflated
during measurement, open the cuff immediately.
• Do not use the unit near to equipment emitting a strong electric or
magnetic field (such as radios). This may impair the correct functioning of the
unit. Keep a distance of at least 3,3 m to such devices during the measurement.
• This device is not designed to be used by persons (including children) with
limited physical, sensory or mental abilities, or by persons with insufficient
experience and/or knowledge, unless under observation by a person
responsible for their safety, or unless they have been instructed in the use of
the device. The tube may present the risk of strangulation.
• Children must be supervised to ensure that they do not play with the device.
• If a fault occurs, do not try to repair the unit yourself. Attempts to do so will
void the warranty. Refer all servicing to authorized service personnel.
• Protect the unit from moisture. Should moisture enter the unit, remove the
batteries and stop using it immediately. In this case, please contact either
your dealer or MEDISANA directly. You can find details of how to contact
us on the address page.
Safety
Information
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
20
2 Useful Information
GB
Thank you for your confidence in us and congratulations on your purchase!
You have acquired a quality product with your purchase of the MEDISANA
MTX Connect upper arm blood pressure monitor.
To ensure the best results and long-term satisfaction with your MEDISANA
MTX Connect with bluetooth, we recommend that you read the following
operating and maintenance instructions carefully.
Please check first of all that the device is complete and is not damaged in any
way. In case of doubt, do not use it and contact your dealer or your service
centre.
The following parts are included:
• 1 MEDISANA MTX Connect with bluetooth
• 1 Preformed sleeve with air pipe
• 4 Batteries (AA type, LR 6) 1,5V
• 1 Storage bag
• 1 Instruction manual
The packaging can be reused or recycled. Please dispose properly of any
packaging material no longer required. If you notice any transport damage
during unpacking, please contact your dealer without delay.
WARNING
Please ensure that the polythene packing is kept away from the
reach of children! Risk of suffocation!
Blood pressure is the pressure created in the blood vessels by each heart beat.
When the heart contracts (= systole) and pumps blood into the arteries, it
brings about a rise in pressure. The highest value is known as the systolic
pressure and is the first value to be measured. When the cardiac muscle
relaxes to take in fresh blood, the pressure in the arteries also falls. Once the
arteries are relaxed, a second reading is taken – the diastolic pressure.
The MTX Connect is a blood pressure unit which is used to measure blood
pressure at the upper arm. The measurement is carried out by a
microprocessor, which, via a pressure sensor, measures the vibrations resulting
from the inflation and deflation of the cuff over the artery.
MEDISANA has had many years of experience in the field of blood pressure
measurement. The high precision of MEDISANA devices has been confirmed
by extensive clinical tests, carried out according to strict international
standards. One strong argument in favour of taking blood pressure at home is
the fact that the readings are taken in familiar surroundings under relaxed
conditions. The most important reading is the socalled „baseline value“, the
2.2
What is blood
pressure?
2.1
Items supplied
and packaging
2.3
How is blood
pressure
measured?
2.4
Why should
you take your
blood pressure
at home?
Thank you
very much
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
21
GB
2 Useful Information
2.5
Blood
pressure
classification
one taken before breakfast, directly after getting up in the morning. As far as
it is possible, you should always try to take your blood pressure at the same
time and under the same conditions each day.
This ensures comparable results and helps in detecting the first signs of high
blood pressure. If high blood pressure remains undetected, it increases the risk
of further cardiovascular diseases.
THEREFORE WE RECOMMEND:
Take a blood pressure reading daily and on a regular basis,
even if you are experiencing no apparent complaints.
The following table shows the classification of high and low blood pressure,
without taking age into consideration. This blood pressure assessment scale
corresponds to the guidelines of the World Health Organisation (WHO).
Low blood pressure
systolic below <100
diastolic below <60
Normal blood pressure (green display area

)
systolic 100 – 139
diastolic 60 – 89
Types of high blood pressure
mild hypertension medium hypertension serious hypertension
(yellow (orange (red
display area

) display area

) display area

)
systolic 140 – 159 systolic 160 – 179 systolic
>_
180
diastolic 90 – 99 diastolic 100 – 109 diastolic
>_
110
WARNING
Blood pressure that is too low represents just as great a health
risk as blood pressure that is too high! Fits of dizziness may lead
to dangerous situations arising (e.g. on stairs or in traffic)!
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
22
2 Useful Information / 3 Getting Started
GB
2.6
Fluctuations in
blood pressure
2.7
Influencing
and evaluating
readings
3.1
Inserting /
removing
batteries
There are many factors that can influence blood pressure. Readings can be
affected seriously by stress, fear, heavy physical exertion or the time of day at
which the measurements are taken. Your personal blood pressure values are
subject to considerable fluctuation during the course of the day and the year.
These fluctuations are particularly pronounced in the case of patients suffering
from hypertension. Blood pressure is usually at its highest during physical
exertion and at its lowest at night when you are asleep.
• Measure your blood pressure several times, then record and compare the
results. Do not draw any conclusions from a single reading.
• Your blood pressure readings should always be evaluated by a doctor who is
also familiar with your personal medical history. When using the unit
regularly and recording the values for your doctor, you should visit the
doctor from time to time to keep him updated.
• When taking readings, remember that the daily values are influenced by
several factors. Smoking, consumption of alcohol, drugs and physical
exertion influence the measured values in various ways.
• Measure your blood pressure before meals.
• Before taking readings, allow yourself at least five minutes rest.
• If the systolic and diastolic readings seem unusual (too high or too low) on
several occasions, despite correct use of the unit, please inform your doctor.
This also applies to the rare occasions when an irregular or very weak pulse
prevents you from taking readings.
Insertion: Open the battery compartment on the underside of the unit by
exerting gentle pressure in the direction indicated by the arrow, and insert the
four batteries included with the unit (alkaline batteries, AA type, LR 6).
Remember to insert the batteries as indicated in the diagram in the battery
compartment. Once the batteries are inserted correctly (or once the power unit
is plugged in), you should hear 2 beeps.
When the ON/OFF switch is activated

, you will hear 2 beeps and all the
symbols appear on the display for one second. Then the following appears:
“
- : - -
“. Replace the lid, pressing it until it snaps into place.
Removal:
If the voltage in the batteries drops below 5 - 4.5 V, the display shows the
battery symbol

. Switch the device off using the ON/OFF switch

and insert new batteries. Should the used batteries be jammed in the
compartment, use a pointed instrument, such as a ballpoint pen, to carefully
prise them out.
As soon as the voltage in the batteries drops below 4.25 - 0.25 V, the display
shows the battery symbol

. The device automatically switches off
immediately after this symbol appears. The device will not function until new
batteries are inserted.
3 Getting Started
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
23
GB
3 Getting Started
3.2
Using a mains
adaptor
Note that you have to reset the date and time each time you replace the
batteries, unless this is set automatically via the radio clock when the device
is switched on using the ON/OFF switch

. The previous measuring results
are saved.
WARNING
BATTERY SAFETY INFORMATION
•
Do not disassemble batteries!
• Clean the battery and device contacts if necessary before
putting in the batteries!
• Remove discharged batteries from the device immediately!
• Increased risk of leakage, avoid contact with skin, eyes and
mucous membranes! If battery acid comes in contact with
any of this parts, rinse the affected area with copious
amounts of fresh water and seek medical attention
immediately!
• If a battery has been swallowed seek medical attention
immediately!
• Replace all of the batteries simultaneously!
• Only replace with batteries of the same type, never use
different types of batteries together or used batteries with
new ones!
• Insert the batteries correctly, observing the polarity!
• Keep the battery compartment well sealed!
• Remove the batteries from the device if it is not going to
be used for an extended period!
• Keep batteries out of children's reach!
• Do not attempt to recharge these batteries! There is a
danger of explosion!
• Do not short circuit! There is a danger of explosion!
• Do not throw into a fire! There is a danger of explosion!
• Keep unused batteries in their packaging away from metal objects
in order to prevent short circuiting!
• Do not throw used batteries into the household refuse;
put them in a hazardous waste container or take them to
a battery collection point, at the shop where they were
purchased!
Alternatively, you can also run the device using a special power supply unit
(MEDISANA item no. 51036 not supplied). Connect it to

on the back of
the device. In this case the batteries remain in the unit.
Inserting the connector into the back of the blood pressure monitor switches
the batteries off mechanically. It is necessary, therefore, to plug the mains
adaptor into the mains socket first of all and then to connect it to the blood
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
24
3 Getting Started
GB
3.3
Setting time
and date
pressure monitor. If the blood pressure monitor is no longer being used, first
remove the connector from the blood pressure monitor and then pull the mains
adaptor out of the mains socket. Doing this means that the date and time will
not have to be reset each time.
1. You can set the date and time by pressing down the TIMER button

for
several seconds until appears on the display.
2. The selected user memory appears on the display together with the
flashing year.
- Press the TIMER button

again to display the date; the month starts
to flash.
- Press the TIMER button

again and the day starts to flash.
- Press the TIMER button

again to display the time. The hour starts to
flash.
- Press the TIMER button

again and the minutes start to flash.
- You can set the current date and time (year, month, day, hours and
minutes) by pressing the MEMO button

and confirm the date and
time with the TIMER button

. Then the date and time you set are
shown twice each in succession before the device returns to standby
mode. In standby mode, the time is displayed permanently.
NOTE
If you do not wish to set the time or you want to take a
measurement immediately, exit the time setting mode by
pressing the START button

.
Your MEDISANA MTX Connect blood pressure meter is equipped with an
alarm function, with which the two users can set two alarm times each, in
order to take medication are regular intervals, for example.
Make sure that the device is in standby mode and your own user memory

has been selected.
1. Press the TIMERbutton

followed by the MEMO button

. Hold down
both buttons until the alarm symbol

appears on the display.
2. Release both buttons. “ 1 “ flashes at the bottom right of the display for
Alarm 1. You can use the MEMO button

to select Alarm 1 or Alarm 2.
3. Press the TIMER button

. The display for entering the alarm time flashes.
Use the MEMO button

to enter the alarm time.
4. Press the TIMER button

and the alarm symbol flashes. Use the MEMO
button

to switch between alarm on
()
and alarm off ( )
.
To confirm
your entries, press the TIMER button .
3.4
Setting the alarm

51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
25
GB
3 Getting Started
NOTE
If you have correctly entered an alarm time and the alarm is
active, the alarm symbol appears in standby mode. If no
alarm is active for either of the user memories, no symbol
appears in standby mode.
3.5
Setting up the
user memory
The MTX gives you the opportunity to assign readings to one of two user
memories, each of which provides 99 memory slots. Switch the device to standby
mode. Press the TIMER button

to select either user memory 1 or user
memory 2

.
With the MTX, you can choose between two measuring modes.
• Single reading
•
3 consecutive readings ( Measurement Average Mode = 3/MAM-mode )
In the case of a single reading, just one blood pressure reading is taken and
stored.
The 3/MAM-mode is to be recommended if you suffer from severely
fluctuating blood pressure. 3 successive readings are taken at 15 second
intervals, thus eliminating widely varying readings. In the case of serious
deviations, a 4th reading may be taken.
From these three measurements, a weighted average value is calculated,
displayed and stored in the selected memory. A reading taken in this way is
indicated in the memory by an

.
The 3/MAM procedure greatly minimizes the risk of an incorrect reading.
1. To switch between single measurement and
3 consecutive measurements
(3MAM procedure), press and hold
the TIMER button

until the symbol
for 3MAM mode and “ “ appear flashing in the display.
2. Press the MEMO button

to select either “ “ or “ “; “ “
for the 3MAM measuring method and “ “ for single measurement.
3. Confirm your selection by pressing the START button

(or press the
TIMER button

to make the next setting: “ Date time setting “).
3.6
Selecting
the 3MAM
measuring
method
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
26
4 Operation
GB
4.1
Fitting the cuff
4.2
Correct position
for taking readings
4.3
Taking blood
pressure readings
First attach the free end of the cuff’s

air tube to the designated connector

on the unit. Open up the preformed cuff and slip it over the bare left upper
arm just above the elbow region. If wearing tight clothing, ensure that the
circulation in the arm is not impaired when rolling up the sleeve.
The tube should run along the centre of the inner arm towards the palm of the
hand. Ensure that the cuff itself is not twisted in any way.
Fasten the pressure cuff in such a way that it fits tightly but does not constrict
the arm.
• Take readings while in a seated position.
• Relax your arm and rest it lightly on a surface such as a table. Relax during
the measuring process: do not move or talk, as this can affect the results.
Single reading
1. Press the START button

. The device carries out a self test. Once this is
complete, “ “ appears on the display. 5 short beeps sound during the
procedure.
1.
2.
3.
4.
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
27
GB
4 Operation
2. Next, the sleeve

automatically inflates until the pressure required for a
measurement has been reached. Alternatively, you can create this pressure
manually by pressing and holding the START button

after the device
has automatically reached a pressure of at least 30 mmHg. When you think
that the device has reached the ideal starting pressure for a measurement,
release the START button

. The increasing pressure is displayed
numerically.
3. The measuring process begins automatically when inflation is complete by
releasing the pressure built up in the cuff. The display shows the current
pressure in the cuff.
NOTE
If you press the START button

when the device is inflating,
the inflation process stops. The device switches off and the air
escapes from the sleeve.
4. Initially, deflation occurs at a constant speed. The pulse symbol

appears on the display. As soon as the pulse can be determined, deflation
is synchronised with the beating of the heart until the measurement ends.
The device beeps for each heartbeat during this procedure. At the same
time, the pulse symbol

flashes.
5. Once the measurement has been taken, the sleeve suddenly deflates
and a longer beep sounds.
6. The measured values

,

and

are shown on the display, together
with the WHO blood pressure bar display

. At the same time, a longer
beep sounds.
7. If the unit detects an arrhythmia during the measuring process, the
arrhythmia symbol

will flash eight times while the reading is
displayed. You may also hear 8 short beeps; these warn the user of an
irregular heartbeat.
8. If you do not press any buttons for approx. 1 minute, the device
automatically switches to standby mode.
3MAM measuring method
(3 consecutive measurements to calculate an average value)
You have selected the 3MAM measuring method (see section 3, subsection
3.6). You first have to complete any settings you require as per section 3,
subsections 3.3, 3.5 and 3.7.
1. Press the START button

. The device carries out a self test. Once this is
complete, “ “ appears on the display. 5 short beeps sound during the
procedure.
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
28
4 Operation
GB
On the display, you can see

(average) and “ 1 “ for the first
measurement.
2. Next, the sleeve

automatically inflates until the pressure required for a
measurement has been reached. Alternatively, you can create this pressure
manually by pressing and holding the START button

after the device
has automatically reached a pressure of at least 30 mmHg. When you think
that the device has reached the ideal starting pressure for a measurement,
release the START button

. The increasing pressure is displayed numeri-
cally.
3. The measuring process begins automatically when inflation is complete by
releasing the pressure built up in the cuff. The display shows the current
pressure in the cuff.
4. Initially, deflation occurs at a constant speed. The pulse symbol

appears on the display. As soon as the pulse can be determined, deflation
is synchronised with the beating of the heart until the measurement ends.
The device beeps for each heartbeat during this procedure, if you did not
switch off the beep function beforehand. At the same time, the pulse
symbol

flashes.
5. The first measurement ends when the pressure has been completely
discharged from the sleeve during a 15 second countdown. The count-
down symbol

flashes during this process. Beeps sound during the last 5
seconds.
6. Instead of “ 1 “, the display now shows “ 2 “ for the second measurement.
The second measurement proceeds in the same way as the first (see points
2 - 5).
7. Instead of “ 2 “, the display now shows “ 3 “ for the third measurement.
The third measurement proceeds in the same way as the first (see points
2 - 4).
8. Once the 3MAM measurement has been taken, the sleeve suddenly
de-flates and a longer beep sounds.
9. The measured values

,

and

are shown on the display, together
with the WHO blood pressure bar display

. At the same time, a longer
beep sounds.
10. If the unit detects an arrhythmia during the measuring process, the
arrhythmia symbol

will flash eight times while the reading is
displayed. You may also hear 8 short beeps; these warn the user of an
irregular heartbeat.
11. If you do not press any buttons for approx. 1 minute, the device
automatically switches to standby mode.
NOTE
When the device is used, the errors Err 1, 2, 3, 5, 6 can occur in
the integrated circuit. Err 6 can only occur in 3MAM mode.
In 3MAM mode, errors Err 1, 2, 3, 5 do not cause the
measurement to be cancelled, unless they occur twice in
succession. The error display is always accompanied by 3 short
beeps.
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
29
GB
4 Operation
Pairing the MTX Connect with your smartphone.
1.
Download the free-of-charge VitaDock App
from the Apple App Store
or Google Play store.
2. Slide the BLUETOOTH switch

on the reverse of the MTX Connect to
"On" to switch on Bluetooth.
3. Hold down the START button

for 5 seconds to activate the pairing mode
of the MTX Connect. The time and the flashing Bluetooth symbol is shown
on the display. The Bluetooth interface is in pairing mode for the next 60
seconds and enables you to connect your smartphone to the MTX Connect
within this time.
4. Activate the Bluetooth interface of your smartphone. Select "MTX-BT"
(Fig. D.) from the list of found Bluetooth
devices and enter the pairing code "0000"
depending on the smartphone to connect
the devices (Fig. D.).
5. Press the START button

to switch off
the MTX Connect again. If you do not
press any buttons for approx. 1 minute, the
device automatically switches to standby
mode.
Note: Pairing the MTX Connect with your smartphone only needs
to be done once, before the first measurement.
Measure and transfer the measurement results to your smartphone.
1. Slide the BLUETOOTH switch

on the reverse of the MTX Connect to
"On" to switch on Bluetooth.
2. Start the VitaDock App (blood pressure) and tap "new" (Fig. A).
3. Tap "connect Bluetooth device", the VitaDock App waits for the MTX
Connect to connect.
4. Press the START button

on MTX Connect to start a new measurement.
5. After measuring blood pressure the result is shown on the MTX Connect,
please press in the VitaDock App on MTX BT - screen C, to connect the two
devices. The MTX Connect display shows the measurement results, the
selected user memory, the time and the Bluetooth symbol. The data transfer
symbol
flashes.
6. Once the MTX Connect has transfers the current measurement values by
Bluetooth to the VitaDock App, press the START button

to switch off the
MTX Connect again. If you do not press any buttons for approx. 1 minute,
the device automatically switches to standby mode.
7. After successfully connecting the MTX Connect to your smartphone the date
and time are automatically synchronised.
Transfer the measurement results to your smartphone.
1. Start the VitaDock App and tap "new" (Fig. A).
2. Tap "Connect Bluetooth Device", the VitaDock App waits for the MTX
Connect to connect (Fig. B)
4.4
Bluetooth
Functions
Fig. D
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
30
4 Operation
GB
4.5
Troubleshooting
Error messages
Possible causes and solutions
Err 1
Measurement of the systolic blood pressure was completed.
Either the connection with the hose has been interrupted or no pulse
could be determined.
Check the connections between the cuff and the main unit.
Err 2 Non-physiological pressure has been detected.
Cause: the arm was moved during the measurement.
Repeat the measurement, keeping your arm still.
Err 3 If the cuff takes too long to inflate, it is possible that the cuff is
incorrectly positioned or that the connection with the hose is not airtight.
Check the connections and repeat the measurement.
Err 5 An average value could not be calculated due to unstable conditions
while the readings were being taken.
Repeat the measuring process, keeping your arm still and not speaking.
Err 6 There were too many errors during the measurement in MAM mode,
making it impossible to obtain a final result.
Read through the checklist for performing reliable measurements and
then repeat the measurement.
HI The pressure in the cuff is too high (over 300 mmHg) or the pulse is too
high (over 200 beats per minute).
Relax for 5 minutes and repeat the measurement.
LO The pulse is too low (less than 40 beats per minute).
Repeat the measurement.
If no buttons are pressed for approx. 1 minute, the device automatically switch-
es to standby mode. You can also switch to standby mode by pressing the
START button

. The blood pressure meter functions are now switched off.
The time and the selected user memory are continuously shown on the display.
The alarm function remains active.
If you do not want to use the time display and the alarm function, you can also
completely switch off the device by sliding the ON/OFF switch

on the back
of the device into the “ OFF ” position.
4.6
Switching off
the unit
3. First hold down the START button

and then the MEMO button

together for more than 5 seconds, the MTX Connect switches to transfer
mode. The display shows the selected user memory, the time and the
Bluetooth symbol.
4. In the VitaDock App press on MTX BT - screen C to connect the two devices
and transfer the stored measurement values. The MTX Connect connects
with your smartphone and starts to transfer the stored measurement values.
The data transfer symbol flashes.
5. Once the measurement values have been transferred to the VitaDock App,
press the START button

to switch off the MTX Connect again. If you do
not press any buttons for approx. 1 minute, the device automatically switches
to standby mode.
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
31
GB
5 Memory
This unit features 2 separate memories, each with a capacity of 99 memory
slots. Results are automatically stored in the memory selected by the user.
When the memory is full, the oldest reading is always the first to be deleted.
Press the MEMORY
button

to display the stored readings. After all the seg-
ments in the display have been shown, an average value of all the measure-
ments in the selected memory is displayed. This average is indicated by “ ”.
When the MEMORY button

is pressed again, the individual readings
(switching between systolic, diastolic, pulse and date/time) of the last mea-
surement stored are displayed. If this measurement was taken in the 3MAM
procedure,

is displayed. If an arrhythmia was detected during
the measuring process, the arrhythmia symbol will be displayed, but not
flashing.
Press the MEMORY button

again to move backwards through the stored
readings.
If you do not press any buttons for approx. 1 minute, the device automatically
switches back to standby mode.
5.2
Displaying
stored results
5.1
Recording
the results
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
32
6 Miscellaneous
GB
6.1
Care and
maintenance
• Remove the batteries before cleaning the unit.
• Never use strong detergents or hard brushes.
• Clean the unit with a soft cloth, moistened with mild soapy water. Do not let
water enter the unit. After cleaning, only use the unit when completely dry.
• Remove the batteries from the unit if you do not wish to use it for an
extended period of time, otherwise there is a risk of battery leakage.
• Do not expose the unit to direct sunlight, and protect it from dust and
moisture.
• Only inflate the cuff when it is in position around the upper arm.
• Servicing and calibration
The unit has been calibrated by the manufacturer for a period of two years’
use. In the case of commercial use, the unit must be serviced and recali-
brated at least once every two years. This calibration will be charged for and
can be carried out by an appropriate authority or an authorised service
centre - in accordance with the regulations for users of medical equipment.
This product must not be disposed of together with domestic waste.
All users are obliged to hand in all electrical or electronic devices, regardless
of whether or not they contain toxic substances, at a municipal or commer-
cial collection point so that they can be disposed of in an environmentally
acceptable manner.
Please remove the batteries before disposing of the device / unit. Do not dis-
pose of old batteries with your household waste, but at a battery collection
station at a recycling site or in a shop.
Consult your municipal authority or your dealer for information about disposal.
This blood pressure monitor meets the requirements of the EU standard for
non-invasive blood pressure monitors. It is certified in accordance with EC
Guidelines and carries the CE symbol (conformity symbol) “CE 0297”.
The blood pressure monitor corresponds to European standards EN 60601-1, EN
60601-1-11 and EN 1060-3. The specifications of EU Guideline “93/42/EEC of
the Council Directive dated 14 June 1993 concerning medical devices” are met,
along with those of the Radio and Telecommunications Terminal Equipment
Directive 1999/5/EC. You can request the declaration of conformity from
Medisana AG, Jagenbergstraße 1, 41468 Neuss, Germany or download it from
the Medisana homepage (www.medisana.de).
Clinical trial results:
Clinical trials for this unit were carried out in the USA and Germany according
to EN 1060-1 /-3 and according to the AAMI – Standard (US).
Electromagnetic compatibility: (see attached leaflet)
Intended use:
This unit is suitable for non-invasive blood pressure measurement in adults
(i.e., suitable for external use).
6.2
Disposal
6.3
Guidelines /
Standards
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
33
GB
6 Miscellaneous
6.4
Technical
specifications
Name and model : MEDISANA MTX Connect
with bluetooth
Display system : Digital display
Memory slots : 2 x 99
Measuring technique : Oscillometric
Power supply : 6 V= , 4 x 1,5 V Mignon AA LR 6, Alkaline
Blood pressure
measuring range : 40 – 255 mmHg
Pulse measuring range : 40 – 200 beats / Min.
Maximum error tolerance
for static pressure : ± 3 mmHg
Maximum error tolerance
for pulse rate : ± 5 % of the reading
Pressure generation : Automatic, using micro pump
Deflation : Automatic
Automatic Power Off : After approx. 1 minute
Operating conditions : +10 °C to + 40 °C, 15 – 90 %
humidity (non condensing)
Storage conditions : - 20 °C to + 55 °C
Protection against dust
and water : IP 21
Dimensions : approx. 125 mm x 163 mm x 65 mm
Cuff size : 22 – 42 cm for adults with an upper arm
circumference from normal to strong
Weight : approx. 535 g incl. batteries
Article number : 51085
EAN number : 40 15588 51085 4
Accessories : Mains adapter
Art.-No. 51036
Cuff, small 17 - 22 cm for adults with an
upper arm of small circumference
Art.-No. 51089
Smartphone Compatibility:
iOS: iPhone 4 and newer, iPod Touch 3th Gen and higher, iPad 2 and higher
Android: Samsung Galaxy S3, S4, S5 – mini S3, S4; HTC One, - One Max,
One M8; Sony Xperia Z, -Z1, -Z2. System requirements: 100 MB memory,
1 GB RAM, from Android Version 2.3.6, Bluetooth 2.0/3.0/4.0.
More information is available at: www.medisana.com/mtxconnect
In accordance with our policy of continual product improvement,
we reserve the right to make technical and optical changes
without notice.
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
34
7 Warranty
GB
Please contact your dealer or the service centre in case of a claim under the
warranty. If you have to return the unit, please enclose a copy of your
receipt and state what the defect is.
The following warranty terms apply:
1. The warranty period for MEDISANA products is three years from date of
purchase. In case of a warranty claim, the date of purchase has to be
proven by means of the sales receipt or invoice.
2. Defects in material or workmanship will be removed free of charge
within the warranty period.
3. Repairs under warranty do not extend the warranty period either for the
unit or for the replacement parts.
4. The following is excluded under the warranty:
a. All damage which has arisen due to improper treatment, e.g. non-
observance of the user instructions.
b. All damage which is due to repairs or tampering by the customer or
unauthorised third parties.
c. Damage which has arisen during transport from the manufacturer to
the consumer or during transport to the service centre.
d. Accessories which are subject to normal wear and tear (cuff, batteries
etc.).
5. Liability for direct or indirect consequential losses caused by the unit are
excluded even if the damage to the unit is accepted as a warranty claim.
MEDISANA AG
Jagenbergstraße 19
41468 NEUSS
GERMANY
Internet: www.medisana.de
The service centre address is shown on the attached leaflet
.
Warranty and
repair terms
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:17 U
51085_MTX_connect_DE_GB_NL_FR_ IT_ES_17042015.qxd:BlutdruckmessgerÀt 17.04.2015 10:18 U
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110

Medisana MTX connect Owner's manual

Category
Blood pressure units
Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI