3M Attest™ Super Rapid Readout Biological Indicator 1492V Operating instructions

Type
Operating instructions
MECHANICAL ART
DESC: 3M™ Attest™ - Super Rapid Readout Biological Indicator - Insert
Printed Colors‑Front: 1
K
Item Spec. #:
34-8720-3534-9
Supersedes #:
Structure #:
Reference:
1492V
Requesters:
Linda Thompson
DIE #:
5" x 4.75"
DATE CHANGES INT.
05.25.16 PU file from 4175212 and update per marked up PDF kmh
06.28.16 Final Release kmh
02.28.17 Updates per proof report. Jer
04.18.17 Updates per proof report. TjS
05.19.17 Final Report kmh
06.29.17 PU from SGS 4538765, add illustration to TH and KR languages per annotated PDF. TN
SGS#: 4614439 FILE: 1492V_34872035349.indd
Printed Colors‑Back: 1
K
Match Colors: 0
Notes to Printer:
This bar code symbol was produced using a bar width adjustment of minus 0.004. If this does not
meet the needs of the printing process or equipment to be used, the printer is responsible to recreate
the file using the appropriate bar width adjustment while meeting 3M quality requirements.
Because of the differences in scanning equipment, SGS Minnesota cannot guarantee the scannability
of UPC Bar codes. The Uniform Code Council recommends a UPC symbol be printed black on white
at 100% with no truncation. All UPC codes are test scanned and verified before leaving our facility.
Special Instructions:
Created in Adobe InDesign CC14
Proof prepared for 3M Company
REV.
06
3M Attest Super Rapid Readout
1
BiologicalIndicator1492V
ProductDescription
The 3M™ Attest™ Super Rapid
Readout Biological Indicator
1492V (brown cap, referred to
hereinafter as the 1492VBI)
is a self-contained biological
indicator specifically designed for
rapid and reliable qualification
testing and routine monitoring
of 270°F (132°C) and 275°F
(135°C) dynamic-air-removal
(pre-vacuum) steam
sterilization processes when
used in conjunction with the
3M™ Attest™ Auto-reader490
(hereinafter referred to as the 490
Auto-reader). The 1492VBI is a
single-usedevice.
A schematic illustrating the
design of the 1492V BI is provided
in Figure 1.The self-contained design includes a carrier with spores of Geobacillus
stearothermophilus and a media ampoule containing bacteriological growth medium
which meets the requirements for growth promoting ability specified in ANSI/AAMI/
ISO 11138-1:2006/(R)2010. The spore carrier and media ampoule are contained in
a plastic vial topped with a brown cap. A chemical process indicator which changes
from pink to light brown or darker upon exposure to steam is located on the top
of the cap.
The 1492VBI utilizes the α-glucosidase enzyme system, which is generated
naturally within growing cells of Geobacillus stearothermophilus. The α-glucosidase
in its active state is detected by measuring the fluorescence produced by the
enzymatic hydrolysis of a non-fluorescent substrate, 4-methylumbelliferyl-α-D-gl
ucoside (MUG). The resultant fluorescent by-product, 4-methylumbelliferone(MU),
is detected in the 490 Auto-reader.The presence of fluorescence within 1 hour of
incubation of the 1492VBI in the 490 Auto-reader indicates a steam sterilization
processfailure.
The 1492VBI can also indicate the presence of G. stearothermophilus organisms by
a visual pH color change reaction. Biochemical activity of the G. stearothermophilus
organism produces metabolic by-products that cause the media to change color
from purple to yellow which also indicates a steam sterilization process failure. Use
of this indication method is optional and is typically restricted to specialstudies.
Readout Times
The 1-hour super rapid readout and the optional 48-hour visual pH color change
incubation times have been correlated with a 7-day incubation period (at 56+/-2°C)
following the FDAs Reduced Incubation Time protocol. Processed indicators were
examined at 48 hours and 7 days for detection of a visual pH color change. The
1-hour fluorescence change readings and the 48-hour visual pH color change
readings were compared to the 7-day visual pH color change readings to determine
the readout time of theindicator.
1-hour Fluorescence ChangeResult
1492VBIs have 1-hour reduced incubation time results that correlate to the
7-day (168 hours) visual readout result ≥ 97% of thetime.
48-hour Visual pH Color ChangeResult
1492VBIs have 48-hour reduced incubation time results that correlate to the
7-day (168 hours) visual readout result ≥ 97% of thetime.
Due to the high reliability of the 1-hour fluorescent result there is no advantage to
incubating 1492VBIs beyond 1hour.
1492VBIs meet ANSI/AAMI/ISO 11138-1:2006/(R)2010, ANSI/AAMI/ISO
11138-3:2006/(R)2010 and BS EN/ISO 11138-1:2006, BS EN/ISO11138-3:2006.
Indications forUse
UnitedStates
Use the 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V in
conjunction with the 3M™ Attest™ Auto-reader490 to qualify or monitor
dynamic-air-removal (pre-vacuum) steam sterilization cycles of 4 minutes at 270°F
(132°C) and 3 minutes at 275°F(135°C).
The 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V provides a final
fluorescent result in 1 hour. An optional visual pH color change result is observed
in 48hours.
Outside the UnitedStates
Use the 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V in
conjunction with the 3M™ Attest™ Auto-reader490 to qualify or monitor
270°F (132°C) to 275°F (135°C) dynamic-air-removal (pre-vacuum) steam
sterilizationcycles.
Contraindications
None.
PROCESS INDICATOR
BI CAP WITH LABEL
VIAL
MEDIA AMPOULE
AMPOULE CRUSHER
NONWOVEN MATERIAL
GROWTH CHAMBER
SPORE CARRIER
Figure 1: 3M™Attest™ Super Rapid Readout
Biological Indicator 1492V Design
Warnings
There is a glass ampoule inside the plastic vial of the biological indicator (BI). To
avoid the risk of serious injury from flying debris due to a rupturedBI:
Allow the BI to cool for the recommended time period before activating.
Activating or excessive handling of the BI before cooling may cause the glass
ampoule toburst.
Wear safety glasses and gloves when removing the BI from thesterilizer.
Wear safety glasses when activating theBI.
Handle the BI by the cap when crushing orflicking.
Do not use your fingers to crush the glassampoule.
Precautions
1. DO NOT use the 1492VBI to monitor sterilization cycles which it is not designed
tochallenge:
a. Gravity-displacement steam sterilizationcycles;
b. 250°F (121°C) dynamic-air-removal (pre-vacuum) steam sterilization cycles;
c. Dry heat, chemical vapor, ethylene oxide or other low temperature
sterilizationprocesses.
2. After BI activation, ensure media has flowed to the spore growthchamber.
3. Do not place tape or labels on 1492VBI prior to sterilization or incubation in the
490 Auto-reader.
MonitoringFrequency
Follow facility Policies and Procedures which should specify a biological indicator
monitoring frequency compliant with professional association recommended
practices and/or national guidelines and standards. As a best practice and to provide
optimal patient safety, 3M recommends that every steam sterilization load be
monitored with a biological indicator in an appropriate Process Challenge Device
(PCD i.e., BI challenge testpack).
Directions forUse
1. Identify the 1492VBI by writing the load number, sterilizer, and processing date
on the indicator label. Do not place another label or indicator tape on the vial
or on thecap.
2. Place the 1492VBI in a representative tray configuration or Process Challenge
Device (PCD) as recommended by professional association guidelines or national
standards for healthcare facility practice. Do not place the 1492VBI in direct
contact with a chemical indicator as residue could transfer to the biological
indicator and affect theresult.
3. Place the PCD in the most challenging area of the sterilizer. This is typically on
the bottom shelf, over the drain, however, the sterilizer manufacturer should be
consulted to identify the area of the chamber least favorable tosterilization.
4. Process the load according to recommendedpractices.
5. After completion of the cycle take the PCD out of the sterilizer, and remove the
1492VBI.
6. Allow the 1492VBI to cool for 10 minutes prior toactivation.
7. Check the process indicator on the top of the cap of the 1492V BI. A color
change from pink to light brown or darker confirms that the 1492V BI has been
exposed to the steam process. This color change does not indicate that the
steam process was sufficient to achieve sterility. If the process indicator is
unchanged, check the sterilizer physical monitors.
8. To activate the biological
indicator, place it in a 490
Auto-reader incubation well
which is color-coded brown
(i.e., configured to incubate
1492VBIs). Press the cap of
the BI down firmly to close
the cap and crush the glass
ampoule. Immediately remove
the BI and flick it (see picture
at right). Visually verify that
media has flowed into the
growth chamber at the bottom
of the vial. If the media hasn’t
filled the growth chamber, hold the BI by the cap and flick it until media fills the
growth chamber. Return the activated 1492VBI to the incubation well and wait
for the result. See the 490 Auto-reader Operator’s Manual for further information
related to itsuse.
9. Each day that a processed 1492VBI is incubated, activate and incubate at least
one non-processed 1492VBI to use as a positive control. Follow the activation
instructions provided in Step 8 above. Write a "C" (for "control") and the date
on the BI label. The positive control should be from the same lot code as the
processed biological indicator. The positive control BI helpsconfirm:
correct incubation temperatures aremet,
viability of spores has not been altered due to improper storage temperature,
humidity or proximity tochemicals,
capability of media to promote rapid growth,and
proper functioning of the 490Auto-reader.
10. Incubation andReading:
Incubate the positive control and steam processed 1492VBIs at 56 ± 2°C in
a 490 Auto-reader. See the 490 Auto-reader Operator's Manual for the proper
use of this equipment. Positive results are available within 1 hour. The 490
Auto-reader will indicate a positive result as soon as it is obtained. The final
negative 1492VBI reading is made at 1 hour. After the results are displayed and
recorded, the 1492VBIs may bediscarded.
Interpretation ofResults
FluorescentResults
The positive control (unprocessed) 1492VBI must provide a positive fluorescent
result (+ on the 490 Auto-reader LCD display). Processed 1492VBI results are
not valid until the positive control reads fluorescent positive. The positive control
should read positive (+ on the LCD display) at or before 1hour. If the positive
control reads negative (- on the LCD display) at 1 hour, check the 490 Auto-reader
Operator’s Manual Troubleshooting Guide. Retest the 490 Auto-reader with a new
positivecontrol.
With processed 1492VBIs, a positive (+ on the LCD display) result indicates a
sterilization process failure. A final negative processed 1492VBI reading (- on the
LCD display) after 1hour of incubation indicates an acceptable sterilizationprocess.
Act immediately on any positive results for processed BIs. Determine the cause of
the positive BI following facility policies and procedures. Always retest the sterilizer
and do not use sterilizer for processing loads until three consecutive BI results
arenegative.
Optional Visual pH Color ChangeResult
The 1492VBI is normally discarded after the fluorescent result has been recorded. If,
however, special studies are desired, 1492V BIs may be further incubated for a visual
pH color change result. After activation and during incubation, the white Nonwoven
Material will absorb the bromocresol purple indicator, the pH-sensitive indicator dye
in the growth media, and appear blue. In the case of the positive control BI a yellow
color change of the growth media and/or Nonwoven Material will appear within 48
hours. Any observation of a yellow color within the vial indicates a positiveresult.
In the case of a processed 1492VBI, a media and/or Nonwoven Material color
change from purple to yellow indicates a sterilization process failure. A negative pH
color change result, i.e., media and Nonwoven Material remain purple/blue, can be
assessed at 48hours.
Storage
Best stored in original box under normal room conditions: 59-86ºF (15-30ºC),
35-60% relative humidity (RH).
Do not store 1492VBIs near sterilants or otherchemicals.
Disposal
Dispose of used 1492VBIs according to your health care facility policy.You may wish
to steam sterilize any positive biological indicators at 132°C (270°F) for 4minutes
or at 275°F (135°C) for 3minutes in a dynamic-air-removal steam sterilizer prior
todisposal.
Explanation of Symbols
Catalogue Number
Caution, see instructions for use
Do not reuse
Use by date
Batch code
Manufacturer
Date of manufacture
Product is designed for use with steam sterilizationcycles.
British Standards Institution (BSI) Kitemark™ License for compliance with
BS EN ISO 11138-1:2006 and BS EN ISO 11138-3:2009
Made in U.S.A.by
3M HealthCare
2510ConwayAve.
St. Paul, MN 55144
1‑800‑228‑3957
3M.com/Medical
3M and Attest are trademarks of3M.
Used under license inCanada.
© 2017, 3M. All rightsreserved.
3M et Attest sont des marques de commerce
de 3M, utilisées sous licence au Canada.
© 2017, 3M. Tous droits réservés.
Issue Date: 2016‑02
34‑8720‑3534‑9
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96

3M Attest™ Super Rapid Readout Biological Indicator 1492V Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI