DIATERMO50‐80
GIMASpA
EN‐25
MA391b_EN
Instruction’sManualManualed’Istruzioni
DEUTSCHPORTUGUES FRANCOIS ESPANOL ENGLISH ITALIANO
MAINTENANCE
General
Nouseradjustablepartsarewithintheequipment,eitherforcalibrationorservice
purposes.
Theequipmenthousingmustnotbeopened:thewarrantyisinvalidatedbyany
unauthorizedentryintotheunit.Intheeventanyrepairoradjustmentworkbeing
necessary,thewholeequipmentshouldbereturnedtotheLEDSpAServiceCentre
04011APRILIA(LT)‐ITALY,ortoanotherAuthorizedCentre,togetherwitha
descriptionofthefault.
Maintenanceworkbytheuserismainlythecleaningoftheexteriorofthecabinet,
cleaningandsterilizationoftheaccessoryitemsandcheckingoftheequipmentbefore
eachuse.Carryingoutfunctionandsafetycheckforverificationoftheparametersis
demandedtospecializedtechnicalpeople.
CleaningoftheCabinet
Switchtheequipmentoffcompletelyanddisconnectthemainssupplybeforeany
cleaningisundertaken.Cleantheoutsideofthecabinetwithadampcloth.Nochemical
shouldbeused;amildnonabrasivecleansermaybeusedwhennecessary.
CleaningandSterilizationoftheAccessoriesItem
Thebestthingtodoistouseonlyonetimeuseaccessoriesanddiscardthemafteruse.
Sincesomeoftheaccessoryitemsaretobeusedmorethanonceitismandatoryto
cleancarefullyandsterilizethoseaccessoriesbeforethenewuse.Thebestwayto
cleanandsterilizethereusableitemsistofollowthedirectionofthesupplierofeach
item.WhenoriginalreusableaccessoriessuppliedbyLEDSpAareapplied,thecleaning
byusingdeepcleanserandsterilizationthroughsteamsterilizationat121°C/134°Cis
recommended.
GuidetotheSolutionoftheProblems
Incaseofproblemsbeforeallitisadvisedtocheckforthecorrectinstallationofthe
unitandforthecorrectconnectionoftheaccessories.
ProblemsProbableCause Solution
Theequipmentdoesn't
switchon.
Interruptionorabsenceofthe
mainfeeding.
Verifytheconnectionofthemain
cable.
Verifythefusesandreplacethem,
wherenecessary,withnewonesof
theproprietype.
AlarmOCalwaysactive
(onlyforDIATERMO80)
Interruptionorlackofcontact
ontheneutralelectrodecircuit.
Checktheconnectionofthecable
totheneutralelectrode.
Replacethecableofconnectionof
theneutralelectrode.
Theunitdoesn't
respondtothe
commandofactivation
Breakdownofthepedal.‐
Wrongconnectionofthepedal.
Replacethepedal.
Verifytheconnectionofthepedal.
ErrorCode91Currentdeliverycontrol
activatedduringswitchingon.
Disconnectthepedalandswitchon
theunitagain.
ErrorCode92Errorinthemanagementboard. CallforService.
ErrorCode93Errorinthemanagementboard. CallforService.