Endres+Hauser TOCII CA72TOC Operating instructions

Category
Measuring & layout tools
Type
Operating instructions
BA448C/07/en/11.09
71098200
Operating Instructions
TOCII CA72TOC
Analyzer for online determination of TOC in aqueous media
using thermic catalytic combustion
Endress+Hauser
TOCII CA72TOC
Endress+Hauser 3
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 5
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Installation, commissioning and operation . . . . . . . . 5
1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.5 Notes on safety icons and symbols . . . . . . . . . . . . . . 6
2 Identification . . . . . . . . . . . . . . . . . . . . 7
2.1 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.1 Product structure . . . . . . . . . . . . . . . . . . . . 7
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Certificates and approvals . . . . . . . . . . . . . . . . . . . . 8
3 Installation . . . . . . . . . . . . . . . . . . . . . . 9
3.1 Incoming acceptance, transport, storage . . . . . . . . . . 9
3.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . . 10
3.2.1 Methods of installation . . . . . . . . . . . . . . . 10
3.2.2 Compressed air and water supply . . . . . . . . 10
3.2.3 Ambient conditions . . . . . . . . . . . . . . . . . . 11
3.3 Installation instructions . . . . . . . . . . . . . . . . . . . . . 11
3.3.1 Connecting the media . . . . . . . . . . . . . . . . 11
3.3.2 Preparing the chemicals . . . . . . . . . . . . . . . 12
3.4 Post-installation check . . . . . . . . . . . . . . . . . . . . . . 17
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . 18
4.1 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . 18
4.1.1 Preparing to wire . . . . . . . . . . . . . . . . . . . . 18
4.1.2 Power distribution . . . . . . . . . . . . . . . . . . . 19
4.1.3 Signal connection . . . . . . . . . . . . . . . . . . . 21
4.1.4 Power unit . . . . . . . . . . . . . . . . . . . . . . . . 22
4.1.5 Distributor connection . . . . . . . . . . . . . . . 23
4.1.6 Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 24
5 Operation . . . . . . . . . . . . . . . . . . . . . . 25
5.1 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . 25
5.2 Display and keypad . . . . . . . . . . . . . . . . . . . . . . . . 26
5.3 Operation during measuring mode . . . . . . . . . . . . . 28
5.3.1 Recording mode . . . . . . . . . . . . . . . . . . . . 28
5.3.2 Spot view . . . . . . . . . . . . . . . . . . . . . . . . . 28
6 Commissioning . . . . . . . . . . . . . . . . . . 30
6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.2 CLR start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.3 Programming mode . . . . . . . . . . . . . . . . . . . . . . . . 31
6.3.1 Menu SETTING - RANGE DATA . . . . . . . . 32
6.3.2 Menu SETTING - SET CLOCK . . . . . . . . . 33
6.3.3 Menu SETTING - ALARM LIMITS . . . . . . . 33
6.3.4 Menu SETTING - BASIC DATA . . . . . . . . . 34
6.3.5 Menu SETTING - SET DEFAULTS . . . . . . . 35
6.3.6 Menu SETTING - MEASURING SITE . . . . 35
6.3.7 Menu LISTS - MIN MAX AVERAGE . . . . . 35
6.3.8 Menu LISTS - MAINTENANCE RECORD . 36
6.3.9 Menu LISTS - ALARM RECORD . . . . . . . . 37
6.3.10 Menu LISTS - SHOW COMPLETE
RECORDS . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.3.11 Menu TEST - TEST OF INPUTS . . . . . . . . . 39
6.3.12 Menu TEST - TEST OF OUTPUTS . . . . . . . 40
6.3.13 Menu TEST - TEST FREQUENCY I/O . . . . 41
6.3.14 Menu TEST - TEST COM2 . . . . . . . . . . . . 41
6.3.15 Menu TEST - MEASUREMENT OFF . . . . . 42
6.4 Optimizing the analyzer . . . . . . . . . . . . . . . . . . . . . 43
6.4.1 Adjusting the measuring range . . . . . . . . . . 43
6.4.2 Influence on dosing pump P2 on the
measuring signal . . . . . . . . . . . . . . . . . . . . 43
6.4.3 Operating the analyzer with high salt loads 44
6.4.4 Adjusting the gas flow . . . . . . . . . . . . . . . . 44
6.5 Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
TOCII CA72TOC
4Endress+Hauser
7 Maintenance . . . . . . . . . . . . . . . . . . . . 48
7.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.2 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.3 Topping up reagents . . . . . . . . . . . . . . . . . . . . . . . 53
7.4 Replacing the standard . . . . . . . . . . . . . . . . . . . . . . 53
7.5 Entering the concentration of the standard solution 54
7.6 Service software . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.7 Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.7.1 Replacing the hoses on pumps P1 and P4 . . 55
7.7.2 Replacing the hose on pump P2 . . . . . . . . . 64
7.7.3 Replacing the hose on pump P3 . . . . . . . . . 67
7.7.4 Adjusting pump P2 . . . . . . . . . . . . . . . . . . 71
7.8 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
7.8.1 Adjusting the analyzer . . . . . . . . . . . . . . . . 74
7.8.2 Calibrating the analyzer . . . . . . . . . . . . . . . 76
7.8.3 Empty volume dosing . . . . . . . . . . . . . . . . 78
7.8.4 Adjusting the pH sensor . . . . . . . . . . . . . . . 81
7.9 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
7.9.1 Screen flushing . . . . . . . . . . . . . . . . . . . . . 85
7.9.2 Power flush . . . . . . . . . . . . . . . . . . . . . . . . 86
7.9.3 Bypass screen . . . . . . . . . . . . . . . . . . . . . . 87
7.9.4 Strip and separation chamber . . . . . . . . . . . 89
7.9.5 Opening the circuit . . . . . . . . . . . . . . . . . . 93
7.9.6 Combustion pipe . . . . . . . . . . . . . . . . . . . . 95
7.9.7 Leakage test . . . . . . . . . . . . . . . . . . . . . . 100
7.10 Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
7.10.1 Replacing the acid filter . . . . . . . . . . . . . . 102
7.10.2 Replacing the gas filter . . . . . . . . . . . . . . . 104
7.10.3 Replacing the O2 filter . . . . . . . . . . . . . . . 106
7.10.4 Replacing the heated salt trap . . . . . . . . . . 108
8 Accessories . . . . . . . . . . . . . . . . . . . . 112
9 Troubleshooting . . . . . . . . . . . . . . . . 113
9.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . 113
9.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
9.3 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
9.4 Decommissioning . . . . . . . . . . . . . . . . . . . . . . . . 118
9.5 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
10 Technical Data. . . . . . . . . . . . . . . . . . 119
10.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
10.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
10.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
10.4 Performance characteristics . . . . . . . . . . . . . . . . . 120
10.5 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
10.6 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
10.7 Mechanical construction . . . . . . . . . . . . . . . . . . . 120
Index . . . . . . . . . . . . . . . . . . . . . . . . . 121
TOCII CA72TOC Safety instructions
Endress+Hauser 5
1 Safety instructions
1.1 Designated use
The analyzer is a compact thermo-catalytic analytical measurement system. It is designed to monitor
the TOC content of industrial and municipal wastewater.
The device is particularly suitable for the following applications:
• Monitoring of industrial wastewater (e.g. in the inlet and outlet)
• Control of process wastewater
• Monitoring of industrial surface waters
• Municipal wastewater monitoring
• Recording the carbon load for nutrient dosing
Any other use than the one described here compromises the safety of persons and the entire
measuring system and is not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
1.2 Installation, commissioning and operation
Please note the following items:
• Installation, commissioning, operation and maintenance of the measuring system must only be
carried out by trained technical personnel.
Trained personnel must be authorized for the specified activities by the system operator.
• Electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood these Operating Instructions and must
adhere to them.
• Before commissioning the entire measuring point, check all the connections. Ensure that
electrical cables and hose connections are not damaged.
• Do not operate damaged products and secure them against unintentional commissioning.
Mark the damaged product as being defective.
• Measuring point faults may only be rectified by authorized and specially trained personnel.
• If faults can not be rectified, the products must be taken out of service and secured against
unintentional commissioning.
• Repairs not described in these Operating Instructions may only be carried out at the
manufacturer’s or by the service organization.
1.3 Operational safety
The analyzer has been designed and tested to the highest standards and left the factory in perfect
functioning order.
Relevant regulations and European standards have been met.
As the user, you are responsible for complying with the following safety conditions:
• Installation instructions
• Local prevailing standards and regulations.
Immunity to interference
This instrument has been tested for electromagnetic compatibility in industrial use according to
applicable European standards.
Protection against interference as specified above is valid only for an instrument
connected according to the instructions in these Operating Instructions.
Safety instructions TOCII CA72TOC
6Endress+Hauser
1.4 Return
If the analyzer has to be repaired, please contact the responsible sales center.
If the analyzer has to be returned, please return it cleaned to the responsible sales center.
Please use the original packaging, if possible.
Please enclose the completed "Declaration of contamination" (copy the second last page of these
Operating Instructions) with the packaging and the transportation documents.
No repair without completed "Declaration of contamination"!
1.5 Notes on safety icons and symbols
Electrical symbols
#Warning!
This symbol alerts you to hazards that can cause serious damage to the instrument or to persons if
ignored.
"Caution!
This symbol alerts you to possible faults which could arise from incorrect operation. They could
cause damage to the instrument if ignored.
!Note!
This symbol indicates important items of information.
%Direct Current (DC)
A terminal at which DC is applied or through which DC flows.
&Alternating Current (AC)
A terminal at which (sine-form) AC is applied or through which AC flows.
)Ground connecting
A terminal which, from the user’s point of view, is already grounded using a grounding system.
*Protective ground terminal
A terminal which must be grounded before other connections may be set up.
bAlarm relay
Input
Output
DC voltage source
Temperature sensor
TOCII CA72TOC Identification
Endress+Hauser 7
2 Identification
2.1 Nameplate
Compare the order code indicated on the nameplate with the ordering structure and your order.
The nameplate bears the following information:
• Manufacturer data
• Order code (device version)
• Serial number
•Measuring range
• Outputs and communication
• Power supply
• Degree of protection
• (Permitted) ambient conditions
2.1.1 Product structure
Measuring range
A 0.25 to 600 mg/l TOC
B 1 to 2400 mg/l TOC
C 2.5 to 6000 mg/l TOC
D 5 to 12000 mg/l TOC
Y Special version according to customer specification
Sample transfer
1 1 measuring point
2 2 measuring points; time controlled
3 2 measuring points; external controlled
Operation mode
A Batch mode
Power supply
0 230 VAC, 50 Hz
1 115 VAC, 60 Hz
2 115 VAC, 50 Hz
3 230 VAC, 60 Hz
9 Special version according to customer specification
Sample preparation
A 1 x PA-2 PVC, 1 to 8 m3/h (4.4 to 35 gpm) wastewater (not with
CA72TOC-*2*********)
B 1 x PA-3 PVC, 0.1 to 1.0 m3/h (0.4 to 4.4 gpm) wastewater (not with
CA72TOC-*2*********)
C 1 x PA-9 PP, 0.1 to 1.0 m3/h (0.4 to 4.4 gpm) wastewater (not with
CA72TOC-*2*********)
D 2 x PA-2 PVC, 1 to 8 m3/h (4.4 to 35 gpm) wastewater (not with
CA72TOC-*1*********)
E 2 x PA-3 PVC, 0.1 to 1.0 m3/h (0.4 to 4.4 gpm) wastewater (not with
CA72TOC-*1*********)
F 2 x PA-9 PVC, 0.1 to 1.0 m3/h (0.4 to 4.4 gpm) wastewater (not with
CA72TOC-*1*********)
Y Special version according to customer specification
Language instrument, documentation
1German
2 English
CA72TOC- Order code
Iden tification TOCII CA72TOC
8Endress+Hauser
!Note!
To complete your order code, simply add the optional features to the end of order code. If you have
any questions, please contact your local sales office.
2.2 Scope of delivery
The scope of delivery comprises:
• 1 analyzer CA72TOC
• Tool kit for leak check
• Tool kit to remove the glass ball and the media
• Acid filter kit
• Kit for commisioning of stripper and separation chamber
• Maintenance kit for furnace
• Hose kit for inlet and outlet
• 1 canister 5 liter
• 2 canisters 2 liter
•3 covers for canisters
• Key for door lock
• Connection box EMV
• Measuring glas 10 ml
• Sponge
• Safety glasses
• Gloves, acid-proof and base proof
• Silicone
• Operating Instructions
If you have any questions, please contact your supplier or your local sales center.
2.3 Certificates and approvals
Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus complies with
the legal requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.
Carrying gas (optinal, select one option only)
E1 CO2-scrubber
E2 CO2-scrubber + compressor
Analyzer mounting (optinal, select one option only)
F1 Wall mounted
F2 Stand
Accessory mounted (optinal, multiple selection possible)
G1 Sample predilution
G4 Heated salt trap
G5 Pipe back flushing
G6 Standby mode PA-2/-3 failure
G7 sample valve for aggressive media
Accessory enclosed (optinal, multiple selection possible)
H1 230 VAC furnace conditioning station (not with CA72TOC-***1******* and not with
CA72TOC-***2*******)
H2 115 VAC furnace conditioning station (not with CA72TOC-***0******* and not with
CA72TOC-***3*******)
H3 Standard furnace (not with CA72TOC-********G4**)
H4 Furnace for heated salt trap
TOCII CA72TOC Installation
Endress+Hauser 9
3 Installation
3.1 Incoming acceptance, transport, storage
• Make sure the packaging is undamaged!
Inform the supplier about any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the contents. Keep the damaged products until the matter
has been settled.
• Check that the order is complete and agrees with your shipping documents.
• The packaging material used to store or to transport the product must provide shock protection
and humidity protection. The original packaging offers the best protection. Also, keep to the
approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your local sales center.
Installation TOCII CA72TOC
10 Endress+Hauser
3.2 Installation conditions
3.2.1 Methods of installation
The analyzer can be mounted in three ways:
• As a standalone desktop unit
• Mounted to the wall (option F1)
• Mounted on a base frame with wheels (option F2)
a0012435
Fig. 1: Dimensions
3.2.2 Compressed air and water supply
When operating the analyzer, only use tonnage oxygen (gas in cylinders) or compressed air free from
carbon dioxide.
The compressed air must be dry and meet the following conditions:
• < 3 ppm CO2
• < 3 ppm hydrocarbons
• The pressure must be 2 bar (29 psi) constantly.
• The pressure tolerance is ±5 %.
• The compressed air volume is 650 to 750 cm3/min (1.37 to 1.6 ft3/hr).
• If compressed air is supplied in the building, the pipe must be fitted with a CO2 scrubber (infeed
pressure 4 to 10 bar (58 to 145 psi) and a pressure regulator.
• Connection: 4/6 mm DN
The water is connected via a 6/8 mm DN or G3/8 coupling. The pressure must be between 2 and
4 bar (29 to 58 psi).
400 (15.8)
600 (23.6)
785 (30.9)
1010 (39.8)
1558 (61.3)
mm (inch)
50
(1.97)
TOCII CA72TOC Installation
Endress+Hauser 11
3.2.3 Ambient conditions
• The analyzer requires a drain beneath the device. Use a 6/8 mm PTFE drain pipe. No
backpressure should form in the drain.
• An exhaust gas connection is required in enclosed spaces. Halogens or other vapors may not build
up in this area. No backpressure should form in the 4/6 mm connection.
• Ambient temperature range: 5 to 35 °C (41 to 95 °F)
• Humidity: 10 to 90 %, non-condensating
• It should be possible to access the analyzer from the front and rear.
3.3 Installation instructions
#Warning!
Danger! Risk of electric shock!
Do not connect the analyzer to the power grid before all the installation work has been completed
and the liquid and gas media are connected.
Follow the instructions in the "Wiring" section to connect the analyzer to the electricity supply.
3.3.1 Connecting the media
a0012643
Fig. 2: Connections on the left side panel
1 Power cable
2 Gas outlet
3 Gas connection
4 Water connection
5 Bypass rinse water
6 External grounding
7 Standard C2 connection
8 Standard C1 connection
9 Acid connection
10 Condensate drainage
11 Sample supply
12 Sample drainage
1
2
3
4
5
6
7
8
9
10
11
12
Installation TOCII CA72TOC
12 Endress+Hauser
Sample supply
Sample is supplied via a 3/8" connection on the left side panel (item 11).
!Note!
Additional connections are required if sample conditioning takes place (see instructions for sample
conditioning).
Analyzer sample drainage
Sample is drained off unpressurized via a DN 6/8 mm hose connection (compression fitting) on the
left side panel (item 12) into an open channel or pipe.
Condensate drainage
Condensate is drained off via a hose gland (PE, DN 1.6/3.2 mm, scope of delivery) on the left side
panel (item 10).
Gas outlet
Gas escapes via a hose gland (DN 4/6 mm) on the left side panel (item 2).
3.3.2 Preparing the chemicals
#Warning!
Accidents with chemicals can kill, blind, burn, or damage lungs. Many chemicals are poisonous or
corrosive; some are explosive, by themselves or in combination. Others are hazardous because they
can be absorbed easily through skin or the respiratory system. When working with any chemicals,
always observe the precautions in the section "Precautions regarding use of chemicals".
The analyzer requires the following chemical solutions:
• Stripper solution
Nitric acid HNO3 conc. 25 %
This acid is used to acidify the sample. In this way, the carbonate ion CO3 is converted to CO2
and the dissolved CO2 is removed from the solution (TIC stripping).
• Parent solution
Potassium hydrogen phthalate (KHP) with a concentration of 5 000 mg TOC/l
This parent solution for calibrating and adjusting the analyzer is used as a diluted standard in the
measuring range from 0 to 600 mg TOC/l. In the event of high KHP concentrations and sample
acidification (pH < 2.5), there is the risk of KHP precipitating in the solution.
• Parent solution
Citric acid with a concentration of 100 000 mg TOC/l
This parent solution for calibrating and adjusting the analyzer is used as a diluted standard in the
measuring range from 600 mg TOC/l onwards.
The stripper solution and the parent solutions are available from Endress+Hauser (see "Accessories"
chapter).
TOCII CA72TOC Installation
Endress+Hauser 13
Specifying the standard concentrations
The right choice of standard concentration is key to the accuracy of the measurement method.
Before you specify the concentrations of the standard solutions, determine what concentration
range the device should measure. The most common concentrations should be covered by the
standard solutions.
Please note, however, that there should be a concentration ratio of between 1:4 and 1:20 between
the two standard solutions.
If a limit value is to be monitored, it makes sense for the concentration of the standard to correspond
to the concentration of a limit value as this guarantees maximum precision when monitoring.
!Note!
Please note that no measurement method can provide accurate measurement results over a span
greater than 1:30.
Example:
20 and 200 mg/l should be selected here as the standard solutions. The system can then measure
accurately in the range from 10 to 300 mg/l (taking the measuring range of the system into
account). A higher measured value deviation can be expected below a concentration level of
10 mg/l and above a concentration level of 300 mg/l.
Precautionary measures when working with chemicals
• Carefully read the "Material Safety Data Sheet" (MSDS) that is supplied with every chemical to
determine the hazards posed and what precautionary measures have to be taken. If you are not
sure what risks are associated with the chemical, please ask appropriately qualified experts.
• Never prepare chemicals on your own. You may need assistance in the event of an accident. For
this reason, always make sure that someone is close-by. Only prepare chemicals in a properly
equipped laboratory.
• When preparing chemicals always wear protective goggles, rubber gloves and a rubber apron. In
addition, wear a dust mask or face shield when working with fine-powder chemicals.
• Chemicals or solutions that are used with the analyzer (or another process analysis device) should
never be inhaled, tasted or swallowed.
• When preparing solutions, you must always label the containers appropriately, indicating the
contents and the date of preparation. Dispose of unlabeled or expired solutions in accordance with
local regulations and guidelines.
• Some chemicals are very reactive when combined with other substances or dissolved in water
and cause explosions, fire or heat to occur, or dangerous gases, foams or sprays to be released.
Avoid the risk of dangerous accidents occurring: If you do not know how certain chemicals react,
make sure you do not mix them with other substances. Never mix chemicals that are known to
react severely.
Concentration to be measured: 3 to 300 mg/l
Most frequent concentration: 50 to 150 mg/l
Limit value to be monitored: 200 mg/l
Installation TOCII CA72TOC
14 Endress+Hauser
Reagent quality
!Note!
Reagent quality can affect the accuracy of measurements. For this reason, follow the steps outlined
below to ensure reagent quality.
• For 10, 100, and 500 mg/l units, use the best grade of reagents. For 1,000, 5,000, and
10,000 mg/l units, ordinary commercial grade acids and persulfate are suitable. For the KHP
calibration standard, use reagent grade or better for all units. For the 1,000, 5,000 and
10,000 mg/l units, you can experiment with quality once you have established repeatability, and
as long as the results are not affected. Lower grades of acid and persulfate may be acceptable.
• Rinse all glassware and plastic containers thoroughly with deionized water. For best results,
acid-wash and rinse them again before use.
• Use good scales and weigh the standard as accurately as possible before mixing. Keep containers
closed to prevent contamination and deterioration.
• Accurate preparation of the calibration standard is essential for accurate calibration of the
analyzer. Inaccurate preparation of the standard will result in an erroneous calibration, which in
turn will yield worthless results.
• The KHP calibration standards for 10 and 100 mg/l deteriorate rapidly in solution. Keep
containers closed when not in use, and store in a cool place (such as a refrigerator). Heat and
exposure to air can ruin sensitive solutions quickly.
Preparing potassium hydrogen phthalate
"Achtung!
Potassium hydrogen phthalate (KHP) can irritate the skin and cause problems with the eyes or
respiratory channel. Do not inhale the powder and do not attempt to ever swallow any of the
solution prepared.
Shelf life: Relatively stable, depending on the concentration. (Stability decreases with decreasing
concentration; refrigeration is recommended.)
Concentrations: Depends on the application.
Procedure:
For an organic carbon solution with a concentration of 5,000 mg/l, dissolve 10.627 g of p.a. (or
better) KHP in 500-700 ml of deionized water in a 1-liter volumetric flask (other concentrations of
the solution are listed in the table below). Once the chemical has dissolved, add 4 ml of H2SO4 and
top up the mixture with water until the 1-liter mark is reached. Stir the solution once more. Label
the container, indicating the contents and the date of preparation.
mg/l carbon g KHP Volume
5,000 10.627 1 l
1,000 2.1254 1 l
500 1.0627 1 l
100 0.21254 1 l
10 0.021254 1 l
TOCII CA72TOC Installation
Endress+Hauser 15
The solution with a concentration of 5,000 mg/l is stable (particularly when stored in a cool dark
place). It is recommended to perform serial dilutions on the basis of a standard substance of
5,000 mg/l to reduce the concentrations if needed. Example:
• If 10 ml of a solution with a concentration of 5,000 mg/l is diluted with 90 ml deionized water,
this results in a concentration of 500 mg/l.
• If 10 ml of a solution with a concentration of 500 mg/l is diluted with 90 ml deionized water,
this results in a concentration of 50 mg/l.
• If 10 ml of a solution with a concentration of 50 mg/l is diluted with 90 ml deionized water, this
results in a concentration of 5 mg/l.
Serial dilution is the preferred method to create lower concentrations instead of diluting 1 ml of the
5,000 mg/l solution with 99 ml deionized water. The latter method can result in measurement
errors.
"Caution!
Please note the following:
• Storable parent solutions with concentrations of 1,000 and 5,000 mg/l are relatively stable for
several weeks at room temperatures. Parent solutions must be stored in cool dark place.
• In the case of a solution with a concentration of 10 mg/l, on the other hand, the quality starts to
deteriorate at room temperature within 3 to 5 days. To give the KHP standard solutions greater
stability, 4 ml H2SO4 (4 mol) can be added to one liter.
• If parent solutions with a high KHP content acidify, there is the risk of the KHP precipitating.
• Make sure to keep the container with the crystalline KHP sealed at all times. If the crystalline KHP
comes into contact with the air, it absorbs water very quickly and must be dried before use.
Otherwise you will get inaccurate measurements since the concentration of carbon is lower in the
hydrous salt. KHP that has come into contact with air must be dried for one hour at 105 °C
(221 °F).
Preparing the parent solution from citric acid
"Caution!
• Nitric acid is highly caustic.
• Citric acid can irritate the skin and eyes and cause respiratory problems. Never attempt to swallow
any of the solution prepared.
• Wear protective goggles, gloves and protective clothing.
• Always add acids to water, not vice versa.
Shelf life: Storable parent solutions with concentrations of 100,000 mg/l are stable for 6 months
if stored in a cool, dark place at 4 to 8 °C (40 to 46 °F). Prepared standard solutions must be used
within four weeks even if stored in a cool, dark place.
Concentrations: Depends on the application.
Material:
• 1-liter volumetric flask
• Citric acid monohydrate C6H8O7 & H2O (purity: p.a.)
• Nitric acid HNO3 (65 %, purity: p.a.)
Procedure:
For an organic carbon solution with a concentration of 100,000 mg/l, dissolve 291.6 g of citric acid
monohydrate in 500 ml of deionized water in a 1-liter volumetric flask (other concentrations of the
solution are listed in the table below). Carefully add 55.0 ml (77.0 g) of nitric acid. Once the
chemicals have dissolved, top up the mixture with water until the 1-liter mark is reached. Stir the
solution once more. Label the container, indicating the contents and the date of preparation.
mg/l carbon Citric acid in g Nitric acid in ml Volume
100,000 291.6 55.0 1 l
50,000 145.8 55.0 1 l
10,000 29.16 55.0 1 l
Installation TOCII CA72TOC
16 Endress+Hauser
The solution with a concentration of 100,000 mg/l is stable (particularly when refrigerated). It is
recommended to perform serial dilutions on the basis of a standard substance of 100,000 mg/l to
reduce the concentrations if needed. Example:
• If 10 ml of a solution with a concentration of 100,000 mg/l is diluted with 90 ml deionized water,
this results in a concentration of 10,000 mg/l.
• If 10 ml of a solution with a concentration of 10,000 mg/l is diluted with 90 ml deionized water,
this results in a concentration of 1,000 mg/l.
• If 10 ml of a solution with a concentration of 1,000 mg/l is diluted with 90 ml deionized water,
this results in a concentration of 100 mg/l.
Serial dilution is the preferred method to create lower concentrations instead of diluting 1 ml of the
100,000 mg/l solution with 99 ml deionized water. The latter method can result in measurement
errors.
Preparing the stripping reagent
Stripping reagent dosing is regulated via the pH sensor. The regulation range of the dosing is from
the minimum feed rate of the acid pump to approximately 1000 times this amount. The necessary
acid quantity varies greatly from measuring place to measuring place. Ideally, the strength of the
acid in the feeder tank should be set in such a way that the system can regulate in both directions,
but the regulation range should be higher for higher volumes of acid dosed.
Regulation aid:
The acid feeder is typically in the region of a few % acid. The aim should be to achieve a feed rate
of 2 to 5 % for pump P3 (the current feed rate can be read in the
PROGRAMMING/TEST/OUTPUT/PUMPS menu).
To set the acid concentration, an acid is prepared as follows as the starter point:
• Fill the acid hose
• Start measuring operation with a real sample
• Allow the acid dosing system to regulate
• Feed rate > 2 % < 5 %, note the acid concentration and use for future mixtures
• Feed rate < 2 % acid concentration, dilute preparation
• Feed rate > 5 % acid concentration, increase the concentration of the preparation
• Replace the contents of the acid hose
• Allow the acid dosing system to regulate, read off the feed rate
Deionized water (ml) HNO3 25 % (ml) HNO3 concentration
Increase concentration 500 + 125 5.0 %
+ 125 8.3 %
+ 125 10.7 %
+125 12.5 %
Dilute 500 +125 5.0 %
+ 500 2.8 %
+1000 1.5 %
+2000 0.8 %
TOCII CA72TOC Installation
Endress+Hauser 17
Use of other acids
• Nitric acid
Nitric acid does not form lyophobic salts in the strip chamber. The resulting nitrogen oxides in
the combustion gas are filtered out with an acid filter upstream from the IR detector.
• Phosphoric acid
Phosphoric acid is a low-volatility, odorless acid that generates practically no precipitate. After
combustion, substances are formed that are filtered out completely in the heatable salt trap.
• Sulphuric acid
Sulphuric acid is a low-volatility, odorless acid. When using sulphuric acid, precipitate can be
expected with alkaline earth metal ions. In addition to the sulfur dioxides, salts also occur as a
reaction product downstream from the furnace. These salts are easily removed in the heatable salt
trap. Gypsum can also occur if the water is very hard.
• Hydrochloric acid
Hydrochloric acid is a volatile acid with a pungent odor. Salts do not precipitate in the strip
chamber. During combustion, salts occur (heatable salt trap) as does HCl gas, which is absorbed
in the condensate.
3.4 Post-installation check
• After installation, check that all connections are tightly fitted and are leak-resistant.
• Check all hoses for damage.
Wiring TOCII CA72TOC
18 Endress+Hauser
4 Wiring
4.1 Electrical connection
#Warning!
• The electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
• Ensure that there is no voltage at the power cable before beginning the connection work.
4.1.1 Preparing to wire
#Warnung!
Before connecting the device, make sure that the line voltage matches the voltage indicated on the
nameplate.
The following versions are available:
• 115 V AC 50 Hz
• 115 V AC 60 Hz
• 230 V AC 50 Hz
• 230 V AC 60 Hz
#Warning!
Danger! Risk of electric shock! Disconnect the device from the power supply system (disconnect the
mains plug). The line filter, the overvoltage module and the main switch are still connected to the
power supply even when the main switch is switched off.
#Warning!
Ensure that the analyzer is sufficiently grounded via the mains connection.
Condition: 50 V < R*Imax
Imax = is the maximum current above which the residual current circuit breaker is triggered.
R = the resistance between protective ground and device ground.
If these conditions cannot be ensured, the device must be grounded on site.
You must make the following connections:
• Alternating current connection via mains plug
• External grounding if necessary
• Analog 0/4 to 20 mA outputs
•Binary outputs
• Binary inputs
•RS-232
The signal connections are located in the EMC screening box on the right-hand side of the cabinet.
The connection for the external ground is located at the bottom on the left-hand side of the cabinet.
TOCII CA72TOC Wiring
Endress+Hauser 19
4.1.2 Power distribution
a0011954-en
Fig. 3: Power distribution wiring diagram
+-
23
12345678
2
3
1
4
44
+
-
13 14 23 24 33 34 43 5344 54 63 64 73 74 8483
1234 6789 1112510 13LNPE
764 12 13
25118
A9 10
14
14
40
39
15
15
terminal strip
relay module
relay 3
relay no.
fuse 16 A
rel. 3A 24VDC
relay 4A
relay 4A
relay 4A
adapter
relay 4A
relay 4A
relay 230V DC
Wiring TOCII CA72TOC
20 Endress+Hauser
Terminal strip assignment:
Relay module assignment:
The power distribution is located at the back in the top door.
Terminal Description
A Main switch, power distribution
4 Solenoid valve 3 - stripping gas
5 Tube furnace regulator
6 Peltier cooler regulator
7 Membrane compressor
9 Heated salt trap regulator
10 Solenoid valve 4, standard 1 + 2
11 Solenoid valve 7, carrier gas
12 Solenoid valve 2, strip chamber rinsing
13 Solenoid valve 5, bypass screen rinsing
14 Solenoid valve 1, sample/standard
15 Solenoid valve 6, channel 1/channel 2 (optional)
40 Solenoid valve 8, dosing
Relay No. Relay type Function
1 4A Solenoid valve 1, sample/standard
2 4A Solenoid valve 2, strip chamber rinsing
3 25A Solenoid valve 3, stripping gas, tube furnace regulator, heated salt trap regulator,
Peltier cooler regulator, membrane compressor
4 4A Solenoid valve 4, standard 1 + 2
5 4A Solenoid valve 5, bypass screen rinsing
6 4A Solenoid valve 6, two-channel changeover (optional)
7 4A Solenoid valve 7, carrier gas
8 3A Solenoid valve 8, dosing
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124

Endres+Hauser TOCII CA72TOC Operating instructions

Category
Measuring & layout tools
Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI