Proactive 715-420 Pulse TENS Electro-Stimulator Operating instructions

Type
Operating instructions
Manufactured in China
Physiotherapy • Physiothérapie
TM
TENS Electro-Stimulator
Instructions
Please read this manual thoroughly
before using this device for the first time.
Model: 715-420
A.M.G. Medical Inc.
8505 Dalton, Montréal, QC H4T 1V5
Canada • 1-800-363-2381
Distributed in the USA by:
AMG Medical, Inc. (USA)
8396 State Route 9, West Chazy, NY 12992
USA1-888-412-4992
Rev4 0122 P/N 861-715420
Generated by : Jean-François Perron
Special Instructions :
This is a double-sided,
saddle-stitched instruction booklet.
Flat Size : 24,811 cm x 18,81 cm
Assembled Size : Folded: 10,5 cm x 15 cm.
Open: 21 cm x 15 cm
Projet : 715-420 TENS Instructions EN Rev4 0122_original
Item No. : 715-420 Part No. : 861-715420
Colours :
B/W CMYK : Spots
TABLE OF CONTENTS
INTRODUCTION ..................................................................................3
What is pain? How does TENS work?
SAFETY INFORMATION......................................................................4
Precautions •Contraindications Adverse reactions • Warnings
DEVICE OVERVIEW.............................................................................6
Components and parts • Pulse duration • Pulse rate
Treatment mode • Time
INSTRUCTIONS FOR USE.................................................................9
Insert the battery Connect electrodes to lead wires
Connect lead wires to the device Place electrodes on skin
Select a treatment mode Set the pulse rate
Set the pulse width Set the timer
Set the intensity and begin treatment
After treatment is finished
CLEANING AND MAINTENANCE....................................................12
Cleaning the device • Electrodes
Cleaning the electrode cords • Maintenance
TROUBLESHOOTING ......................................................................13
ELECTRODE POSITIONING ............................................................14
CHECK / REPLACE THE BATTERY.................................................15
TECHNICAL SPECIFICATIONS .......................................................16
WARRANTY ....................................................................................17
2
3 4
INTRODUCTION
The Proactive™ Pulse™ Tens is intended to be used by adults for
the temporary relief of muscle and joint pain by applying electrical
nerve stimulation to the surface of the skin near the site
requiring therapy.
This device provides two independent, controllable output channels.
A pair of electrodes which can be connected to each output channel
and the current's parameters and intensity level can be adjusted
according to your needs.
1. What is pain?
Pain warns our body of injury to prevent additional damage. This
sensation is important because without it, vital parts of our bodies
might be injured without our knowledge. However, long-lasting,
persistent (chronic) pain, once diagnosed serves no apparent
purpose and reduces quality of life.
2. How does TENS work?
TENS (Transcutaneous Electrical Nerve Stimulation) refers to
the transmission of small electrical pulses through the skin to
the underlying peripheral nerves. The theory of TENS suggests two
different modes of operation:
In conventional (high frequency) TENS, continuous mild electrical
activity may block the pain signal traveling to the brain. If the pain
signal does not get through to the brain, the pain is not “felt”.
In low frequency TENS, short bursts of electrical activity may
stimulate the release of endorphins, the body's own pain-control
substance. Ask your physician or therapist for more details. No
matter what pain theory is used, TENS has been proven useful in pain
management for many patients in helping to make their lives better.
For external use only. The long-term effects of frequent electrical
stimulation are unknown. This device should be used only under
the continued supervision of a licensed medical practitioner.
Do not apply stimulation over the thyroid or carotid sinus regions,
as this could disrupt breathing, cardiac frequency or blood
pressure. Do not use while connected to high-frequency surgical
equipment or near shortwave or microwave therapy equipment.
Never use in environments with high humidity (ex.: bathroom).
Never place the electrodes anywhere on the front of the thorax or
transthoracically as it can increase the risk of ventricular fibrillation,
cause cardiac arrhythmia and lead to cardiac arrest. Never place
the electrodes in a way that would cause the current to go through
the head. Never use near the eyes, the genitals, the heart or on
areas which lack normal sensation. This stimulator should never
be used by patients who are noncompliant or are emotionally,
cognitively or mentally impaired. Keep the stimulator out of reach
of children.
WARNING
SAFETY INFORMATION
1. TENS is a symptomatic treatment that suppresses the sensation
of pain. It has no curative value and is not effective on pain that is
of central origin.
2. Do not use in the following situations: Undiagnosed pain
syndromes (until etiology is established). • Cancerous lesions that
are present in the treatment area. Swollen, infected, inflamed
areas or skin eruptions (e.g. phlebitis, thrombophlebitis, varicose
veins, etc.). • Demand type implanted pacemaker or defibrillator,
or any metallic implant. • Epilepsy • Serious arterial circulatory
problems in the lower limbs • Abdominal or inguinal hernia
Safety has not been established for use during pregnancy.
65
4. Precautions
Do not adjust controls while operating machinery or vehicles.
Turn the stimulator off before applying or removing electrodes.
TM
Only use with ProActive lead wires and electrodes.
Long-term stimulation at the same electrode site may cause
skin irritation. Use only as prescribed by a physician.
Never use in rooms where aerosols (sprays) are used
or pure oxygen is being administered. Do not use it near
highly flammable substances, gases or explosives.
Apply the electrodes to clean, dry, and unbroken skin only.
Keep electrodes separate during treatment. Electrodes coming
in contact with each other could result in improper stimulation
or skin burns.
• Should the treatment become unpleasant, discontinue the use
of this device until its use has been reassessed by a physician or
therapist.
5. Possible adverse reactions
Skin irritation or electrode burn under the electrodes.
Allergic skin reaction to electrode gel may also occur.
DEVICE OVERVIEW
Front
Side
Top
Belt clip
Slide cover
Intensity control
(on / off switch)
Lead wire
socket
Lead wire
socket
Pulse rate
Pulse width
Timer
Lead wire
socket
Cap
Mode
87
DEVICE OVERVIEW DETAILS
Display LED
Lights up when a current
pulse is sent through the
corresponding channel.
At frequencies above 30
Hz, the LED appears to be
constantly lit.
On / off switch and intensity
controls:
When both controls are in the OFF
position, the device is turned off.
Turning the controls clockwise turns
on the corresponding channel
according to the chosen settings.
Turn clockwise to augment intensity
and counterclockwise to reduce the
intensity, or all the way to the OFF
position to stop the treatment.
Slide cover
Reveals the controls for
pulse width, pulse rate,
mode selector and
modulation, as well as
the battery. Your medical
professional may wish to
set these controls for you
and request that you leave
the cover in place.
Pulse duration (width)
A wider pulse will deliver stronger stimulation.
Using a combination of intensity and pulse
duration can stimulate different nerve groups.
Pulse duration is partially dependent on the
treatment mode and protocol.
Pulse rate
The pulse rate (hz or pulses / second) is
set depending on electrode placement.
For contiguous and dermatome electrode
placements (i.e. stimulating the pain site
directly), a quick pulse rate (over 80 Hz) is desired. You should not
perceive individual pulses but rather have the sensation of steady,
continuous stimulation.
Treatment modes
N - Normal (or conventional) TENS offers
complete control of the various parameters.
B - Burst mode provides a combination of
conventional and low-frequency TENS by
replacing the individual pulses by short “bursts” of 7 - 10 pulses.
In this mode, the treatment frequency cannot be adjusted.
M - Modulated mode attempts to prevent nerve accommodation by
continuously cycling the treatment intensity. In this mode, increase
the intensity only when at the cycle’s maximum intensity.
Time
Pain relief should begin shortly after the onset
of treatment. However, in some cases, it may
take as long as 30 minutes to achieve. TENS
units are typically operated for long periods of time,
with a minimum of 20 - 30 minutes and in some post-operation
protocols, as long as 36 hours. In general, pain relief will diminish
within 30 minutes of the cessation of stimulation.
9
INSTRUCTIONS FOR USE
1. Insert the battery
Slide the cover completely to uncover the battery cavity.
Insert a 9V battery (included), matching the positive
and negative ends of the battery to the markings, then close.
2. Connect electrodes to lead wires
Insert the lead wire connector into
the electrode connectors, ensuring
that no bare metal remains exposed.
3. Connect lead wires to the device
This device has 2 output channels,
which can be used simultaneously
(with 2 electrode pairs) or individually
(1 channel with 1 electrode pair).
After making sure that the device
is turned off, insert the wire plugs
into the sockets at the top of the unit.
Caution: Always use the lead
wires supplied with this unit.
4. Place electrodes on skin
Ensure that the skin surface
where the electrodes will be
placed is clean and dry.
Apply electrodes to the exact site
indicated by your physician.
The electrodes should be placed firmly
and evenly on the skin to ensure good contact.
Caution:
Always wash and dry the treatment area before applying the
electrodes. Do not turn the device on when the electrodes are not
applied to the treatment area. Never adjust, reposition, or remove
the electrodes while the device is still on. Always use this device
with 4cm x 4cm self-adhesive square electrodes, or larger. The
electrodes are for single-patient use only and should be discarded
when they no longer adhere properly.
5. Select a treatment mode
Expose the controls by sliding
front cover down from top of unit.
This switch has 3 positions: B for
burst stimulation, N for constant
stimulation, and M for
modulation stimulation. Push the
mode selector until engaged in
position desired.
6. Set the pulse rate
Turn the pulse rate dial to the
setting recommended by your
doctor or therapist. Unless
otherwise instructed, the pulse
rate control should be set in the
70 -120 Hz range.
7. Set the pulse width
Turn the pulse width dial to the
setting recommended by your
doctor or therapist. Unless
otherwise instructed, the pulse
width control should be set in the
70 -120 µs range.
10
8. Set the timer
Select the appropriate treatment
time (15 minutes, 30 minutes or
continuous) as recommended by
your doctor or therapist.
1211
9. Set the intensity and begin treatment
Slowly turn the intensity control dials clockwise. This will
turn the corresponding channels on and begin treatment.
To further increase the intensity, turn the dials clockwise.
To decrease the intensity or turn the device off, rotate the
dials counterclockwise to the desired setting or to the
"OFF" position.
10. After treatment is finished
Once the treatment time has elapsed, turn the intensity
control dials counterclockwise to the "OFF" position.
Unplug the lead wires from the device.
To remove the electrodes, lift one edge and gently peel
(do not pull on the lead wires because it may
damage the electrodes). Place the
electrodes on the liner and remove
the lead wire by twisting and
pulling at the same time.
CLEANING AND MAINTENANCE
1. Stimulator
Remove the battery before cleaning. Wipe the stimulator with a
soft, slightly moistened cloth. If a more thorough cleaning is
needed, you can also moisten the cloth with mild soapy water.
Do not submerge the stimulator or expose it to a large amount
of water. Do not use any chemical cleaners or abrasive agents
to clean this device. After use, store the stimulator in the storage
pouch to protect it from shock and dust. If you plan not to use it
for a prolonged period, remove the battery from the unit and store
in a cool, dry place.
2. Electrodes
• Use the device only with the leads and electrodes provided
with this device or sold under the ProActive brand, following
the placements and settings prescribed by your practitioner.
It is recommended to use at minimum 4cm x 4cm self-adhesive,
square electrodes.
Inspect your electrodes before every use and replace as needed.
Reusing electrodes too many times may cause slight skin irritation,
low adhesion, or ineffective stimulation.
• Between uses, store the electrodes in the resealed bag in a cool
dry place. It may be helpful, between applications, to moisten
their adhesive surface with a few drops of water (be careful not to
over-saturate) and then let them air-dry to help them last longer.
Reusable, self-adhesive electrodes
1413
TROUBLESHOOTING
Should any malfunction occur while using this device:
Verify that the controls are set to the appropriate therapy settings
and adjust accordingly.
Verify that the lead wires are correctly connected to the device.
They should be inserted completely into the sockets.
Verify whether the LEDs light up when the unit is in use. If
necessary, insert a new battery.
Inspect the lead wires for damage and replace if necessary.
Verify that the electrodes are properly applied to the skin in a
position that promotes proper functioning of the device. See the
notes on electrode positioning on the next page for more details.
If the above fails, do not attempt any repairs to the device or
any of its accessories. Contact us at 1-800-363-2381 for repair
information.
4. Maintenance
Do not attempt any repairs to the device or any of its accessories.
Contact us at 1-800-363-2381 for repair information.
The manufacturer will not be held responsible for the results
of maintenance or repairs by unauthorized persons.
Check the device before each use for signs of wear and/or
damage. Replace worn items (electrodes, lead wires) as required.
3. Cleaning the electrode cords
Clean the electrode cords by wiping them with a damp cloth.
Coating them lightly with talcum powder will reduce tangles
and help them last longer.
ELECTRODE POSITIONING
Electrode positioning can be one of the most important parameters
in achieving success with TENS therapy. Of utmost importance is
the willingness of the clinician to try the various styles of electrode
placement to find which method best fits the needs of each
individual patient.
Everyone responds to electrical stimulation differently and,
accordingly, their needs may vary from the conventional settings
suggested here. If the inital results are not positive, feel free to
experiment. Once an effective positioning has been achieved,
mark down the electrode sites and the settings at the end of this
manual so that it can easily be replicated at home.
Contiguous placement is the most common technique. It involves
placing the electrodes alongside the site of localized pain, in such a
way as to direct the flow of current through or around the pain site.
In a single channel application, this would involve placing the
electrodes on either side of the pain site in the case of limb or deep
tissue pain. Electrode placement on the posterior and anterior
aspects of the affected limb will allow the current to flow completely
through the limb, and thus through the endogenous pain site. With
a two channels application, the clinician may either direct the
current flow to cross through the pain site or, in what is called the
“bracket” method allows the current flow on either side of the
painful area, generally through the nerve branches that feed into
the pain site.
1615
Over time, in order to ensure the functional safety
of this device, changing the battery is necessary.
1. Make sure that both intensity controls
are switched to the off position.
2. Slide the battery compartment cover
and remove.
3. Remove the battery from the compartment.
4. Insert the battery into the compartment.
Note the polarity indicated on the battery
and in the compartment.
5. Replace the battery compartment cover
and slide to close.
CH
1
CH2
TECHNICAL SPECIFICATIONS
Mechanism Technical description
Channel Dual, isolated between channels
Pulse amplitude Adjustable, 0 - 80 mA
at 500 ohm load each channel
Pulse rate Adjustable, from 2 to 150 Hz
Pulse width Adjustable, from 30 to 260 microseconds
Modulation mode Pulse rate is automatically varied in a
cyclic pattern over an interval of nominally
10 seconds (in max 150 Hz). Pulse rate
decreases linearly over a period of
4 seconds from the control setting value
to a value which is 40% less. The lower
pulse rate will continue for 1 second. Then
increase linearly over a 4 seconds period
to its original value. The original pulse rate
will continue for 1 second. The cycle is
then repeated.
Burst mode Bursts occur twice every second.
Pulse width (adjustable), frequency:
100 Hz
Wave form Asymmetrical bi-phasic square pulse
Timer 15, 30 minutes or continuous
Voltage 0 to 40 V (load: 500 ohm)
Max charge per pulse 20 micro-coulombs
Power supply One 9 V battery (alkaline or
nickel-cadminum rechargeable)
Battery life Approximately 50 hours at nominal settings
Size 95 mm (H) x 65 mm (W) x 23.5 mm (T)
Weight 115 grams (battery included)
CHECK / REPLACE THE BATTERY
Operating Conditions 5°C ~ 40°C; 30% RH ~ 75% RH
Storage Conditions -10°C ~ 55°C; 10% RH ~ 90% RH
17
LIMITED WARRANTY
A.M.G. Medical Inc. warrants the stimulator to be free from defects in
material and workmanship for a period of one (1) year, to be proven by
means of the sales receipt or invoice. This warranty is valid for the
original purchaser only. Any alterations, abuse, misuse or accidental
damage voids this warranty. Repairs under warranty do not extend the
warranty period. For service under warranty, call us at:
1-800-363-2381, between 8:30 AM and 5:00 PM EST.
• The following is excluded under the warranty:
A) All damage which has arisen due to improper treatment,
e.g. nonobservance of the user instructions.
B) All damage which is due to repairs or tampering by the customer
or unauthorized third parties.
C) Damage which has arisen during transport from the
manufacturer to the consumer or during transport to
the service centre.
D) Accessories which are subject to normal wear and tear.
• Liability for direct or indirect consequential losses caused by the
unit are excluded even if the damage to the unit is accepted as a
warranty claim.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10

Proactive 715-420 Pulse TENS Electro-Stimulator Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI