Micro Motion 1500 Owner's manual

Type
Owner's manual
Configuration and Use Manual
MMI-20019023, Rev AB
March 2018
Micro Motion
®
Model 1500 Transmitters with
Analog Outputs
Configuration and Use Manual
Safety messages
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully
before proceeding to the next step.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration
manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government
transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept
your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer
Service department.
Emerson Flow customer service
Email:
Worldwide: [email protected]
Asia-Pacific: [email protected]
Telephone:
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 704 136 666 New Zealand 099 128 804
Mexico +41 (0) 41 7686 111 France 0800 917 901 India 800 440 1468
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000
Central & Eastern +41 (0) 41 7686 111 Japan +81 3 5769 6803
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684
Contents
Part I Getting started
Chapter 1 Before you begin ............................................................................................................3
1.1 About this manual ....................................................................................................................... 3
1.2 Transmitter model code .............................................................................................................. 3
1.3 Communications tools and protocols .......................................................................................... 3
1.4 Additional documentation and resources .................................................................................... 4
Chapter 2 Quick start .....................................................................................................................5
2.1 Power up the transmitter .............................................................................................................5
2.2 Check meter status ......................................................................................................................6
2.2.1 Transmitter status reported by LED ...............................................................................6
2.3 Make a startup connection to the transmitter ..............................................................................6
2.4 (Optional) Adjust digital communications settings ......................................................................7
2.5 Verify mass flow measurement ....................................................................................................7
2.6 Verify the zero ............................................................................................................................. 7
2.6.1 Terminology used with zero verification and zero calibration ........................................9
Part II Configuration and commissioning
Chapter 3 Introduction to configuration and commissioning ....................................................... 13
3.1 Configuration flowchart ............................................................................................................ 13
3.2 Default values and ranges ..........................................................................................................15
3.3 Disable write-protection on the transmitter configuration ........................................................ 15
3.4 Restore the factory configuration .............................................................................................. 15
Chapter 4 Configure process measurement ..................................................................................17
4.1 Configure mass flow measurement ........................................................................................... 17
4.1.1 Configure Mass Flow Measurement Unit .................................................................... 17
4.1.2 Configure Flow Damping ........................................................................................... 19
4.1.3 Configure Mass Flow Cutoff ........................................................................................20
4.2 Configure volume flow measurement for liquid applications ..................................................... 22
4.2.1 Configure Volume Flow Type for liquid applications ....................................................22
4.2.2 Configure Volume Flow Measurement Unit for liquid applications .............................. 23
4.2.3 Configure Volume Flow Cutoff ................................................................................... 25
4.3 Configure GSV flow measurement .............................................................................................27
4.3.1 Configure Volume Flow Type for gas applications ....................................................... 27
4.3.2 Configure Standard Density of Gas .............................................................................27
4.3.3 Configure Gas Standard Volume Flow Unit .................................................................29
4.3.4 Configure Gas Standard Volume Flow Cutoff ..............................................................32
4.4 Configure Flow Direction .......................................................................................................... 33
4.4.1 Options for Flow Direction ......................................................................................... 34
4.5 Configure density measurement .............................................................................................. 38
4.5.1 Configure Density Measurement Unit ........................................................................ 38
4.5.2 Configure two-phase flow parameters ........................................................................ 39
Contents
Configuration and Use Manual i
4.5.3 Configure Density Damping ....................................................................................... 41
4.5.4 Configure Density Cutoff ............................................................................................42
4.6 Configure temperature measurement .......................................................................................43
4.6.1 Configure Temperature Measurement Unit ................................................................43
4.6.2 Configure Temperature Damping .............................................................................. 43
4.6.3 Effect of Temperature Damping on process measurement ......................................... 44
4.7 Configure pressure compensation .............................................................................................44
4.7.1 Configure pressure compensation using ProLink III .................................................... 45
4.7.2 Configure pressure compensation using the Field Communicator ..............................46
4.7.3 Options for Pressure Measurement Unit .....................................................................48
Chapter 5 Configure device options and preferences ....................................................................49
5.1 Configure response time parameters ........................................................................................ 49
5.1.1 Configure Update Rate ...............................................................................................49
5.1.2 Configure Response Time ........................................................................................... 51
5.2 Configure alert handling ............................................................................................................52
5.2.1 Configure Fault Timeout ............................................................................................ 52
5.2.2 Configure Status Alert Severity ...................................................................................53
5.3 Configure informational parameters ......................................................................................... 56
5.3.1 Configure Sensor Serial Number .................................................................................56
5.3.2 Configure Sensor Material ..........................................................................................57
5.3.3 Configure Sensor Liner Material ................................................................................. 57
5.3.4 Configure Sensor Flange Type .................................................................................... 57
5.3.5 Configure Descriptor ..................................................................................................58
5.3.6 Configure Message .................................................................................................... 58
5.3.7 Configure Date ...........................................................................................................59
Chapter 6 Integrate the meter with the control system ................................................................61
6.1 Configure the transmitter channels ........................................................................................... 61
6.2 Configure the mA Output ..........................................................................................................62
6.2.1 Configure mA Output Process Variable ...................................................................... 62
6.2.2 Configure Lower Range Value (LRV) and Upper Range Value (URV) .............................64
6.2.3 Configure AO Cutoff ...................................................................................................65
6.2.4 Configure Added Damping ........................................................................................ 66
6.2.5 Configure mA Output Fault Action and mA Output Fault Level ...................................68
6.3 Configure the Frequency Output ............................................................................................... 69
6.3.1 Configure Frequency Output Polarity ......................................................................... 70
6.3.2 Configure Frequency Output Scaling Method .............................................................70
6.3.3 Configure Frequency Output Fault Action and Frequency Output Fault Level ............. 72
6.4 Configure the Discrete Output .................................................................................................. 73
6.4.1 Configure Discrete Output Source ............................................................................. 73
6.4.2 Configure Discrete Output Polarity ............................................................................ 75
6.4.3 Configure Discrete Output Fault Action ......................................................................76
6.5 Configure events ....................................................................................................................... 77
6.5.1 Configure a basic event ...............................................................................................77
6.5.2 Configure an enhanced event ..................................................................................... 78
6.6 Configure digital communications ............................................................................................ 80
6.6.1 Configure HART/Bell 202 communications ................................................................ 80
6.6.2 Configure Modbus/RS-485 communications .............................................................. 85
6.6.3 Configure Digital Communications Fault Action .........................................................86
Chapter 7 Complete the configuration ......................................................................................... 89
Contents
ii Micro Motion Model 1500 Transmitters with Analog Outputs
7.1 Test or tune the system using sensor simulation ........................................................................89
7.1.1 Sensor simulation ....................................................................................................... 90
7.2 Back up transmitter configuration ............................................................................................. 91
7.3 Enable write-protection on the transmitter configuration ......................................................... 91
Part III Operations, maintenance, and troubleshooting
Chapter 8 Transmitter operation ................................................................................................. 95
8.1 Record the process variables ..................................................................................................... 95
8.2 View process variables ...............................................................................................................96
8.2.1 View process variables and other data using ProLink III ...............................................96
8.2.2 View process variables using the Field Communicator ................................................96
8.3 View transmitter status using the status LED ............................................................................. 96
8.4 View and acknowledge status alerts .......................................................................................... 97
8.4.1 View and acknowledge alerts using ProLink III ............................................................ 97
8.4.2 View alerts using the Field Communicator ................................................................. 98
8.4.3 Alert data in transmitter memory ............................................................................... 98
8.5 Read totalizer and inventory values ........................................................................................... 99
8.6 Start and stop totalizers and inventories ....................................................................................99
8.7 Reset totalizers ........................................................................................................................100
8.8 Reset inventories .....................................................................................................................100
Chapter 9 Measurement support ............................................................................................... 103
9.1 Options for measurement support .......................................................................................... 103
9.2 Use Smart Meter Verification (SMV) ........................................................................................ 104
9.2.1 SMV requirements .................................................................................................... 104
9.2.2 SMV test preparation ................................................................................................ 104
9.2.3 Run SMV ................................................................................................................... 105
9.2.4 View test data ...........................................................................................................106
9.2.5 Schedule automatic execution of the SMV test ......................................................... 109
9.3 Use PVR, TBR, and TMR ............................................................................................................109
9.3.1 PVR, TBR, and TMR applications ................................................................................110
9.4 Piecewise linearization (PWL) for calibrating gas meters ..........................................................111
9.5 Zero the meter ........................................................................................................................ 111
9.6 Validate the meter ...................................................................................................................112
9.6.1 Alternate method for calculating the meter factor for volume flow ...........................113
9.7 Perform a (standard) D1 and D2 density calibration .................................................................114
9.7.1 Perform a D1 and D2 density calibration using ProLink III .......................................... 115
9.7.2 Perform a D1 and D2 density calibration using the Field Communicator ................... 116
9.8 Perform a D3 and D4 density calibration (T-Series sensors only) .............................................. 117
9.8.1 Perform a D3 or D3 and D4 density calibration using ProLink III ................................ 117
9.8.2 Perform a D3 or D3 and D4 density calibration using the Field Communicator ......... 118
9.9 Perform temperature calibration .............................................................................................119
Chapter 10 Troubleshooting ........................................................................................................ 121
10.1 Status LED states ..................................................................................................................... 122
10.2 Status alerts, causes, and recommendations ........................................................................... 122
10.3 Flow measurement problems ................................................................................................. 130
10.4 Density measurement problems ............................................................................................. 132
10.5 Temperature measurement problems .....................................................................................133
10.6 Milliamp output problems ....................................................................................................... 134
10.7 Frequency Output problems ....................................................................................................136
Contents
Configuration and Use Manual iii
10.8 Using sensor simulation for troubleshooting ........................................................................... 136
10.9 Check power supply wiring ......................................................................................................137
10.10 Check sensor-to-transmitter wiring ......................................................................................... 138
10.11 Check grounding .....................................................................................................................138
10.12 Perform loop tests ...................................................................................................................139
10.12.1 Perform loop tests using ProLink III ...........................................................................139
10.12.2 Perform loop tests using the Field Communicator ....................................................140
10.13 Check the HART communication loop ..................................................................................... 141
10.14 Check HART Address and mA Output Action ........................................................................... 142
10.15 Check HART burst mode ..........................................................................................................143
10.16 Check the trimming of the mA Output .................................................................................... 143
10.17 Check Lower Range Value and Upper Range Value ..................................................................143
10.18 Check mA Output Fault Action ................................................................................................143
10.19 Check for radio frequency interference (RFI) ............................................................................144
10.20 Check Frequency Output Scaling Method ............................................................................... 144
10.21 Check Frequency Output Fault Action .....................................................................................144
10.22 Check Flow Direction .............................................................................................................. 145
10.23 Check the cutoffs .................................................................................................................... 145
10.24 Check for two-phase flow (slug flow) .......................................................................................145
10.25 Check the drive gain ................................................................................................................ 146
10.25.1 Collect drive gain data .............................................................................................. 147
10.26 Check the pickoff voltage ........................................................................................................ 147
10.26.1 Collect pickoff voltage data ...................................................................................... 148
10.27 Check for internal electrical problems ..................................................................................... 149
10.27.1 Check the sensor coils ...............................................................................................149
10.28 Check the core processor LED ..................................................................................................151
10.28.1 Core processor LED states .........................................................................................151
10.29 Perform a 700 core processor resistance test ...........................................................................153
Appendices and reference
Appendix A Using ProLink III with the transmitter .........................................................................155
A.1 Basic information about ProLink III ..........................................................................................155
A.2 Connect with ProLink III .......................................................................................................... 156
A.2.1 Connection types supported by ProLink III ................................................................156
A.2.2 Connect with ProLink III to the service port ............................................................... 157
A.2.3 Make a HART/Bell 202 connection ............................................................................ 159
A.2.4 Connect with ProLink III to the RS-485 port ...............................................................163
Appendix B Using a Field Communicator with the transmitter ...................................................... 167
B.1 Basic information about the Field Communicator ................................................................... 167
B.2 Connect with the Field Communicator ................................................................................... 168
Appendix C Default values and ranges .......................................................................................... 171
Appendix D Transmitter components and installation wiring ........................................................177
D.1 Installation types ..................................................................................................................... 177
D.2 Power supply terminals ...........................................................................................................179
D.3 Input/output (I/O) wiring terminals ......................................................................................... 180
Appendix E NE 53 history ..............................................................................................................181
Contents
iv Micro Motion Model 1500 Transmitters with Analog Outputs
Part I
Getting started
Chapters covered in this part:
Before you begin
Quick start
Getting started
Configuration and Use Manual 1
Getting started
2 Micro Motion Model 1500 Transmitters with Analog Outputs
1 Before you begin
Topics covered in this chapter:
About this manual
Transmitter model code
Communications tools and protocols
Additional documentation and resources
1.1 About this manual
This manual helps you configure, commission, use, maintain, and troubleshoot the
Model 1500 transmitter.
Important
This manual assumes that the following conditions apply:
The transmitter has been installed correctly and completely according to the instructions in
the transmitter installation manual
The installation complies with all applicable safety requirements
The user is trained in local and corporate safety standards
1.2
Transmitter model code
You can verify that this manual pertains to your transmitter by ensuring the model code on
the transmitter tag matches the format.
Example:
The transmitter has a model number of the following form: 1500DEB**A******
D
4-wire remote DIN rail–mount
E
4-wire remote DIN rail transmitter with 9-wire remote enhanced core processor
B
4-wire remote DIN rail transmitter with 9-wire remote core processor
A
Analog outputs option board
1.3 Communications tools and protocols
You must have a communications tool to interface with the transmitter. Several different
communications tools and protocols are supported. You may use different tools in
different locations or for different tasks.
Before you begin
Configuration and Use Manual 3
Communications tool Supported protocols
ProLink III HART/Bell 202
Modbus/RS-485
Service port
Field Communicator HART/Bell 202
Tip
You may be able to use other communications tools, such as AMS Suite: Intelligent Device Manager,
or the Smart Wireless THUM
Adapter. Use of AMS or the Smart Wireless THUM Adapter is not
discussed in this manual. For more information on the Smart Wireless THUM Adapter, refer to the
documentation available at www.emerson.com.
1.4 Additional documentation and resources
The following additional documentation supports the installation and operation of the
transmitter.
Topic
Document
Hazardous area installation See the approval documentation shipped with the transmit-
ter, or download the appropriate documentation at
www.emerson.com.
Product Data Sheet
Micro Motion Series 1000 and Series 2000 Transmitters with
MVD
Technology Product Data Sheet (PDS)
Production Volume Reconciliation
(PVR), Transient Bubble Remedia-
tion (TBR), and Transient Mist Re-
mediation (TMR) applications
Micro Motion Oil and Gas Production Supplement
Sensor Sensor documentation
Transmitter installation
Micro Motion Model 1500 and Model 2500 Transmitters: Instal‐
lation Manual
All documentation resources are available at www.emerson.com or on the user
documentation DVD.
Before you begin
4 Micro Motion Model 1500 Transmitters with Analog Outputs
2 Quick start
Topics covered in this chapter:
Power up the transmitter
Check meter status
Make a startup connection to the transmitter
(Optional) Adjust digital communications settings
Verify mass flow measurement
Verify the zero
2.1 Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for
process measurement.
The Model 1500:
Is DC powered only
Has a minimum 19.2 to 28.8 VDC, 6.3 watts
At startup, the transmitter power source must provide a minimum of 1.0 amperes of
short-term current per transmitter
The length and conductor diameter of the power cable must be sized to provide
19.2 VDC minimum at the power terminals, at a load current of 330 mA
Procedure
1. Ensure that all transmitter and sensor covers and seals are closed.
WARNING!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers
and seals are tightly closed. For hazardous area installations, applying power while
housing covers are removed or loose can cause an explosion.
2. Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. During this period,
Alert 009 is active. The diagnostic routines should complete in approximately
30 seconds. The status LED will turn green when the startup diagnostics are
complete. If the status LED exhibits different behavior, an alert is active.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics
can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial
startup, or if power has been off long enough to allow components to reach ambient
Quick start
Configuration and Use Manual 5
temperature, allow the electronics to warm up for approximately 10 minutes before
relying on process measurements. During this warm-up period, you may observe minor
measurement instability or inaccuracy.
2.2 Check meter status
Check the meter for any error conditions that require user action or that affect
measurement accuracy.
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions. During the power-up sequence, Alert A009 is active.
This alert should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Related information
View and acknowledge status alerts
2.2.1 Transmitter status reported by LED
Status LED statesTable 2-1:
LED state Alarm condition Description
Solid green No alarm Normal operation
Flashing yellow No alarm Zero calibration procedure is in progress
Loop test is in progress
Solid yellow Low-severity alarm Alarm condition that will not cause measure-
ment error (outputs continue to report proc-
ess data)
Solid red High-severity alarm Alarm condition that will cause measurement
error (outputs in fault)
2.3 Make a startup connection to the transmitter
To configure the transmitter, you must have an active connection from a communications
tool. Follow this procedure to make your first connection to the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in
the appropriate appendix. Use the default communications parameters shown in the
appendix.
Quick start
6 Micro Motion Model 1500 Transmitters with Analog Outputs
Communications tool Connection type to use Instructions
ProLink III Modbus/RS-485 Appendix A
Field Communicator HART Appendix B
2.4 (Optional) Adjust digital communications
settings
Change the communications parameters to site-specific values.
Important
If you are changing communications parameters for the connection type that you are using, you will
lose the connection when you write the parameters to the transmitter. Reconnect using the new
parameters.
Procedure
1. To change the communications parameters using ProLink III, choose Device Tools >
Configuration > Communications.
2. To change the communications parameters using the Field Communicator, choose
On-Line Menu > Configure > Manual Setup > Inputs/Outputs > Communications.
2.5
Verify mass flow measurement
Check to see that the mass flow rate reported by the transmitter is accurate. You can use
any available method.
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in
the Process Variables panel.
Connect to the transmitter with the Field Communicator and read the value for Mass
Flow Rate.
On-Line Menu > Overview > Primary Purpose Variables
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues.
2.6
Verify the zero
Verifying the zero helps you determine if the stored zero value is appropriate to your
installation, or if a field zero can improve measurement accuracy.
Quick start
Configuration and Use Manual 7
The zero verification procedure analyzes the Live Zero value under conditions of zero flow,
and compares it to the Zero Stability range for the sensor. If the average Live Zero value is
within a reasonable range, the zero value stored in the transmitter is valid. Performing a
field calibration will not improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one
of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Procedure
1. Allow the flowmeter to warm up for at least 20 minutes after applying power.
2. Run the process fluid through the sensor until the sensor temperature reaches the
normal process operating temperature.
3. Stop flow through the sensor by shutting the downstream valve, and then the
upstream valve if available.
4. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
5. From ProLink III, choose Device Tools > Calibration > Zero Verification and
Calibration > Verify Zero and wait until the procedure completes.
6. Observe the drive gain, temperature, and density readings. If they are stable, check
the Live Zero or Field Verification Zero value. If the average value is close to 0, you
should not need to zero the meter.
7. If the zero verification procedure fails:
a. Confirm that the sensor is completely blocked in, that flow has stopped, and that
the sensor is completely full of process fluid.
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Remove or reduce sources of electromechanical noise if appropriate.
d. Repeat the zero verification procedure.
e. If it fails again, zero the meter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Related information
Zero the meter
Quick start
8 Micro Motion Model 1500 Transmitters with Analog Outputs
2.6.1 Terminology used with zero verification and zero
calibration
Term Definition
Zero In general, the offset required to synchronize the left pickoff and the right
pickoff under conditions of zero flow. Unit = microseconds.
Factory Zero The zero value obtained at the factory, under laboratory conditions.
Field Zero The zero value obtained by performing a zero calibration outside the fac-
tory.
Prior Zero The zero value stored in the transmitter at the time a field zero calibration
is begun. May be the factory zero or a previous field zero.
Manual Zero The zero value stored in the transmitter, typically obtained from a zero
calibration procedure. It may also be configured manually. Also called
“mechanical zero” or “stored zero”.
Live Zero The real-time bidirectional mass flow rate with no flow damping or mass
flow cutoff applied. An adaptive damping value is applied only when the
mass flow rate changes dramatically over a very short interval. Unit = con-
figured mass flow measurement unit.
Zero Stability A laboratory-derived value used to calculate the expected accuracy for a
sensor. Under laboratory conditions at zero flow, the average flow rate is
expected to fall within the range defined by the Zero Stability value (0 ±
Zero Stability). Each sensor size and model has a unique Zero Stability val-
ue. Statistically, 95% of all data points should fall within the range defined
by the Zero Stability value.
Zero Calibration The procedure used to determine the zero value.
Zero Time The time period over which the Zero Calibration procedure is performed.
Unit = seconds.
Field Verification Zero A 3-minute running average of the Live Zero value, calculated by the
transmitter. Unit = configured mass flow measurement unit.
Zero Verification A procedure used to evaluate the stored zero and determine whether or
not a field zero can improve measurement accuracy.
Quick start
Configuration and Use Manual 9
Quick start
10 Micro Motion Model 1500 Transmitters with Analog Outputs
Part II
Configuration and commissioning
Chapters covered in this part:
Introduction to configuration and commissioning
Configure process measurement
Configure device options and preferences
Integrate the meter with the control system
Complete the configuration
Configuration and commissioning
Configuration and Use Manual 11
Configuration and commissioning
12 Micro Motion Model 1500 Transmitters with Analog Outputs
3 Introduction to configuration and
commissioning
Topics covered in this chapter:
Configuration flowchart
Default values and ranges
Disable write‐protection on the transmitter configuration
Restore the factory configuration
3.1 Configuration flowchart
Use the following flowchart as a general guide to the configuration and commissioning
process.
Some options may not apply to your installation. Detailed information is provided in the
remainder of this manual.
Introduction to configuration and commissioning
Configuration and Use Manual 13
Configuration flowchartFigure 3-1:
Integrate device with control system
Configure device options and
preferences
Configure process measurement
Configure mass flow
measurement
Configure volume flow
meaurement
Configure temperature
measurement
Volume flow type
Liquid
Gas
Define gas properties
Configure fault handling
parameters
Configure sensor
parameters
Configure device
parameters
Configure the channel(s)
Configure digital
communications
Test and move to production
Done
Test or tune transmitter
using sensor simulation
Back up transmitter
configuration
Enable write-protection on
transmitter configuration
Configure the mA
output(s)
Configure the frequency
output(s)
Configure the discrete
output(s)
Configure events
Configure density
measurement
Configure flow direction
Configure PVR, TMR,
or TBR (if available)
Configure pressure
compensation (optional)
Introduction to configuration and commissioning
14 Micro Motion Model 1500 Transmitters with Analog Outputs
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194

Micro Motion 1500 Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI